Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K242019

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-07

    (181 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K221345, K210316, K223930, K210314, K182537

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GXR-Series Diagnostic X-ray System is intended for use in obtaining human anatomical images for medical diagnostic by using X-rays.

    Device Description

    The GXR Series Diagnostic X-ray System consists of a combination of an x-ray generator, and associated equipment such as tube stand, patient table, and, digital imaging system.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies for the GXR-series diagnostic x-ray system:

    It's important to note that the provided document is a 510(k) Summary, which is a regulatory filing for a medical device seeking clearance from the FDA based on substantial equivalence to a predicate device. It typically focuses on demonstrating that the new device is as safe and effective as a legally marketed device, rather than proving absolute performance against specific clinical acceptance criteria in a comprehensive clinical study. Therefore, the details provided often lean towards non-clinical testing and comparison with established standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" for clinical performance. Instead, it relies on demonstrating adherence to recognized safety and performance standards, and comparison of technical characteristics to a predicate device. The "performance" is implicitly deemed acceptable if the device meets these standards and is substantially equivalent to the predicate.

    Here's a generalized interpretation based on the document's content, focusing on what would typically be implied performance requirements for an X-ray system:

    Acceptance Criteria (Implied)Reported Device Performance
    Safety (Electrical, Mechanical, Radiation)Meets international safety and EMC standards (IEC 60601-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, IEC 60601-1-2), and 21CFR 1020.30. "No negative impact on safety or effectiveness" reported for differences.
    Essential Performance (X-ray Generation Parameters)Output Power Rating (32kW-82kW), Line Voltage (220-230VAC, 380/400/480VAC) are equivalent or within acceptable ranges of predicate.
    Image Quality (Digital Diagnostic X-ray System)"State-of-the-art image quality," "excellent spatial resolution, MTF, DQE and stability based on fine pixel pitch" reported. Non-clinical performance data for new flat panel detectors.
    Software Functionality (Image Processing, User Interface)"System imaging software 'RADMAX'" updated GUI for "better visibility & faster workflow." Image Processing Module 4 added; "performance verification...concluded no impact on safety and effectiveness."
    UsabilityAdheres to IEC 60601-1-6 (Usability). Operator control console designed to be "simple and user-friendly."
    Risk ManagementAdheres to ISO 14971 (Risk Management).
    Software Life Cycle ProcessesAdheres to IEC 62304 (Software Life Cycle Processes).
    Compliance with DICOM and Image Compression StandardsAdheres to NEMA PS 3.1-3.20 (DICOM) and ISO IEC10918-1 (Image Compression).
    Exposure Index of Digital X-ray Imaging SystemsAdheres to IEC 62494-1 (Exposure Index).
    Substantial Equivalence to Predicate Device (Overall)"Substantially equivalent in the areas of technical characteristics, general function, application, and intended use," and "does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a "test set" sample size in terms of clinical images or patient cases for performance evaluation against specific acceptance criteria. The testing discussed is primarily non-clinical, related to hardware and software verification and validation.
    • Data Provenance: The document implies that the testing data is generated from laboratory testing and verification during the development and modification of the device. There is no mention of clinical data or patient data being used for the performance evaluation in this 510(k) summary. Given the context of a 510(k), particularly for an X-ray system, the primary focus is on engineering and performance testing against standards, rather than large-scale clinical studies. The data is thus likely prospective in terms of being generated specifically for this submission but is non-clinical in nature.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. As the evaluation is non-clinical, there is no mention of "ground truth established by experts" in the context of diagnostic performance evaluation. The "ground truth" for non-clinical testing would typically be the expected technical output or adherence to a standard, rather than expert interpretation of images.

    4. Adjudication Method for the Test Set

    This information is not provided as the testing described and implied is non-clinical and does not involve expert adjudication of diagnostic findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or performed as part of this 510(k) submission. The document focuses on demonstrating substantial equivalence through technical comparisons and compliance with standards, not on proving improved reader performance with or without AI assistance.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    This is not applicable in the sense of an AI algorithm's standalone performance. The device is a diagnostic X-ray system, which is inherently designed to be used with a human interpreter (a medical professional). While it has image processing software ("RADMAX"), this software enhances images for human diagnosis, not to provide an automated diagnosis itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical testing is adherence to technical specifications and international standards. For example, for radiation output, the ground truth is that the device delivers the specified kVp and mA, and for electrical safety, that it meets the requirements of IEC 60601-1. For image quality, it refers to intrinsic properties like spatial resolution, MTF, and DQE, which are measured objectively, not subjective expert consensus on diagnostic findings.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. The device is an X-ray imaging system, not an AI/ML diagnostic algorithm that requires a training set of medical images in the conventional sense. The "training" for the device would involve calibration and configuration during manufacturing and installation to ensure it meets its technical specifications. The "RADMAX" software has image processing modules, but the document does not suggest these are deep learning models trained on vast datasets.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as #8. If any parameters for the image processing modules are "learned" or optimized, the document does not elaborate on this process or the ground truth used for such optimization.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242015

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-16

    (159 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K221345, K210316, K223930, K202572, K210314, K223930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 'TOPAZ Mobile X-ray System' is intended for use in obtaining human anatomical images of patients who cannot be moved to the radiology department for medical diagnosis.

    Device Description

    "TOPAZ" system is a system providing state-of-the-art image quality, user interface. "TOPAZ" system may be moved quietly and smoothly with motor drive mechanism "TOPAZ" system has a basic type column, and a collapsible type column option with a trendy design that allows driving without disturbing the front view. The core part of x-ray source adopts high quality tube assembly, motorized x-ray collimator. HV cable assembly and High Voltage X-Ray Generator. Touch screen LCD based x-ray control console provides user-friendly interface and easy technique selection. Collimator supports high accuracy for selected x-ray field size over any SID. Direct radiography via flat panel detector improves-exam speed and comfort with efficiency. Digital flat panel detector with Csl screen provides spatial resolution, MTF, DQE and stability based on fine pixel pitch. Selection of an anatomical study on the Digital Imaging Software automatically sets up the x-ray generator's preprogrammed exposure technique. The types of "TOPAZ" system are divided into TOPAZ-32D, and TOPAZ-40D according to maximum power and mA. The higher the maximum output, the wider the mA range to choose from, giving the user more technical options to choose from. The "TOPAZ Mobile X-ray System" consists of a tube assembly. x-ray collimator. High Voltage X-Rav Generator, detector and mechanical parts for mobility.

    AI/ML Overview

    The provided text is a 510(k) Summary for the TOPAZ Mobile X-ray System, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria in the format typically used for AI/CADe devices. This document describes the device, its intended use, technological characteristics, and differences from the predicate, along with non-clinical testing for safety and EMC standards.

    Therefore, the specific information about "acceptance criteria and the study that proves the device meets the acceptance criteria" as requested for AI/CADe devices (including details like sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, and training set details) is not present in this 510(k) Summary.

    This document primarily asserts that the "TOPAZ Mobile X-ray System" is substantially equivalent to the predicate device "TOPAZ Mobile DR System (K201124)" based on:

    • Identical intended use.
    • Similar technological characteristics, with modifications thoroughly tested for safety and effectiveness against international standards.
    • Nonclinical testing results provided in the 510(k) demonstrating that predetermined acceptance criteria were met for safety (electrical safety, EMC, radiation protection) and software validation.

    The "study that proves the device meets the acceptance criteria" in this context refers to the nonclinical testing against various recognized international and FDA standards, not a clinical performance study with human readers or pathology, as would be expected for AI/CADe systems.

    Here's a summary of the available information regarding acceptance criteria and testing, tailored to what is provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria and reported device performance in the typical format for clinical accuracy for AI/CADe. Instead, it states that the device was assessed, tested, and passed predetermined testing criteria during validation testing, aligning with the risk analysis. It also confirms that the device meets "all the requirements listed in the Standards" (see the Standards table below). The "device performance" reported is its conformance to these standards and its substantial equivalence to the predicate.

    Nonclinical Standards Met (acting as acceptance criteria for safety and effectiveness):

    StandardDescriptionFDA Rec. StandardReported Device Performance
    IEC 60601-1Medical electrical equipment, Part 1: General requirements for basic safety and essential performance19-46Met all requirements
    IEC 60601-1-2 (EMC)Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances Requirements and tests.19-36Met all requirements
    IEC 60601-1-3Medical electrical equipment Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray Equipment12-336Met all requirements
    IEC 60601-1-6Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability5-132Met all requirements
    IEC 60601-2-28Medical electrical equipment Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis12-309Met all requirements
    IEC 60601-2-54Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy12-348Met all requirements
    IEC 62304:2006Medical device software - Software life cycle processes13-79Met all requirements
    ISO 14971:2019Medical devices - Applications of risk management to medical devices.5-125Met all requirements
    ISO 15223-1Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements.5-134Met all requirements
    NEMA PS 3.1 - 3.20 (2016).Digital Imaging and Communications in Medicine (DICOM) Set DICOM Standard.12-349Met all requirements
    IEC/ISO10918-1Information technology - Digital compression and coding of continuous-tone still images: Requirements and guidelines12-261Met all requirements
    IEC 62494-1Medical electrical equipment - Exposure index of digital X-ray imaging systems - Part 1: Definitions and requirements for general radiography.12-215Met all requirements
    TR 60601-4-2Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems19-19Met all requirements
    FDA Guidance (various)Pediatric Information for X-ray Imaging Device, Format for Traditional and Abbreviated 510(k)s, Submission of 510(k)s for Solid State X-ray Imaging Devices, Content of Premarket Submissions for Device Software Functions, Content of Premarket Submissions for Software contained in Medical Devices, Cybersecurity in Medical Devices.N/AComplies/Addressed
    21 CFR 1020.30-31Applicable requirements for X-ray equipmentN/AConforms

    2. Sample size used for the test set and the data provenance:

    • Not Applicable/Not Provided. The document describes non-clinical engineering and software validation testing against standards, not a clinical study involving a "test set" of patient data for diagnostic performance. The focus is on the device's hardware, software (RADMAX), and new flat panel detectors meeting safety and electrical standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not Applicable/Not Provided. Ground truth establishment by experts is relevant for clinical performance studies, which this document does not describe.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable/Not Provided. This is relevant for clinical performance studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This document describes a 510(k) for an X-ray system, not an AI/CADe system. No MRMC study was performed or is mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable/No. The device itself is an X-ray system, not an algorithm, and its performance is assessed in terms of meeting engineering and regulatory standards, not standalone diagnostic performance. The imaging software (RADMAX) is mentioned as identical to the predicate and has a "Basic Documentation Level" of concern, implying human interpretation of images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable/Not Provided. For the non-clinical testing described, "ground truth" would relate to the correct functioning of the hardware and software according to specifications and standards, not clinical diagnostic accuracy.

    8. The sample size for the training set:

    • Not Applicable/Not Provided. This is relevant for AI/ML models. While the device contains software, it is not described as an AI/ML diagnostic algorithm needing a training set. The software changes are primarily GUI and image processing module updates, verified for impact on safety and effectiveness (not AI training).

    9. How the ground truth for the training set was established:

    • Not Applicable/Not Provided. As above, this is not an AI/ML submission requiring a training set.

    In summary, the provided 510(k) document is for a mobile X-ray system and demonstrates substantial equivalence through nonclinical testing against recognized performance, safety, and EMC standards, rather than a clinical performance study with acceptance criteria related to diagnostic accuracy, which would be typical for AI-powered diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1