510(k) Data Aggregation

    K Number
    K250178

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (268 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K223714, K112585, K081751, K223677, K153111, K140293

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adin Customized Abutments are intended for use with Adin dental implants in partially or fully edentulous mandibles and maxillae, as a support for single-unit or multiple-unit screw-retained or cement-retained restorations.

    All digitally designed abutments for use with Adin Ti-base and Adin Ti-blanks are intended to be sent to an FDA-registered Adin Dental Implants validated milling center for manufacture or to be manufactured according to the digital dentistry workflow, which integrates multiple components: scans from desktop and intraoral scanners, CAD and CAM software and a milling machine with associated accessories.

    Device Description

    The purpose of this submission is to obtain marketing clearance for Adin Customized Abutments, a series of components compatible with dental implants from Adin Dental Implant Systems Ltd., intended for use in the fabrication of patient-specific abutments. The subject device components include CAD CAM abutments (Ti-bases, Ti-blanks). All patient-specific abutment fabrication is by prescription on the order of the clinician. Fabrication of patient-specific abutments will be performed at validated milling centers or, by using a validated digital dentistry workflow, at point-of-care facilities.

    The design and fabrication of the zirconia superstructure for Ti-bases or of patient-specific abutments from Ti-blanks will be conducted using a digital dentistry workflow requiring the use of the following equipment and software:

    Scanner: iTero Intra Oral Scanner or Medi Corp. Identica 3D Desktop scanner
    Design Software: EXOCAD AbutmentCAD design software (K193352)
    CAM Software: WORKNC Dental
    Milling Machine: Ceramill Motion 2.

    The superstructures to be used on Ti-bases are to be made from the following material and are to be cemented as follows:

    Zirconia Material: ArgenZ Ultra (K071410)
    Cement: PANAVIA V5 (K150704).

    The abutments are provided non-sterile and are intended to be cleaned and steam-sterilized by the clinician prior to use according to the instructions given in the Instructions for Use (IfU) accompanying the device.

    No compatibility with implants other than those from Adin Dental Implant Systems Ltd. is claimed or intended. Ti-bases are compatible with Adin Dental implants cleared in K081751 and K223714. Ti-blanks are compatible with Adin Dental implants cleared in K081751, K153111, K140293, and K112585.

    Ti-base abutments are two-piece abutments, with the base component cemented to a CAD CAM fabricated zirconia superstructure. The final two-piece abutment (base and cemented superstructure) is the finished device used for the prosthetic restoration. Each patient-specific superstructure is individually prescribed by the clinician.

    Ti-base abutments are provided with a compatible abutment screw that retains the base to the implant (the restoration is screw-retained on the implant level). The manufactured superstructure is attached to the Ti-base.
    Engaged Ti-bases include anti-rotational connections that engage with the dental implant and are intended for single tooth restorations. Non-engaged Ti-bases include round connections that do not limit rotation and are intended for multi-unit/bridge restorations. Ti-base Angular are provided only in engaging designs.

    The all-metal Ti-base components are straight with regard to the implant axis. Angular Ti-base includes an angled upper portion that allows the screw channel either to be straight relative to the implant axis or to be angled to allow for moving the screw channel in an oral direction (up to 20°) for esthetic purposes.

    All Ti-bases are color coded per platform by an anodizing process for easier identification. Adin Ti-bases are available for Adin implant platform connections (standard internal hex – RS, and UniFit - UF)

    The all-metal Ti-base components are provided in engaging and non-engaging versions with gingival heights ranging from 0.65 mm to 4.0 mm. They are provided with post heights of 4.0, 6.0, and 8.0 mm. Ti-base Angular is a component with gingival heights ranging from 1.0 mm to 3.0 mm, provided in an engaging design. Ti-base Angular includes an angular cutout in the post to allow for moving the screw channel in an oral direction for esthetic purposes. It is provided with post heights from 4.0 mm to 8.0 mm.

    The design parameters for the CAD CAM zirconia superstructure to be used on Ti-base and Ti-base Angular, are:

    • Minimum wall thickness – 0.5 mm
    • Minimum post height for single unit restorations – 4.0 mm
    • Minimum gingival height – 0.0 mm
    • Maximum gingival height – 4.5 mm
    • Maximum angulation of the final abutment - 20°

    Adin Ti-blank abutments include on one side a pre-fabricated connection for the dental implant and on the other side a pre-fabricated connection to the milling machine blank holder. They are available in diameters of 11.5 mm and 15.8 mm and are provided with a compatible abutment screw.

    All Ti-blank designs are subject devices to be manufactured using a CAD CAM workflow. Ti-blank abutments are provided in engaging and non-engaging designs, are available for all Adin implant platform connections, and are made of titanium alloy (Ti 6Al 4V).

    The design parameters for Ti-blanks are:

    • Minimum wall thickness – 0.7 mm for RS; 0.5 mm for UNP, NP, RP, WP
    • Minimum post height for single unit restorations – 4.0 mm
    • Maximum milled abutment height above implant level – 10 mm
    • Minimum gingival height – 0.5 mm
    • Maximum gingival height – 4.0 mm for RS; 1.0 mm for UNP; 3.0 mm for NP, RP, WP
    • Maximum angulation of the final abutment – 25° for RS, NP, RP, WP; 15° for UNP
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233208

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-01-24

    (118 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NobelProcera® Titanium ASC Abutments:

    NobelProcera® Abutment Titanium is a patient-matched CAD/CAM prosthetic component directly connected to endosseous dental implants and is indicated for use as an aid in prosthetic rehabilitation.

    Omnigrip Clinical Screw Titanium:

    Clinical and Abutment Screws are indicated for use to secure a dental abutment or framework to a dental implant in the maxilla or mandible for supporting tooth replacements and are indicated as an aid in prosthetic rehabilitation.

    Device Description

    The Subject Device NobelProcera® Titanium ASC Abutment is composed of two device lines: NobelProcera® Titanium ASC Abutment and the Clinical Screw Omnigrip Titanium.

    NobelProcera® Titanium ASC Abutment is a patient matched CAD/CAM dental prosthesis which is connected to the Nobel Biocare implants featuring an internal conical and/or internal tri-channel connection, is available in the platform sizes NP, RP WP and 6.0 (for internal tri-channel connection only) and is intended for use as an aid in prosthetic rehabilitation to restore chewing function and esthetic appearance.

    NobelProcera® Titanium ASC Abutments undergo patient matched customization at a Nobel Biocare production facility for the final, finished abutment device manufacturing.

    NobelProcera® Titanium ASC Abutment is connected to the implant with a clinical screw and features an angulated screw channel which can be defined by the customer in an angulation (to the implant's axis) between 0° and 25°, in addition the abutment can be angulated to a maximum of 30°. The clinical screw features the Omnigrip Interface which allows tightening up to 25°.

    NobelProcera® Titanium ASC Abutment and Omnigrip Clinical Screw Titanium are composed of titanium vanadium alloy Ti6Al4V ELI (ISO 5832-3, ASTM F136) and the surface of the abutments are provided with and without anodization and the Omnigrip Clinical Screw Titanium are provided with and without DLC coating.

    The finished NobelProcera® Titanium ASC Abutment supports the placement of a cement-retained dental prosthesis.

    Omnigrip Clinical Screw Titanium is available for the NP, RP, WP and 6.0 (for internal tri-channel connection only) platform, the devices connect the NobelProcera Titanium ASC Abutments to the dental implants. The devices feature an Omnigrip interface.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and device performance.

    Important Note: The provided document is an FDA 510(k) Premarket Notification letter. This type of document is a submission seeking regulatory clearance, not a standalone study report. As such, it outlines the basis for demonstrating substantial equivalence to a predicate device, which includes non-clinical testing. It does not typically contain detailed acceptance criteria and performance data for a study proving a device meets those criteria in the way a clinical trial report or a comprehensive validation study would.

    The document primarily focuses on demonstrating that the new device is substantially equivalent to legally marketed predicate devices, meaning it is as safe and effective. It refers to non-clinical tests that were performed, but does not present the specific acceptance criteria or the numerical results of those tests in a "performance table" format.

    Therefore, many of the requested points below cannot be fully extracted from this document, as it's not the type of report that contains that level of detail for acceptance criteria and specific study outcomes. I will highlight what can be inferred or directly stated from the provided text, and what cannot.


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" alongside "reported device performance" in a quantitative manner as one might find in a detailed engineering validation report or a clinical study. Instead, it states that "non-clinical tests demonstrate that the device is substantial equivalent."

    The document mentions several tests and standards the device was subjected to:

    • Packaging system performance testing: per ASTM D4169. (Specific acceptance criteria and reported performance not detailed.)
    • Dynamic loading testing: conducted according to ISO 14801 and FDA Guidance Document ("Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" (May 12, 2004)). (Specific acceptance criteria and reported performance not detailed.)
    • Magnetic Resonance compatibility testing: according to ASTM F2052, ASTM F2213, ASTM F2119, and ASTM F2182. Status: "MR Conditional." (This is a performance outcome, but the specific metrics and acceptance thresholds for each ASTM standard are not detailed.)
    • Verification of biocompatibility: in accordance with ISO 10993-1. (The conclusion is that "no new issues regarding biocompatibility were raised," implying acceptance, but specific criteria and test results not detailed.)
    • End user cleaning and sterilization validation: in accordance with ISO 17665-1. (Implied acceptance, but specific criteria and results not detailed.)

    The "acceptance criteria" here are generally referred to as meeting the requirements of the standards and demonstrating substantial equivalence to the predicate device. The "reported device performance" is summarized as favorable results sufficient for substantial equivalence determination.

    2. Sample sizes used for the test set and the data provenance

    • Sample sizes: Not specified in the provided text for any of the non-clinical tests mentioned (packaging, dynamic loading, MR compatibility, biocompatibility, sterilization validation).
    • Data provenance: Not explicitly stated (e.g., country of origin). The testing is non-clinical/bench testing, not patient data. It is implied to be internal testing by the manufacturer, Nobel Biocare AB (Sweden).
    • Retrospective or prospective: N/A, as these are non-clinical/bench tests, not clinical studies involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Applicability: This question is not directly applicable to the type of non-clinical, bench testing described for the NobelProcera® Titanium ASC Abutment and Omnigrip Clinical Screw Titanium. "Ground truth" established by experts (e.g., radiologists for image interpretation) is relevant for AI/ML device evaluations, particularly in diagnostics. This document refers to physical device testing (mechanical, biocompatibility, etc.).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Applicability: Not applicable. Adjudication methods are typically used in clinical studies or expert reviews to resolve disagreements in interpretations or diagnoses for establishing "ground truth," which is not the nature of the non-clinical tests described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Applicability: Not applicable. This device is a dental abutment and screw, not an AI/ML-enabled diagnostic device for image interpretation. Therefore, no MRMC study or AI assistance evaluation would have been performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Applicability: Not applicable. See point 5. This is a physical medical device, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Applicability: The concept of "ground truth" as typically applied in AI/ML validation (e.g., for diagnostic accuracy) does not directly apply to the non-clinical device testing described. For these tests, the "ground truth" is typically defined by the specified parameters and performance requirements of the relevant international standards (e.g., ISO 14801 for dynamic loading, ISO 10993-1 for biocompatibility). The devices are expected to meet the performance criteria defined by these standards.

    8. The sample size for the training set

    • Applicability: Not applicable. This refers to the training of an AI/ML algorithm. The document describes a physical medical device.

    9. How the ground truth for the training set was established

    • Applicability: Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223535

    Validate with FDA (Live)

    Date Cleared
    2023-06-20

    (209 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMARTbase Abutment System is designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially or fully edentulous patient. The SMARTbase Abutment System is intended for use in the mandible or maxilla in support of single or multiple unit restorations.

    The SMARTbase Abutment System integrates multiple components for use in both a traditional and digital dentistry workflow: scan files from Intra-oral Scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The SMARTbase Abutment System consists of two major parts: the titanium base and zirconia top components make up a two-piece abutment.

    • SMARTbase abutment for narrow (3.2mmD) Legacy implants: Indicated for single-tooth replacement of mandbular central and lateral incisors and maxillary lateral incisors.
    • SMARTbase abutment for short (8mm) 3.7mmD Legacy implants: Indicated for tooth replacement of mandbular and maxillary central and lateral incisors.
    • SMARTbase abutment for narrow diameter (3.2. 3.3mm) InterActive implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. Also indicated for multiple tooth replacements or denture stabilization.
    Device Description

    The SMARTbase Abutment System is a two-piece engaging and non-engaging dental implant abutment comprised of a titanium base and a zirconia top (which can be supplied with the base or acquired separately by the customer). There are three device lines offered in the SMARTbase Abutment System: Legacy™ SMARTbase Abutment, InterActive™ SMARTbase Abutment, and SMARTbase Cylinder.

    The abutments are offered in three widths (narrow, regular, and wide), platform diameters of 3.0mm, 3.5mm, 4.5mm and 5.7mm for Legacy™ and 3.0mm, 3.4mm for InterActive™, and collar (titanium base) heights of 0.25, 1.0, and 2.0 mm in order to accommodate different patient anatomies. The SMARTbase Cylinder is a two-piece non-engaging dental implant and multi-unit abutment cylinder comprised of a titanium base and a zirconia top (which can be supplied with the base or acquired separately by the customer). The SMARTbase Cylinder is offered in two heights 9.0mm (that can be shortened to 4.0mm) and 4.0mm and in one width, platform diameters and collar (titanium base) height in order to accommodate different patient anatomies.

    The subject device is supplied with fixation screws that function as an extension of the implant or multi-unit abutment to which the SMARTbase Abutment or SMARTbase Cylinder is secured and is used with several accessories in digital workflows to fabricate the patient-specific restorations, including scan adapters, implant analogs, and off-axis tools.

    The available design options for the zirconia top components to be provided either as a superstructure (to then receive a separate crown or bridge) or hybrid abutment-crown. There are three workflow options for fabricating the zirconia top component which fits the titanium abutment base:

    (1) end user creation of a press-ceramic material by conventional wax-up technique,

    (2) Implant Direct design and milling of zirconia in stock sizes using ceramic material of ZirCAD Prime (K142233) and provision of same to the end user, and

    (3) digital workflow using 3Shape or Exocad software where CAD design and milling of the superstructure or hybrid crown component is done at the end user's dental laboratory/office; the CAD design requires loading of Implant Direct's abutment design library to the 3Shape or Exocad software to design the superstructure or hybrid crown component within the established design limitations and specifications. The digital workflow includes use of the following products (not subject devices of this submission):

    • Ceramic material: ZirCAD Prime (K142233) .
    • . Cement: Maxcem Elite Self-Etch/Self-Adhesive Resin Cement (K060469)
    • . Composite: Kerr Harmonized (K151332)
    • Intra oral scanners: Medit Scanner, ITero Scanner Trios Scanner, CareStream . Scanner
    • Abutment design software: 3Shape Abutment Designer™ Software (K151455) and • Exocad AbutmentCAD Software (K193352)
    • . Milling machine: Wieland-Zenotec Select, Zenotec CAM, iCAM V5, and imes icore
    AI/ML Overview

    This document describes a premarket notification for the "SMARTbase Abutment System," a dental implant abutment. The document focuses on demonstrating that the new device is substantially equivalent to existing predicate devices through various performance tests.

    Here's an analysis of the acceptance criteria and the studies performed, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results in a single, clear format for each test. Instead, it states that "The test results support that the Subject Device met the performance specifications as intended." and "The difference does not introduce a fundamentally new scientific technology and the nonclinical tests demonstrate that the device is substantial equivalent."

    However, we can infer the types of performance tests and the general acceptance (i.e., compliance) from the "x. Performance Testing Data" section.

    Acceptance Criteria (Inferred Standard Compliance)Reported Device Performance (Implied)
    Fatigue Resistance (according to ISO 14801 and FDA Guidance Document: "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments")The device met performance specifications as intended, demonstrating substantial equivalence.
    Magnetic Resonance Compatibility (according to ASTM F2052, ASTM F2213, ASTM F2119, and ASTM F2182)The device met performance specifications as intended, demonstrating substantial equivalence.
    Biocompatibility (according to ISO 10993-1)The device is biocompatible as per the standard.
    Steam Sterilization Validation (according to ISO 17665-1)Sterilization process was validated.
    Shipping Validation (according to ASTM D4169)Shipping validation was performed.
    Software Verification & Validation (Abutment Design Library) (demonstrating use with 3Shape Abutment Designer Software and Exocad AbutmentCAD software)Demonstrated that design restrictions prevent creation of components outside allowable limitations; established design limitations are locked and unmodifiable.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample sizes used for each of the performance tests (e.g., number of abutments subjected to fatigue testing).
    • Data Provenance: The document does not explicitly state the country of origin or whether the studies were retrospective or prospective. Given that this is a premarket notification for regulatory clearance, the testing would generally be conducted by the manufacturer or a contract research organization on manufactured devices, following established protocols. It is non-clinical testing, meaning it's not performed on patients.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This question is not applicable as the document describes non-clinical performance testing for a medical device (dental abutment), not studies involving expert evaluation of diagnostic outputs or ground truth for AI algorithms. The "ground truth" here is defined by engineering standards (e.g., ISO, ASTM).

    4. Adjudication Method for the Test Set

    This question is not applicable for the same reasons as #3. Adjudication methods (like 2+1) are typically used for establishing ground truth in clinical or diagnostic studies involving human interpretation or pathology.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    This question is not applicable. The document describes non-clinical performance testing of a physical dental abutment and associated design software. It does not involve human readers, cases, or AI assistance in a diagnostic context.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    For the software component:

    • Standalone Performance: Yes, standalone software verification and validation testing was performed for the abutment design library. This testing demonstrated that the software's inherent design restrictions and specifications for creating the top half of the two-piece abutment prevent designs outside allowable limitations and that the design limitations are locked and cannot be modified. This is essentially an "algorithm only" test to ensure the software functions as designed within its specified parameters.

    7. The Type of Ground Truth Used

    • For Physical Device Performance (Fatigue, MR Compatibility, Biocompatibility, Sterilization, Shipping): The ground truth is based on established international and national standards (e.g., ISO 14801, ISO 10993-1, ISO 17665-1, ASTM F2052, ASTM D4169). Compliance with these standards serves as the "ground truth" for acceptable mechanical, safety, and operational performance.
    • For Software Verification & Validation: The ground truth is the defined design limitations and specifications established by the manufacturer for the abutment components. The software's ability to enforce these limitations (preventing designs outside them and locking specifications) constitutes meeting this "ground truth."

    8. The Sample Size for the Training Set

    This question is not applicable. The device is a physical dental abutment and associated design software. It is not an AI/ML algorithm that requires a "training set" in the conventional sense for learning patterns from data to make predictions or classifications. The "training" for the software is its programmed adherence to design rules and specifications.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons as #8. There is no training set mentioned for an AI/ML model. The software's functional parameters are established through engineering design and specifications.