Found 5 results

510(k) Data Aggregation

    K Number
    K243287

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-24

    (129 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K223545 APRO 70 Catheter and Alembic Aspiration Tubing

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 70 Swift Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    The Alembic Aspiration Tubing is intended to connect the APRO 70 Swift Catheter to the aspiration pump.

    Device Description

    The APRO 70 Swift Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 70 Swift Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 70 Swift Catheter targets aspiration from a suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 70 Swift Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 70 Swift Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 70 Swift Catheter is available in lengths of 125 cm. 132 cm, and 135 cm and is provided with an introducer sheath.

    The Alembic Aspiration Tubing connects the APRO 70 Swift Catheter to the aspiration pump. The flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study results for the APRO 70 Swift Catheter and Alembic Aspiration Tubing. It is important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical performance data, rather than a clinical trial report. Therefore, some of the requested information, particularly regarding clinical studies, human readers, and effect sizes, is not applicable or available in this document.

    Here's the breakdown of the information based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and performance data are primarily from non-clinical bench testing.

    Test CategoryAcceptance CriteriaReported Device Performance
    Visual and Dimensional CharacteristicsCatheter meets visual and dimensional specifications.Introducer Sheath meets visual and dimensional specifications.The APRO 70 Swift Catheter met the acceptance criteria.The Introducer Sheath met the acceptance criteria.
    ParticulateCatheter meets the acceptance criteria.Subject device was evaluated with a predicate device under the same test conditions.The APRO 70 Swift Catheter particulates were comparable to the predicate device.
    Vacuum IntegrityCatheter is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 70 Swift Catheter met the acceptance criteria.
    Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 70 Swift Catheter met the acceptance criteria.
    Catheter Hub LeakageCatheter does not leak into hub assembly during aspiration, with methods specified in ISO 10555-1, Annex D.The APRO 70 Swift Catheter met the acceptance criteria.
    Hub CompatibilityHub meets the requirements per EN ISO 80369-7.The APRO 70 Swift Catheter met the acceptance criteria.
    Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 70 Swift Catheter met the acceptance criteria.
    Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 70 Swift Catheter met the acceptance criteria.
    Fluid LeakageCatheter must withstand pressure with methods per ISO 10555-1 Annex C.The APRO 70 Swift Catheter met the acceptance criteria.
    Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 70 Swift Catheter met the acceptance criteria.
    Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 70 Swift Catheter met the acceptance criteria.
    Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 70 Swift Catheter met the acceptance criteria.
    Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 70 Swift Catheter met the acceptance criteria.
    Tip Buckling ForceCatheter tip buckling force must be acceptable. Forces were compared to a predicate.The APRO 70 Swift Catheter met the acceptance criteria.
    Simulated Use including Clot RetrievalWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications. Clot retrieval from various locations of the anatomical model was performed using the subject device system.The APRO 70 Swift Catheter and Alembic Aspiration Tubing met the acceptance criteria.
    Biocompatibility (Sensitization)Did not elicit a sensitization response.Non-sensitizing
    Biocompatibility (Irritation)Demonstrated no evidence of irritation.Non-irritant
    Biocompatibility (Cytotoxicity)Did not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
    Biocompatibility (Hemolysis – Indirect)No significant differences between the test article extract and negative control article results.Non-hemolytic
    Biocompatibility (Hemolysis - Direct)No differences between the hemolytic index of the test article and the negative control.Non-hemolytic
    Biocompatibility (Thrombogenicity - PTT)Average clotting time of the test article was greater than vehicle control and negative control.Acceptable clotting times
    Biocompatibility (Thrombogenicity - Platelet Leukocyte Count)Performed similar to comparator and negative controls.Non-thrombogenic
    Biocompatibility (Thrombogenicity - Comparative Surface and Geometry Assessment)Comparative surface assessment using 40X optical microscopy at representative locations of the subject device and the reference device APRO 70 Catheter did not find any differences in roughness or presence of any defects. The subject device has the same geometry as the reference device APRO 70 Catheter.Acceptable results
    Biocompatibility (SC5b9 Complement Activation)Sc5b9 concentration of the test article was statistically less than the positive control and was not statistically higher than the negative control.Acceptable
    Biocompatibility (Acute Systemic Toxicity)No weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice.Non-toxic
    Biocompatibility (Material-Mediated Pyrogenicity)All individual rabbits for both the test article and negative control showed a total rise in temperature of < 0.5 ℃ and were determined to be nonpyrogenic.Non-pyrogenic

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state specific sample sizes for each non-clinical bench test. The testing involved various repetitions per test to demonstrate meeting acceptance criteria. The data provenance is non-clinical bench testing performed by Alembic, LLC. No human (clinical) test set is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable (N/A) as the studies described are non-clinical bench tests and do not involve expert interpretation or ground truth establishment in the context of medical imaging or diagnosis.

    4. Adjudication Method for the Test Set

    This information is not applicable (N/A) for non-clinical bench testing. Adjudication methods are typically relevant for human reader studies or clinical trials where expert consensus is needed to establish ground truth for complex interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The document states: "Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary." This device is a physical catheter, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable (N/A). This is a 510(k) submission for a physical medical device (catheter and tubing), not a software algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical bench tests, the "ground truth" or reference for performance was established by predefined engineering specifications, international standards (e.g., ISO 10555-1, EN ISO 80369-7, ISO 11135, ISO 10993-1), and comparisons to predicate devices. For example, particulate levels were compared to predicate devices, and clot retrieval was assessed against "functionality specifications" using an anatomical neurovascular model. Biocompatibility tests followed established methodologies like those in ISO 10993-1.

    8. The Sample Size for the Training Set

    Not applicable (N/A). This submission is for a physical medical device, not a machine learning or AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable (N/A). As there is no training set for an algorithm, there is no ground truth established for it in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243297

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-14

    (88 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K223545

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 70 Swift Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 70 Swift Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The APRO 70 Swift Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 70 Swift Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 70 Swift Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 70 Swift Catheter is provided with an Introducer Sheath.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the APRO 70 Swift Catheter. It focuses on demonstrating the device's substantial equivalence to a predicate device, primarily through non-clinical performance data.

    Here's a breakdown of the information requested, based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaReported Device Performance (Conclusion)
    Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications.The APRO 70 Swift Catheter met the acceptance criteria.
    ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 70 Swift Catheter particulates were comparable to the predicate device.
    Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 70 Swift Catheter met the acceptance criteria.
    Hub Air LeakageCatheter does not leak air into hub assembly with methods specified in ISO 10555-1, Annex D.The APRO 70 Swift Catheter met the acceptance criteria.
    Hub CompatibilityCatheter meets the requirements specified in ISO 80369-7.The APRO 70 Swift Catheter met the acceptance criteria.
    Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 70 Swift Catheter met the acceptance criteria.
    Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 70 Swift Catheter met the acceptance criteria.
    Liquid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 70 Swift Catheter met the acceptance criteria.
    Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 70 Swift Catheter met the acceptance criteria.
    Hub and Shaft Tensile StrengthCatheter hub and shaft must meet tensile strength specification.The APRO 70 Swift Catheter met the acceptance criteria.
    Tip Tensile StrengthCatheter tip must meet tip tensile strength specification.The APRO 70 Swift Catheter met the acceptance criteria.
    Tip Buckling ForceCatheter tip buckling force must be acceptable. Forces were compared to a predicate.The APRO 70 Swift Catheter was comparable to the predicate device.
    Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 70 Swift Catheter met the acceptance criteria.
    Simulated UseWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model and during simulated clot retrieval, the Catheter must meet functionality specifications including compatibility with a stent retriever.The APRO 70 Swift Catheter met the acceptance criteria.
    Sterilization (Ethylene Oxide)Sterility assurance level of 1 x 10^-6 in accordance with ISO 11135.Verified to a sterility assurance level of 1 x 10^-6.
    Shelf-Life (Aging Studies)Subject device and packaging remain functional for the labeled expiration date; packaging integrity, seal strength, and device functionality meet acceptance criteria.Met the acceptance criteria.
    Sensitization (Guinea Pig Maximization)Not elicit a sensitization response.Non-sensitizing.
    Irritation/Intracutaneous ReactivityNo evidence of irritation.Non-irritant.
    Cytotoxicity (MEM Elution, L929 cells)Not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic.
    Hemolysis - IndirectNo significant differences between the test article extract and negative control.Non-hemolytic.
    Hemolysis - DirectNo differences between the hemolytic index of the test article and the negative control.Non-hemolytic.
    Thrombogenicity - Partial Thromboplastin Time (PTT)Acceptable clotting times (average clotting time of test article greater than vehicle control and negative control).Acceptable clotting times.
    Thrombogenicity - Platelet Leukocyte CountPerformed similar to comparator and negative controls.Non-thrombogenic.
    Thrombogenicity - Comparative Surface and Geometry AssessmentNo differences in roughness or presence of any defects compared to reference device. Same geometry as reference device.Acceptable results.
    SC5b9 Complement ActivationSc5b9 concentration statistically less than positive control and not statistically higher than negative control.Acceptable.
    Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity.Non-toxic.
    Material-Mediated PyrogenicityTotal rise in temperature of < 0.5 °C for all individual rabbits.Non-pyrogenic.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document extensively details non-clinical bench testing, sterility testing, shelf-life testing, and biocompatibility testing. It does not specify sample sizes for each individual test or the data provenance in terms of country of origin or retrospective/prospective data collection. The general statement "Alembic performed non-clinical bench, sterility, shelf-life, and biocompatibility testing" implies these tests were conducted by the manufacturer for this submission. Given that these are bench and lab tests for a medical device, they would be considered prospective in nature, as they are experiments conducted specifically to evaluate the device. The location of these tests (country of origin of the data) is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the entire study described is based on non-clinical bench, sterility, shelf-life, and biocompatibility testing. There is no "ground truth" derived from expert review of clinical data (like medical images) as there are no human studies or AI components described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this study is based on non-clinical testing, not human expert evaluations of a test set, so no adjudication method for establishing ground truth is relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is a 510(k) submission for a physical medical device (catheter), not an AI/Software as a Medical Device (SaMD). Therefore, no MRMC study or AI assistance evaluation was performed or is relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This is not an AI/SaMD device, so no standalone algorithm performance testing was conducted.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance evaluation is established through:

    • Predefined engineering specifications and standards: For bench tests like dimensions, tensile strength, burst pressure, kink resistance, etc.
    • Industry consensus standards (e.g., ISO): For tests like hub air leakage (ISO 10555-1), hub compatibility (ISO 80369-7), sterilization (ISO 11135), and biocompatibility (ISO 10993-1).
    • Comparative data to a predicate device: For certain performance aspects like particulate levels, tip buckling force, and delivery/retrieval force, where the device must be "comparable."

    There is no "expert consensus," "pathology," or "outcomes data" as ground truth because this is a physical device submission without clinical data.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K234115

    Validate with FDA (Live)

    Device Name
    APRO 55 Catheter
    Manufacturer
    Date Cleared
    2024-03-15

    (79 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K223545 APRO 70 Catheter and Alembic Aspiration Tubing

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 55 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 55 Catheter is provided with an Introducer Sheath.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (APRO 55 Catheter) seeking FDA clearance, not a study report detailing acceptance criteria for an AI/ML device. Therefore, it does not contain information typically found in such a study, such as acceptance criteria for machine learning performance metrics (e.g., sensitivity, specificity, AUC), sample sizes for test sets in an AI context, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, or training set details.

    The document describes non-clinical bench testing for a physical medical catheter, including mechanical, material, and biocompatibility tests. The "acceptance criteria" mentioned are for these engineering and biological tests, not for an AI/ML diagnostic or predictive model.

    Therefore, I cannot fulfill your request for information related to AI/ML acceptance criteria and study details based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232971

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-29

    (161 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K223545 APRO® 70 Catheter and Alembic Aspiration Tubing

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 55 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    The Alembic Aspiration Tubing is intended to connect the APRO 55 Catheter to the aspiration pump.

    Device Description

    The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 55 Catheter targets aspiration from a suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 55 Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 55 Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 55 Catheter is available in lengths of 125 cm and 137 cm and is provided with an introducer sheath.

    The Alembic Aspiration Tubing connects the APRO 55 Catheter to the aspiration pump. The flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

    AI/ML Overview

    Here's a summary based on the provided FDA 510(k) document, focusing on acceptance criteria and study details:

    Device: APRO 55 Catheter and Alembic Aspiration Tubing

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria listed are primarily for non-clinical bench tests. The reported device performance uniformly states that the APRO 55 Catheter (and Alembic Aspiration Tubing where applicable) met the acceptance criteria.

    TestAcceptance CriteriaReported Device Performance
    Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications. Introducer Sheath meets the visual and dimensional specifications.The APRO 55 Catheter met the acceptance criteria. The Introducer Sheath met the acceptance criteria.
    ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 55 Catheter met the acceptance criteria.
    Vacuum IntegrityCatheter is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 55 Catheter met the acceptance criteria.
    Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 55 Catheter met the acceptance criteria.
    Catheter Hub LeakageCatheter does not leak into hub assembly during aspiration, with methods specified in ISO 10555-1, Annex D.The APRO 55 Catheter met the acceptance criteria.
    Hub CompatibilityHub meets the requirements per EN ISO 80369-7.The APRO 55 Catheter met the acceptance criteria.
    Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 55 Catheter met the acceptance criteria.
    Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 55 Catheter met the acceptance criteria.
    Fluid LeakageCatheter must withstand pressure with methods per ISO 10555-1 Annex C.The APRO 55 Catheter met the acceptance criteria.
    Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 55 Catheter met the acceptance criteria.
    Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 55 Catheter met the acceptance criteria.
    Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 55 Catheter met the acceptance criteria.
    Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 55 Catheter met the acceptance criteria.
    Tip Buckling ForceCatheter tip buckling force must be acceptable. Forces were compared to a predicate.The APRO 55 Catheter met the acceptance criteria.
    Simulated Use including Clot RetrievalWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications. Clot retrieval from various locations of the anatomical model was performed using the subject device system.The APRO 55 Catheter and Alembic Aspiration Tubing met the acceptance criteria.

    Biocompatibility Tests:

    TestResultsConclusions
    Sensitization (Guinea Pig Maximization)The APRO 55 Catheter did not elicit a sensitization response.Non-sensitizing
    Irritation/Intracutaneous ReactivityThe APRO 55 Catheter demonstrated no evidence of irritation.Non-Irritant
    Cytotoxicity (MEM Elution, L929 cells)The APRO 55 Catheter did not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
    Hemolysis – IndirectThere were no significant differences between the test article extract and negative control article results.Non-hemolytic
    Hemolysis – DirectThere were no differences between the hemolytic index of the test article and the negative control.Non-hemolytic
    Partial Thromboplastin Time (PTT)The average clotting time of the test article was greater than vehicle control and negative control.Acceptable clotting times
    SC5b9 Complement ActivationThe Sc5b9 concentration of the test article was statistically less than the positive control and was not statistically higher than the negative control.Acceptable
    Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice.Non-toxic
    Material-Mediated PyrogenicityAll individual rabbits for both the test article and negative control showed a total rise in temperature of < 0.5 °C and were determined to be nonpyrogenic.Non-pyrogenic

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size for the "test set" in terms of patient data. All testing described is non-clinical bench testing and biocompatibility testing.

    • Data Provenance: Not applicable as no human patient data was used for effectiveness evaluation. The studies were laboratory-based bench tests and animal biocompatibility studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not provided as the studies were non-clinical bench tests and biocompatibility testing, which do not typically involve human expert ground truth for performance evaluation in the way imaging AI or diagnostic devices do. Standardized test methods and acceptance criteria are used.

    4. Adjudication Method for the Test Set

    • This information is not provided and is not applicable, as there was no test set requiring human adjudication for performance evaluation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary."

    6. Standalone (Algorithm only without human-in-the-loop) Performance

    • This is not applicable. The APRO 55 Catheter and Alembic Aspiration Tubing is a physical medical device (catheter for thrombus retrieval) and not an AI algorithm or software-only device. Therefore, standalone algorithm performance is not a relevant concept for this submission.

    7. Type of Ground Truth Used

    • For non-clinical bench testing, the "ground truth" is defined by the specifications, industry standards (e.g., ISO 10555-1, EN ISO 80369-7), and acceptable ranges or comparisons to predicate devices outlined in the acceptance criteria. For example, for "Kink Resistance," the ground truth is "Catheter shaft shall not kink at clinically relevant radii."
    • For biocompatibility testing, the ground truth is established by standardized biological response criteria (e.g., non-sensitizing, non-irritant, non-cytotoxic) defined by ISO 10993-1.

    8. Sample Size for the Training Set

    • This information is not applicable as this is a physical medical device and not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable as this is a physical medical device and not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230695

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-10

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K223545 APRO 70 Catheter and Alembic Aspiration

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APRO 70 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    The Alembic Aspiration Tubing is intended to connect the APRO 70 Catheter to the aspiration pump.

    Device Description

    The APRO 70 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 70 Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 70 Catheter targets aspiration from the suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 70 Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 70 Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 70 Catheter is available in lengths of 125 cm, and 135 cm and is provided with an introducer sheath.

    The Alembic Aspiration Tubing connects the APRO 70 Catheter to the aspiration pump. The flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

    AI/ML Overview

    The provided text is a 510(k) summary for the APRO 70 Catheter and Alembic Aspiration Tubing. This type of submission is for medical devices, not AI/ML-based software as a medical device (SaMD). Therefore, the questions regarding acceptance criteria and study data typical for AI/ML performance (e.g., ground truth establishment, expert adjudication, MRMC studies, training set details) are not applicable to this document.

    The document describes the device, its intended use, and demonstrates substantial equivalence to a predicate device through non-clinical bench testing. The performance data is focused on the physical characteristics and functional integrity of the catheter and tubing.

    Here's an analysis based on the information provided for this medical device:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance (Conclusion)
    Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications. Introducer Sheath meets the visual and dimensional specifications. Aspiration Tubing meets the visual and dimensional specifications.The APRO 70 Catheter met the acceptance criteria. The Introducer Sheath met the acceptance criteria. The Alembic Aspiration Tubing met the acceptance criteria.
    ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 70 Catheter met the acceptance criteria.
    Vacuum IntegrityCatheter with Aspiration Tubing is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 70 Catheter and Alembic Aspiration Tubing met the acceptance criteria.
    Kink ResistanceCatheter distal shaft shall not kink.The APRO 70 Catheter met the acceptance criteria.
    Catheter Hub LeakageCatheter does not leak into hub assembly during aspiration, with methods specified in ISO 10555-1, Annex D.The APRO 70 Catheter met the acceptance criteria.
    Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage.The APRO 70 Catheter met the acceptance criteria.
    Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 70 Catheter met the acceptance criteria.
    Fluid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 70 Catheter met the acceptance criteria.
    Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 70 Catheter met the acceptance criteria.
    Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 70 Catheter met the acceptance criteria.
    Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 70 Catheter met the acceptance criteria.
    Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 70 Catheter met the acceptance criteria.
    Tip Buckling ForceCatheter tip buckling force must be acceptable. Forces were compared to a predicate.The APRO 70 Catheter met the acceptance criteria.
    Simulated UseWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications.The APRO 70 Catheter and Alembic Aspiration Tubing met the acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of devices tested) for each non-clinical bench test. It states "Alembic performed non-clinical bench testing in accordance with design controls, protocol, and test methods". The data provenance is non-clinical bench testing; it's not patient-derived data, so country of origin or retrospective/prospective distinctions are not applicable in the typical sense for clinical studies.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Ground truth as understood in AI/ML is not relevant for this type of medical device bench testing. The "ground truth" for these tests is defined by established engineering and material science specifications, and ISO standards (e.g., ISO 10555-1), not expert consensus on medical images or patient outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication process described for these bench tests. The results are based on quantitative measurements and observations against pre-defined engineering acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This is a medical device, not an AI/ML-based software. Therefore, an MRMC study related to human readers improving with AI assistance is not applicable. The study presented here is focused on demonstrating the physical and functional performance of the catheter and tubing, and its substantial equivalence to a predicate device through non-clinical bench testing.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI/ML algorithm. The performance documented is that of the physical device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluation of this medical device is based on:

    • Pre-defined engineering and material specifications.
    • Compliance with recognized standards (e.g., ISO 10555-1).
    • Functional performance criteria observed during bench testing (e.g., freedom from kinking, maintaining vacuum).
    • Comparison to the performance of a legally marketed predicate device in some tests (e.g., Particulate, Delivery and Retrieval Force, Tip Buckling Force sections mention comparison to a predicate).

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.