510(k) Data Aggregation
(72 days)
The SIGNA™ Bolt system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA™ Bolt system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
SIGNA™ Bolt is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times, and is designed for improved patient comfort and workflow. The system features a 3.0T superconducting magnet with a 70 cm bore size and can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. SIGNA™ Bolt is designed to conform to NEMA DICOM standards.
The SIGNA™ Bolt system will be offered as two commercial configurations with the following features:
- Magnet: 3.0T superconducting magnet with a wide (70 cm) bore size and active shielding
- Maximum Gradient Strength: 80 mT/m (SuperXG Gradient), 65 mT/m (SuperXF Gradient)
- Maximum Slew Rate: 200 T/m/s (SuperXG Gradient and SuperXF Gradient)
- RF Transmit: A liquid cooled In-Scan-Room RF transmit architecture with a peak power capability of 36 kW and 3.0T Platform Body Coil
- RF Receive Chain: 162 Ch available (SuperXG Gradient), 130 Ch available (SuperXF Gradient)
- Patient Table: Detachable SIGNA One Patient Table with embedded 3.0T AIR PA XL coil and up to four 32-channel high density auto-coil sensing connection ports
- Power Rating: 113 kVA (SuperXG Gradient), 90 kVA (SuperXF Gradient)
- Software: Software platform featuring various productivity enhancement features, designed to improve workflow and reduce scan time
- AIRx (previously cleared in K183231) – AI-based automated slice prescription tool now extended with new deep learning models for spine and prostate imaging
- SIGNA One Camera – Real-time AI-enabled image guidance that assists with automated patient positioning
- Gating Options: Wired, wireless, and contactless physiological gating options
This document outlines the acceptance criteria and supporting studies for the SIGNA™ Bolt device, based on the provided FDA 510(k) clearance letter.
Key Features and AI/ML Components of SIGNA™ Bolt:
The SIGNA™ Bolt system includes several AI/Machine Learning components:
- AIRx: An AI-based automated slice prescription tool, previously cleared for brain and knee imaging (K183231), now extended with new deep learning models for spine and prostate imaging.
- SIGNA One Camera: Real-time AI-enabled image guidance that assists with automated patient positioning.
- Contactless Gating: This feature leverages underlying physiological signal detection that might involve advanced signal processing or AI techniques, though the document primarily describes its functional outcome.
Acceptance Criteria and Reported Device Performance
The following table summarizes the acceptance criteria and reported performance for the AI/ML components of the SIGNA™ Bolt device:
| Feature/Component | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| SIGNA One Camera | Landmark Inference Accuracy: 90% successful detection of camera-predicted anatomical landmarks compared to ground truth annotations. | Landmark Inference Accuracy: Achieved up to 99% successful detection across all evaluated anatomical regions. |
| Landmark Acceptance (with obstructions): 95% success rate. | Landmark Acceptance (with obstructions): Achieved 97% success rate. | |
| AIRx Spine | All deep learning models met their predefined acceptance criteria (specific criteria not detailed, but implied to be related to accuracy, variability reduction, and successful adaptation to spinal curvatures and complex scan setups). | Model Performance: All models met their predefined acceptance criteria. |
| Reduced scan prescription times and minimized inter-operator variability compared to manual workflows. | Demonstrated reduced scan prescription times and minimized inter-operator variability (confirmed by SSIM analysis and visual comparisons). Successfully adapted prescriptions to patient-specific spinal curvatures and automated Pars Interarticularis and Cervical Foramina scans. | |
| AIRx Prostate | All deep learning models met predefined acceptance criteria (specific criteria not detailed, but implied to be related to accuracy and robustness to variations in anatomy, pathology, and implants). | Model Performance: All models met predefined acceptance criteria, confirming robustness to variations in anatomy, pathology, and presence of implants. |
| Contactless Gating | Accurately detecting and displaying respiratory and peripheral cardiac waveforms without physical accessories. Supporting use of these waveforms for triggering MR acquisitions across multiple anatomical regions. | Verified and validated to accurately detect and display respiratory and peripheral cardiac waveforms without physical accessories. Supports use of these waveforms for triggering MR acquisitions across multiple anatomical regions (meeting performance specifications). |
Study Details for AI/ML Components:
1. SIGNA One Camera
- Sample Size for Test Set: Data collected from 80 volunteers.
- Data Provenance: US and China (to ensure diverse datasets).
- Number of Experts & Qualifications for Ground Truth: Not explicitly stated for this component. Ground truth is described as "MR system coordinates of the camera-predicted anatomical landmarks against ground truth annotations," suggesting a technical or measurement-based ground truth rather than expert reads.
- Adjudication Method: Not specified.
- MRMC Comparative Effectiveness Study: Yes, a "time on task study" was conducted with 11 MR Scan Operators comparing the AI-powered workflow to conventional laser landmarking.
- Effect Size: The camera workflow "consistently enabled faster setup times for landmarking." Specific quantitative improvement (e.g., % reduction in time) is not provided in text.
- Standalone Performance: Yes, "Accuracy was evaluated by comparing the MR system coordinates of the camera-predicted anatomical landmarks against ground truth annotations." This indicates an algorithm-only evaluation.
- Type of Ground Truth: MR system coordinates.
- Sample Size for Training Set: Not explicitly stated, but the test dataset was "entirely separate from the training and validation datasets."
- Ground Truth for Training Set: Not specified, but likely established in a similar manner to the test set (MR system coordinates or similar technical measurements).
2. AIRx Spine
- Sample Size for Test Set: 376 subjects.
- Data Provenance: Multiple clinical sites and internal GE HealthCare sites.
- Number of Experts & Qualifications for Ground Truth: Not explicitly stated. Ground truth is implied to be established for "accurate multi-slice, multi-angle prescriptions."
- Adjudication Method: Not specified.
- MRMC Comparative Effectiveness Study: Yes, "Comparative studies demonstrated that AIRx Spine reduced scan prescription times compared to manual workflows and minimized inter-operator variability."
- Effect Size: "Reduced scan prescription times" and "minimized inter-operator variability" (confirmed by Structural Similarity Index (SSIM) analysis and visual comparisons). Specific quantitative improvement is not provided.
- Standalone Performance: Yes, "Performance testing was conducted on the AIRx Spine deep learning models," indicating an algorithm-only evaluation.
- Type of Ground Truth: Not explicitly stated but implied to be based on accurate anatomical prescriptions suitable for diagnostic imaging. SSIM analysis and visual comparisons suggest a comparison against an ideal or expert-defined prescription.
- Sample Size for Training Set: Not explicitly stated, but the test dataset was "held separate from training and validation data."
- Ground Truth for Training Set: Not specified, but likely established to enable the model to learn "patient-specific spinal curvatures" and "accurate multi-slice, multi-angle prescriptions."
3. AIRx Prostate
- Sample Size for Test Set: 785 exams.
- Data Provenance: Clinical sites in the US and Europe.
- Number of Experts & Qualifications for Ground Truth: Not explicitly stated.
- Adjudication Method: Not specified.
- MRMC Comparative Effectiveness Study: Not explicitly mentioned for this specific feature in the provided text.
- Standalone Performance: Yes, "Performance testing was conducted on the six deep learning models that comprise the AIRx Prostate feature," evaluating automated prostate scan plane prescription, indicating an algorithm-only evaluation.
- Type of Ground Truth: Not explicitly stated but implied to be based on accurate anatomical prescriptions for the prostate, using SSFSE localizer images.
- Sample Size for Training Set: Not explicitly stated, but the test dataset was "kept separate from the training and validation data."
- Ground Truth for Training Set: Not specified, but likely established to enable the model to learn "automated prostate scan plane prescription."
4. Contactless Gating
- Sample Size for Test Set: Not explicitly stated for this particular feature's performance validation.
- Data Provenance: Not specified.
- Number of Experts & Qualifications for Ground Truth: Not specified.
- Adjudication Method: Not specified.
- MRMC Comparative Effectiveness Study: Not mentioned.
- Standalone Performance: Yes, "Verification and validation testing confirmed that the contactless gating feature meets its performance specifications by accurately detecting and displaying respiratory and peripheral cardiac waveforms," indicating a system performance evaluation.
- Type of Ground Truth: Underlying physiological waveforms (respiratory and cardiac).
- Sample Size for Training Set: Not specified.
- Ground Truth for Training Set: Not specified, but likely established from physiological signal data.
Overall Conclusion from Performance Testing:
GE HealthCare concludes that the SIGNA™ Bolt is as safe and effective, with performance substantially equivalent to the predicate device, based on the nonclinical testing, including extensive software verification and validation, as well as specific performance evaluations for its new AI-enabled features. No clinical studies were required to support substantial equivalence.
Ask a specific question about this device
(71 days)
The SIGNA™ Sprint Select is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by SIGNA™ Sprint Select reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
SIGNA™ Sprint Select is a whole-body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time. The system uses a combination of time-varying magnet fields (Gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system features a 1.5T superconducting magnet with 70cm bore size. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).
N/A
Ask a specific question about this device
(59 days)
The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.
Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time. The system uses a combination of time-varying magnet fields (Gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system features a 1.5T superconducting magnet with 70cm bore size. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).
The provided text does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/human reader performance study. The document is a 510(k) premarket notification for a Magnetic Resonance Diagnostic Device (SIGNA™ Champion).
The relevant sections state:
- "The subject of this premarket submission, the SIGNA™ Champion, did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission."
- "The sample clinical images demonstrate acceptable diagnostic image performance of the SIGNA™ Champion in accordance with the FDA Guidance 'Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices' issued on October 10, 2023. The image quality of the SIGNA™ Champion is substantially equivalent to that of the predicate device."
This indicates that the FDA clearance for the SIGNA™ Champion MR system was based on demonstrating substantial equivalence to a previous predicate device (SIGNA™ Voyager) through non-clinical testing and the review of sample clinical images, rather than a prospective clinical study involving human readers and AI assistance for diagnostic tasks.
Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI-assisted diagnostic device performance study because such a study was not conducted or reported in this 510(k) submission.