Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241202

    Validate with FDA (Live)

    Date Cleared
    2024-07-01

    (62 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K222547

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electronic Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.

    Device Description

    Electric nasal aspirator consists of main unit, and the suction portion working together as one unit. The Electronic Nasal Aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. The motor pump operates on a rechargeable battery. The rechargeable battery can be charged from the external power adapter (not included in this device) through the provided charging line. The user interface consists of buttons and LED display, and the user can control the vacuum pressure through the button.

    AI/ML Overview

    The provided text is a 510(k) summary for an Electric Nasal Aspirator. This document describes its substantial equivalence to a predicate device based on non-clinical testing. It explicitly states that "No clinical test data was used to support the decision of substantial equivalence."

    Therefore, the requested information about acceptance criteria, study details, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment for a clinical study to prove the device meets acceptance criteria cannot be extracted from this document, as no such clinical study was conducted or presented.

    The document focuses on demonstrating substantial equivalence through non-clinical testing, which includes various safety and performance benchmarks.

    However, based on the non-clinical tests mentioned, we can infer the acceptance criteria and reported performance for those specific tests:

    1. Table of Acceptance Criteria and Reported Device Performance (Based on Non-Clinical Tests)

    Acceptance Criteria (Implied)Reported Device Performance
    Product service life meets predefined requirement.Meets requirements; 5 years (longer than predicate)
    Software functions as validated.Validation performed (implies success)
    Electromagnetic compatibility (EMC) meets standards.Complies with ANSI AAMI IEC 60601-1-2:2014 & AMD 1:2021
    Electrical safety meets standards.Complies with ANSI AAMI ES60601-1:2005 & AMD2:2021
    Usability meets standards.Complies with IEC 60601-1-6 Ed 3.2 2020-07
    Function tests (Negative pressure, Max flow, Auto power-off, Noise) meet requirements.Negative pressure <68kPa; Max flow ≥1.6L/min; Noise <60dB (all tests successful)
    Biocompatibility meets standards (ISO 10993 series).Complies with ISO 10993-5, ISO 10993-10, ISO 10993-23
    Water resistance meets standard.IP22

    Points 2 through 9 are not applicable as no clinical study was performed or reported in this document.

    2. Sample sizes used for the test set and the data provenance: Not applicable. No clinical test set discussed. Non-clinical tests would use a sample of devices for testing, but no specific numbers are given, nor is "data provenance" in the sense of patient data relevant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth established.

    4. Adjudication method for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable for clinical ground truth. For non-clinical tests, the "ground truth" is defined by the technical specifications and international standards the device must meet (e.g., a certain decibel level for noise, a certain pressure for suction).

    8. The sample size for the training set: Not applicable. No AI model or training set discussed.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1