Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242049

    Validate with FDA (Live)

    Device Name
    SureCath Set
    Manufacturer
    Date Cleared
    2024-11-26

    (137 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K221401

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The product is intended for use in male, female, and pediatric patients requiring bladder drainage as determined by their physician. This device is indicated for those individuals unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode.

    Device Description

    SureCath Set is a sterile, single-use hydrophilic coated intermittent catheter with integrated urine bag. An ampoule of saline solution is provided inside the urine bag. This saline solution from the ampoule is poured over the catheter to activate the coating before use.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study:

    The document (K242049) is a Traditional 510(k) Summary for the Coloplast SureCath Set, a sterile, single-use hydrophilic coated intermittent catheter with an integrated urine bag. It demonstrates substantial equivalence to predicate devices. The information provided primarily focuses on non-clinical performance testing rather than human clinical studies involving AI assistance.

    Based on the provided text, the device SureCath Set is a medical device, not an AI/ML powered device. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study, effect size of human readers with/without AI, or standalone algorithm performance are not applicable.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document broadly states that "Bench performance testing and usability testing were conducted to verify the proposed subject devices met the pre-determined acceptance criteria per specified requirements." It also mentions various ISO and ASTM standards to which the device was tested. However, it does not explicitly list specific quantitative acceptance criteria or detailed reported performance values for each test. It only states that the device "met the pre-determined acceptance criteria."

    Implicit Acceptance Criteria (based on standards and test descriptions):

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityMet acceptance criteria per ISO 10993 series for cytotoxicity, irritation/intracutaneous reactivity, sensitization, material mediated pyrogenicity, acute systemic toxicity, and subacute toxicity.
    Catheter Performance (ISO 20696, ASTM F623-19, Coloplast Test Methods)Met acceptance criteria for: - Strength of connection between catheter and urine bag. - Aseptic insertion (non-touch approach) of catheter through urine bag. - Water amount of 20 ml in ampoule must activate coating on catheter. - Friction of the coated catheter after 30 seconds activation time. - Tearing and emptying of the urine bag. - Volume of the urine bag. - Strength of the urine bag.
    UsabilityMet acceptance criteria per EN/IEC 62366-1.
    Packaging IntegrityMet acceptance criteria per ISO 11607-1, DS EN ISO 20696, ASTM D4169-22 for maintenance of sterile barrier and impact of transportation.
    Shelf Life/Aging StabilityMet acceptance criteria after accelerated aging per ASTM F1980-21, demonstrating stability for the defined shelf life (24 months at time of submission, planned 36 months).

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each non-clinical test. It states "Testing was performed on final, finished, and sterilized devices." The provenance of the data is from Coloplast A/S, Denmark. The tests are non-clinical, bench performance tests, so the concepts of "retrospective or prospective" human data don't directly apply here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of non-clinical device testing. There are no "experts" establishing a "ground truth" for a test set in the way it would be for an AI/ML clinical study (e.g., radiologists interpreting images). The "ground truth" for non-clinical tests is based on objective measurements against established engineering and biological standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically used in clinical studies where expert consensus is needed to establish ground truth for ambiguous cases. Non-clinical bench tests rely on objective measurements against defined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not conducted. The device is a physical medical device (intermittent catheter), not an AI-powered diagnostic or assistive tool. Therefore, there is no "AI assistance" for human readers to improve upon.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm-only performance study was not done. The device is not an algorithm or software requiring human-in-the-loop performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests is based on:

    • Compliance with recognized international standards (ISO, ASTM, EN/IEC).
    • Pre-determined acceptance criteria defined by Coloplast for specific performance aspects (e.g., strength, friction, volume, activation).
    • Objective measurements of physical and chemical properties.

    8. The sample size for the training set

    This information is not applicable. "Training set" is a concept for AI/ML models. This document describes testing for a physical medical device.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1