Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221231

    Validate with FDA (Live)

    Device Name
    Drug Relief v1
    Manufacturer
    Date Cleared
    2022-06-06

    (38 days)

    Product Code
    Regulation Number
    882.5896
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211971

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Drug Relief v1 is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination.

    Device Description

    The Drug Relief v1 is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over five days / 120 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 120 hours. There are three stimulation electrode and one ground electrode which constitute of a titanium needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. The device is rectangular in shape and designed to fit comfortably on the neck. The biocompatible adhesive ensures that the device maintains contact with the skin during normal use. The adhesive fasteners ensure that the electrode needles and the entire device stay in place in a secure manner. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin. The stimulation pattern consists of rectangular pulses with differing interpulse intervals. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

    AI/ML Overview

    The provided text describes the regulatory clearance of the "Drug Relief v1" device (K221231) by DyAnsys, Inc. It explicitly states that the device is substantially equivalent to a previously cleared predicate device, also named "Drug Relief v1" (K211971). The document primarily focuses on demonstrating this substantial equivalence rather than presenting an independent clinical study to prove the device meets specific performance criteria against a predefined ground truth in a clinical setting.

    Therefore, many of the requested items related to clinical study design, expert involvement, and statistical measures of performance (like effect size with AI assistance) are not directly applicable or available in this regulatory submission. The "acceptance criteria" discussed are largely in the context of demonstrating equivalence to the predicate and safety/performance through non-clinical testing.

    Here's an attempt to answer the questions based solely on the provided text, noting where information is not available:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria and corresponding reported device performance against a clinical ground truth for the subject device (K221231). Instead, it focuses on demonstrating that the subject device has "equivalent performance specifications" and "identical fundamental technological and operational characteristics" to the predicate device (K211971). The "Performance Data" section discusses non-clinical testing.

    Acceptance Criteria (Implicit - related to predicate equivalence and non-clinical performance):

    Acceptance Criteria (Implicit)Reported Device Performance (K221231)
    Intended Use / Indications for Use: Identical to predicateIdentical to predicate: "aid to reduce the symptoms of opioid withdrawal, through application to branches of cranial nerves V, VII, IX and X, and the occipital nerves identified by transillumination."
    Technological Characteristics: Equivalent to predicate"identical intended use, indications for use, population, application of the device, environment of use" to predicate
    Product Dimensions (mm x mm x mm): 38 * 21 * 1038 * 21 * 10
    Mass (g): 8g (including battery)8g (including battery)
    Frequency (Hz): 1 - 101 - 10
    Waveform: Biphasic with Rectangular PulseBiphasic with Rectangular Pulse
    Battery Type: P13 Zinc Air batteries, non-rechargeableP13 Zinc Air batteries, non-rechargeable
    Battery Capacity: 310 mAh310 mAh
    No. x Voltage (V): 3 x 1.4 V3 x 1.4 V
    Pulse Width (ms): 0.9800.980
    Duty Cycle: 2 hours ON / 1 min OFF2 hours ON / 1 min OFF
    Typical Battery Operating Time (hours): 120120
    Operating Temperature: 5 °C to 45 °C5 °C to 45 °C
    Operating Humidity: 40% to 80%40% to 80%
    Environmental Use: Clinics, Hospital, HomeClinics, Hospital, Home
    Sterilization of Electrodes: EtO SterilizationEtO Sterilization
    Re-use: Single use DeviceSingle use Device
    Max Charge Density (µC / cm²) per needle: 67.16 @ 1 kΩ, 8.01 @ 10 kΩ67.16 @ 1 kΩ, 8.01 @ 10 kΩ
    Max Average Power Density (W / cm²): 0.363 @ 1 kΩ, 0.0516 @ 10 kΩ0.363 @ 1 kΩ, 0.0516 @ 10 kΩ
    Net Charge (µC per pulse): 0 (biphasic)0 (Due to biphasic nature of the waveform)
    Software Controlled: YesYes
    Shelf Life: Increased from 6 months (predicate) to 12 months12 months (demonstrated through sterilization and non-clinical performance testing)
    Biocompatibility: PassSubjected to biocompatibility testing (implied pass for clearance)
    Electrical Safety (EMC and safety): PassSubjected to electrical safety (electromagnetic compatibility and safety) testing (implied pass for clearance)
    Performance Bench Testing (120 hours continuous performance): Pass for pulse width, pulse duration, amplitude, current valuesFunctional test performed for 120 hours to monitor continuous performance. Captured pulse width, pulse duration, amplitude, and current values. Performance equivalent to predicate.
    Software Validation: PassSubjected to software validation (implied pass for clearance)

    2. Sample size used for the test set and the data provenance
    The document does not describe a clinical test set with human subjects or patient data. The "test set" discussed refers to non-clinical bench testing.

    • Sample size: Not specified for individual bench tests, but implies a sufficient number of devices were tested for validation.
    • Data provenance: Non-clinical bench testing data, likely conducted by DyAnsys, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. There was no clinical test set requiring expert ground truth establishment for diagnostic performance.

    4. Adjudication method for the test set
    Not applicable. No clinical test set and no expert adjudication were described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is a nerve stimulator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This device is a nerve stimulator, not an algorithm. Its function is direct stimulation.

    7. The type of ground truth used
    For the non-clinical performance data, the "ground truth" was likely defined by engineering specifications and direct measurements against those specifications (e.g., measuring actual pulse width against the specified 0.980 ms). For establishing substantial equivalence, the "ground truth" was the characteristics, intended use, and indications for use of the predicate device (K211971).

    8. The sample size for the training set
    Not applicable. There is no mention of a training set for an algorithm in this submission.

    9. How the ground truth for the training set was established
    Not applicable. No training set was described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1