    K Number
    K253008

    Validate with FDA (Live)

    Device Name
    maiLink
    Manufacturer
    Date Cleared
    2026-01-05

    (108 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211733, K211678, K241439, K243696, K201710, K212365

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    maiLink, a PACS and radiology workflow management software available as both a desktop/laptop application and a web viewer, used for viewing and assessing DICOM images Ex: DX, DR, CR, CT, MR, US, RF and 2D mammography. It gathers digital images and information from a variety of sources that adhere to the DICOM standard. These sources encompass a range of devices, including digital and computed radiographic equipment, CT and MR scanners, ultrasound and RF machines, secondary capture tools, imaging gateways. maiLink enables the storage, transmission, processing, and visualization of images and data within the system itself or over computer networks spanning different locations. Only pre-processed DICOM images specifically intended for presentation are suitable for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretation on the maiLink. To ensure accurate interpretation, mammographic images should only be evaluated on a monitor that adheres to the technical specifications outlined by the FDA. This system is designed exclusively for use by proficient and certified medical experts, including physicians, radiologists, and medical technicians.

    Device Description

    maiLink is software designed to receive and transmit medical images from various medical imaging devices through SDK, DICOM communication, or user image uploads. It sends these images to an AI medical imaging reading solution software, receives the interpretation results, and allows for storage or transmission to the hospital's PACS system. maiLink acts as an intermediary service between medical imaging modalities and AI-assisted reading solutions.

    The medical imaging devices and AI medical imaging reading solution software are not included in this product and require separate authorization. This product complies with the DICOM standard protocol, a standard for medical imaging, to facilitate integration with PACS for the storage and communication of medical images.

    maiLink's integrated FDA-cleared algorithm list is exclusively managed by maiLink. AI Customers do not have the ability technically or administratively to add or modify which FDA-cleared AI algorithms are integrated with maiLink. Only FDA-cleared algorithms which have passed rigorous regulatory and quality standards review by maihub Corp. to guarantee functionality, safety and security are made available in maiLink.

    maiLink simply presents the simple AI response output, and the initial anonymized image remains consistently available. The duty of evaluating the AI output, validating the results, and conducting the diagnosis lies with competent medical professionals.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K231470

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-11-06

    (168 days)

    Product Code
    Regulation Number
    892.2090
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K211678

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis. the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population.

    Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician's review or their clinical judgment that takes into account other relevant information from the image or patient history.

    Device Description

    Lunit INSIGHT DBT is a computer-assisted detection/diagnosis (CADe/x) software as a medical device that provides information about the presence, location and characteristics of lesions suspicious for breast cancer to assist interpreting physicians in making diagnostic decisions when reading digital breast tomosynthesis (DBT) images. The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.

    For each DBT case, Lunit INSIGHT DBT generates an artificial intelligence analysis results that include the lesion type, location, lesion-level case-level score, and outline of the regions suspected of breast cancer. This peripheral information intends to augment the physician's workflow to better aid in detection and diagnosis of breast cancer.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device, Lunit INSIGHT DBT, meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device PerformanceStatistical Significance / Comment
    Standalone Performance
    AUROC> 0.903 (mean AUROC of predicate device K211678)0.928 (95% Cl: 0.917 - 0.939)p < 0.0001 (exceeded criteria)
    Clinical Assessment (MRMC Study with CAD Assistance)
    Patient-level LOS AUROCCAD-assisted performance superior to CAD-unassisted performance with statistical significanceCAD-unassisted AUROC: 0.897 (95% Cl 0.858 - 0.936)CAD-assisted AUROC: 0.915 (95% Cl: 0.874 - 0.955)Inter-test difference: 0.017 (95% Cl: 0.000 - 0.034, P = 0.0498) - met criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Standalone Performance Test Set: 2,202 DBT exams (1,100 negative/benign, 1,102 cancer cases).
      • Data Provenance: Collected at multiple imaging facilities in the US. The data was collected consecutively.
    • Clinical Assessment (MRMC) Test Set: 258 DBT exams (65 cancer cases, 193 non-cancer cases, comprising 128 normal and 65 benign cases).
      • Data Provenance: Acquired from US clinical centers.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Standalone Performance Test Set: The text states ground truth localization was "derived based on the radiologic review and annotation by multiple MQSA qualified ground truthers." The exact number of experts is not specified.
      • Qualifications: "MQSA qualified ground truthers."
    • Clinical Assessment (MRMC) Test Set: The ground truth for the cases used in the MRMC study is implicitly established by the case classification (cancer vs. non-cancer). It's not explicitly stated how many experts established the underlying ground truth for these 258 cases, beyond the radiologists participating in the MRMC study itself. The readers for the MRMC study were "a total of 15 MQSA qualified and US board-certified radiologists."

    4. Adjudication Method for the Test Set

    • Standalone Performance Test Set: Ground truth was established through "binary classification of each case based on clinical supporting data, particularly pathology reports for cancer and biopsy-proven benign cases, followed by localization which was derived based on the radiologic review and annotation by multiple MQSA qualified ground truthers." This suggests an expert consensus/review process for localization, likely involving reconciliation or multiple reads. The specific type of adjudication (e.g., 2+1, 3+1) is not explicitly detailed.
    • Clinical Assessment (MRMC) Test Set: The ground truth for the MRMC study seems to rely on the pre-established classification of cases (cancer/non-cancer) based on clinical and pathological data. The radiologists in the MRMC study were evaluating cases against this existing ground truth, not establishing it in an adjudicated reading session.

    5. Was a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Done? If so, what was the effect size of how much human readers improve with AI vs. without AI assistance?

    • Yes, an MRMC comparative effectiveness study was done.
    • Effect Size of Improvement:
      • CAD-unassisted AUROC: 0.897
      • CAD-assisted AUROC: 0.915
      • Inter-test difference (effect size): 0.017 (Absolute difference in AUROC). This indicates an improvement in AUROC of 0.017 when radiologists were assisted by Lunit INSIGHT DBT. The p-value of 0.0498 indicates this improvement was statistically significant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance study was done.
      • Performance: AUROC of 0.928 (95% Cl: 0.917 - 0.939).

    7. The Type of Ground Truth Used

    • Standalone Performance and Clinical Assessment Test Sets: The ground truth was established primarily through clinical supporting data, specifically pathology reports for cancer and biopsy-proven benign cases, further supplemented by radiologic review and annotation by MQSA qualified ground truthers for localization. This can be categorized as a combination of pathology and expert consensus/review.

    8. The Sample Size for the Training Set

    • The document states that the "dataset used in the standalone performance test was independent from the dataset used for development of the artificial intelligence algorithm." However, it does not specify the sample size of the training set used for the development of Lunit INSIGHT DBT.

    9. How the Ground Truth for the Training Set Was Established

    • The document mentions that the training dataset was separate from the test dataset. While it details how the ground truth was established for the test set (pathology, biopsy, radiologic review/annotation), it does not explicitly describe how the ground truth was established for the training set. Being a deep learning model, it's highly probable the training data also relied on verified diagnoses, likely similarly derived from pathology/biopsy given the nature of CADe/x devices for breast cancer.