Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K242975

    Validate with FDA (Live)

    Date Cleared
    2025-05-07

    (223 days)

    Product Code
    Regulation Number
    890.3800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201196

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended / indications for use for the FreeStyle Libre 3 system are as follows:

    • The FreeStyle Libre 3 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real-time alarms capability indicated for the management of diabetes in persons aged 4 years and older.
    • It is intended to replace blood glucose testing for diabetes treatment decisions unless otherwise indicated.
    • The System performs a scan every minute to measure interstitial fluid glucose concentrations.
    Device Description

    Google Wifi point and AC2200 router with Ethernet port and USB-C port

    AI/ML Overview

    This document is an FDA 510(k) Clearance Letter for a Mobility Scooter, not an AI/Software as a Medical Device (SaMD). Therefore, the acceptance criteria and study proving device performance are based on physical and mechanical performance standards, safety, and substantial equivalence to a predicate device, not on artificial intelligence algorithms or human-in-the-loop performance.

    The request asks for information typically associated with AI/SaMD submissions, such as:

    • Table of acceptance criteria and reported device performance (in terms of AI metrics)
    • Sample size for test set and provenance
    • Number of experts for ground truth and their qualifications
    • Adjudication method
    • MRMC comparative effectiveness study results
    • Standalone (algorithm only) performance
    • Type of ground truth used
    • Sample size and ground truth for training set

    None of this information related to AI/SaMD studies is applicable or present in the provided 510(k) clearance letter for a mobility scooter.

    Instead, the document details non-clinical testing against established international standards (ISO, IEC, EN) for wheelchairs and mobility scooters, and biocompatibility testing.

    Here's a breakdown of the relevant acceptance criteria and performance proof from the provided document for the Mobility Scooter:


    Acceptance Criteria and Device Performance for Mobility Scooter (K242975)

    The acceptance criteria for this Class II Mobility Scooter are primarily based on demonstrating substantial equivalence to a legally marketed predicate device (K201196) and compliance with recognized international performance and safety standards for motorized wheelchairs and scooters. The study's proof of performance relies on non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly met by demonstrating compliance with the listed ISO, IEC, and EN standards, and by showing comparable performance characteristics to the predicate device. The performance data is presented as reported values from the tests.

    Acceptance Criterion (Standard or Characteristic)Reported Device Performance (Subject Device)Predicate Device Performance (K201196)Remark / Compliance
    Non-Clinical Standards Compliance:All listed tests demonstrated compliance.
    ISO 7176-1: Static Stability15.6° (forward), 9.5° (rearward), 14.1° (sideways)30° (forward), 20.1° (rearward), 15.3° (sideways)Subject device values are lower than predicate, but explicitly stated "Both of the devices are evaluated according to standard ISO 7176-1:2014, so the different static stability will not impact the safety and effectiveness." (Implies meeting the standard's minimums).
    ISO 7176-2: Dynamic StabilityNot explicitly stated numerical value but testedNot explicitly stated numerical valueTest results demonstrated compliance.
    ISO 7176-3: Brakes EffectivenessBraking time: 1.93s, Braking distance: 1.16m (at 7km/h)1.1m for 6km/h, 1s"Similar" performance; tested successfully to the standard.
    ISO 7176-4: Energy Consumption/Range15 km (theoretical distance range)Not Publicly Available (N/A)Compliance demonstrated by testing.
    ISO 7176-5: Dimensions, Mass, ManeuveringDimensions: 1080 X650 X940mm; Weight: 44kg; Min. Turning radius: 1400mmDimensions: 1050 X 550 X 870mm; Weight: 29kg; Min. Turning radius: 1200mm"Minor differences in the dimensions will not impact the safety and effectiveness." "The difference [in weight] will not raise any new safety and effectiveness concerns." "Little difference in turning radius will bring more convenience when it turns. The difference will not raise any new safety and effectiveness concerns." All imply meeting the standard's requirements for documented parameters.
    ISO 7176-6: Max. Speed, Acceleration, DecelerationMax Speed Forwards: 1.94m/s (7 km/h); Max. Speed Backward: 0.93 m/s (3.3 km/h)Max Speed Forwards: 1.67m/s (6 km/h); Max. Speed Backward: 0.83m/s (3 km/h)"The devices are evaluated according to standard ISO 7176-6:2018, so the different will not impact the safety and effectiveness." (Implies meeting regulatory limits).
    ISO 7176-7: Seating & Wheel DimensionsFront wheel: 8 inch; Rear tire: 8 inch; Anti-tip Wheels: 2.0 inchFront wheel: 7 inch; Rear tire: 8 inch; Anti-tip Wheels: 2.5 inch"Larger size of the front wheel. The difference will not raise any new safety and effectiveness concerns." and "Same" for other wheels. Implies meeting standard's measurement requirements.
    ISO 7176-8: Strengths (Static, Impact, Fatigue)Not explicitly stated numerical values but testedNot explicitly stated numerical valuesTest results demonstrated compliance.
    ISO 7176-9: Climatic testsNot explicitly stated numerical values but testedNot explicitly stated numerical valuesTest results demonstrated compliance.
    ISO 7176-10: Obstacle-climbing ability40mm60mm"The smaller height in the obstacle climbing will not impact the safety and effectiveness of the subject device." (Implies meeting the standard's performance as relevant for intended use).
    ISO 7176-14: Power and Control SystemsBritish PG Controller R-seriesBritish PG Controller PG45ADifferent controllers, but "both the control system... are similar. Both devices have the same user interface... Both of the control systems are evaluated according to standard ISO 7176-14:2008 and software validation requirement and there are no new safety and effectiveness concerns due to the difference." (Implies meeting functional and safety requirements of the standard).
    ISO 7176-21: EMC (Electromagnetic Comp.)Not explicitly stated numerical values but testedNot explicitly stated numerical valuesTest results demonstrated compliance.
    IEC 60601-1-2: EMC (Medical Electrical Equip.)Not explicitly stated numerical values but testedNot explicitly stated numerical valuesTest results demonstrated compliance.
    Biocompatibility (ISO 10993-1):All listed tests demonstrated compliance.
    Cytotoxicity (ISO 10993-5)ComplyComplyBiocompatibility evaluation carried out per standard.
    Skin Sensitization (ISO 10993-10)ComplyComplyBiocompatibility evaluation carried out per standard.
    Irritation (ISO 10993-23)ComplyComplyBiocompatibility evaluation carried out per standard.
    Maximum Loading Capacity265lbs (120kg)265lbs (120kg)Same.
    Ground Clearance40mm50mm"The device has been tested according to ISO7176 series standards and the test records support its safety and effectiveness." (Implies meeting requirements for safe operation given the specified clearance).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a numerical sample size (e.g., number of scooters) for the non-clinical tests. These types of tests are typically performed on a limited number of production units, as specified by the respective ISO standards.
    • Data Provenance: The device manufacturer, Wuyi Aichi Industry & Trade Co., Ltd., is located in Wuyi, Zhejiang Province, China. The non-clinical testing would have been conducted by the manufacturer or a contracted test lab(s), likely in China, to demonstrate compliance with international standards for regulatory submission. The data is retrospective in the context of the 510(k) submission, meaning the tests were completed prior to the submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable in the context of a mobility scooter's
      non-clinical performance testing. The "ground truth" is defined by the
      established parameters and limits within the specified international
      standards (e.g., ISO 7176 series). Performance is measured objectively
      against these defined criteria.

    4. Adjudication Method for the Test Set

    • Not applicable. Performance is measured objectively against
      international standards. There is no concept of "adjudication" as there
      would be for human-annotated data in an AI/SaMD context.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, this is not applicable. MRMC studies are used for evaluating the interaction between human readers (e.g., radiologists) and AI algorithms in diagnostic tasks. This submission is for a physical medical device (mobility scooter), not an AI/SaMD.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • No, this is not applicable. There is no "algorithm" in the sense of AI/SaMD for this device's regulatory pathway. The "standalone" performance is the scooter's physical and mechanical performance as measured by the non-clinical tests.

    7. The Type of Ground Truth Used

    • The "ground truth" is established by internationally recognized performance and safety standards (e.g., ISO 7176 series, IEC 60601-1-2, EN 12184) and biocompatibility standards (ISO 10993 series). These standards define the test methodologies, acceptance limits, and evaluation criteria for safe and effective operation of mobility devices.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical medical device. There is no "training set" in the context of machine learning. The design and manufacturing processes are validated against quality system regulations (21 CFR Part 820), and performance is validated through non-clinical testing on a sample of the manufactured device.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for a machine learning model, there is no ground truth established in that context. The "ground truth" for the device's design and manufacturing is adherence to engineering principles, regulatory requirements, and international consensus standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222711

    Validate with FDA (Live)

    Device Name
    HS186B Scooter
    Date Cleared
    2023-01-09

    (123 days)

    Product Code
    Regulation Number
    890.3800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201196

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

    Device Description

    The subject device, Model HS186B Mobility Scooter, is an indoor/outdoor electric scooter that is intended to be used by individuals that are able to walk but suffer from mobility limitations. It has a base with frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a tiller, a control panel, an electric motor, an electromagnetic brake, a rechargeable Lithium-lon Battery with a charger. The movement of the scooter is controlled by the rider who operates the throttle Control Lever, speed knob and hand grips. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter is suitable for use indoors and on flat roads near buildings.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (HS186B Scooter) and an FDA clearance letter. It details a comparison between the proposed device and a predicate device, focusing on non-clinical testing and safety features to demonstrate substantial equivalence. However, it does not describe acceptance criteria for an AI device or a study involving AI performance metrics.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance for an AI device.
    • Sample size used for the test set and data provenance for an AI device.
    • Number of experts and their qualifications for establishing ground truth for an AI test set.
    • Adjudication method for an AI test set.
    • MRMC comparative effectiveness study results (AI improvement vs. without AI assistance).
    • Standalone AI performance.
    • Type of ground truth used for AI.
    • Sample size for AI training set.
    • How ground truth for AI training set was established.

    This document pertains to a physical medical device (a motorized scooter) and its performance is evaluated against ISO standards and through comparison with a predicate device, not through AI-specific metrics. The "Non-Clinical Testing" section describes adherence to various ISO standards relevant to wheelchairs and scooters, and a "Biocompatibility" assessment. The document explicitly states "No clinical study implemented for the scooter." It does not mention any AI components or related studies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1