510(k) Data Aggregation
(25 days)
STAR Apollo™ Mapping System assists users in the interpretation and manual annotation of 3D anatomical and electrical maps of human atria using data from multipolar, intracardiac, atrial, electrograms during atrial fibrillation. The clinical significance of utilizing the STAR Apollo Mapping System, to help identify areas with intracardiac atrial electrograms, of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
The STAR Apollo Mapping System (v1.8) is a software driven system designed to assist operators in identifying Early Sites of Activation (ESA) and Repetitive Patterns of Activation (RPA) in patients undergoing a cardiac mapping procedure for Atrial Fibrillation (AF). The software is designed for use with FDA cleared electroanatomic mapping systems specifically:
- CARTO™ 3 EP Navigation System (V8.1) (K252302) (Biosense Webster) and
- OPTRELL™ Mapping Catheter with TRUEref™ Technology (K230253) (Biosense Webster)
for exporting geometry data, electrograms and electrode locations over ethernet connection during the electrophysiology procedure with CARTO 3 API (K231207) to provide input data for the STAR Apollo Mapping System.
- Ensite Precision Model EE 3000 Cardiac Mapping System (V2.6) (K201148) and
- Advisor™ HD Grid Mapping Catheter, Sensor Enabled™ (K172393) (Abbott Medical) or
- EnSite X EP System (V 1.1.1, V 2.0, V 3.0) (K213364) (K221213) (K231415) (Abbott Medical) and
- Advisor™ HD Grid Mapping Catheter, Sensor Enabled™ (K172393) (Abbott Medical) or
- Ensite X EP System (V 3.1) (K242016) (Abbott Medical) and either:
for exporting geometry data, electrograms and electrode locations via a portable external storage device or over ethernet data connection (Ensite X EP System (V 3.0, V 3.1) with LiveSync module) (K231415) (K242016) during the electrophysiology procedure to provide input data for the STAR Apollo Mapping System.
The principle of STAR Apollo Mapping System analysis is to use data on multiple individual wavefront trajectories to identify Repetitive Patterns of Activation (RPA) or regions of the atrium that represent Early Sites of Activation (ESA) which most often precede activation of neighboring areas, with the aim of helping clinicians to identify regions of the atria that may be the origins for AF activation. The system consists of proprietary STAR Apollo Mapping System software and a hardware component. STAR Apollo Mapping System software consists of 3 main components: Electroanatomic data import, the STAR Apollo Mapping System engine (C++ code) and Graphics User Interface (GUI). The STAR Apollo Mapping System is designed to run on a laptop computer running Windows 11 Operating System. STAR Apollo Mapping System software is pre-installed onto the laptop.
The STAR Apollo Mapping System uses export data from the compatible Mapping System that has been collected with the compatible Mapping Catheter during the electrophysiology procedure. The Mapping Catheter is used to collect anatomy and electrogram data in the atria. Recordings are made for at least 30 seconds with the Mapping Catheter in a stable position and in contact with the atrial wall. These ≥30 second acquisitions are made in multiple, non-overlapping locations, to generate recordings over the entire atrial chamber. The data is exported via an external portable storage device or by streaming via an ethernet data cable connected to the data ethernet port of the EnSite X or CARTO 3 workstation. It is transferred to the laptop computer running the STAR Apollo Mapping System. The export data accepted from the Mapping Systems consists of electrograms, electrode coordinates, ECG recordings and the geometry model. The data is imported utilizing the portable external data storage device or via ethernet into the STAR Apollo Mapping System and then processed by the STAR Apollo Engine to generate a STAR Apollo Map visualized by the GUI. The STAR Apollo Map will highlight sites deemed to be Early Sites of Activation (ESA), as a red sphere at the endocardial locations corresponding to the recording electrode position. These sites are areas where the myocardium has initiated activation earlier than its neighboring sites on multiple occasions and therefore may be a potential site of AF initiation or maintenance. The more repetitive these sites are, the larger the red sphere appears on the STAR Apollo Map. The system will rank the ESA according to their repetition frequency and cycle length and identify the most relevant 3 sites. The system is designed to show the physician Repetitive Patterns of Activation (RPA). These are shown as colored arrows, which start from the leading electrode position, following the summarized activation sequence. The more repetitive or consistent that activation pattern is, the wider the white arrow. Based on this information the physicians may then use this as an additional guide for further mapping of the AF, using FDA cleared mapping system catheters.
The STAR Apollo Mapping System operates outside the sterile field and is only connected to the EnSite Precision, EnSite X EP or CARTO 3 workstation and not to the amplifier, patient, or any other devices used in the procedure. No data is transferred from the STAR Apollo Mapping System back to the EnSite Precision, EnSite X EP mapping system, or CARTO 3 i.e., data transfer is only in one direction. No modifications to the EnSite Precision, EnSite X EP mapping systems or CARTO 3 are made to accommodate the STAR Apollo Mapping System. The STAR Apollo Maps may be used to give physicians additional information about the AF activations. The physician may use them as an additional aid to identify areas within the atria that may warrant further and close examination using the mapping system, and the compatible Mapping Catheter. The STAR Apollo System is never directly connected to a patient, nor does it deliver therapy. It is used as a software tool that provides supplementary information to the physician in an electrophysiology procedure.
N/A
Ask a specific question about this device
(259 days)
STAR Apollo™ Mapping System assists users in manual annotation of 3D anatomical and electrical maps of human atria using data from multipolar, intracardiac, atrial, electrograms during atrial fibrillation. The clinical significance of utilizing the STAR Apollo Mapping System, to help identify areas with intracardiac atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
The STAR Apollo 100 Mapping System is a software driven system designed to assist operators in identifying Early Site of Activations (ESA) and Repetitive Patterns of Activation (RPA) in patients undergoing a cardiac mapping procedure for Atrial Fibrillation (AF). The software is designed for use in association with a commercially approved electroanatomic mapping system, specifically, Ensite Precision Cardiac Mapping System Model EE 3000 (V2.6) or Ensite X EP System (V 1.1.1) (Abbott Medical), and the commercially approved catheter, Advisor™ HD Grid Mapping Catheter, Sensor Enabled™ (Abbott Medical). The system consists of proprietary Star Apollo Mapping System software and a hardware component. Star Apollo Mapping System software consists of 3 main components: Electroanatomic data import, the STAR Apollo Mapping System engine (C++ code) and Graphics User Interface (GUI). The STAR Apollo Mapping System is designed to run on a laptop computer running Windows 10 or later operating system. Star Apollo Mapping System software is pre-installed onto the laptop. The STAR Apollo Mapping System uses export data from the Ensite Precision or Ensite X EP system that has been collected with the HD Grid catheter during the mapping procedure. The HD Grid is used to collect anatomy, localization and electrogram data in the atria. Recordings are made for 30 seconds with the HD Grid in a stable position and in contact with the atrial wall. These 30 seconds acquisitions are made in multiple, non-overlapping locations, so as to generate recordings over the entire atrial chamber. Once the data has been acquired, the data is anonymized and exported, via a portable external data storage device, to the laptop computer running the STAR Apollo Mapping System. The export data from the Ensite Precision or Ensite X EP systems consists of electrograms, electrode coordinates, ECG recordings and the geometry model. The data are imported, utilizing the portable external data storage device, into the STAR Apollo Mapping System and then processed by the STAR Apollo engine to generate a STAR Apollo map visualized by the GUI.
The provided FDA 510(k) summary for the STAR Apollo™ Mapping System outlines the device's acceptance criteria and the studies conducted to demonstrate its performance.
Here's an analysis of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily derived from "Software design verification testing" and "Verification of algorithm calculations and graphic output." The performance is reported by stating that the software met these criteria.
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Software design verification testing | (Implied: Passed) No specific metrics are detailed, but the document states testing was conducted to demonstrate substantial equivalence. |
| Identification of each algorithm and testing to demonstrate mathematical calculations match predetermined output. (Algorithm description and testing) | "testing to demonstrate that the mathematical calculations performed by the software match that of the correct, predetermined output." (Implied: Passed, with output matching predetermined results) |
| Software requirement specifications met for test cases, numerical and graphical output correct. (Verification of algorithm calculations and graphic output) | "demonstrated that the software performance met the acceptance criteria for each of the test cases and both numerical and graphical output was correct." (Implied: Passed, with correct outputs) |
| Manual annotation yielded similar results to STAR Apollo Mapping System output. (Blinded Physician Validation) | "Testing demonstrated that manual annotation yielded similar results to the STAR Apollo Mapping System output." (Implied: Passed, with similarity to human experts) |
| Overall use of STAR Apollo Mapping System software rated as appropriate for use. (STAR Apollo Physician Use) | "Testing demonstrated that the STAR Apollo System Mapping System software was appropriate for use." (Implied: Passed, with positive user feedback) |
2. Sample Sizes and Data Provenance
- Test Set Sample Size: Not explicitly stated in the provided document. The descriptions refer to "test cases" and "electrophysiologists performing manual annotation of ECGs" and "electrophysiologists and rated the overall use." No specific number of cases or patient data points is given.
- Data Provenance: Not explicitly stated. The studies are described as "simulated testing." It is not clear if these simulations used real patient data (retrospective/prospective) or synthetically generated data. There is no mention of country of origin for the data.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Not explicitly stated. The document mentions "electrophysiologists" (plural), but not a specific count.
- Qualifications of Experts: They are identified as "electrophysiologists," which implies medical doctors specializing in cardiac electrophysiology. No further details regarding their experience level (e.g., years of experience, board certification) are provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. For the "Blinded Physician Validation," it says "electrophysiologists performing manual annotation of ECGs and comparing the results with the STAR Apollo Mapping System." This suggests a comparison to individual expert annotation rather than a formal consensus or adjudication process among multiple experts to establish a single ground truth.
5. Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, a formal MRMC comparative effectiveness study demonstrating how human readers improve with AI vs. without AI assistance was not reported. The studies described are:
- Blinded Physician Validation: This compares the device's output to manual human annotation, suggesting a standalone performance aspect relative to human interpretation, but not the human-in-the-loop improvement.
- STAR Apollo Physician Use: This assessed the usability and appropriateness of the device, which is a qualitative assessment of user experience, not an effectiveness study.
- Effect Size: Not applicable, as no MRMC study was performed or reported.
6. Standalone (Algorithm Only) Performance
- Standalone Performance: Yes, aspects of standalone performance were evaluated.
- "Algorithm description and testing" and "Verification of algorithm calculations and graphic output" describe testing the core algorithms' accuracy against predetermined outputs, which is a standalone assessment.
- The "Blinded Physician Validation (Manual Annotation versus STAR Apollo Mapping System)" also assesses the device's output independently, comparing it to human annotation, rather than measuring human improvement when using the device.
7. Type of Ground Truth Used
- Ground Truth Type:
- For algorithm and software verification, the ground truth was "correct, predetermined output" and "acceptance criteria for each of the test cases" set during development. This implies a ground truth defined by the developers based on internal specifications.
- For the "Blinded Physician Validation," the ground truth was "manual annotation" performed by electrophysiologists, essentially using expert interpretation as a reference. There is no mention of pathology or outcomes data as ground truth.
8. Sample Size for the Training Set
- Training Set Sample Size: Not explicitly stated. The document describes the device as using "fixed algorithms and equations" rather than deep learning or machine learning that would typically require a training set. The comparison to the predicate (VX1) explicitly states: "VX1 is a software that utilizes deep learning based algorithms to analyze data sets, whereas the STAR Apollo Mapping System software uses fixed algorithms and equations to analyze the data sets." Therefore, a traditional "training set" as understood in machine learning may not apply in the same way, or its size is not relevant for a fixed-algorithm system.
9. How Ground Truth for Training Set was Established
- Training Set Ground Truth Establishment: Not applicable given that the device uses "fixed algorithms and equations" and not deep learning, negating the need for a labeled training dataset in the conventional sense. The "ground truth" for its development would be the fundamental principles of electrophysiology and the expected mapping outcomes that the algorithms are designed to mimic.