510(k) Data Aggregation

    K Number
    K223422

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-30

    (136 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201034

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emory Cardiac Toolbox™ 4.0 – 4.3 software program should be used for the quantification of myocardial perfusion for the display of wall motion and quantification of left-ventricular function parameters from SPECT & PET myocardial perfusion studies (EGS™) including to quantitatively evaluate the wall motion and wall thickening of the heart using longitudinal, radial, and circumferential strain measurements, for the 3D alignment of coronary artery models from CT coronary angiography onto the left ventricular 3D epicardial surface and analysis of the CT data to evaluate calcified plaques in the coronary arteries, for the assessment of cardiac mechanical dyssynchrony using phase analysis, for generation of the short axis, vertical, and horizontal long axis tomograms from the SPECT raw data using either filtered back projection (FBP) or iterative reconstruction (MLEM/OSEM), for the quantification of myocardial blood flow and coronary flow reserve, and for the decision support in interpretation (LVX) and automatic structured reporting of the study.

    Device Description

    The Emory Cardiac Toolbox™ 4.3 is used to display gated wall motion and for quantifying parameters of left-ventricular perfusion and function from gated SPECT & PET myocardial perfusion studies and for the evaluation of dynamic PET studies.

    AI/ML Overview

    The Emory Cardiac Toolbox™ 4.3 device's performance validation is described in section 8 of the provided document. The device includes two main new features: Emory-CACS for calcium plaque evaluation and ECTb™ Strain for wall motion and thickening analysis using strain measurements.

    Here's a breakdown of the acceptance criteria and study details for each feature:


    1. Table of acceptance criteria and reported device performance:

    FeatureAcceptance Criteria (Proxy)Reported Device Performance
    Emory-CACSHigh correlation with the predicate device (Siemens SyngoVia™ toolbox) for both plaque volume and Agatston score. (Implied: r² > 0.99 for both based on predicate's performance)Plaque Volume: Linear regression (m = 1.10; b = 0.46) produced an r² = 0.99. Agatston Score: Linear regression (m = 1.01; b = 1.06) produced an r² > 0.99. "The values obtained by the Emory-CACS module demonstrated excellent correlations with the Syngo.CT CaScoring developed by Siemens Medical Solutions."
    ECTb™ Strain AnalysisEstablish normal limits for longitudinal, radial, and circumferential strains from a low-risk cohort. Demonstrate strong correlation between PET MPI and TTE for longitudinal, radial, and circumferential strains at both stress and rest in a CAD cohort. (Implied: r > 0.5 for all correlations based on predicate's performance or clinical significance)Normal Limits: Established using a low-risk cohort (n=22), defined as mean radial strain - 2SD, and mean circumferential/longitudinal strain + 2SD. Correlation (CAD Cohort, n=80): Longitudinal Strain: Stress (r=0.68, p<.001), Rest (r=0.58, p<.001) Radial Strain: Stress (r=-0.70, p<.001), Rest (r=-0.59, p<.001) Circumferential Strain: Stress (r=0.67, p<.001), Rest (r=0.69, p<.001) "The results demonstrate that all values between ECTb™ Strain and TTE correlated strongly and were found to be highly reproducible."

    2. Sample size used for the test set and the data provenance:

    • Emory-CACS:

      • Sample Size: 52 patients, 110 vessels with calcified plaques.
      • Data Provenance: Retrospective, as the lesions had been "previously scored by Emory radiologists." The country of origin is not explicitly stated but can be inferred as the U.S. (Emory is in Atlanta, GA).
    • ECTb™ Strain Analysis:

      • Sample Size:
        • Low-risk cohort (for establishing normal limits): n=22
        • High-risk / CAD cohort (for validation): n=80
      • Data Provenance: Not explicitly stated whether retrospective or prospective, but clinical comparison tests are often retrospective utilizing existing patient data. Country of origin is not explicitly stated but can be inferred as the U.S. (Emory is in Atlanta, GA).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Emory-CACS:

      • Number of experts: Not explicitly stated, but lesions were "previously scored by Emory radiologists." This implies at least one, and likely more, radiologists.
      • Qualifications: "Emory radiologists." Specific experience level (e.g., "10 years of experience") is not provided.
    • ECTb™ Strain Analysis:

      • Number of experts: Not applicable in the same way as Emory-CACS, as the ground truth was established by another accepted reference standard (TTE echo).
      • Qualifications: Not applicable for establishing ground truth from experts, but rather from the TTE echo measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Emory-CACS: Not explicitly described. The text states lesions were "previously scored by Emory radiologists," suggesting consensus or a single expert's reading, but no formal adjudication method like "2+1" is mentioned.

    • ECTb™ Strain Analysis: Not applicable, as the ground truth was established by an existing clinical measurement method (TTE echo), not by expert consensus on the device's output.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study focused on human reader improvement with AI assistance was not explicitly described for either new feature in this validation summary. The studies focused on comparing the device's performance directly against a predicate device or an accepted clinical reference standard.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, for both features, the validation described appears to be a standalone (algorithm only) performance assessment.
      • Emory-CACS: The module was compared directly against the SyngoVia™ module, analyzing the same lesions. This implies the device's output was generated and then compared to the predicate's output without human modification or interpretation of the device's results as part of the primary validation.
      • ECTb™ Strain Analysis: The PET MPI measurements (presumably generated by the ECTb™ Strain algorithm) were correlated with TTE echo measurements, which is a comparison of two different measurement modalities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Emory-CACS: The ground truth for the reference lesions was established by the "previously scored by Emory radiologists" and the output of the predicate device (Siemens SyngoVia™ toolbox). In essence, the predicate device's output served as the reference standard against which Emory-CACS was compared.

    • ECTb™ Strain Analysis: The ground truth (reference standard) for strain measurements was established using Transthoracic Echocardiography (TTE echo), which is an accepted clinical reference standard for cardiac function assessment.


    8. The sample size for the training set:

    • The document does not explicitly state the sample size for the training set for either the Emory-CACS or ECTb™ Strain analysis features. The described methodology focuses solely on the validation/test sets used for performance assessment.

    9. How the ground truth for the training set was established:

    • Since the training set size and details are not provided, the method for establishing its ground truth is also not described in the document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213725

    Validate with FDA (Live)

    Device Name
    CardIQ Suite
    Date Cleared
    2022-03-11

    (105 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K073138, K201034

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CardIQ Suite is a non-invasive software application designed to provide an optimized application to analyze cardiovascular anatomy and pathology based on 2D or 3D CT cardiac non contrast and angiography DICOM data from acquisitions of the heart. It provides capabilities for the visualization and measurement of vessels and visualization of chamber mobility. CardIQ Suite also aids in diagnosis and determination of treatment paths for cardiovascular diseases to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging. CardIQ Suite provides calcium scoring, a non-invasive software application, that can be used with non-contrasted cardiac images to evaluate calcified plaques in the coronary arteries, heart valves and great vessels such as the aorta. Calcium Scoring may be used to monitor the progression of calcium in coronary arteries overtime, which may aid in the prognosis of cardiac disease.

    Device Description

    CardlQ Suite is a non-invasive software application designed to work with DICOM CT data acquisitions of the heart. It is a collection of tools that provide capabilities for generating measurement's both automatically and manually, displaying images and associated measurements in an easy-to-read format and tools for exporting images and measurements in a variety of formats.

    CardIQ Suite provides an integrated workflow to seamlessly review calcium scoring and coronary CT angiography (CCTA) data. Calcium Scoring has the capability to automatically segment and label the calcifications within the coronary arteries, and then automatically compute a total and per territory calcium score. The calcium segmentation/labeling is using algorithm. The calcium scoring is based on the standard Agatston/Janowitz 130 (AJ 130) and Volume scoring methods for the segmented calcific regions. The software also provides the users a manual calcium scoring capability that allows them to edit (add/delete or update) auto scored lesions. It also allows the user to manually score calcific lesions within coronary arteries, aorta, aortic valve as well as other general cardiac structures. Calcium scoring offers quantitative results in the AJ 130 score, Volume and Adaptive Volume scoring methods.

    Calcium Scoring results can be exported as DICOM SR to assist with integration into structured reporting templates. Images can be saved and exported for sharing with referring physicians, incorporating into reports and archiving as part of the CT examination.

    CardlQ Suite provides the Coronary 2D Review toolset which allows interactive review of cardiac exams. Coronary CTA datasets can be reviewed utilizing the double oblique angles to visually track the path of the coronary arteries as well as to view the common cardiac chamber orientations. Cine capability for multi-phase data may be useful for visualization of cardiac structures such as chambers, valves and arteries in motion. Distance measurement and ROI tools are available for quantitative evaluation of the anatomy.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the CardIQ Suite, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    The algorithm successfully passed the defined acceptance criteria for automatically segmenting, labeling, and scoring calcific regions in coronary arteries.The validation study demonstrated that the algorithm successfully passed these defined acceptance criteria. (Specific quantitative metrics for "successful passing" are not detailed in this summary, but the clinical testing section implies a qualitative assessment of "very high correlations" with manual methods).
    Equivalent performance to the predicate device (SmartScore 4.0) for computing total calcium score."Very high correlations were found between manual and automated methods for computing the total calcium score, demonstrating equivalent performance of the CardIQ Suite software to the predicate device SmartScore 4.0."

    Study Details

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The summary mentions "a representative set of clinical sample images" for the clinical testing. A specific number is not provided.
    • Data Provenance: The study used a "database of retrospective CT exams." The country of origin is not specified. It is likely internal data from GE Medical Systems SCS or collaborating institutions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Three.
    • Qualifications of Experts: "Three board certified radiologists." Specific experience (e.g., years) is not mentioned.

    4. Adjudication method for the test set

    • The summary states that "Three board certified radiologists manually scored a representative set of clinical sample images using the predicate device." It then compares these manual scores to the automated scores. This implies that the manual scores (generated by the three radiologists using the predicate device) served as the reference standard or ground truth against which the automated algorithm's performance was compared. There is no explicit mention of an adjudication method like 2+1 or 3+1 among the radiologists themselves to reach a single consensus before comparison with the AI.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study with human readers improving with AI assistance vs. without AI assistance was not explicitly described.
    • The clinical study performed was a comparison of the algorithm's standalone performance against manual scoring performed by radiologists using the predicate device. It did not evaluate human reader performance with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone algorithm performance study was done for the Calcium Scoring algorithm. The study compared the automated score outputted by CardIQ Suite with manual scores provided by radiologists using the predicate device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth used was expert manual scoring (by three board-certified radiologists) using the predicate device (SmartScore 4.0). This serves as a form of expert consensus or reference standard for comparison.

    8. The sample size for the training set

    • The summary does not provide a specific sample size for the training set. It mentions that the algorithm was validated using "a database of retrospective CT exams" which is "representative of the clinical scenarios" but does not differentiate between training and test sets by size.

    9. How the ground truth for the training set was established

    • The summary does not explicitly detail how the ground truth for the training set was established. It only describes the validation process (testing set) where radiologists manually scored images. It is generally understood that for deep learning algorithms, training data would also require some form of expert-annotated ground truth, but this document does not provide those specifics for the training phase.