Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223415

    Validate with FDA (Live)

    Date Cleared
    2023-01-26

    (77 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K192954

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs: Test Chemotherapy Drugs are as follows: Carmustine 3.3mg/ml, Cisplatin 1 mg/ml, Cyclophosphamide 20 mg/ml, Dacarbazine 10 mg/ml, Doxorubicin, HCl 2 mg/ml, Etoposide 20 mg/ml, Fluorouracil 50 mg/ml, Methotrexate 25 mg/ml, Mitomycin 0.5 mg/ml, Oxaliplatin 5 mg/ml, Paclitaxel 6 mg/ml, Thiotepa 10 mg/ml, Vincristine 1 mg/ml. Test Fentanyl Citrate are as follows: Fentanyl Citrate Injection 100mcg/2ml. Warning: Do not use with Carmustine and Thiotepa. Please note that the following drugs have low permeation times: (1)Carmustine - 14.1 minutes (2)Thiotepa - 57.2 minutes.

    Device Description

    Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate meets all the requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250. Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs are Class I patient examination gloves bear the product code Nitrile - LZA, LZC, OPJ and QDO (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are blue in color and are powder free. The gloves are ambidextrous i.e., can be worn on right hand or left hand, single use disposable devices that come in five sizes (XS, S, M, L and XL). The physical properties of glove, i.e., tensile strength meet ASTM D 6319-19.

    AI/ML Overview

    The provided document is a 510(k) summary for Nitrile Powder Free Blue Patient Examination Gloves. It details non-clinical testing to demonstrate the device meets acceptance criteria. No information regarding clinical effectiveness studies (such as MRMC studies or standalone algorithm performance for AI devices) is present, as this is not an AI/ML device.

    Here's an analysis of the acceptance criteria and the studies that prove the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Characteristic and ParameterStandard / Test MethodAcceptance CriteriaProposed Device Performance (Result)Comparison Analysis (where available)
    Residual PowderASTM D6124-06 (Reapproved 2017)Less than 2mg / gloveXS: 0.25mg/glove, S: 0.20mg/glove, M: 0.28mg/glove, L: 0.28mg/glove, XL: 0.38mg/gloveSimilar
    Freedom of HolesASTM D5151-19Inspection level, G-I AQL 2.5 (In accordance with ASTM D6319-19)Passed G-I, AQL 1.5Similar (meets AQL 1.5 which is better than 2.5)
    Physical Dimensions (ASTM D6319-19)
    LengthXS: 220mm, min; S: 220mm, min; M: 230mm, min; L: 230mm, min; XL: 230mm, minXS: 245-251mm; S: 247-255mm; M: 245-255mm; L: 247-255mm; XL: 247-252mmSimilar
    WidthXS: 70 ± 10mm; S: 80 ± 10mm; M: 95 ± 10mm; L: 110 ± 10mm; XL: 120 ± 10mmXS: 76-78mm; S: 82-88mm; M: 95-99mm; L: 111-118mm; XL: 120-127mmSimilar, subject device meets ASTM D6319 requirements
    Thickness (Palm)0.05mm, minXS: 0.06-0.06mm; S: 0.06-0.07mm; M: 0.06-0.07mm; L: 0.06-0.06mm; XL: 0.06-0.06mm (average)Similar (all above min)
    Thickness (Finger)0.05mm, minXS: 0.07-0.07mm; S: 0.07-0.08mm; M: 0.07-0.08mm; L: 0.07-0.08mm; XL: 0.07-0.08mm (average)Similar (all above min)
    Physical Properties (ASTM D6319-19)
    Tensile Strength (Before aging)14 MPa, minXS: 23.97 MPa, S: 26.35 MPa, M: 25.15 MPa, L: 25.11 MPa, XL: 25.85 MPa (average)Similar (all above min)
    Tensile Strength (After accelerated aging)14 MPa, minXS: 27.11 MPa, S: 26.78 MPa, M: 29.22 MPa, L: 25.94 MPa, XL: 26.50 MPa (average)Similar (all above min)
    Ultimate Elongation (Before aging)500%, minXS: 571%, S: 571%, M: 538%, L: 552%, XL: 525% (average)Similar (all above min)
    Ultimate Elongation (After accelerated aging)400%, minXS: 462%, S: 458%, M: 434%, L: 457%, XL: 434% (average)Similar (all above min)
    Chemotherapy Drug Permeation (ASTM D6897-05)
    Carmustine 3.3mg/ml>240 minutes (General criteria mentioned in Table 2, but specific values for individual drugs are listed in the Indications for Use and Summary of Technological Characteristics)14.1 minutesSimilar; Below 240 minutes permeation times (Warning included)
    Cisplatin 1 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Cyclophosphamide 20 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Dacarbazine 10 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Doxorubicin, HCl 2 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Etoposide 20 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Fluorouracil 50 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Methotrexate 25 mg/mlNot specified in predicate, but >240 min in proposed device>240 minutesOptional, subject device performs additional Chemo drug test
    Mitomycin 0.5 mg/mlNot specified in predicate, but >240 min in proposed device>240 minutesOptional, subject device performs additional Chemo drug test
    Oxaliplatin 5 mg/mlNot specified in predicate, but >240 min in proposed device>240 minutesOptional, subject device performs additional Chemo drug test
    Paclitaxel 6 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Thiotepa 10 mg/ml>240 minutes (General criteria)57.2 minutesSimilar; Below 240 minutes permeation times (Warning included)
    Vincristine 1 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Fentanyl Citrate Injection 100mcg/2ml>240 minutes (General criteria)>240 minutesSimilar
    Biocompatibility Testing
    Animal Irritation TestISO 10993-10Under the condition of study not an irritant.No observable irreversible alteration; PII = "0"; not corrosive; negligible primary irritation. Met requirement.Similar
    Dermal Sensitization AssayISO 10993-10Under the condition of the study not a sensitizer.No sensitization induced. Met requirement.Similar
    Acute Systemic ToxicityISO 10993-11Not induce systemic toxicityNon-toxic effects of extracts. Met requirement.Similar

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes used for each individual test (e.g., number of gloves tested for holes, number of material samples for tensile strength). However, the tests are conducted in accordance with the specified ASTM and ISO standards, which typically include defined sampling plans.

    • Data Provenance: The origin of the data is from non-clinical laboratory testing performed by "Mah Sing Healthcare Sdn Bhd" or their contracted laboratories, as part of the 510(k) submission process. The submission is from Malaysia. This is prospective data generated for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the device is a physical medical glove and the testing involves objective measurement against pre-defined physical, chemical, and biological standards, not interpretation by human experts or AI algorithms. Ground truth is established by the test procedures and acceptance criteria defined in the ASTM and ISO standards.

    4. Adjudication Method for the Test Set:

    This is not applicable. As stated above, the device performance is evaluated against objective standards, not through adjudication of expert interpretations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This is not applicable. The device is a physical medical glove. The submission does not involve Artificial Intelligence (AI) or Machine Learning (ML) components, nor does it relate to diagnostic image interpretation where MRMC studies would typically be conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This is not applicable. The device is a physical medical glove and does not involve an algorithm.

    7. The Type of Ground Truth Used:

    The ground truth used for these tests is based on defined performance specifications and measurement methodologies established by widely recognized consensus standards such as ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). For example:

    • Physical dimensions: Measured against specified millimeter ranges.
    • Tensile strength and elongation: Measured against minimum MPa and percentage values.
    • Freedom from holes: Assessed against an Acceptable Quality Limit (AQL) per ASTM D5151-19.
    • Chemotherapy drug permeation: Measured as breakthrough detection time in minutes per ASTM D6978-05.
    • Biocompatibility: Assessed based on observable reactions in animal models according to ISO 10993 series and reported as "Passed" or "Not an irritant/sensitizer/systemic toxin."

    8. The Sample Size for the Training Set:

    This is not applicable. There is no "training set" as this device is not an AI/ML product. The manufacturing process is controlled and monitored, but there isn't a data-driven training phase for the device itself.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable. As there is no training set for an AI/ML algorithm, there is no ground truth established for one. The "ground truth" for glove manufacturing is established through quality control and adherence to standardized manufacturing practices and raw material specifications.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1