Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231557

    Validate with FDA (Live)

    Date Cleared
    2023-08-24

    (86 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K192263, K211090

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZENEX FreeMilling & CCM Cast Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

    Device Description

    ZENEX FreeMilling & CCM Cast Abutment consists of FreeMilling Abutment and CCM Cast Abutment.

    ZENEX FreeMilling & CCM Cast Abutment is compatible with the fixtures below:

    K number: K211090
    Device Name: ZENEX Implant System (ZENEX MULTI&PLUS Fixture)
    Dimension Ranges: Diameter 3.75 x 8.5, 10, 11.5, 13, 15; Diameter 4.25 x 7, 8.5, 10, 11.5, 13, 15; Diameter 4.6 x 7, 8.5, 10, 11.5, 13, 15; Diameter 5.05 x 7, 8.5, 10, 11.5, 13, 15; Diameter 5.4 x 7, 8.5, 10, 11.5, 13; Diameter 5.9 x 7, 8.5, 10, 11.5, 13; Diameter 6.75 x 7, 8.5, 10, 11.5, 13

    Tolerance of dimension shall be within ± 1% range.

    The dimensions of abutments are as following:

    Device Name: FreeMilling Abutment
    Dimension Ranges: Diameter 4.0/4.5/5.0/5.2/5.7/6.0/6.5/7.0
    Gingival Heights (mm): 1.3/1.8/2.8/3.8
    Minimum Post height (mm): 4
    Angulation: 0°

    Device Name: CCM Cast Abutment
    Dimension Ranges: Diameter 4.0 / 4.5
    Gingival Heights (mm): 0.8 / 2.8
    Minimum Post height (mm): 4
    Angulation: 0°

    The Abutments have below featured:

    Name: FreeMilling Abutment
    Uses: The Abutment is connected with fixture, and it supports prosthesis which restores tooth function.
    Surface: Partial TiN coated in upper
    Connection: Screw retained

    Name: CCM Cast Abutment
    Uses: It is an abutment that is used when making a retained / customized prosthesis.
    Surface: NA
    Connection: Screw retained

    Tolerance of dimensions for Abutments shall be within ± 1% range.

    ZENEX FreeMilling & CCM Cast Abutment is provided non-sterile.

    Materials:

    • FreeMilling Abutment is fabricated of Ti-6Al-4V ELI according to ASTM F136-13
    • CCM Cast Abutment is fabricated of Co-Cr-Mo according to ASTM F1537-11
    AI/ML Overview

    I'm sorry, but your prompt asks for details about acceptance criteria and a study proving device performance, including a table of acceptance criteria and reported device performance, sample size, ground truth establishment, expert qualifications, adjudication methods, MRMC study information, standalone performance, and training set details.

    However, the provided text is a 510(k) summary for a dental abutment, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and the results of a primary clinical or non-clinical performance study. It mentions non-clinical test data for biocompatibility and sterilization, but it doesn't present performance metrics against specific acceptance criteria in the format you've requested for a device that relies on a performance study to prove its criteria.

    The 510(k) summary establishes equivalence based on:

    1. Indications for Use: Same as predicate.
    2. Technological Characteristics: Similar design, dimensions (with minor, deemed non-critical differences), materials, and sterilization methods.
    3. Non-Clinical Test Data: Biocompatibility and sterilization (often leveraged from predicate devices or performed according to recognized standards).

    Therefore, I cannot extract the specific information you've requested about acceptance criteria and a study proving those criteria from the given text. The document's purpose is to argue substantial equivalence, not to present a standalone performance study with detailed acceptance criteria and results.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1