510(k) Data Aggregation

    K Number
    K243659

    Validate with FDA (Live)

    Date Cleared
    2025-04-02

    (126 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K240625, K192140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FLASH™ Facet Fusion Instruments are intended to be used during spinal surgery to assist the surgeon in precisely locating and preparing anatomical structures, specifically the lumbar facet joints, in either open or minimally invasive procedures. The FLASH™ Facet Fusion Instruments are specifically designed for use with the 7D Surgical System™, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure can be identified.

    The IsoTis® Facet Fusion Instruments are intended to be used during spinal surgery to assist the surgeon in precisely locating and preparing anatomical structures, specifically the lumbar facet joints, in either open or minimally invasive procedures. The IsoTis® Facet Fusion Instruments are specifically designed for use with the 7D Surgical System™, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure can be identified.

    Device Description

    The FLASH™ Facet Fusion Instruments are a manually operated disposable instrument set to be used with the 7D Surgical System to assist the surgeon in precise site preparation during open or minimally invasive spinal surgery. The FLASH™ Facet Fusion Instruments consist of a Lumbar Inserter, a Navigation T-Handle, an Array Adaptor, a Lumbar Tamp and Lumbar Facet Drill. The product is provided sterile in a ready-to-use, single patient use container. The subject instruments are made of a combination of stainless steel and medical grade plastics commonly used in orthopedic procedures. The surgical instruments are intended for use as an adjunct to established posterior lumbar fixation procedures, where the Instrumentation enables the location and preparation of a facet joint defect with or without the use of Navigation and delivery of bone graft into the defect.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the FLASH™ Facet Fusion Instruments (K243659) do not contain information about acceptance criteria or a detailed study proving the device meets those criteria, particularly not in the context of an AI/algorithm-driven device performance evaluation.

    The document states:

    • "Clinical Testing: A clinical trial was not required to demonstrate the safety and effectiveness of FLASH™ Facet Fusion Instruments. Clinical validation is unnecessary as FLASH™ Facet Fusion Instruments does not introduce new indications for use, and device features are substantially equivalent to the previously cleared predicate device identified. The clinical safety and effectiveness of image guided surgery systems are historically accepted for both the predicate and subject device."

    This indicates that the clearance was based on substantial equivalence to predicate devices, and the testing performed was primarily non-clinical bench performance testing, sterilization and packaging validation, and compliance conformity assessments. The device itself is described as a "manually operated disposable instrument set" used with a surgical system, not an AI or algorithm-driven diagnostic or assistive technology in the way implied by the question's requirements (e.g., MRMC studies, standalone algorithm performance, AI assistance effect size).

    Therefore, based solely on the provided text, I cannot fulfill the request for information related to acceptance criteria and a study proving an AI/algorithm-driven device meets them. The document does not describe such studies or acceptance criteria because the device is not an AI/algorithm device that would typically undergo such evaluations.

    Response based on the provided document:

    1. A table of acceptance criteria and the reported device performance:
    * Acceptance Criteria: Not explicitly stated in terms of quantitative performance metrics for device output as would be seen for an AI/algorithm. The criteria were based on functional verification ("design outputs meet design inputs") and compatibility.
    * Reported Device Performance:
    * "The functionality of the new or modified features was tested to ensure that system requirements and user needs were met as defined."
    * "Specifically, the navigation compatibility of the subject tool with the 7D Surgical System and facet decortication capability were verified and validated through testing."
    * "The results show that the subject device is substantially equivalent to the cleared predicates."
    * "non-clinical performance data demonstrate that the FLASH™ Facet Fusion Instruments perform as expected and in a manner that is substantially equivalent to its predicate devices."

    2. Sample sized used for the test set and the data provenance:
    * Not applicable/Not provided. The testing described is bench performance testing of mechanical instruments, not an evaluation of an AI algorithm on a data set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    * Not applicable/Not provided. Ground truth in the context of AI/algorithms is not relevant for this device's described testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    * Not applicable/Not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * No. The document explicitly states: "A clinical trial was not required to demonstrate the safety and effectiveness of FLASH™ Facet Fusion Instruments." This type of study is for AI-assisted diagnostic or decision support systems, which this device is not.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    * No. This is a manually operated surgical instrument set, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    * Not applicable/Not provided. The "ground truth" for this device would relate to its mechanical function and compatibility with the surgical system, verified through engineering tests, not medical expert consensus on patient data.

    8. The sample size for the training set:
    * Not applicable/Not provided. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:
    * Not applicable/Not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222753

    Validate with FDA (Live)

    Date Cleared
    2022-12-09

    (88 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K192140, K162375

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SeaSpine 7D Navigation Instruments are intended to be used during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are specifically designed for use with the 7D Surgical System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure can be identified.

    Device Description

    The SeaSpine 7D Navigation Instruments are used during the preparation and placement of SeaSpine screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The SeaSpine 7D Navigation Instruments are wireless optically tracked and specifically designed for use with the 7D Surgical system.

    The SeaSpine 7D Navigation Instruments utilize commercially available passive reflective marker spheres to determine the position and orientation of instruments. Each tracked Instruments requires a unique marker position configuration to enable the tracking system to distinguish the tools from one to the other. Position and orientation data of tracked Spinal Instruments are linked to the preoperative scan data using the 7D Surgical System workstation.

    AI/ML Overview

    The provided text concerns the FDA 510(k) premarket notification for the "SeaSpine 7D Navigation Instruments." This submission focuses on the mechanical compatibility and accuracy of newly designed instruments within an existing navigated surgical system (7D Surgical System), rather than an AI/ML-driven diagnostic or prognostic device.

    Therefore, many of the requested criteria related to AI/ML device validation (e.g., number of experts for ground truth, adjudication methods, MRMC studies, training set details) are not applicable to this type of device and the information is not provided in the submission. The device is a surgical instrument that is optically tracked and mechanically connects to an existing navigation system.

    Here's an analysis of the provided information relative to your request:

    Acceptance Criteria and Device Performance (Mechanical & Positional Accuracy)

    The acceptance criteria are primarily focused on system compatibility, mechanical design, and positional accuracy of the instruments.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryDescription from DocumentReported Device Performance / Conclusion
    System VerificationScope of the test is to verify the design requirement specifications of the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe compatibility with the 7D Surgical System under test case protocols.Verification successful, all design requirements have been fulfilled.
    System ValidationScope of the test is to validate the Indications for Use and Customer Requirements of the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe compatibility with the 7D Surgical System under simulated use case situations.Validation successful, all user needs met.
    Safety Regarding Risk AnalysisImplementation and effectiveness of all risk control requirements specified in the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe risk analysis are tested and verified.Risk Control requirements are effective and mitigate the associated risks to an acceptable level.
    Product Safety StandardsThe SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe was tested to the following recognized standards: ISO 10993-1, and ISO 17665-1. (Note: These standards relate to biocompatibility and sterilization, not directly positional accuracy).Compliance with recognized standards have been verified in the previous application. Previous test results have not been affected by this change.
    Non-Clinical Accuracy (Positional)System's accuracy is tested using the SeaSpine Navigation Adapter, Reference Clamp, Pedicle Probe compatibility with the 7D Surgical System on phantom models following the ASTM F2554-10 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems in addition to Target Registration Error (TRE). Also, Angular Trajectory Error (ATE) was used.All accuracy specifications have been met for the SeaSpine® Navigation Adapter, Reference Clamp, and Pedicle Probe. Accuracy testing for Universal Tracking / non-tracked surgical probe Array/Navigation Adapter assembly have been verified in previous application K192140 and K162375. (Specific numerical values for TRE/ATE or other accuracy metrics from ASTM F2554-10 are not explicitly provided in this summary, only that "all accuracy specifications have been met").

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated as a number of distinct "cases" or "patients" for positional accuracy testing. The document states "phantom models" were used. It implies repeatable tests were run to demonstrate accuracy, but the number of phantom models or repetitions is not specified.
    • Data Provenance:
      • Country of Origin: Not specified but likely conducted in North America (given the submitter's location in Canada and the FDA clearance).
      • Retrospective/Prospective: Neither. The studies were non-clinical performance surgical simulations conducted on phantom models. This is a prospective testing methodology for device performance on inanimate objects.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of those Experts

    • Not Applicable. This is not an AI/ML diagnostic device requiring expert interpretation for ground truth. Ground truth for positional accuracy testing (TRE/ATE) would be established by precise physical measurements using metrology equipment on the phantom models, not human expert consensus.

    4. Adjudication Method for the Test Set

    • Not Applicable. As above, no human expert consensus or adjudication was involved in establishing ground truth for mechanical accuracy on phantom models.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI/ML diagnostic or image-based assistance device. It's a suite of surgical instruments used with a navigation system. MRMC studies are not relevant for this type of device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Partially Applicable / Adapted Interpretation: While not an AI algorithm, the device's accuracy (its "standalone" performance within the larger navigation system) was assessed through "Non-Clinical Accuracy" tests on phantom models. This involved evaluating the positional and angular accuracy of the SeaSpine instruments when used with the 7D Surgical System. The ASTM F2554-10 standard is for measuring the positional accuracy of computer-assisted surgical systems. The "standalone" performance here refers to the system's ability to accurately track and report the position of the instruments relative to the "ground truth" on the phantom.

    7. The Type of Ground Truth Used

    • Precision Measurement/Metrology on Phantom Models: Ground truth for Target Registration Error (TRE) and Angular Trajectory Error (ATE) is established by precise physical measurements of the actual position of fiducials or targets on the phantom model, compared against the position reported by the navigation system. This is a form of engineered ground truth for spatial accuracy.

    8. The Sample Size for the Training Set

    • Not Applicable. This is not a machine learning device that requires a "training set." The instruments are designed and verified based on engineering principles and compatibility with an existing optical tracking system.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As above, no training set for a machine learning model.

    In summary, the provided document details the non-clinical testing of a surgical navigation instrument for its compatibility, safety, and mechanical positional accuracy with an existing stereotactic system. It does not describe the validation of an AI/ML-driven diagnostic or prognostic device, hence many of the requested criteria are not present or applicable. The key validation data provided relates to meeting ASTM standards for positional accuracy and demonstrating compatibility with the predicate device's system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201966

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-09-03

    (50 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180352, K192140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 7D Surgical System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons. The system is also intended to be used as the primary surgical luminaire during image guided surgery. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure can be identified.

    Device Description

    The 7D Surgical System is intended for use as a stereotaxic image guided surgical navigation system during spine surgery. The system provides image registration between preoperative scan data and data captured intraoperatively from the 7D Surgical System structured light scanner and/or user selected points. The system provides guidance data by tracking and displaying the position and orientation of wireless optically tracked Spinal Instruments including the 7D Surgical Pedicle Probe and Awl, now including the FlashLock and Flex Rod, relative to the patient. Position and orientation data of tracked Spinal Instruments are linked to the preoperative scan data using the 7D Surgical System workstation. The system is intended to be used as the primary surgical luminaire for image guided surgery. Similar to the previously cleared 7D Surgical System, the system tracks the position and orientation of a FlashLock and Flex Rod.

    The system is intended to be used for both image fusion and navigation for neurological applications where reference to a rigid structure can be identified relative to a preoperative image data of the anatomy.

    The Tracking System enables the surgeon to view the position and orientation of 7D Surgical System Spinal Instruments relative to registered preoperative image data while performing the surgical procedure. Each of the 7D Surgical System instruments, including the FlashLock and Flex Rod, utilizes commercially available passive reflective marker spheres to determine the position and orientation of instruments. Each tracked Instruments requires a unique marker position configuration to enable the tracking system to distinguish the tools from one to the other.

    The Software links all system components and displays navigational data to the surgeon. lt provides methods for loading preoperative scans and guides the surgeon through the process of surface model creation, structured light acquisition, registration, registration verification, and navigation.

    AI/ML Overview

    The provided text describes a 510(k) submission for the 7D Surgical System with new instruments (FlashLock and Flex Rod). It does not contain information about an AI-powered device or a multi-reader, multi-case (MRMC) comparative effectiveness study. Therefore, I cannot fully answer all aspects of your request as they pertain to AI device performance.

    However, I can extract the information provided regarding the device's acceptance criteria and the study conducted, focusing on the available details.

    Here's a breakdown of the information provided in the document:

    1. A table of acceptance criteria and the reported device performance

    The document mentions "Non-Clinical Accuracy" testing as proof that "All accuracy specifications have been met for the FlashLock and Flex Rod." However, it does not provide specific numerical acceptance criteria or the exact reported performance values (e.g., specific TRE or ATE values). It only states that the testing was performed using ASTM F2554-10 and Target Registration Error (TRE).

    Acceptance CriteriaReported Device Performance
    Non-Clinical Accuracy (ASTM F2554-10 & TRE): System's accuracy testing on phantom models. Specific numerical criteria not provided in document."All accuracy specifications have been met for the FlashLock and Flex Rod." Specific numerical performance not provided.

    2. Sample size used for the test set and the data provenance

    The document mentions "phantom models" for non-clinical performance surgical simulations. It does not specify the sample size of these phantom models or their provenance (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish ground truth for the test set. The ground truth appears to be established by physical measurements on phantom models, as indicated by the description of TRE and ATE.

    4. Adjudication method for the test set

    Not applicable, as expert adjudication is not mentioned for establishing ground truth or evaluating the device's accuracy on phantom models.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a stereotaxic surgical navigation system, not an AI-powered diagnostic or assistive tool for human readers in the sense of image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The performance testing (Non-Clinical Accuracy) was conducted on the "System's accuracy... using the 7D Surgical System on phantom models." This can be considered a form of standalone performance evaluation for the navigation system, focusing on its accuracy in guiding instruments. However, it's not an "algorithm-only" performance in the context of an AI device interpreting data. The system itself is the "standalone" entity being evaluated for its navigational accuracy.

    7. The type of ground truth used

    The ground truth used for accuracy evaluation (TRE and ATE) was based on: "the error discrepancy between the position reported by the image guided surgery system and the ground truth position measured physically or otherwise." This implies a physically measured ground truth on phantom models, not derived from expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    The document does not mention a "training set" as would be applicable to machine learning or AI models. This system is described as a "stereotaxic image guidance system" and its functionality seems to be based on established engineering principles of tracking and navigation, rather than a learned model.

    9. How the ground truth for the training set was established

    Not applicable, as a training set for an AI model is not mentioned.