Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241339

    Validate with FDA (Live)

    Date Cleared
    2024-11-14

    (185 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K191909

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Besmed HME Filter (Heat and Moisture Exchange) is intended to decrease the transmission of bacteria and viruses to/from a patient, and to maintain moisture levels in the patient's respiratory tract while administering anesthesia gas, providing artificial respiration, or assisting with ventilation.

    The Besmed HME Filter is typically placed at the patient's end of the breathing system, specifically between the circuits Y - piece and either the catheter mount or patient airway device. It is essential to note that the Besmed HME Filter is intended for single-use only and should be used on a single patient for up to 24 hours.

    Besmed HME Filter is intended to be used within critical care and hospital/ institutional environments by qualified personnel.

    Besmed Bacterial Filter

    The purpose of the Besmed Bacterial Filter is to diminish the passage of bacteria and viruses to patients receiving anesthesia gas. Besmed Bacterial Filter is designed for utilization with ventilators, anesthesia machines, and open flow systems where there is a need for filtration of inhaled gases.

    The Besmed Bacterial Filter is a disposable device intended for the exclusive use of a single patient within a 24-hour timeframe. Bacterial Filter is designed for utilization in hospital environments by medical professionals who possess adequate training in the operation of mechanical ventilators, respiratory systems, and humidification systems.

    Device Description

    Besmed HMEF

    The Besmed HME Filter absorbs the heat and moisture exhaled by the patients. Upon inhalation, it releases the previously retained heat and humidity to the dry inhalation gases. The HME filter is designed with electrostatic cotton for filtration, as well as a heat-preserved and moisture-absorbed paper roll for maintaining the warmth and humidity of the airway in a respiratory system. Additionally, it eliminates bacterial and viral contaminants from the breathing circuit gas. To ensure passive humidification treatment, it is recommended to install the Heat and Moisture Exchanger Filter (HMEF) on the patient's end. Besmed HME Filter is intended for adult and child patients with the tidal volume between 150ml-1000ml.

    Besmed Bacterial Filter

    Besmed Bacterial Filter is for single-use purpose. The intended use is achieved through the implementation of an electrostatic filtration mechanism. Transparent and compact housing material makes it easy to check the surface of the filter media for a foreign substance.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of non-clinical testing for the "Besmed Bacterial Filter and HMEF." This summary focuses on the non-clinical test results, as no clinical testing was performed or required.

    Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated against a set of predetermined acceptance criteria across various standards. The table below summarizes these criteria and the reported results.

    Test StandardsTest PerformedAcceptance CriteriaReported Device Performance
    Packaging for Terminally Sterilized Medical Devices
    ISO 11737-2:2020Sterility testNegativePass
    ISO 10993-7:2008EO/ ECH/ EG residual testNon-DetectedPass
    ASTM F1929-23Dye Penetration TestNo PenetrationPass
    ASTM F1140/F1140M-13(2020)e1Burst testPackage successfully held for creep durationPass
    Creep testIt should maintain the specified pressure for 30 secondsPass
    ASTM F88/F88M-23Seal strength of flexible barrier materials testSeal strength shall not be greater than ± 2SD compare to pre transit packagePass
    ASTM F1608-21Microbial Ranking TestLRV>3.0 (99.9%)Pass
    ISTA 2A:2011Packaged products testNo visible damagePass
    Shelf Life Test
    ASTM F1980-21Accelerated aging testNo visible damage after aging simulationDevice meets its performance specification
    Performance Test
    Measurement of moisture loss<16 mg/l at 250 ml/500 ml/1000 ml tidal volumeThe average moisture loss met the acceptance criteria for the specified tidal volumes
    HME moisture outputGreater than 30 mg/LPass
    ISO 9360-1:2000Measurement of pressure dropPressure drop at 30L/min, 60L/min and 90L/min should not be greater than 5hPa.Pass
    HMEF gas leakageLeakage less than 4ml/min @ 7kpa, and 15kpaPass
    ComplianceMethodology only, no acceptance criteriaMeets requirements
    ISO 23328-1:2008Salt test method to assess filtration performanceBacterial filter and HMEF efficiency shall be > 99.9% @15 LPM and 30 LPM under 0 hour and 24 hour conditioningPass
    ISO 80369-7:2021Leakage by pressure decayLeakage rate should not exceed 0.005 Pa·m³/s @ 330 kPa for 20 sPass
    Sub-atmospheric pressure air leakageLeakage rate should not exceed 0.005 Pa·m³/s @ 101kPa for 20 sPass
    Stress crackingNo signs of seal break and leakage @101 kPa after being stressPass
    Resistance to separation from axial loadNo disconnections observed at an axial force of 35 NPass
    Resistance to separation from unscrewingNo disconnections observed at unscrewing torque of 0.020NmPass
    Resistance to overridingNo override the threads observed at torque of 0.17 NmPass
    ISO 5356-1:2015Security of engagementNo disconnections observed at axial separation force of (50 ± 5) N for 10 sPass
    Leakage from 22 mm latching socketsNo signs of leakage at (8 ± 0.5) kPaPass
    Drop procedure for 22 mm latching socketsNo signs of damage during the testPass
    ASTM F2101-19Bacterial / Virus Filtration Efficiency testEvaluate BFE/VFE filter efficiency shall be greater than 90%Pass
    Biocompatibility Test
    ISO 10993-5:2009In Vitro Cytotoxicity testNot more than 50 % of the cells are roundPass
    ISO 10993-10:2021Skin sensitization tests (Maximization Test)Did not produce skin sensitization on Guinea pigPass
    ISO 10993-23:2021Tests for irritationDid not cause intracutaneous irritationPass
    ISO 10993-11:2017Acute systemic toxicity studyDid not cause systemic toxicity reaction or deathPass
    USP 151Pyrogenicity TestDid not induce pyrogenic responsePass
    ISO 18562-2:2024Emissions of particulate matterMeasured concentrations of particulate matter PM2.5 ≤ 35µg/m³; PM10 ≤ 150µg/m³Pass
    ISO 18562-3:2024Emissions of VOCsThe MOS (Margin Of Safety) value shall be higher than 1 for identified substances.Pass

    Study Details

    The provided document describes a non-clinical testing summary to support the substantial equivalence of the "Besmed Bacterial Filter and HMEF."

    1. Sample Size used for the test set and the data provenance: The document does not specify the exact sample sizes (number of devices) used for each individual test. The data provenance is implied to be from internal testing conducted by Besmed Health Business Corp., likely in Taiwan (New Taipei City, Taiwan is their address). The nature of the tests (performance, packaging, biocompatibility) indicates that these were prospective evaluations of manufactured devices.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not applicable as the studies are non-clinical, laboratory-based tests against established international and ASTM standards. Ground truth would be defined by the criteria within these standards, which are met if the device passes the specified thresholds. No human expert consensus for "ground truth" in clinical interpretation is relevant here.

    3. Adjudication method for the test set: Not applicable for non-clinical, objective laboratory tests against defined standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not applicable as the device is a physical medical device (breathing circuit bacterial filter and HMEF), not an AI-powered diagnostic or assistive technology.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is not applicable as the device is a physical medical device, not an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for these tests is based on the acceptance criteria defined by international standards (ISO) and ASTM standards. These standards provide objective, measurable thresholds for performance, safety, and biocompatibility.

    7. The sample size for the training set: This is not applicable as the device is a physical medical device, not a machine learning model that requires a training set.

    8. How the ground truth for the training set was established: This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1