    Reference Devices :

    K163387, K180698, K182323, K191760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PICO Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. PICO Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a hospital and homecare setting. Appropriate wound types include:

    • Chronic
    • Acute
    • Traumatic
    • Subacute and dehisced wounds
    • Partial-thickness burns
    • Ulcers (such as diabetic or pressure)
    • Flaps and grafts
    • Closed surgical incisions

    When used on closed surgical incisions, PICO Single Use Negative Pressure Wound Therapy System is intended to aid in reducing the incidence of:

    • Superficial and deep incisional surgical site infections for high risk patients in Class II wounds
    • Post-operative seroma
    • Dehiscence

    Note: When used on closed incisions for the reduction of SSI, the safety and effectiveness for Class III (contaminated) and Class IV (Dirty/Infected) wounds have not been demonstrated. Furthermore. Class IV surgical wounds are not expected to be closed primarily. The device has not been demonstrated to reduce organ space surgical site infections. The device is intended to aid in reducing the incidence of, but not treat, seroma, dehiscence, or infected wounds - the use of PICO does not preclude the need to develop and follow a comprehensive infection management protocol.

    PICO 7 Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. PICO 7 Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a hospital and homecare setting. Appropriate wound types include:

    • Chronic
    • Acute
    • Traumatic
    • Subacute and dehisced wounds
    • Partial-thickness burns
    • Ulcers (such as diabetic or pressure)
    • Flaps and grafts
    • Closed surgical incisions

    When used on closed surgical incisions, PICO 7 Single Use Negative Wound Therapy System is intended to aid in reducing the incidence of:

    • Superficial and deep incisional surgical site infections for high risk patients in Class II wounds
    • Post-operative seroma
    • Dehiscence

    Note: When used on closed incisions for the reduction of SSI, the safety and effectiveness for Class III (contaminated) and Class IV (Dirty/Infected) wounds have not been demonstrated. Furthermore, Class IV surgical wounds are not expected to be closed primarily. The device has not been demonstrated to reduce organ space surgical site infections. The device is intended to aid in reducing the incidence of, but not treat, seroma, dehiscence, or infected wounds - the use of PICO does not preclude the need to develop and follow a comprehensive infection management protocol.

    PICO 7Y Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. PICO 7Y Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a hospital and homecare setting. Appropriate wound types include:

    • Chronic
    • Acute
    • Traumatic
    • Subacute and dehisced wounds
    • Partial-thickness burns
    • Ulcers (such as diabetic or pressure)
    • Flaps and grafts
    • Closed surgical incisions

    When used on closed surgical incisions, PICO 7Y Single Use Negative Pressure Wound Therapy System is intended to aid in reducing the incidence of:

    • Superficial incisional surgical site infections for high risk patients in Class I wounds
    • Post-operative seroma
    • Dehiscence

    Note: When used on closed incisions for the reduction of SSI, the safety and effectiveness for Class II (Clean/ Contaminated), Class III (contaminated) and Class IV (Dirty/Infected) wounds have not been demonstrated. Furthermore, Class IV surgical wounds are not expected to be closed primarily. The device has not been demonstrated to reduce organ space surgical site infections. The device is intended to aid in reducing the incidence of, but not treat, seroma, dehiscence, or infected wounds - the use of PICO does not preclude the need to develop and follow a comprehensive infection management protocol.

    PICO 14 Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. PICO 14 Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a hospital and homecare setting. Appropriate wound types include:

    • Chronic
    • Acute
    • Traumatic
    • Subacute and dehisced wounds
    • Partial-thickness burns
    • Ulcers (such as diabetic or pressure)
    • Flaps and grafts
    • Closed surgical incisions

    When used on closed surgical incisions for up to 7 days, PICO 14 Single Use Negative Pressure Wound Therapy is intended to aid in reducing the incidence of:

    • Superficial and deep incisional surgical site infections for high risk patients in Class II wounds
    • Post-operative seroma
    • Dehiscence

    Note: When used on closed incisions for the reduction of SSI, the safety and effectiveness for Class III (contaminated) and Class IV (Dirty/Infected) wounds have not been demonstrated. Furthermore, Class IV surgical wounds are not expected to be closed primarily. The device has not been demonstrated to reduce organ space surgical site infections. The device is intended to aid in reducing the incidence of, but not treat, seroma, dehiscence, or infected wounds - the use of PICO does not preclude the need to develop and follow a comprehensive infection management protocol.

    Device Description

    The PICO Family of devices, PICO (cleared under K163387), PICO 7 (cleared under K180698), PICO 7Y (cleared under K182323), PICO 14 (cleared under K191760) all have the same main function of management of fluid through dressing absorbency and evaporation with added benefit of negative pressure. The pump provides a -80mmHg nominal pressure under the dressing, applying Negative Pressure Wound Therapy (NPWT) to the wound. The PICO Single Use Negative Therapy Systems consist of:

    • PICO Pump
    • Dressing (s)
    • Fixation strips
    • Batteries
    • Connection tubing
    • Instructions for Use

    The system is a canister-less system - fluid is managed by the dressing. The pump that delivers the NPWT is a portable, battery-powered, software-controlled system that can provide continuous application of negative pressure to the wound as a delivery pressure at a nominal value of -80mmHg. The PICO Systems are designed to be used at home or within a healthcare setting by an appropriate healthcare professional.

    AI/ML Overview

    The PICO family of devices (PICO, PICO 7, PICO 7Y, and PICO 14) are Negative Pressure Wound Therapy Systems intended for use in wound treatment.

    1. A table of acceptance criteria and the reported device performance

    The provided document does not explicitly state "acceptance criteria" with numerical thresholds for performance. Instead, it demonstrates the device's effectiveness through a systematic literature review and meta-analyses, aiming to show a "statistically significant reduction" in the incidence of certain wound complications compared to standard care. The reported device performance is presented as Odds Ratios (OR) with 95% Confidence Intervals (CI). A favorable effect is suggested by an OR less than 1.

    Here's a summary of the meta-analysis results as reported:

    OutcomeDevice Performance (Odds Ratio [95% CI])Interpretation (Favorable Effect if OR < 1)
    Surgical Site Infections (SSI) - Overall0.36 [0.27; 0.49]Favorable
    Superficial SSI0.30 [0.17; 0.53]Favorable
    Deep SSI0.67 [0.46; 0.96]Favorable
    Post-operative Seroma0.37 [0.16; 0.86]Favorable
    Dehiscence0.63 [0.43; 0.92]Favorable

    2. Sample sizes used for the test set and the data provenance

    The "test set" for this submission is comprised of studies identified through a systematic literature review. This means the data is retrospective, as it relies on previously published research. The provenance of the data is global, as the literature review included studies from various researchers and institutions, not limited to a single country.

    • Overall SSI Meta-analysis:

      • Evaluable patients: 2,737 in the PICO Family therapy group, 2,936 in the SOC (control) group.
      • Total studies: 17 (12 RCTs, 5 prospective observational).
    • Superficial SSI Meta-analysis:

      • Evaluable patients: 356 in the PICO Family therapy group, 367 in the SOC (control) group.
      • Total studies: 8 (5 RCTs, 3 prospective observational).
    • Deep SSI Meta-analysis:

      • Evaluable patients: 1,146 in the PICO Family therapy group, 1,138 in the SOC (control) group.
      • Total studies: 6 (4 RCTs, 2 prospective observational).
    • Post-operative Seroma Meta-analysis:

      • Evaluable patients: 608 in the PICO Family therapy group, 618 in the SOC (control) group.
      • Total studies: 10 (7 RCTs, 3 prospective observational).
    • Dehiscence Meta-analysis:

      • Evaluable patients: 551 in the PICO Family therapy group, 656 in the SOC (control) group.
      • Total studies: 7 (6 RCTs, 1 prospective observational).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for the test set (the outcomes of the studies included in the meta-analysis) was established by the original researchers and clinicians who conducted each individual study. The document states:

    • "Two (2) independent reviewers performed the study selection."
    • "In case of disagreement a third reviewer made the final decision after reading the full text paper or conference abstract."
    • "Data was extracted from included studies by one reviewer using a predefined and standardized data extraction form and checked by a second reviewer for accuracy."

    The qualifications of these specific reviewers are not provided in the document. However, the outcomes themselves (SSI, seroma, dehiscence) are medical diagnoses and complications identified by the healthcare professionals directly involved in the patient care and study follow-up within the original research for each of the 25 articles. These would typically be surgeons, nurses, and other medical personnel, under the guidance of the study investigators.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    For the selection of studies for inclusion in the meta-analysis:

    • "Two (2) independent reviewers performed the study selection."
    • "If either reviewer deemed an article as potentially relevant, then the article progressed to full text screening."
    • "In case of disagreement a third reviewer made the final decision after reading the full text paper or conference abstract."
      This indicates a 2-reviewer initial screening with a third reviewer for adjudication in case of disagreement (2+1 method) for study selection.

    For data extraction:

    • "Data was extracted from included studies by one reviewer using a predefined and standardized data extraction form and checked by a second reviewer for accuracy."
      This implies a single reviewer for initial extraction with a second reviewer for verification, which is a form of 1+1 review.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done, nor is this device an AI-powered diagnostic tool. The document describes a medical device (Negative Pressure Wound Therapy System) and its clinical effectiveness compared to standard care through a meta-analysis of existing clinical trials, not an AI system. Therefore, details regarding human reader improvement with/without AI assistance are not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a medical device, a wound therapy system, not an algorithm or AI system. Therefore, the concept of "standalone algorithm performance" (without human-in-the-loop) is not applicable. The device operates independently of human interpretation in its direct therapeutic action (applying negative pressure), but its application and management involve healthcare professionals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the effectiveness claims of the PICO family of devices (reduction in SSI, seroma, and dehiscence) is based on outcomes data from various clinical studies. These outcomes (e.g., incidence of infection, presence of seroma, occurrence of dehiscence) were reported by the original study investigators, presumably determined through clinical assessment, and potentially supported by other diagnostic methods (e.g., microbiology for infection, imaging for seroma, clinical examination for dehiscence), depending on the specific protocols of each individual study included in the meta-analysis. For infection, the CDC guidelines for defining SSI were used, which involve clinical criteria.

    8. The sample size for the training set

    This document describes a medical device and its clinical effectiveness evaluated through a meta-analysis of clinical studies, not an AI/machine learning algorithm that requires a "training set." Therefore, the concept of a training set sample size is not applicable here.

    9. How the ground truth for the training set was established

    As explained in point 8, the concept of a "training set" is not applicable to this submission.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K180698, K182323, K191760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System. PICO 14 Single Use Negative Pressure Wound Therapy System. and the PICO Fluid Management Pack are indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.

    Appropriate wound types include:

    • Chronic
    • · Acute
    • · Traumatic
    • Subacute and dehisced wounds
    • · Partial-thickness burns
    • · Ulcers (such as diabetic or pressure)
    • · Flaps and grafts
    • Closed surgical incisions

    PICO 7, PICO 7 Y and PICO 14 Single Use Negative Pressure Wound Therapy Systems are suitable for use in both a hospital and homecare setting.

    Device Description

    All of the PICO devices (PICO 7 (K180698), PICO 7Y (K182323), and PICO 14 (K191760) are canister-free single-use Negative Pressure Wound Therapy (NPWT) Systems and use an absorbent dressing connected to a small NPWT pump by a tubing and port. The dressing manages wound exudate by a combination of absorption and evaporation. The pump provides for a -80mmHg nominal pressure under the dressing, applying Negative Pressure Wound Therapy (NPWT) to the wound. PICO Fluid Management Pack are packs of 5 individually packaged PICO Dressings designed for use with PICO devices.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device, specifically the PICO Single Use Negative Pressure Wound Therapy System. This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the device meets specific acceptance criteria through a standalone clinical study on its performance in treating wounds.

    Therefore, the information typically found in a clinical study report proving device performance against acceptance criteria (such as sample size, expert adjudication, MRMC studies, ground truth establishment for training, etc.) is not present in this type of regulatory submission. The document explicitly states:

    "Performance data provided in previously cleared 510(k)s for PICO 7, PICO 7Y and PICO 14 Single Use Negative Pressure Wound Therapy Systems (including PICO Fluid Management Pack) continue to support substantial equivalence."

    Instead of a new clinical performance study, the document focuses on:

    • Comparison of Technological Characteristics: Demonstrating that the subject device's features, indications for use, environment of use, patient population, materials, sterilization methods, biocompatibility, pump type, and electrical safety testing methods are the same as the predicate device.
    • Minor Changes and Supporting Data: Addressing specific minor changes in the device (updated magnet warning text, removal of pump warning label) and providing additional data to support these changes. This additional data includes:
      • Bench testing: For magnetic field strength.
      • Analysis of postmarket surveillance: Including a systematic literature review on IMDs and magnetic interference, and a review of MAUDE database and company complaint data.
      • Risk analysis: Updating the risk assessment based on the above.

    Based on the provided text, here's what can be extracted and what information is not available:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a table of acceptance criteria in the traditional sense of a clinical trial (e.g., Sensitivity X%, Specificity Y%). Instead, the "acceptance" for this 510(k) is based on demonstrating substantial equivalence to the predicate device for various technical and clinical characteristics. The "performance" is primarily demonstrated by showing that the characteristics are identical or that any changes do not raise new questions of safety or effectiveness.

    CharacteristicPredicate Device (PICO 7)Subject Devices (PICO 7, 7Y, 14, Fluid Management Pack)Comparison/ "Performance"
    Indications For UseIndicated for patients who would benefit from a suction device (NPWT) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include: Chronic, Acute, Traumatic, Subacute and dehisced wounds, Partial-thickness burns, Ulcers (diabetic or pressure), Flaps and grafts, Closed surgical incisions. Suitable for use both in a hospital and homecare setting.Same as predicate, extended to PICO 7 (K180698), PICO 7Y (K182323), PICO 14 (K191760) and PICO Fluid Management Pack.Same
    Environment of useHospital, homeHospital, homeSame
    Patient PopulationPatients who would benefit from a suction device (negative pressure wound therapy)Patients who would benefit from a suction device (negative pressure wound therapy)Same
    MaterialsPU Plastic pump casingPU Plastic pump casingSame
    Single-use or ReusableSingle useSingle useSame
    Method of SterilizationPump, dressing, and fixation strips sterilized by ethylene oxidePump, dressing, and fixation strips sterilized by ethylene oxideSame
    BiocompatibilityDressing complies with ISO 10993Dressing complies with ISO 10993Same
    Type of PumpCustom designed "voice-coil" pump controlled by microprocessorCustom designed "voice-coil" pump controlled by microprocessorSame
    Electrical Safety TestingComplies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11Same
    Instructions for UseOld text for magnet warning, no MRI warning.Updated/added text for magnet warning and MRI warnings ("MR Unsafe. You must remove the PICO 7 pump from the dressing before entering the MRI suite. Do not bring PICO 7 into the MRI scan room. The device presents a projectile hazard."). Text amended in "Important Information Section" and "Glossary of Symbols" regarding magnet proximity warnings, specifically removing language about "death" and adding "As with all electrical medical equipment, failure to maintain appropriate distance may disrupt the operation of nearby medical devices."Changed (supported by additional testing)
    Warning Label on front of PumpPresentNo warning label on the front of the pump.Removed (supported by analysis)
    Magnetic Field Strength(Implied to be similar or stronger than current)Bench testing demonstrated that the average magnetic field strength of PICO Family pump is weaker than multiple everyday devices (e.g. laptops, phones, tablets, and headphones).Assessed (Supported changes)
    Post Market and Risk AnalysisN/ASystematic literature review of IMDs and magnetic interference conducted, confirming risk not as high as originally assessed. Comprehensive review of over 60,000 records from MAUDE database and company complaint data confirmed literature review outcome (no reported deaths/serious harms related to magnetic interference from PICO devices).Assessed (Supported changes)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set (for the current K202157 submission):
      • Magnetic Field Testing: This was bench testing, not involving human subjects. The sample size for devices tested is not specified, but it implies a measurement on the PICO Family pumps.
      • Post Market and Risk Analysis: Reviewed over 60,000 records from the MAUDE database (USA, retrospective) and Smith +Nephew PICO devices complaint data (provenance not specified, but likely global, retrospective).
    • Provenance: MAUDE data is from the United States. The literature review is not geographically specified but typically covers international publications. Smith +Nephew complaint data is likely global. All are retrospective analyses.
    • The document states that clinical performance data was provided in previously cleared 510(k)s. This suggests the initial clinical data was obtained for those initial clearances, not for this specific submission which focuses on minor labelling changes and supporting evidence for those changes through non-clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable to this 510(k) submission. This is not a study assessing diagnostic or clinical performance against expert ground truth. The "ground truth" for the current changes is based on:
      • Bench testing results (magnetic field strength)
      • Regulatory requirements (for labeling updates)
      • Analysis of real-world adverse event data (MAUDE database, company complaints) and scientific literature review. This analysis would have been performed by internal regulatory/safety/scientific experts, but the specifics of their number or qualifications are not provided as it's not a reader study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This document describes a substantial equivalence claim for a physical medical device, not an AI/imaging algorithm requiring adjudication of interpretations. The analysis of post-market data and literature would involve methods of synthesizing evidence and risk assessment, but not "adjudication" in the context of clinical interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a Negative Pressure Wound Therapy System, not an AI-powered diagnostic tool, so an MRMC study is not relevant to its clearance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device, not software or an algorithm with standalone performance metrics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the magnetic field testing, the "ground truth" is measured physical properties based on calibrated equipment.
    • For the post-market surveillance and risk analysis, the "ground truth" is derived from:
      • Reported adverse events/complaints (MAUDE database, company data). These are real-world reported events, which serve as evidence of actual outcomes (or lack thereof).
      • Published scientific literature on magnetic interference with implantable medical devices. This represents the established scientific understanding and reported cases in medical literature.

    8. The sample size for the training set

    • Not applicable. This device does not use machine learning where a "training set" would be relevant.

    9. How the ground truth for the training set was established

    • Not applicable.