Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K223019

    Validate with FDA (Live)

    Date Cleared
    2023-06-23

    (267 days)

    Product Code
    Regulation Number
    890.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K191568

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The sam 2.0 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, the local increase in circulation, and the relief of pain associated with limited mobility and function related to soft tissue injuries (e.g., knee osteoarthritis, chronic myofascial pain, and shoulder, elbow and ankle tendinopathy).

    Device Description

    The sam 2.0 Long Duration Ultrasound Device consists of ultrasound: Power Controller, Cables, Applicators, Coupling Patches. The sam 2.0 Long Duration Ultrasound Device contains firmware for logging usage time. There is no control function of this firmware. The Power Controller and cables can be used to power one or two Applicators simultaneously to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per Applicator. In single Applicator mode, the patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with a onetime use Coupling Patches. The system is intended to apply ultrasonic energy for a long duration (4 hours) to generate deep heat within body tissues. The system is intended for prescription home use after proper instruction from a healthcare professional.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from ZetrOZ Systems, LLC for their sam 2.0 Long Duration Ultrasound Device. It seeks to demonstrate substantial equivalence to a previously cleared predicate device, K191568. The key aspect of this submission is an expansion of indications for use for an identical device design.

    The document states that the substantial equivalence is supported by clinical performance testing which involved a systematic literature review and meta-analysis rather than a new clinical study. This means there is no traditional "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the way one would analyze a new AI medical device's performance.

    Based on the provided text, here's an analysis of the requested information:


    1. A table of acceptance criteria and the reported device performance

    The document does not present quantitative acceptance criteria in the way one would for a new AI or diagnostic device (e.g., target sensitivity/specificity). Instead, the "acceptance criteria" for this 510(k) submission are implicitly met by demonstrating substantial equivalence to the predicate device and by providing clinical evidence from published literature supporting the expanded indications.

    The 'performance' is described qualitatively as "efficacy of SAM therapy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation." For the expanded indications, it is stated: "the relief of pain associated with limited mobility and function related to the soft tissue injuries that have been clinically validated include knee osteoarthritis, shoulder, elbow and ankle tendinopathy, and chronic myofascial pain."

    Given that this is a K-submission for an ultrasound device with expanded indications and not a novel AI/diagnostic device, the table structure won't fit perfectly with typical AI acceptance criteria. The most relevant 'performance' is the equivalency of the device's technical specifications and the clinical efficacy of ultrasound therapy as demonstrated by the literature review.

    Table of Device Comparison (as presented in the document to demonstrate equivalence):

    Feature/ParameterSubject Device: sam 2.0 Long Duration Ultrasound DevicePredicate Device: sam 2.0 Long Duration Ultrasound Device (K191568)Comparison (Acceptance Criteria implicitly met by "Identical" or "Similar")
    Classification NameUltrasonic Diathermy DeviceUltrasonic Diathermy DeviceIdentical
    Service TypePhysical MedicinePhysical MedicineIdentical
    Classification21 CFR 890.530021 CFR 890.5300Identical
    ClassIIIIIdentical
    Product CodePFWPFWIdentical
    Indications For UseExtended (see document for full text)Original (see document for full text)Similar (expanded indications supported by clinical evidence)
    ManufacturerZetrOZZetrOZIdentical
    Console/Generator Dimensions6.10 cm L x 7.09 cm H x 1.88cm W6.10 cm L x 7.09 cm H x 1.88cm WIdentical
    Treatment Head Dimensions3.81 cm L x 3.30 cm W x 1.14 cm H3.81 cm L x 3.30 cm W x 1.14 cm HIdentical
    Console/Generator Weight0.10 kg0.10 kgIdentical
    Treatment Head Weight0.01 kg0.01 kgIdentical
    Power Supply120/240 VAC with 5V DC Input Power Jack and Lithium Battery Powered120/240 VAC with 5V DC Input Power Jack and Lithium Battery PoweredIdentical
    Leakage Current0.3 mA0.3 mAIdentical
    Crystal MaterialLead Zirconate-TitanateLead Zirconate-TitanateIdentical
    Technology of ultrasound generationPiezoelectricPiezoelectricIdentical
    Treatment Mode(s)Two discrete settings of power at same FrequencyTwo discrete settings of power at same FrequencyIdentical
    Beam TypeDivergentDivergentIdentical
    Transducer Diameter5 cm5 cmIdentical
    Acoustic Working Frequency and Accuracy3MHz ± 20%3MHz ± 20%Identical
    Effective Radiating AreaOne: 6 cm² ± 20% (Two: 10 cm²)One: 6 cm² ± 20% (Two: 10 cm²)Identical
    Beam Nonuniformity Ratio and AccuracyBNR: <5:1 ± 20%BNR: <5:1 ± 20%Identical
    Output ModeContinuous Wave - 100% duty cycleContinuous Wave - 100% duty cycleIdentical
    Maximum Timer Setting and Accuracy4 Hours +/- 1 minute4 Hours +/- 1 minuteIdentical
    Beam Maximum Intensity and Accuracy0.132 W/cm² ± 20%0.132 W/cm² ± 20%Identical
    Max Output Power and AccuracySingle: 0.65W ± 20%, Dual: 1.3W ± 20%Single: 0.65W ± 20%, Dual: 1.3W ± 20%Identical
    Max Effective Intensity and Accuracy0.264 W/cm² ± 20%0.264 W/cm² ± 20%Identical
    For Amplitude Modulated WavesNot Amplitude ModulatedNot Amplitude ModulatedIdentical
    Peak Temperature Rise vs. Time and Tissue Depth8°C at 1 cm, 6°C at 3 cm, 3°C at 5 cm (Max 4 hrs)8°C at 1 cm, 6°C at 3 cm, 3°C at 5 cm (Max 4 hrs)Identical
    Maximum Patient Contact Surface Temperature44 °C44 °CIdentical
    Applicator TypeUltrasound Coupling Patch, Up to two circular Applicators with 3 MHz outputUltrasound Coupling Patch, Up to two circular Applicators with 3 MHz outputIdentical
    Applicator Emitting Surface AreasUp to two circular Applicators: One: 5 cm², Two: 10 cm²Up to two circular Applicators: One: 5 cm², Two: 10 cm²Identical
    Coupling BandageABS Plastic with integrated coupling mediumABS Plastic with integrated coupling mediumIdentical
    Applicator Lens MaterialTPXTPXIdentical
    Environmental - Operating Temperature Range0°C to +50°C (32°F to +122°F)0°C to +50°C (32°F to +122°F)Identical
    Performance Standards21 CFR 1050.10, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-1121 CFR 1050.10, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11Identical
    SterilityNon SterileNon SterileIdentical
    BiocompatibilityYesYesIdentical
    Mechanical safetyYesYesIdentical
    Radiation safetyNot RadioactiveNot RadioactiveIdentical
    Software/FirmwareYes (firmware for logging usage time, no control function)Yes (firmware for logging usage time, no control function)Identical
    Output ChannelsTwo Independent Power ChannelsTwo Independent Power ChannelsIdentical

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This submission relies on a systematic literature review and meta-analysis to support the expanded indications, not a new clinical trial with a defined "test set" sample size. Therefore, there isn't a single "sample size" for a test set in the traditional sense.

    • Sample Size: Not applicable as a new test set was not generated. The evidence comes from accumulated studies in the literature.
    • Data Provenance: The systematic literature review was conducted using databases such as PubMed, EBSCOhost, Academic Search Complete, Google Scholar, and ClinicalTrials.gov. This implies global data sources, reflecting a diverse range of studies published in scientific literature. The studies themselves would have their own provenance (e.g., country of origin, retrospective/prospective design), but these details are not provided for the individual studies included in the review. The review itself is retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This question is not applicable as this submission is for an ultrasound therapy device and relies on clinical efficacy from published literature, not on expert adjudication of diagnostic images or interpretations to establish ground truth for a test set. There's no mention of experts involved in reviewing the literature for the purpose of establishing a "ground truth" for a device performance test. The "ground truth" is implied by the clinical outcomes reported in the accumulated scientific literature on SAM treatment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This question is not applicable as there was no new "test set" and no image-based diagnostic task requiring adjudication. The evidence comes from a systematic literature review and meta-analysis of clinical studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable. The device is an ultrasound therapy device, not an AI-assisted diagnostic or imaging interpretation tool that would involve human readers or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. The device is a therapeutic ultrasound device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for the expanded indications for this therapeutic device is based on outcomes data from various clinical studies summarized in the systematic literature review. These outcomes relate to "the relief of pain, the relief of muscle spasms, the treatment of joint contractures, the local increase in circulation," and specifically for the expanded indications, "relief of pain associated with limited mobility and function related to soft tissue injuries (e.g., knee osteoarthritis, chronic myofafascial pain, and shoulder, elbow and ankle tendinopathy)." The efficacy is based on the therapeutic effect reported in the aggregated clinical literature.

    8. The sample size for the training set

    • This question is not applicable. The device is not an AI algorithm requiring a training set. The clinical evidence is derived from existing scientific literature.

    9. How the ground truth for the training set was established

    • This question is not applicable. The device is not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211513

    Validate with FDA (Live)

    Date Cleared
    2021-08-18

    (96 days)

    Product Code
    Regulation Number
    890.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K191568

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The sam X I Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation

    Device Description

    The sam X1 Long Duration Ultrasound Device consists of:

    • Rechargeable Power Controller and Timer
    • Ultrasound Generating Applicators
    • Single use disposable accessories
    • Ultrasonic Coupling Patch
      The Power Controller can be used to power the Applicator to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per Applicator. In single Applicator mode, the patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with onetime use Ultrasonic Coupling Patches.
      The sam X1 Long Duration Ultrasound Device contains firmware that controls the timing display and inputs from the power button to activate/deactivate the device. There is no control function of this firmware.
      The system is intended for prescription home use to apply ultrasonic energy for a long duration (1 hours) to generate deep heat within body tissues and treat selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
    AI/ML Overview

    This document is a 510(k) Premarket Notification from ZetrOZ Systems, LLC for their sam X1 Long Duration Ultrasound Device. It seeks to demonstrate substantial equivalence to a predicate device (sam 2.0 Long Duration Ultrasound System, K191568).

    The document does not describe an AI/ML-based device or a study involving human readers and AI assistance for diagnostic purposes. Instead, it details a physical medical device (an ultrasonic diathermy device) used for therapeutic purposes (deep heat generation for pain relief, muscle spasms, joint contractures, and circulation increase).

    Therefore, I cannot provide the information requested in the prompt, as the prompt's questions pertain to acceptance criteria and studies for AI/ML diagnostic devices, such as those that might involve:

    1. A table of acceptance criteria and reported device performance (for AI metrics like sensitivity, specificity, AUC).
    2. Sample sizes and data provenance for test sets (relevant to AI model validation).
    3. Number and qualifications of experts for ground truth establishment (critical for AI imaging diagnostics).
    4. Adjudication methods for test sets (common in AI imaging studies).
    5. MRMC comparative effectiveness studies (specifically for AI assistance to human readers).
    6. Standalone algorithm performance.
    7. Types of ground truth (e.g., pathology, outcomes data - often used in AI validation).
    8. Training set sample size and ground truth establishment (fundamental to AI development).

    The provided text focuses on the physical and functional equivalence of the sam X1 device to its predecessor, particularly noting a smaller battery and a maximum 1-hour treatment duration compared to the predicate's 4 hours. The testing mentioned (power output, treatment time, battery conformance, electrical safety, usability, software, cybersecurity) are standard for medical device hardware and firmware safety and performance, not for AI model validation.

    In summary, as the provided document is for a non-AI medical device, the requested information regarding AI acceptance criteria and study design is not applicable and cannot be extracted from the text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1