510(k) Data Aggregation

    K Number
    K241080

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-21

    (124 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K190528

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Management of wounds including:

    • Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Chronic vascular ulcers
    • Diabetic ulcers
    • Trauma wounds (abrasions, lacerations, partial thickness burns, skin tears)
    • Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence)
    • Draining wounds
    Device Description

    Kerecis "Parvus™ is a lyophilized, terminally sterilized, acellular, particulate fish skin medical device comprised of biocompatible, non-crosslinked, resorbable, acellular fish skin (North Atlantic Cod) for wound management. The device is intended for single use only.

    The subject device is packaged in the following weights:
    100mg (4 cm2) 200mg (8 cm2) 500mg (19 cm²) 1000mg (38 cm²) 2,500mg (95 cm2 ) 3,000mg (114 cm2)

    AI/ML Overview

    This document is a 510(k) Premarket Notification summary for Kerecis® Parvus™, a wound dressing. It is not an AI/ML device, and therefore does not contain acceptance criteria for device performance related to AI/ML or a study proving those criteria are met. The document states that the Kerecis® Parvus™ is substantially equivalent to a predicate device, Kerecis MariGen Wound Extra (K190528), in terms of indications for use, intended use, raw material origin and composition, device performance, packaging material, and sterilization methods. The primary difference is an additional cutting and sieving step in the manufacturing process to convert the fish skin from intact sheets to fragmented pieces ≤2.0mm.

    Here's a breakdown of the requested information based on the provided document, noting that many items are not applicable (N/A) due to the nature of the device:

    1. A table of acceptance criteria and the reported device performance

      This document does not provide specific performance acceptance criteria or reported performance data in the traditional sense of a clinical or analytical study with defined metrics (e.g., accuracy, sensitivity, specificity). Instead, it asserts substantial equivalence to a predicate device based on material properties and manufacturing processes. The "Performance Characteristics" section for the subject device largely mirrors the predicate device, with differences primarily in the physical form (fragmented vs. sheets) and nominal sizes.

      CharacteristicAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (as described)
      Source OriginWild Caught Atlantic Cod FishWild Caught Atlantic Cod Fish (Same as predicate)
      Tissue sourceFish SkinFish Skin (Same as predicate)
      Nominal SizesFragmented, size controlled to ≤2.00mmIrregular shaped 3D fragmented fish skin, size controlled to ≤2.00mm and packaged in various weights (100mg to 3,000mg). (Differs from predicate's sheet sizes, but subject device is cut from predicate device sizes).
      PresentationLyophilized, sterilized, fragmented fish skinLyophilized, sterilized, Fragmented, fish skin in a Tyvek peel pouch. (Differs from predicate's sheet presentation due to cutting and sieving).
      PackagingTyvek Single and Double Peel PouchTyvek Single and Double Peel Pouch (Same as predicate)
      SterilizationEthylene OxideEthylene Oxide (Same as predicate)
      Sterility Assurance Level (SAL)SAL 10^-6^SAL 10^-6^ (Same as predicate)
      Endotoxin limit≤20 EU/device≤20 EU/device (Same as predicate)
      Shelf LifeAt least 1 year (matching current reported)1 year (Real-time shelf life testing in progress for the subject device; predicate has 3 years).
      BiocompatibilityDemonstratedAdditional testing for biocompatibility completed (for the subject device).
      Fragmented size characterizationDemonstratedAdditional testing for fragmented size characterization completed (for the subject device).
      Ethylene oxide residualsDemonstratedAdditional testing for ethylene oxide residuals completed (for the subject device).
      Metal contaminationDemonstratedAdditional testing for metal contamination completed (for the subject device).
      Protein analysisDemonstratedAdditional testing for protein analysis completed (for the subject device).
      BioburdenDemonstratedAdditional testing for bioburden completed (for the subject device).
      Residual limitsDemonstratedAdditional testing for residual limits completed (for the subject device).
    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      N/A. This is a 510(k) submission for a non-AI/ML medical device establishing substantial equivalence through material and manufacturing process comparisons, not a clinical study involving a test set for performance evaluation. The "additional testing" mentioned is likely laboratory bench testing on the device materials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      N/A. Not an AI/ML device requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      N/A. Not an AI/ML device requiring adjudication of a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      N/A. This device is not an AI/ML product and does not involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      N/A. This device is not an AI/ML product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      N/A. Not an AI/ML device requiring ground truth for performance evaluation. The substantial equivalence argument relies on material characteristics and manufacturing processes.

    8. The sample size for the training set

      N/A. This is not an AI/ML device with a training set.

    9. How the ground truth for the training set was established

      N/A. This is not an AI/ML device with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213904

    Validate with FDA (Live)

    Device Name
    Kerecis Oral
    Manufacturer
    Date Cleared
    2022-09-30

    (290 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K192612, K190528, K153364

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants;
    • Covering of implants placed in immediate extraction sockets;
    • Covering of implants placed in delayed extraction sockets;
    • Covering of bone defects after root resection and removal of retained teeth; and
    • Guided tissue regeneration procedures in periodontal and recession defects.
    Device Description

    The subject device is an acellular resorbable fish dermal matrix, intended for use in periodontal surgical procedures to aid in soft tissue and bone regeneration. It is obtained from cod fish skin by a standardized controlled manufacturing process, and supplied in terminally sterile peel-pouch packaging in the following solid sizes:

    • 15mm x 20mm
    • 20mm x 30mm
    • . 30mm x 40mm
      It is biocompatible, non-cross linked, and therefore resorbable, strong, flexible, and supports fixation by sutures and pins.
    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the Kerecis® Oral device. It details the device's characteristics, intended use, and a comparison to predicate and reference devices to demonstrate substantial equivalence.

    Based on the provided text, the device itself (Kerecis® Oral) is a biological material (acellular resorbable fish dermal matrix) used for various oral surgical procedures to aid in soft tissue and bone regeneration. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the bench testing and animal studies performed to demonstrate that the Kerecis® Oral device is substantially equivalent to existing predicate devices, thereby clearing it for market.

    Here's an analysis of the provided information to answer your questions:

    1. Table of acceptance criteria and the reported device performance

    The document doesn't present a single, explicit "acceptance criteria" table with pass/fail thresholds for the entire device in the way a software or AI model might have. Instead, it outlines a series of tests performed on the material properties and functional characteristics of the Kerecis® Oral device and compares them to predicate devices. The "acceptance" is implied by demonstrating substantial equivalence.

    Here's a table summarizing the comparison and performance points, where "acceptance" means demonstrating substantial equivalence or meeting/exceeding predicate device performance:

    Feature/TestAcceptance Criteria (Implied: Substantial Equivalence to Predicate/Reference)Reported Device Performance (Kerecis® Oral)
    Technological Characteristics
    Device NameN/A (Identification)Kerecis Oral
    Regulation21 CFR 872.3930 (Bone Grafting Material)21 CFR 872.3930 (Same Regulation as Predicate)
    Device ClassClass IIClass II (Same Class as Predicate)
    Product CodeNPLNPL (Same Product Code as Predicate)
    Device ClassificationBarrier, Animal Source, IntraoralBarrier, Animal Source, Intraoral (Same Classification as Predicate)
    Intended UseIncludes elements of predicate and reference devices, supported by performance testing.Covers localized gingival augmentation, implant covering (immediate and delayed), bone defect covering, GTR.
    Animal Origin MaterialSimilar origin to reference device (fish) or acceptable alternative to predicate (porcine).Fish: skin tissue, single layer sheet (Same animal species as reference)
    BiocompatibilityYes (demonstrated per ISO 10993 series)Yes (Biocompatibility testing performed per ISO 10993 series standards)
    Non-PyrogenicYes (demonstrated safe per ISO 10993 series)Yes (Materials-mediated pyrogenicity safe per ISO 10993 series standard)
    Tensile StrengthComparable to predicate device (4.6 MPa for predicate)14.3 MPa (Meets/Exceeds tensile strength value of the predicate device per ASTM D638)
    ResorbableYesYes (All devices are resorbable per comparative performance data)
    SizesEquivalent to predicate/reference sizes (15x20mm, 20x30mm, 30x40mm)15x20mm, 20x30mm, 30x40mm (Equivalent sizes by dimensional analysis)
    SterilizationTraditional Sterilization Methods (EO or Gamma)Ethylene Oxide (Traditional Sterilization Methods per ISO 11135 and ISO 11737-1. EO residual testing per ISO 10993-7)
    Sterility Assurance Level (SAL)10^-610^-6 (Equivalent SAL per ISO 11137 and ISO 11737)
    Shelf Life3 years3 years (Equivalent shelf life per ASTM F1980 and Q5C (R2)[ICH])
    Mode of ActionFixationFixation (Equivalent mode of action per ANSI/AAMI/ISO 7198 and ASTM F-1839-08)
    Performance Testing - Bench
    Morphology Observation (H&E, SEM)Rich in collagen, porous, favoring cellular infiltration, preserved collagen structure.Rich in collagen and porous, favoring cellular infiltration. SEM shows equivalent preserved collagen structure.
    Cellular Ingrowth Comparison (Fibroblast)Favorable cellular infiltration after 14 days.Favorable cellular infiltration of fibroblasts after 14 days.
    Heavy Metal AnalysisAcceptable limits per ICH guidelines (Q3D Elemental Impurities).Limits of Cd, Pb, As, Hg acceptable under ICH guidelines.
    Dimensional ValidationQuality limits for capability index > 1.33.All tested parameters met or exceeded the set goal (>1.33) on three lots.
    Stability in Simulated Physiological Env.Structurally stable compared to predicate, dissolves slower at neutral pH.Structurally more stable than predicate, dissolved slower than predicate at neutral pH 7.
    Suture Pull-Out StrengthMeets or exceeds predicate with >95% confidence.Meets or exceeds predicate with >95% confidence (Consistent with ANSI/AAMI/ISO 7198).
    Pin Pull-Out StrengthExceeds predicate with >95% confidence.Exceeds predicate with >95% confidence (Consistent with ASTM F-1839-08).
    Compression (Peak-Load, Load-at-Break, etc.)Meet or exceed predicate with >95% confidence.Meet or exceed predicate with >95% confidence (Consistent with ASTM-F-1306).
    User Evaluation (Cutting and Shaping)Favorable usability, substantially equivalent to predicate.Favorable usability, substantially equivalent to predicate for cutting and shaping (via questionnaire).
    PackagingCompliant with ISO 11607-series, ASTM F88 and ASTM F1886.Compliant.
    Animal Origin and Viral InactivationCompliant with ISO 22422 series.Compliant.
    Performance Testing - Animal
    New Bone Formation, Xenograft Resorption, Soft Tissue InfiltrationNo qualitative or significant differences compared to predicate.No noticeable or significant differences between the two membranes at any time point (Histologic and micro-CT volumetric analysis).

    2. Sample sizes used for the test set and the data provenance

    • Bench Testing: The specific sample sizes for each bench test (e.g., number of samples for tensile strength, compression, dimensional validation) are not explicitly stated in the provided text. It mentions "representative product samples" and "three lots" for dimensional validation. The provenance is implied to be laboratory testing of the manufactured Kerecis® Oral device.
    • Animal Testing: The sample size for the animal study is a "canine (beagle bilateral mandibular bony defect model)." This implies multiple beagle dogs, but the exact number is not specified. The study was conducted in AAALC accredited facilities in compliance with GLP (Good Laboratory Practices) 21 CFR §58, suggesting a controlled, prospective study. The location (country) is not specified, but GLP compliance implies a rigorous study design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • For the animal study, the evaluation involved "gross pathology and histological, and quantitative micro-CT volumetric analysis." This type of analysis would typically be performed by veterinary pathologists and possibly radiologists/imaging specialists. However, the number and specific qualifications of these experts are not mentioned in the text.
    • For the user evaluation of cutting and shaping, it was done "using a questionnaire." This implies that "ground truth" was established by user feedback, not by experts in the context of diagnostic accuracy. The number and qualifications of these users are not stated.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • None explicitly mentioned. For the animal studies, "no qualitative or significant differences was observed" suggests direct comparison by the evaluators rather than an explicit adjudication process among multiple independent reviewers, though this is not definitively stated. For the bench tests, the data itself (e.g., tensile strength, dimensional measurements) would be objective, not typically requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This product is a biological medical device (a tissue graft), not a diagnostic AI device that assists human readers. Therefore, the concept of "human readers improving with AI assistance" is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device, not an algorithm. The "performance" assessment is based on its material properties and biological interaction in an in vitro and in vivo (animal) setting, not an algorithm's output.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For Bench Testing: The "ground truth" is defined by objective material property measurements (e.g., tensile strength in MPa, dimensional measurements in mm, pH and conductivity values) and established international standards (e.g., ISO, ASTM, ICH).
    • For Animal Testing: The "ground truth" was established through histologic and micro-CT volumetric analysis of animal tissues collected at different time points post-surgery. This involves pathological assessment (histology) and quantitative imaging analysis (micro-CT) to assess bone formation, xenograft resorption, and soft tissue infiltration. While the specific experts aren't named, this is a form of expert assessment of biological outcomes.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set" in the context of an AI/ML algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. As a physical device, there is no "training set" or "ground truth for the training set."
    Ask a Question

    Ask a specific question about this device

    K Number
    K192612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-13

    (420 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K190528, K153364

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerecis Gingiva Graft is indicated for:

    • Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants.
    Device Description

    The subject device is fish skin medical device indicated gingiva augmentation.
    The subject device is obtained from cod fish skin by a standardized controlled manufacturing process and supplied in a peel-pouch terminally sterile packaging in the following rectangular sizes:

    • 15mm x 20mm
    • 20mm x 30mm
    • 30mm x 40mm
      The subject device becomes completely integrated into the surrounding tissue over time, with corresponding new host tissue deposition. The physical properties of the subject device allow cellular ingrowth for augmentation of keratinized tissue.
      The subject device is biocompatible, non-crosslinked, bioresorbable, strong, pliable and supports fixation by sutures.
    AI/ML Overview

    The provided text describes the Kerecis Gingiva Graft device, a collagen membrane intended for localized gingival augmentation. This is a 510(k) submission, meaning the device seeks clearance based on its substantial equivalence to a legally marketed predicate device, not necessarily on meeting quantitative acceptance criteria established by the FDA for novel devices.

    Therefore, the information typically requested in questions 1 through 9 (acceptance criteria, sample sizes, expert ground truth, MRMC studies, standalone performance, training sets) is not applicable in the context of this 510(k) summary, as it describes a different type of regulatory submission process. The study presented here aims to demonstrate equivalence, not to quantify the device's performance against pre-defined thresholds.

    However, I can extract information related to the "study that proves the device meets the acceptance criteria" in the context of proving substantial equivalence to the predicate device.

    Here's the relevant information based on the provided text, reinterpreting "acceptance criteria" as "criteria for demonstrating substantial equivalence" for this 510(k) submission:

    1. A table of (Substantial Equivalence) Criteria and the Reported Device Performance:

    The document establishes substantial equivalence by comparing the Kerecis Gingiva Graft (subject device) to the MUCOGRAFT® Collagen Matrix (predicate device) and other reference devices across various characteristics and performance tests. Rather than explicit "acceptance criteria" with numerical thresholds, the "performance" is the demonstration of comparability or superiority to the predicate device.

    Characteristic / Performance ElementSubstantial Equivalence Criterion (Implicit)Reported Device Performance
    Intended UseSame as, or subset of, predicate device."Subset of the intended use of the predicate device." (Kerecis: "a biocompatible, sterile collagen membrane intended for augmentation and regeneration of soft tissue in oral surgical settings." Predicate: includes broader indications like "guided tissue regeneration and multiple oral tissue defect regeneration in oral surgical settings.")
    Indications for UseSame as, or subset of, predicate device."Subset of the indications of the predicate device." (Kerecis: "Localized gingival augmentation to increase Keratinized tissue (KT) around teeth and implants." Predicate: includes this, plus "Covering of implants placed in immediate extraction sockets," "Alveolar ridge reconstruction for prosthetic treatment," and other GTR procedures.)
    Regulation, Product Code, ClassSame as predicate."Same as predicate." (21 CFR 872.3930, NPL, Class II)
    BiocompatibilityMeets ISO 10993 series standards. Comparable to predicate."Yes" (Same as predicate). Leveraged testing from applicant's own predicate devices (K190528 and K153364) for Cytotoxicity, Sensitization, Irritation, Acute/Subacute/Sub-chronic/Chronic Toxicity, Genotoxicity, Implantation, Materials-Mediated Pyrogenicity, Carcinogenicity.
    NON-PyrogenicYes."Yes" (Same as predicate). Endotoxin validation (<20 EU/device) per LAL turbidimetric kinetic method following ISO 10993-11.
    ResorbableYes."Yes" (Same as predicate).
    Sterilization & SALValidated sterilization method, SAL 10-6.Ethylene Oxide sterilization with SAL 10-6. "Traditional Sterilization Method." Ethylene Oxide residual test following ISO 10993-7.
    Shelf LifeValidated, comparable to predicate."3 years" (Same as predicate). Validated per ASTM F1980 and Q5C (R2)[ICH] using accelerated and real-time aged samples. Packaging per ISO 11607-series, ASTM F88 and ASTM F1886.
    Animal Origin MaterialSafe and suitable (via viral inactivation, etc.).North Atlantic Cod fish: skin tissue, single layer sheet. Predicate is Porcine skin and connective tissue, double layer sheet. "Different animal source, same anatomical tissue." Viral inactivation per ISO 22422 series.
    Morphology (H&E, SEM)Rich in collagen, porous, favoring cellular infiltration, preserved collagen structure.Both subject and predicate "rich in collagen and porous, therefore favoring cellular infiltration." SEM "shows equivalent preserved collagen structure." Cross section shows porous surfaces allowing tissue adherence and promoting cellular ingrowth.
    Cellular IngrowthFavorable cellular infiltration in vitro.Both materials "showed favorable cellular infiltration of fibroblasts after 14 days which is a key component for tissue augmentation and re-epithelization of defected keratinized tissue."
    Tensile StrengthComparable to predicate."Determined to be comparable to the predicate device measured by ultimate tensile strength."
    Heavy Metal AnalysisWithin ICH guidelines: Q3D Elemental Impurities.Limits of cadmium (Cd), lead (Pb), arsenic (As), and mercury (Hg) were "acceptable under the ICH guidelines."
    Stability (Simulated Physiological)Stable in artificial saliva buffer, comparable to predicate's effect on pH/conductivity.Buffer incubated with both products was "stable over time." Subject device "raises the pH slightly," predicate "decreases it slightly." Subject device "is structurally more stable than the predicate device since it dissolved slower than the predicate device at neutral pH 7."
    Suture Pull-Out StrengthMeets or exceeds predicate with 95% confidence."Meets or exceeds that of the predicate with a confidence of greater than 95%." Deemed equivalent for oral surgery and gingival augmentation.
    Pin Pull-Out StrengthExceeds predicate with 95% confidence."Exceeds that of the predicate at a confidence level of greater than 95%." Deemed equivalent for oral surgery and gingival augmentation.
    CompressionCompressive Peak-Load, Load-at-Break, Probe Penetration-at-Break, Energy-to-Break meet or exceed predicate with 95% confidence."Meet or exceed those of the predicate device, with a confidence of greater than 95%." Deemed equivalent for oral surgery and gingival augmentation.
    User Evaluation (Cutting/Shaping)Favorable usability, substantially equivalent to predicate.Evaluated by four dental clinicians using questionnaire. "Showed a favorable usability that was substantially equivalent to the predicate for cutting and shaping the device for use as a dental barrier membrane." Assessed ease of placement, stability, robustness, handling satisfaction.
    Animal (Canine) Study - SafetyAcceptable safety profile for the subject device."All animals remained in good general health throughout the duration of the study and gross pathological findings suggest an acceptable safety profile for the subject device."
    Animal (Canine) Study - EfficacyNo statistically significant difference from predicate in key outcomes.For all parameters (KT thickness, KT length, root coverage, inflammation, membrane degradation), "no statistically significant difference was detected between the subject and predicate devices after 90 days of healing." Both devices "were able to increase keratinized tissue around teeth."
    Clinical Study - SafetyNo adverse effects or complications."There were no adverse effects or complications during the duration of the clinical study."
    Clinical Study - EfficacyIncreased KT, similar to predicate's published results, long-term stability."An increased in the in the width of the KT was noticed for all treated sites. This average gain has been reported as adequate to maintain long-term periodontal health." "The device showed stability as the KT slowly replaced the piscine xenograft." Results "similar to published results reported for the predicate device."

    2. Sample sizes used for the test set and the data provenance:

    • Bench Testing: "representative products" used for various tests. Specific numbers for each test are not provided, only statements like "with a confidence of greater than 95%" (implying statistical analysis on a certain sample size).
    • Animal Study: 10 canines (3 terminated at 30 days, 3 at 60 days, 4 at 90 days). Data provenance: Preclinical testing in canines (implied prospective). No country of origin specified.
    • Clinical Study: 6 human individuals. Data provenance: Open label, noncomparative study (prospective). No country of origin specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Bench Testing (User Evaluation): Four "dental clinicians" evaluated the device for cutting and shaping. Their specific qualifications (e.g., years of experience, specialization) are not detailed.
    • Animal Study: Outcomes included "Histomorphometric measurements" and "Histological analysis." This implies expert pathologists/histologists reviewed the samples, but the number and qualifications are not specified.
    • Clinical Study: Ground truth was the observation of clinical outcomes (KT increase) by clinicians. The number and qualifications of the clinicians performing assessments in the clinical study are not specified beyond the study being "clinical."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • No specific adjudication method is mentioned for any of the studies (bench, animal, or clinical). For a 510(k) submission focused on equivalence, such detailed adjudication methods are less commonly emphasized compared to novel device approvals.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical gingival graft, not an AI or imaging diagnostic device. Therefore, MRMC studies and the concept of human readers improving with AI assistance are not relevant to this product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical gingival graft. There is no "algorithm" or standalone software component. Its performance is inherent to its physical properties and biological interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Bench Testing: Laboratory measurements (tensile strength, compression, etc.), and subjective "user evaluation" feedback from dental clinicians.
    • Animal Study: Histomorphometric measurements (objective), histological analysis (pathology, expert interpretation), in vivo and postmortem measurements by periodontal probe.
    • Clinical Study: Clinical observations of "increased in the width of the KT" and overall "stability" reported by the study investigators/clinicians as outcomes data.

    8. The sample size for the training set:

    • Not applicable. This is a physical medical device, not a machine learning or AI algorithm, so there is no concept of a "training set" in the traditional sense for algorithm development. The "training" here would be the manufacturing process and quality control methods to ensure consistent device properties.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as point 8.