Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K213799

    Validate with FDA (Live)

    Date Cleared
    2022-07-12

    (218 days)

    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161531, K131414, K180881, K210906, K193391, K182979

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

    • · BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only. CCO using FloTrac is intended for adult patients only;
    • · C.O. monitoring and A-Fib are intended for adult patients only;
    • · ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
    • · rSO2 monitoring is intended for use in individuals greater than 2.5kg.

    The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.

    The BeneVision N1 Patient Monitor is intended for monitoring, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO₂), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

    • PAWP, PNP, and PNC are intended for adult and pediatric patients only;
    • A-Fib is intended for adult patients only;

    The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment, whereas N1 configured with WMTS technology can be used inside the hospital only. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.

    Device Description

    The subject BeneVision N Series Patient Monitors includes six monitors:

    • . BeneVision N12 Patient Monitor
    • BeneVision N15 Patient Monitor
    • BeneVision N17 Patient Monitor ●
    • BeneVision N19 Patient Monitor ●
    • BeneVision N22 Patient Monitor ●
    • BeneVision N1 Patient Monitor

    Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring.

    AI/ML Overview

    This document is a 510(k) Summary for the Mindray BeneVision N Series Patient Monitors, which focuses on demonstrating substantial equivalence to a previously cleared predicate device (K202405).

    The information provided primarily details the device's technical specifications and comparisons to a predicate device, rather than a full study proving the device meets acceptance criteria for a specific medical condition or AI diagnostic output.

    Therefore, I cannot fully answer all parts of your request as the document does not contain the detailed clinical study results (like sample sizes for test sets, number of experts for ground truth, adjudication methods, MRMC studies, or specific AI performance metrics) that would typically be found for a device requiring those types of studies (e.g., an AI-powered diagnostic tool).

    However, I can extract the relevant information regarding the device's functional and technical performance as demonstrated in this 510(k) submission.

    Here's a breakdown of what can be inferred and what is missing:


    Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" in this context are related to meeting the performance specifications of the predicate device and relevant consensus standards. The "reported device performance" is demonstrated through functional and system-level testing, ensuring the device meets its accuracy specifications for the various physiological parameters it monitors.

    Table of "Acceptance Criteria" (Implied Specifications) and "Reported Device Performance" (Conformance):

    Parameter / FeatureImplied Acceptance Criteria (from Predicate/Standards)Reported Device Performance (as stated in document)
    General Device PerformanceReliability, accuracy, and safety equivalent to the predicate device (K202405). Compliance with general controls and specific standards."Mindray conducted functional and system level testing on the subject device. The testing provided an evaluation of the performance of the device relevant to each of the differences between the subject device and the predicate device. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate." "The results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device."
    ECG (HR)HR Measurement range: 15350 bpm (neonate, pediatric), 15300 bpm (adult); Accuracy: ±1 bpm or ±1%, whichever is greater.No specific deviation reported from these specifications. The device supports intelligent arrhythmia alarms, adjustment of QT calculation, SVT and SVCs/min high arrhythmia alarm, and Multi-lead ECG synchronization analysis.
    ECG (ST)ST Measurement range: -2.0mV~+2.0mV; Accuracy: -0.8mV~+0.8mV, ±0.02mV or ±10%, whichever is greater, other range: not specified.No specific deviation reported from these specifications.
    ECG (QT)QT Measurement range: 200~800ms; Accuracy: ±30ms.No specific deviation reported from these specifications. Adjustment of QT calculation is a new feature.
    Respiration Rate (Resp)Measurement range: Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm. Accuracy: 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater; 0 to 6 rpm: Not specified.No specific deviation reported from these specifications.
    Temperature (Temp)Measurement range: 0 to 50°C (32 to 122°F). Accuracy: ±0.1°C or ±0.2°F (without probe).No specific deviation reported from these specifications.
    Pulse Oxygen Saturation (SpO2)Mindray SpO2 module: Range: 0~100% Accuracy: 70%100%: ±2% ABS (Adult/pediatric); ±3% ABS (neonate); 069%: not specified. Masimo SpO2: With motion 70%~100%: ±3% ABS. Nellcor SpO2: Accuracy: 70%~100%: ±2% ABS (Adult/pediatric); ±3% ABS (neonate).No specific deviation reported from these specifications.
    Pulse Rate (PR)Accuracy: ±3 bpm (Mindray SpO2), ±3 bpm without motion, ±5 bpm with motion (Masimo SpO2); 20250 bpm ±3 bpm; 251300 bpm, not specified (Nellcor SpO2); ±1 bpm or ±1%, whichever is greater (IBP).No specific deviation reported from these specifications.
    Non-invasive Blood Pressure (NIBP)Max mean error: ±5mmHg; Max standard deviation: 8mmHg. PR: ±3 bpm or ±3%, whichever is greater.No specific deviation reported from these specifications.
    Invasive Blood Pressure (IBP)Accuracy of module: ±2% or ±1mmHg, whichever is greater (without sensor).No specific deviation reported from these specifications. New feature: Artifact flag of Arterial Blood Pressure (ABP) shields alarms monitoring support. BeneVision N1 now supports up to 4 IBP channels (vs. 2 in predicate).
    Cardiac Output (C.O.)Accuracy: ±5% or ±0.1L/min, whichever is greater. TB, TI: ±0.1°C (without sensor).No specific deviation reported from these specifications.
    Continuous Cardiac Output (FloTrac)(Not supported in predicate) Measurement range: CCO: 1.0-20.0L/min; Reproducibility: ±6% or 0.1 L/min, whichever is greater. PR:0~220bpm, Arms ≤3bpm. Live pressure display range: -34 to 312 mmHg. MAP/DIA/SYS display range: 0-300 mmHg. Accuracy: ±4% or ±4 mmHg, whichever is greater, in the range of -30 mmHg to 300 mmHg.Added feature with stated performance specifications as listed. (This is a new feature compared to the predicate, and its performance data is listed as its "acceptance criteria" and "reported performance" upon introduction.)
    Carbon Dioxide (CO2)Sidestream: CO2: 040 mmHg: ±2mmHg; 4176 mmHg: ±5% of reading; 7799 mmHg: ±10% of reading; 100150mmHg: ±(3mmHg + 8% of reading). awRR: <60rpm, ±1rpm, 60150rpm, ±2rpm. Microstream: CO2: 038mmHg: ±2mmHg; 3999mmHg: ±5% of the reading+0.08% of (the reading-38). awRR: 070rpm: ±1rpm, 71120rpm: ±2rpm, 121150rpm: ±3rpm. Mainstream: CO2: 040mmHg: ±2mmHg, 4170mmHg: ±5% of the reading, 71100mmHg: ±8% of the reading, 101150mmHg: ±10% of the reading; awRR: ±1rpm.No specific deviation reported from these specifications. For BeneVision N1: component changes for Sidestream CO2 1.0 and 2.0 modules.
    Oxygen (O2)Accuracy: 025%, ±1%; 2680%, ±2%; 81~100%, ±3%.No specific deviation reported from these specifications.
    Anesthetic Gas (AG)Various specific accuracies for N2O, HAL, ENF, ISO, SEV, DES based on concentration ranges. awRR: 2~60rpm, ±1rpm, >60rpm, not specified.No specific deviation reported from these specifications. Internal electronic component changes for the AG module.
    Electroencephalograph (EEG/aEEG)Frequency response: 0.5Hz ~ 50 Hz (-3 dB). Input range: 4 mVpp. DC offset: ±500 mV. CMRR: ≥ 100 dB @ 51 KΩ and 60 Hz. Noise level: ≤ 0.5 uV rms (1Hz to 30 Hz). Differential input resistance: > 15 MΩ @ 10 Hz. Electrode resistance: 0 to 90 KΩ, resolution: ±1 KΩ or 10%, whichever is the greater.Added EEG-1 module and aEEG module, with their performance specifications stated as meeting these criteria.
    A-Fib Overview(Not supported in predicate) The A-Fib overview function only collects atrial fibrillation, atrial fibrillation with rapid ventricular rate, atrial fibrillation with R-R long interval and other related events, and the A-Fib overview screen displays the A-Fib specifications.This is a newly added feature with its intended functionality described.
    Electromagnetic Compatibility & Electrical SafetyCompliance with relevant standards (ANSI/AAMI ES 60601-1:2005, IEC 60601-1-2:2014).Assessed for conformity and found to comply with ANSI/AAMI ES 60601-1:2005/(R) 2012 and IEC 60601-1-2:2014. Also meets IEC 60601-1-8:2020 for alarm systems.
    Software Verification & ValidationMeets FDA Guidance for Software Contained in Medical Devices."Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance... Verification... was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product."

    Detailed Study Information (Based on Document Content):

    1. Sample sizes used for the test set and the data provenance:

      • The document states that "functional and system level testing" and "bench testing" were conducted.
      • However, no specific sample sizes for test sets (e.g., number of patients, number of data recordings) are provided for any of the performance evaluations.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The tests described are generic "bench testing" to ensure compliance with technical specifications and standards, not clinical studies.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable / Not provided. This document describes engineering and bench testing against pre-defined technical specifications and industry standards for physiological measurement accuracy. It does not describe a clinical study involving human experts establishing ground truth for diagnostic interpretation (e.g., for an AI algorithm interpreting medical images).
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable / Not provided. Same reason as above.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This device is a patient monitor. It detects physiological parameters and provides alarms, and some interpretations of ECG (e.g., 12-lead ECG interpretation, arrhythmia detection). It is not an AI-assisted diagnostic device in the sense of image interpretation for which MRMC studies are typically performed. The document details that "optimized auditory ALARM SIGNALS" and "alarm highlight" were added, suggesting improvements to the human-device interface, but not a formal MRMC study on diagnostic improvement.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The document implies that algorithms for ECG (Mindray or Mortara algorithm for arrhythmia and ST-segment analysis) are embedded in the device. The listed accuracy specifications for these measurements (e.g., HR, ST, QT) reflect the standalone performance of these measurement algorithms and sensors against established benchmarks. However, a formal "standalone study" with detailed methodology, distinct from the general bench testing, is not specifically described or provided with separate results. The performance data listed (e.g., accuracy for HR, ST, QT) serves as the "standalone" performance verification for these integrated functionalities.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For physiological measurements (ECG, SpO2, NIBP, etc.), the "ground truth" would typically refer to reference measurement devices or calibrated simulators used during bench testing to verify the accuracy of the monitor's readings against a known, accurate value.
      • For the ECG interpretation (e.g., 12-lead ECG interpretation, arrhythmia detection), the ground truth for the algorithms would have been established during their development and previous clearance processes (Mindray or Mortara algorithms). This document focuses on demonstrating that the integration and revised features maintain that established accuracy rather than re-proving the core algorithms.
    7. The sample size for the training set:

      • Not provided. This document pertains to the 510(k) clearance of updates to an existing patient monitor series. It does not detail the development or training of new AI/ML algorithms, which would typically involve substantial training datasets. The ECG algorithms (Mindray or Mortara) were presumably "trained" (or developed and validated) previously as part of their initial predicate clearances.
    8. How the ground truth for the training set was established:

      • Not provided. (See point 7). For existing algorithms like Mortara or Mindray ECG algorithms, ground truth for their original development would likely have been established using large, diverse ECG databases with expert cardiologist interpretations and/or correlation with clinical outcomes where relevant. This particular 510(k) document is concerned with demonstrating equivalence and continued performance with minor changes, not the original algorithm development.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202336

    Validate with FDA (Live)

    Device Name
    Patient Monitor
    Date Cleared
    2021-01-23

    (159 days)

    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K160981,K182075,K182979

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitors are intended to be used for monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple physiological parameters of adults and pediatrics. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), and Anaesthesia gas(AG). The arrhythmia detection and ST Segment analysis are intended for adult patients. The monitors are not intended for MRI environments.

    Device Description

    The iM series Patient Monitor including iM50, iM60, iM70 and iM80 can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate.

    AI/ML Overview

    Based on the provided text, the device in question is a Patient Monitor (Model: iM50, iM60, iM70, iM80), which monitors various physiological parameters. The document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a standalone study for a novel AI device. Therefore, much of the requested information regarding AI-specific evaluation (e.g., sample size for AI test sets, expert adjudication, MRMC studies, AI effect size, ground truth establishment for training) is not applicable or not present in this 510(k) summary.

    However, I can extract information related to the device's self-contained performance testing and regulatory acceptance criteria.


    Acceptance Criteria and Device Performance for Patient Monitor (iM Series)

    The document primarily relies on bench testing and software verification and validation to demonstrate that the iM series Patient Monitor meets its accuracy specifications and relevant consensus standards, thereby establishing substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present explicit "acceptance criteria" in a quantitative table for this specific device in the manner typically seen for novel AI models. Instead, it compares the technical specifications of the subject device to a predicate device and states that "the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards."

    The comparison table (pages 5 & 6) implicitly indicates that "acceptance" for the subject device's performance corresponds to ranges/specifications that are identical or comparable to the cleared predicate device.

    Parameter/FeatureAcceptance Criteria (Predicate Device K192514)Reported Device Performance (Subject Device iM50, iM60, iM70, iM80)Comparison Result
    ECG Module
    Lead Mode3, 5, 6, 10 Electrodes3, 5, 6, 10 ElectrodesSame
    Arrhythmia analysesASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNPASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNPSame
    ST value Measurement Range-2.0 mV to +2.0 mV-2.0 mV to +2.0 mVSame
    Pace Pulse Indicator (Amplitude)±2 mV to ±700 mV±2 mV to ±700 mVSame
    Pace Pulse Indicator (Width)0.1 ms to 2.0 ms0.1 ms to 2.0 msSame
    Pace Pulse Indicator (Ascending time)10 $μ$s to 100 $μ$s10 $μ$s to 100 $μ$sSame
    PVC Range (ADU)0 to 300 PVCs/min0 to 300 PVCs/minSame
    PVC Range (PED/NEO)0 to 350 PVCs/min0 to 350 PVCs/minSame
    HR Measurement Range (ADU)15 bpm to 300 bpm15 bpm to 300 bpmSame
    HR Measurement Range (PED/NEO)15 bpm to 350 bpm15 bpm to 350 bpmSame
    QT Range200 ms ~ 800 ms200 ms ~ 800 msSame
    QTc Range200 ms ~ 800 ms200 ms ~ 800 msSame
    $\Delta$ QTc Range-600 ms ~ 600 ms-600 ms ~ 600 msSame
    RESP Module
    Principle of OperationImpedance between RA-LL, RA-LAImpedance between RA-LL, RA-LASame
    Measurement Range (Adult)0 to 120 rpm0 to 120 rpmSame
    Measurement Range (Pediatric/neonate)0 to 150 rpm0 to 150 rpmSame
    NIBP Module
    TechniqueOscillometryOscillometrySame
    Measurement Range (Systolic Adult)25-29025-290Same
    Measurement Range (Systolic Pediatric)25-24025-240Same
    Measurement Range (Systolic Neonate)25-14025-140Same
    Measurement Range (Diastolic Adult)10-25010-250Same
    Measurement Range (Diastolic Pediatric)10-20010-200Same
    Measurement Range (Diastolic Neonate)10-11510-115Same
    Measurement Range (Mean Adult)15-26015-260Same
    Measurement Range (Mean Pediatric)15-21515-215Same
    Measurement Range (Mean Neonate)15-12515-125Same
    PR from NIBP Measurement Range40 bpm to 240 bpm40 bpm to 240 bpmSame
    SpO2 Module
    SpO2 Measurement Range0% to 100%0% to 100%Same
    Pulse Rate Measurement Range25 to 300 bpm25 to 300 bpmSame
    Temperature Module
    Number of channels22Same
    Measurement Range0 °C to 50 °C (32 °F to 122 °F)0 °C to 50 °C (32 °F to 122 °F)Same
    IBP Module
    PA/PAWP Range-6 to +120 mmHg-6 to +120 mmHgSame
    CVP/RAP/LAP/ICP Range-10 to +40 mmHg-10 to +40 mmHgSame
    P1/P2 Range-50 to +300 mmHg-50 to +300 mmHgSame
    C.O. Module
    TechniqueThermodilution TechniqueThermodilution TechniqueSame
    C.O. Measurement Range0.1 to 20 L/min0.1 to 20 L/minSame
    TB Range23 °C to 43 °C (73.4 °F to 109.4 °F)23 °C to 43 °C (73.4 °F to 109.4 °F)Same
    TI Range-1 °C to 27 °C (30.2 °F to 80.6 °F)-1 °C to 27 °C (30.2 °F to 80.6 °F)Same
    CO2 Module
    Intended PatientAdult, pediatric, neonatalAdult, pediatric, neonatalSame
    Measure ParametersEtCO2, FiCO2, AwRREtCO2, FiCO2, AwRRSame
    CO2 Measuring Range0 mmHg to 150 mmHg (0% to 20%)0 mmHg to 150 mmHg (0% to 20%)Same
    AwRR Measuring Range2 rpm to 150 rpm2 rpm to 150 rpmSame
    AG Module (EDAN G7)Not present in primary predicateCO2, N2O, O2, HAL, ISO, ENF, SEV, DES, AwRR, MACDifferent (but similar to referenced predicate K160981)
    WI-FI
    IEEE802.11a/b/g/n802.11a/b/g/nSame
    Frequency Band2.4 GHz ISM band & 5 G ISM band2.4 GHz ISM band & 5 G ISM bandSame
    Power Supply
    AC requirement100-240V, 50/60Hz100-240V, 50/60HzSame
    Rechargeable BatteryYesYesSame

    Notes on the 'AG Module': The document explicitly states for the AG (Anesthesia Gas) module that its "indication is not present in the primary predicate, but is present in Edan Patient Monitor V series K160981." This implies that while it differs from the immediate primary predicate, it is substantially equivalent to a different, already cleared, predicate device from the same manufacturer.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify a distinct "test set" sample size in terms of patient data or number of tests. The performance data section refers to "functional and system level testing" and "bench testing." This implies testing against specifications and standards rather than a clinical dataset of a specific size.
    • Data Provenance: Not specified. Given it's a bench test, it would typically be conducted in a laboratory setting. There's no mention of country of origin for test data, nor whether it's retrospective or prospective patient data, as clinical data was deemed "Not applicable."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a physiological monitor, not an AI diagnostic device requiring expert consensus for ground truth on images or signals. The "ground truth" for the device's performance would be derived from calibrated measurement references and established engineering principles in bench testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As above, this is not an AI diagnostic device relying on human expert review for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physiological monitor, not an AI-assisted diagnostic tool that requires human readers for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device's performance in terms of its physiological measurements and alarm detection is inherently "standalone" in that it performs these functions without direct human intervention in the measurement process itself, generating outputs for healthcare professionals. The bench testing performed would be considered evaluating this standalone performance against technical specifications and standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance (bench testing), the "ground truth" is based on:

    • Calibration standards: Using known, precise inputs (e.g., electrical signals simulating ECG, precise pressures for NIBP, known gas concentrations for CO2/AG) to verify the accuracy of the device's measurements.
    • Consensus Standards: Adherence to recognized international standards for medical electrical equipment (e.g., IEC 60601 series, ISO 80601 series). These standards define acceptable performance limits and test methodologies.

    8. The sample size for the training set

    Not applicable. This document does not describe an AI/ML device that requires a "training set" in the conventional sense. The device's algorithms are likely based on established physiological signal processing, not deep learning models trained on large datasets.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for an AI/ML model described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1