    K Number
    K242589

    Validate with FDA (Live)

    Device Name
    Scarlet® AL-T
    Manufacturer
    Date Cleared
    2024-10-23

    (54 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K181818, K192993

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SCARLET® AL-T system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous and /or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    When used with the integrated fixation by the mean of the bone screws provided, the SCARLET® AL-T is a stand-alone system and requires no additional supplemental fixation system.

    When used as a lumbar intervertebral fusion device (i.e. without the bone screws provided), the SCARLET® AL-T interbody device must be used with supplemental internal spinal fixation system that has been cleared by the FDA for use in the lumbosacral spine.

    Device Description

    The SCARLET® AL-T spinal system is an anterior lumbar interbody fusion device with integrated fixation intended to provide mechanical support to the lumbar spine and maintain adequate disc space until fusion occurs. The SCARLET® AL-T system comprises a range of intervertebral spacers implanted via an anterior approach, and having various sizes, heights, footprints and lordosis so as to adapt individual pathology and different patient's anatomical conditions. The interbody device is a box-shaped spacer with a large central cavity that can receive bone graft intended to promote intervertebral fusion. The SCARLET® AL-T spacers are all made from medical grade titanium alloy conforming to ASTM F136 standard and are produced by additive manufacturing (SLM) according to ASTM F3001. Subsequently the spacer is machined (thread tapping) and polished. The Scarlet® AL-T interbody spacer has a monolithic design that incorporates solid, lattice and porous structures along with superior and inferior rough surfaces intended to increase implant stability into the intervertebral space and bony integration throughout the implant. When used with its integrated fixation, the spacer is crossed by three (3) bone screws protruding into the vertebral endplates. The bone screws are secured by the mean of two (2) cam locks that prevent backing out. The SCARLET® AL-T spinal implants are delivered sterile (gamma sterilization) and supplied with dedicated surgical instruments (reusable - provided non-sterile). Bacterial endotoxin testing as specified in USP standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / device.The purpose of this submission is to add a 5° range of lordotic intervertebral spacers in 3 footprints (small, medium, a 20mm height to the existing 10° and 15° footprints to the previously cleared SCARLET® AL-T range of devices (K1818) K192993). The solid, lattice and porous structures of the same. No modifications have been made to the intended use, indications for use, materials, manufacturing process, sterilization process.

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA to Spineart SA regarding their Scarlet® AL-T intervertebral body fusion device. It primarily focuses on the substantial equivalence determination for additional lordotic intervertebral spacers to their previously cleared devices.

    Based on the provided text, there is no information available regarding the acceptance criteria, the specific study that proves the device meets acceptance criteria, or any of the detailed aspects of a clinical or analytical study as requested in the prompt.

    The document states:

    • "The present submission introduces only additional references for the interbody spacer and trials." (Page 5)
    • "The engineering analysis has demonstrated that SCARLET® AL-T line extension (present submission) does not introduce new worst cases regarding the mechanical testing, so the tests performed on SCARLET® AL-T predicatedevices (K181818 / K192993) are still valid and cover the added cage references." (Page 6)
    • "There is no significant difference between the SCARLET® AL-T line extension) and the previously cleared SCARLET® AL-T predicate devices (K1818 / K192993) which would adversely affect the use and performances of the product or raise any issue of safety and effectiveness. Consequently, no new mechanical tests were carried out." (Page 6)

    This implies that the current submission relies on the testing and acceptance criteria established for the predicate devices (K181818 and K192993). The document does not provide details about those previous tests or acceptance criteria. It only lists a summary of non-clinical tests conducted on the predicate devices, such as:

    • Static and dynamic axial compression (per ASTM F2077)
    • Static and dynamic shear compression (per ASTM F2077)
    • Static and dynamic torsion (per ASTM F2077)
    • Expulsion
    • Subsidence (per ASTM F2267)
    • Mass loss determination after dynamic testing that passed 5 million cycles.

    Therefore, I cannot provide the requested table or answer the specific questions about sample size, expert ground truth, adjudication, MRMC studies, standalone performance, or training set details because this information is not contained within the provided FDA 510(k) summary. The document is for a "line extension" of an already cleared device, asserting substantial equivalence based on prior testing, rather than presenting new comprehensive study data for the current submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192993

    Validate with FDA (Live)

    Device Name
    SCARLET AL-T
    Manufacturer
    Date Cleared
    2020-01-22

    (89 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181818

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scarlet® AL-T system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These spinal implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

    When used with the integrated fixation by the mean of the bone screws provided, the Scarlet® AL-T is a stand-alone system and requires no additional supplemental fixation system.

    When used as a lumbar intervertebral fusion device (i.e. without the bone screws provided), the Scarlet® AL-T interbody device must be used with supplemental internal spinal fixation system that has been cleared by the FDA for use in the lumbosacral spine.

    The Scarlet® AL-T Hyperlordotic (≥ 20°) cages are to be used with BOTH the integrated fixation by the mean of the bone screws provided and ALSO additional supplemental fixation system that has been cleared by the FDA for use in the lumbosacral spine.

    Device Description

    The Scarlet® AL-T spinal system is an anterior lumbar interbody fusion device with integrated fixation intended to provide mechanical support to the lumbar spine and maintain adequate disc space until fusion occurs. The Scarlet® AL-T system comprises a range of intervertebral spacers implanted via an anterior approach, and having various sizes, heights, footprints and lordosis so as to adapt individual pathology and different patient's anatomical conditions. The interbody device is a box-shaped spacer with a large central cavity that can receive bone graft intended to promote intervertebral fusion. The Scarlet® AL-T spacers are all made from medical grade titanium alloy conforming to ASTM F136 standard and are produced by additive manufacturing (SLM) according to ASTM F3001. Subsequently the spacer is machined (thread tapping) and polished.

    The Scarlet® AL-T interbody spacer has a monolithic design that incorporates solid, lattice and porous structures along with superior and inferior rough surfaces intended to increase implant stability into the intervertebral space and bony integration throughout the implant.

    When used with its integrated fixation, the spacer is crossed by three (3) bone screws protruding into the vertebral endplates. The bone screws are secured by the mean of two (2) cam locks that prevent backing out.

    The Scarlet® AL-T spinal implants are delivered sterile (gamma sterilization) and supplied with dedicated surgical instruments (reusable - provided non-sterile). Bacterial endotoxin testing as specified in USP standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / device.

    The subject product line extension consists of an extended range of titanium alloy spacers intended to be used with the three unicortical cancellous bone screws previously approved (K181818).

    Scarlet® AL-T product line extension comprises a range of hyperlordotic intervertebral spacers implanted via an anterior approach, and having various sizes, heights, footprints and lordosis so as to adapt individual pathology and different patient's anatomical conditions. The Scarlet® AL-T Hyperlordotic (≥ 20°) cages are to be used with BOTH the integrated fixation by the mean of the bone screws provided and ALSO additional supplemental fixation system that has been cleared by the FDA for use in the lumbosacral spine.

    AI/ML Overview

    The provided text describes a medical device, the Scarlet® AL-T intervertebral body fusion device, and its substantial equivalence to predicate devices, supported by non-clinical testing. However, it does not include information about acceptance criteria or a study that proves the device meets those criteria in the context of AI performance, human reader improvement with AI assistance, or standalone algorithm performance.

    The document is a 510(k) summary for a spinal implant, which means the "acceptance criteria" are related to mechanical and biological performance, and the "study" is non-clinical testing. It explicitly states, "Based on the static testing results and engineering analysis it has been determined that dynamic testing was not necessary." This further confirms the absence of clinical trials or AI-related performance studies.

    Therefore, I cannot fulfill the request for information regarding AI performance, human reader improvement, standalone performance, or sample sizes related to such studies, as this information is not present in the provided text.

    Here is a partial answer based only on the information available in the document, focusing on the mechanical acceptance criteria and the non-clinical study conducted:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific TestAcceptance Criteria (Implied)Reported Device Performance
    Mechanical PerformanceStatic Axial Compression (ASTM F2077-18)Comparable to predicate deviceResults demonstrate comparable mechanical properties to the predicate device Scarlet® AL-T (K181818).
    Mechanical PerformanceStatic Shear-compression (ASTM F2077-18)Comparable to predicate deviceResults demonstrate comparable mechanical properties to the predicate device Scarlet® AL-T (K181818).
    Mechanical PerformanceStatic Torsion (ASTM F2077-18)Comparable to predicate deviceResults demonstrate comparable mechanical properties to the predicate device Scarlet® AL-T (K181818).
    Mechanical PerformanceSubsidence (ASTM F2267-04)Comparable to predicate deviceResults demonstrate comparable mechanical properties to the predicate device Scarlet® AL-T (K181818).
    Mechanical PerformanceExpulsion (ASTM Draft F-04.25.02.02)Comparable to predicate deviceResults demonstrate comparable mechanical properties to the predicate device Scarlet® AL-T (K181818).
    BiocompatibilityBacterial Endotoxin Testing (USP standard)Endotoxin limit of < 20 EU/deviceAchieves the Endotoxin limit of 20 EU / device.

    2. Sample sized used for the test set and the data provenance
    The document does not specify the sample size (number of units tested) for each non-clinical test. The "data provenance" is implied to be laboratory testing conducted by the manufacturer (Spineart). This is a non-clinical, in-vitro study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This is a non-clinical mechanical and biocompatibility study, not an AI or diagnostic study involving expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as this is a non-clinical mechanical testing report.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    No, an MRMC comparative effectiveness study was not done. This device is a physical spinal implant, not an AI or diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    No, a standalone algorithm performance study was not done. This device is a physical spinal implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" for this non-clinical study is defined by the established mechanical testing standards (e.g., ASTM F2077-18, ASTM F2267-04, ASTM Draft F-04.25.02.02, USP standard) and the performance of the predicate device (Scarlet® AL-T K181818).

    8. The sample size for the training set
    Not applicable. This is a non-clinical mechanical and biocompatibility study, not an AI study involving a training set.

    9. How the ground truth for the training set was established
    Not applicable, as there is no training set for this type of non-clinical study.