Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K243509

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-12

    (181 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Archer^R PSI System is indicated as an orthopedic instrument to assist the physician in the intraoperative positioning of total shoulder replacement components and in guiding the drill and the cut of the bone.

    Archer^R PSI System must only be used conjointly with Archer™ CSR Total Shoulder (K152825, K173812, K181287, K182500, K191811), Catalyst EA Convertible Stemmed Shoulder (K222317) and Archer™ R1 Reverse (K202611, K211991, K213349, K223655, K232583) components in the context of primary total shoulder replacement and following a delto-pectoral approach only. Archer^R PSI System is manufactured from a pre-operative planning validated by the surgeon in the 'Archer™ 3D Targeting' platform (K213779). Archer^R PSI System is indicated for patient population fulfilling the Archer™ CSR Total Shoulder, Catalyst EA Convertible Stemmed Shoulder and Archer™ R1 Reverse indications and for which CT images are available with identifiable placement anatomical landmarks and compliant with imaging protocol provided by Archer 3D Targeting.

    The device is intended for single use only.

    The device is intended for adult patients.

    The device has to be used by a physician trained in the performance of surgery.

    Device Description

    The "Archer PSI System" device is a patient-matched additively manufactured single use surgical instrument (PSI). Archer PSI System is an instrument set containing a glenoid guide and its bone model and/or a humeral guide and its bone model. This patient-specific medical device is designed to fit the patient's anatomy to transfer a patient-specific pre-operative plan to the operating room. It is intended for surgical interventions in orthopaedic procedures for total shoulder arthroplasty.

    The Archer PSI system instruments are designed from a draft treatment plan available via the Archer™ 3D Targeting' platform. Based on computed tomography (CT) of the shoulder anatomy, 3D CAD models of the bones and positioning and sizing of the glenoid and humeral components are submitted for evaluation to the surgeon. Upon the surgeon's approval, the guides and bone models are designed based on the validated planning and are manufactured using additive manufacturing.

    AI/ML Overview

    The provided FDA 510(k) summary for the "Archer PSI System" does not contain the detailed acceptance criteria or the specific study that directly proves the device meets those criteria in a quantitative manner as typically expected for medical device performance studies involving sensitivity, specificity, accuracy, etc. However, it does outline the types of non-clinical and cadaveric testing performed to demonstrate substantial equivalence to a predicate device.

    Here's an attempt to structure the information based on the request, extracting what is available and noting what is not:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not overtly state quantitative "acceptance criteria" (e.g., "accuracy must be > 95%") nor does it provide "reported device performance" in terms of explicit metrics like sensitivity, specificity, or error rates. Instead, the "performance" is described in terms of demonstrating "substantial equivalence" through various engineering and cadaveric tests.

    A more accurate representation, based on the provided text, would be:

    Acceptance Criteria CategoryDescription (from document)Reported Device Performance (from document)
    Mechanical IntegrityDemonstrate mechanical integrity post-processing.Testing was conducted.
    Debris GenerationAssess debris generation.Testing was conducted.
    Intra-Designer VariabilityAssess variability within a single designer's output.Testing was conducted.
    Inter-Designer VariabilityAssess variability between different designers' outputs.Testing was conducted.
    BiocompatibilityEnsure material biocompatibility.Assessment conducted.
    Cleaning & SterilizationValidate cleaning and sterilization processes.Validations conducted.
    Manufacturing CleaningValidate manufacturing cleaning processes.Validation conducted.
    Packaging & Shelf-lifeValidate packaging integrity and shelf-life.Validation conducted.
    Functional EquivalenceDemonstrate functional equivalence to manual techniques for positioning and guiding drill/cut.Cadaveric testing executed to demonstrate substantial equivalence between two techniques (manual and PSI, for both anatomic and for reverse techniques).
    Pre-operative PlanningManufactured from a pre-operative planning validated by the surgeon in the 'Archer™ 3D Targeting' platform (K213779).The device design is based on surgeon-validated plans within the Archer™ 3D Targeting platform. This implies an acceptance of the planning accuracy by the surgeon.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Cadaveric Testing: The document states that "Cadaveric testing was executed" but does not specify the sample size (number of cadavers or procedures) used for this testing.
    • Data Provenance: The cadaveric testing is implied to be prospective in nature, as it was "executed to demonstrate the substantial equivalence." There is no information provided regarding the country of origin of the cadaveric data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: The document states that "a pre-operative planning validated by the surgeon" is part of the process. For the cadaveric testing, it does not explicitly state the number of experts (e.g., surgeons) involved in establishing the "ground truth" or assessing the "substantial equivalence."
    • Qualifications of Experts: The document mentions that the device is to be used by a "physician trained in the performance of surgery." For the validation of the pre-operative plan, the expert is identified as "the surgeon." While this indicates a medical professional, specific qualifications (e.g., years of experience, subspecialty) are not provided.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the cadaveric testing or any other performance evaluation. The "validation by the surgeon" for the pre-operative plan suggests a form of single-expert consensus at the planning stage, but not for the overall performance assessment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No.
    • Effect Size: As no MRMC study was performed, no effect size of human readers improving with AI vs. without AI assistance is reported. The Archer PSI System is a patient-specific instrument, not an AI diagnostic or assistive tool in the MRMC sense. The comparison was between manual surgical techniques and PSI-assisted techniques in cadavers.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Not explicitly in terms of an "algorithm only" performance study. The device itself is a physical, patient-specific instrument derived from a digital plan. The document describes "Cadaveric testing" which evaluates the combined PSI system (planning software output + physical guide) in a simulated surgical environment, not just the planning algorithm in isolation from its physical manifestation or use.
    • The "Intra- and Inter-Designer Variability testing" and "Mechanical Integrity" tests are standalone evaluations of aspects of the device's design and manufacturing, but not of the surgical guidance algorithm's performance on its own.

    7. Type of Ground Truth Used

    • For Pre-operative Planning: The ground truth for the design of the PSI System is based on a "pre-operative planning validated by the surgeon" using CT images and anatomical landmarks. This can be considered a form of expert consensus/validation on the desired surgical outcome/instrument design.
    • For Cadaveric Testing: The "ground truth" for the cadaveric study would be the actual anatomical targets and the achieved drill/cut placements, compared to the planned placements and traditional manual techniques. While experts (surgeons) would perform and assess these, the ultimate "truth" is the physical reality within the cadaver. The document implies comparison to "manual techniques" as a reference.

    8. Sample Size for the Training Set

    The document does not provide any information regarding a "training set sample size." The Archer PSI System is described as being "designed from a draft treatment plan" and "manufactured from a pre-operative planning validated by the surgeon." This suggests a patient-specific design process rather than a machine learning model trained on a large dataset. The underlying "Archer™ 3D Targeting' platform (K213779)" (a separate cleared device) would be the system performing the planning, and its own 510(k) might contain training data details if it uses AI/ML. However, for the Archer PSI System itself, no training set information is present.

    9. How Ground Truth for the Training Set Was Established

    As no training set is mentioned for the Archer PSI System in this document, no information is provided on how its ground truth would have been established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182500

    Validate with FDA (Live)

    Date Cleared
    2019-01-11

    (121 days)

    Product Code
    Regulation Number
    888.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K152825, K181287, K131298, K003044

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.

    • Osteoarthritis
    • Avascular Necrosis ●
    • Rheumatoid Arthritis ●
    • Post-traumatic Arthritis
    • Correction of functional deformity

    The Catalyst CSR humeral and glenoid implants are intended for cemented use.

    The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.

    Device Description

    The Catalyst CSR Shoulder System is a bone preserving total shoulder prosthesis designed for use in patients where the humeral head, neck and glenoid vault are of sufficient bone stock and there is an intact or reconstructable rotator cuff. The design requires minimal bone resection, thus giving the patient an alternative to other total shoulder designs where more bone is removed.

    This submission adds Press-Fit Humeral Components to the CSR Shoulder System. The press-fit humeral components have a non-spherical polished surface for articulation with the glenoid component or the glenoid cavity of the scapula. The humeral components incorporate 4 pegs which assist with alignment and provide rotational stability. The four plane geometry of the back side of the humeral component matches four cut surfaces on the humeral head to recreate the geometry and thickness of the removed bone. The CSR Press-Fit Humeral Components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. The undersurface of the head and the proximal portion of the alignment pegs are coated with a plasma sprayed coating of CP Ti conforming to ASTM F1580.

    The CSR humeral components are compatible with previously cleared CSR 3 Peg glenoid components.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the Catalyst CSR Press-Fit Humeral Components:

    Summary of Device and Context:

    The document is an FDA 510(k) clearance letter and a 510(k) Summary for the Catalyst CSR Press-Fit Humeral Components, which are an addition to the existing Catalyst CSR Shoulder System. This device is a shoulder prosthesis intended for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint, where the humeral head, neck, and glenoid vault have sufficient bone stock, and the rotator cuff is intact or reconstructable. The key new feature here is the "Press-Fit" capability, allowing for uncemented or cemented use, and the device incorporates a plasma sprayed coating of CP Ti. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

    Detailed Analysis of Acceptance Criteria and Study Information:

    Based on the provided text, the focus is on non-clinical testing to demonstrate substantial equivalence, as clinical testing was deemed unnecessary.

    1. Table of Acceptance Criteria and Reported Device Performance:
    Acceptance CriteriaReported Device Performance
    Fixation strength in shearMet the pre-determined acceptance criterion through Static Shear Testing.
    Endotoxin limit (Bacterial Endotoxin Testing)Met the acceptable endotoxin limit of <10EU/component or <20EU/device construct.
    Torque-out (engineering analysis)Deemed applicable from predicate Catalyst CSR Humeral Components.
    1. Sample Size Used for the Test Set and Data Provenance:

      • Static Shear Testing: The document states "Static shear testing was conducted," but it does not specify the sample size for this test.
      • Bacterial Endotoxin Testing: The document states "Bacterial Endotoxin Testing was performed," but it does not specify the sample size for this test.
      • Data Provenance: The document does not provide information on data provenance (e.g., country of origin, retrospective/prospective). Given these are non-clinical lab tests, it's implied they were conducted in a controlled lab environment by the manufacturer or a contracted lab.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • This question is not applicable as the document describes non-clinical engineering and bench tests, not studies requiring expert interpretation for ground truth.
    3. Adjudication Method for the Test Set:

      • This question is not applicable for the same reasons as above. No human expert adjudication was involved in the non-clinical tests described.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done.
      • The document explicitly states: "Clinical testing was not necessary to demonstrate substantial equivalence of the Catalyst CSR Press-Fit Humeral Components to the predicate devices."
      • Therefore, there is no information on human reader improvement with or without AI assistance.
    5. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

      • This question is not applicable. The device is a physical shoulder implant, not an algorithm or AI system. The testing performed was related to its physical properties and biocompatibility.
    6. Type of Ground Truth Used:

      • For the non-clinical tests mentioned:
        • Static Shear Testing: The "ground truth" would be the universally accepted principles of mechanical engineering and material science, measured against a pre-determined acceptance criterion (likely based on established standards for orthopedic implants).
        • Bacterial Endotoxin Testing: The "ground truth" is the predefined acceptable endotoxin limit (<10EU/component or <20EU/device construct), which is based on regulatory standards for medical devices to ensure safety.
        • Torque-out: "Ground truth" from an engineering analysis, likely based on established mechanical principles and data from the predicate device.
    7. Sample Size for the Training Set:

      • This question is not applicable. The device is a physical implant, not an AI model requiring a training set. The term "training set" is typically used in the context of machine learning or AI development.
    8. How the Ground Truth for the Training Set Was Established:

      • This question is not applicable for the same reasons as above.

    Conclusion:

    The FDA clearance for the Catalyst CSR Press-Fit Humeral Components relied on non-clinical testing demonstrating substantial equivalence to predicate devices. The key tests reported were static shear testing and bacterial endotoxin testing, both of which met pre-determined acceptance criteria. Clinical testing, which would involve human performance or outcomes, was explicitly stated as "not necessary." Therefore, many of the questions related to clinical study design, expert involvement, and AI performance are not relevant to this specific clearance document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1