Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192315

    Validate with FDA (Live)

    Date Cleared
    2020-01-10

    (137 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K180696

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs) are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are powder-free, ambidextrous with beaded cuff, green colored nitrile gloves featuring an inner coating of colloidal oatmeal USP. The gloves are offered in five sizes, extra-small, small, medium, large and extra-large.

    The gloves are designed and manufactured in accordance with the ASTM D6319-10 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013).

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of various tests conducted for the "Medline Green Ambidextrous Powder-Free Nitrile Examination Gloves with Colloidal Oatmeal USP (Tested for use with Chemotherapy Drugs)". This is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, rather than a study proving the device meets acceptance criteria in a clinical context. Therefore, many of the requested elements for an AI/medical device study (like sample size for test set, data provenance, expert adjudication, MRMC studies, standalone performance, training set sample size, etc.) are not applicable to this type of submission.

    However, I can extract the acceptance criteria and reported device performance from the provided biocompatibility and performance testing summaries.


    1. Table of acceptance criteria and the reported device performance

    Biocompatibility Testing Summary

    Test StandardTest Standard DescriptionAcceptance CriteriaReported Device Performance
    ISO 10993-10Primary Skin IrritationCutaneous macroscopical examinations are performed after removal of test material. The appearance of each application site and skin reaction observed are scored.No skin reactions observed. Not a skin irritant.
    ISO 10993-10Dermal SensitizationChallenge sites are visually assessed for allergic reactions and the intensity of the reaction is scored.No positive allergic reactions observed. Not a skin sensitizer.
    ISO 10993-5CytotoxicityTest articles scoring '0', '1', or '2' are considered 'non-cytotoxic.' Test articles scoring '3' or '4' are considered 'cytotoxic.'At the neat extraction, the test article is considered cytotoxic.
    ISO 10993-11Acute Systemic ToxicityAbnormal clinical signs indicative of toxicity during the 72 hours test period are observed and reported.No signs of toxicity at any of the observation periods were observed.

    Performance Testing Summary

    Test StandardTest Standard DescriptionAcceptance CriteriaReported Device Performance
    ASTM D6319-10Standard Specification for Nitrile Examination Gloves for Medical ApplicationSamples shall meet specified physical dimensions in accordance with the standard.All device dimensions meet the standard.
    ASTM D6124-06 (Reaffirmation 2011)Standard Test Method for Residual Powder on Medical Gloves< 2mg of residual powder per glove.Complies with standard.
    ASTM D5151-06 (Reapproved 2011)Standard Test Method for Detection of Holes in Medical GlovesAny glove that shows a droplet, stream or other type of water leakage shall be considered to have failed.Complies with standard.
    ASTM D6978-05 (Reapproved 2013)Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy DrugsThe absorbance of test chemicals, which permeate through the test articles, is measured and reported. Note: Specific breakthrough times vary by drug.Refer to the detailed list of chemotherapy drugs and their breakthrough times (2 records below 240 mins, the rest >240 mins)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify sample sizes for each test mentioned (e.g., how many gloves were tested for holes, or how many samples for biocompatibility). The tests are performed according to established ASTM and ISO standards, which typically define the required sample sizes for quality control and performance evaluation.

    Data provenance is not explicitly stated as "country of origin of data" but refers to tests being conducted according to international standards (ASTM, ISO). These are bench tests, not clinical studies. The data would be considered prospective in the sense that the tests were specifically performed for this submission to evaluate the device.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the tests described are laboratory/bench tests, not evaluations requiring expert interpretation of clinical data in the context of ground truth establishment. The "ground truth" here is determined by chemical and physical measurements against established standards.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The tests are objective measurements against defined standards, not requiring human adjudication.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for medical gloves, not an AI-powered diagnostic or assistive device.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a submission for medical gloves.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For physical and chemical properties, the "ground truth" is based on objective measurements defined by international standards (ASTM and ISO). For example:

    • Physical dimensions are measured against the specifications in ASTM D6319-10.
    • Residual powder is measured against a limit in ASTM D6124-06.
    • Detection of holes is defined by the absence of leakage in ASTM D5151-06.
    • Chemotherapy drug permeation is measured by breakthrough time according to ASTM D6978-05.
    • Biocompatibility assessments (irritation, sensitization, cytotoxicity, systemic toxicity) are assessed by standardized laboratory protocols with established biological endpoints according to ISO 10993 series.

    8. The sample size for the training set

    Not applicable. This is a device (medical glove) and not an AI or machine learning algorithm that requires a training set. The device's characteristics are inherent to its manufacturing process and materials.


    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a medical glove.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1