510(k) Data Aggregation

    K Number
    K232173

    Validate with FDA (Live)

    Date Cleared
    2023-10-06

    (77 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K222028, K180480, K211805, K161363, K192938, K222973

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ascend VBR System is intended for use in skeletally mature patients in the cervical spine (C2-T1) and in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e., fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.

    The Ascend VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.

    The Ascend VBR System is intended to be used with supplemental spinal fixation systems cleared for use in the cervical, thoracic, and/or lumbar spine. The use of bone grafting material with the Ascend VBR System is optional.

    The Ascend NanoTec VBR System with advanced NanoTec surface treatment is intended for use in skeletally mature patients in the cervical spine (C2-T1) and in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e., fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.

    The Ascend NanoTec VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.

    The Ascend NanoTec VBR System is intended to be used with supplemental spinal fixation systems cleared for use in the cervical, thoracic, and/or lumbar spine. The use of bone grafting material with the Ascend NanoTec VBR System is optional.

    Device Description

    The Ascend VBR Systems are cervical and thoracolumbar vertebral body replacement systems designed to be inserted in the anterior, anterolateral, and posterior approaches. The implants consist of VBR cores and endplates offered in a range of diameters and heights, endplates, footprints, and lordotic angles to accommodate individual patient anatomy.

    The cores are made of titanium alloy per ASTM F136 and the endplates are made of titanium alloy per ASTM F136 or are made of hybrid titanium consisting of commercially pure porous titanium CP-Ti Grade 2 per ASTM F67 and titanium alloy per ASTM F136.

    The modular endplates are offered with an optional 20-40 nanometer thin hydroxyapatite surface treatment. The surface treatment presents nano-scale topography on the entirety of the implant surface, in addition to macro-/micro-scale topography existing from prior to treatment.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device – the Ascend™ VBR System and Ascend™ NanoTec™ VBR System, which are vertebral body replacement devices. The content focuses on demonstrating substantial equivalence to legally marketed predicate devices, as required for FDA clearance.

    The provided text does not contain any information about an AI/ML-based device or its performance criteria. The entire document pertains to physical medical devices (spinal implants) and their mechanical, chemical, and biological performance characteristics.

    Therefore, it is not possible to describe acceptance criteria and a study proving an AI/ML device meets those criteria based on the provided input. The document details traditional medical device testing, such as:

    • Static and Dynamic Axial Compression per ASTM F2077
    • Static and Dynamic Torsion per ASTM F2077
    • Static Subsidence per ASTM F2267
    • Static Push-out
    • Gravimetric Analysis per ASTM F1714
    • Particulate Analysis per ASTM F1877
    • Bacterial endotoxin testing (BET) per ANSI/AAMI ST72:2019
    • Biocompatibility Testing per ISO 10993-1

    These tests are standard for evaluating the mechanical and biocompatibility properties of physical implants, not the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IdentiTi Porous Ti Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the IdentiTi Porous Ti System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The IdentiTi Porous Ti Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with IdentiTi LIF interbody spacers to provide integrated fixation. IdentiTi LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. IdentiTi ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.

    The IdentiTi Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the IdentiTi NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The IdentiTi NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with IdentiTi NanoTec LIF interbody spacers to provide integrated fixation. IdentiTi NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.

    The Transcend PEEK Interbody System is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the Transcend PEEK Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Transcend PEEK Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with Transcend LIF interbody spacers to provide integrated fixation. Transcend LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation. Transcend ALIF interbody spacers with >20° lordosis must be used with an anterior plate as the form of supplemental fixation.

    The Transcend PEEK Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from T1 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis, spinal stenosis, and/or thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain) at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the Transcend NanoTec Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Transcend NanoTec Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate and supplemental fixation systems that are cleared by FDA for use in the thoracic and lumbar spine. AMP™ Anti-Migration Plate may be used with Transcend NanoTec LIF interbody spacers to provide integrated fixation. Transcend NanoTec LIF spacers with >20° lordosis must be used with AMP Anti-Migration Plate in addition to supplemental fixation.

    The IdentiTi ALIF Standalone Interbody System is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. Additionally, the IdentiTi ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate. The IdentiTi ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    The IdentiTi ALIF Standalone Interbody System with advanced NanoTec surface treatment is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. Additionally, the IdentiTi NanoTec ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi Nanotec ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The IdentiTi NanoTec ALIF Standalone Interbody System is intended for use on patients who have had at least six months of non-operative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate. The IdentiTi NanoTec ALIF Standalone Interbody System implants of ≤20° are a standalone system. The IdentiTi NanoTec ALIF Standalone Interbody System implants of >20° must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    Device Description

    The IdentiTi and Transcend Interbody Systems are thoracolumbar intervertebral body fusion systems designed to be inserted through anterior and posterior surgical approaches. The interbody spacers are manufactured from PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, tantalum per ASTM F560, titanium alloy (Ti-6Al-4V ELI), and commercially pure titanium (CPTi Grade 2) per ASTM F67. The interbody spacers are available in the following material options: (1) PEEK (polyetheretherketone) with tantalum and titanium alloy markers, or (2) commercially pure porous titanium (PTi), or (3) a combination of commercially pure porous titanium (CPT) Grade 2) per ASTM F67 and titanium alloy (Ti-6Al-4V ELI) per ASTM F136. The subject IdentiTi and Transcend Interbody Systems implants consist of various lengths, widths, heights and lordotic options to accommodate individual patient anatomy. To mitigate risk of expulsion, the interbody endplates feature teeth. All interbody spacers feature an internal graft aperture for placement of graft material to promote fusion through the cage. Additionally, the IdentiTi implants are offered with a microstructure due to the layering of material that forms the porous architecture. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation. The IdentiTi and Transcend NanoTec Interbody Systems implant surfaces have been treated with a 20-40 nanometer thin hydroxyapatite (HA) surface treatment. The surface treatment presents nano-scale topography on the entirety of the implant surface, in addition to macro-/micro-scale topography existing from prior to treatment. The purpose of this Traditional 510(k) is to receive clearance for expanded indications for the treatment of degenerative spondylolisthesis, multilevel degenerative scoliosis, spinal stenosis, and sagittal deformity, and use of allogeneic bone consisting of cortical bone, as well as allogeneic bone consisting of demineralized allograft bone with bone marrow aspirate.

    AI/ML Overview

    This document pertains to the Alphatec Spine interbody systems (IdentiTi™ Porous Ti, IdentiTi™ NanoTec™, Transcend™ PEEK, Transcend™ NanoTec™, IdentiTi™ ALIF Standalone, and IdentiTi™ NanoTec™ ALIF Standalone Interbody Systems) and their expanded indications for use.

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state specific acceptance criteria values (e.g., "device must withstand 1000N axial compression"). Instead, it lists the types of non-clinical tests performed, implying that the devices met the requirements of these standard tests. The reported "performance" is implicitly that the devices passed these tests.

    Test CategorySpecific Test / StandardReported Device Performance
    Mechanical TestingASTM F2077Static and dynamic axial compression, compression-shear and torsion tests performed. (Implicitly passed)
    Subsidence TestingASTM F2267Static subsidence test performed. (Implicitly passed)
    Expulsion ResistanceASTM Draft F-04.25.02.02Static push-out test performed. (Implicitly passed)
    Material CharacterizationF1714Gravimetric analysis performed. (Implicitly passed)
    Particulate AnalysisF1877Particulate analysis performed. (Implicitly passed)
    Integrated FixationScrew push-outScrew push-out test performed. (Implicitly passed)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for each non-clinical test (e.g., number of interbody devices tested for axial compression). It refers to the tests generally.

    Regarding data provenance, the tests are non-clinical (laboratory tests) and therefore do not involve patient data or data from specific countries of origin. The data is prospective as it was generated specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable to this submission. The "ground truth" for the non-clinical performance tests is based on established engineering standards (ASTM) and the physical properties of the devices themselves, not on expert medical opinion or consensus.

    4. Adjudication Method for the Test Set

    This question is not applicable to this submission. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters, where discrepancies in interpretation need to be resolved. The non-clinical tests outlined here are objective measurements against predefined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable to this submission. The device is an intervertebral body fusion system, a physical implant, not an AI diagnostic or assistance tool. Therefore, MRMC studies involving human readers or AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to this submission. The device is a physical interbody fusion system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance tests, the "ground truth" is based on engineering standards and material science specifications. The tests are designed to assess the physical and mechanical properties of the interbody systems against recognized benchmarks for safety and effectiveness in their intended use.

    8. The sample size for the training set

    This question is not applicable to this submission. As the device is a physical implant and not an AI/ML algorithm, there is no "training set."

    9. How the ground truth for the training set was established

    This question is not applicable to this submission for the same reason as above (no training set).