Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K173840, K043144

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert® CT/NG test, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro real-time PCR test for the automated detection and differentiation of genomic DNA from Chlamvdia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) to aid in the diagnosis of chlamydial and gonorrheal disease in the urogenital tract and extragenital sites (pharynx and rectum). The assay may be used to test the following specimens from asymptomatic and symptomatic individuals: female and male urine, patient-collected vaginal swabs (collected in a clinical setting), clinician-collected endocervical swabs, and female and male pharyngeal and rectal swabs.

    Device Description

    The Xpert CT/NG test is an automated in vitro diagnostic test for qualitative detection and differentiation of DNA from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG). The test is performed on the Cepheid GeneXpert Instrument Systems. The Xpert CT/NG test on the GeneXpert Instrument System automates and integrates sample purification, nucleic acid amplification and detection of the target sequences in simple or complex samples using real-time PCR. The system consists of an instrument, personal computer, and preloaded software for running the tests and viewing the results. The system requires the use of single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, crosscontamination between samples is minimized.

    The Xpert CT/NG test includes reagents for the detection and differentiation of CT and NG. A Sample Processing Control (SPC), a Sample Adequacy Control (SAC), and a Probe Check Control (PCC) are also included. The SPC is present to control for adequate processing of the target bacteria and to monitor the presence of inhibitors in the PCR reaction. The SAC reagents detect the presence of a single copy human gene and monitor whether the specimen contains human cells. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

    The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems, the GeneXpert Infinity-48 System. GeneXpert Infinity-48s. and the GeneXpert Infinity-80 System, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

    The ancillary specimen collection kits for use with the Xpert CT/NG test are the Xpert Vaginal/Endocervical Specimen Collection Kit, Xpert Swab Specimen Collection kit and the Xpert Urine Specimen Collection kit.

    AI/ML Overview

    This looks like a 510(k) summary for the Cepheid Xpert CT/NG test. The document provides information on analytical and clinical performance studies.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" with numerical targets for clinical performance (sensitivity, specificity) in the same way it defines LoD for analytical sensitivity. However, based on the presentation and context, the reported sensitivities and specificities from the clinical performance study appear to be the performance demonstrated to justify substantial equivalence.

    Since the submission is for a 510(k), the implied "acceptance criteria" for clinical performance is that the device's performance is substantially equivalent to a predicate device. The document states "the Xpert CT/NG test performance is equivalent to the predicate" in the conclusions.

    Here's a table summarizing the reported clinical performance:

    Specimen TypeTarget OrganismPerformance Metric% (95% CI)
    Pharyngeal SwabCTSensitivity95.9% (86.3-98.9)
    Pharyngeal SwabCTSpecificity99.7% (99.4-99.8)
    Rectal SwabCTSensitivity86.0% (80.9-89.9)
    Rectal SwabCTSpecificity99.4% (98.9-99.6)
    Pharyngeal SwabNGSensitivity94.7% (90.7-97.0)
    Pharyngeal SwabNGSpecificity98.8% (98.3-99.2)
    Rectal SwabNGSensitivity91.2% (86.6-94.4)
    Rectal SwabNGSpecificity99.6% (99.3-99.8)

    Analytical Sensitivity (LoD) - (Example of explicit acceptance criteria and performance)

    • Acceptance Criteria for LoD: The lowest concentration at which 95% of at least 20 replicates are positive.
    • Reported Device Performance (LoD - Pharyngeal Swab Matrix):
      • CT ATCC vr885 serovar D: 161 EB/mL
      • CT ATCC vr879 serovar H: 225 EB/mL
      • NG ATCC 19424: 7.1 CFU/mL
      • NG ATCC 49226: 6.4 CFU/mL
    • Reported Device Performance (LoD - Rectal Swab Matrix):
      • CT ATCC vr885 serovar D: 88 EB/mL
      • CT ATCC vr879 serovar H: 161 EB/mL
      • NG ATCC 19424: 4.9 CFU/mL
      • NG ATCC 49226: 5.3 CFU/mL

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set:
      • Pharyngeal Swabs: 2577 specimens (eligible for inclusion in data analyses)
      • Rectal Swabs: 2538 specimens (eligible for inclusion in data analyses)
    • Data Provenance:
      • Country of Origin: United States ("multi-site prospective investigational study at 9 US institutions").
      • Retrospective or Prospective: Prospective ("multi-site prospective investigational study").

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that the ground truth was established using an "anatomic site infected status (ASIS) algorithm based on combined results from two NAAT tests, with a tiebreaker NAAT test if applicable." This suggests a reliance on laboratory test results rather than human expert interpretation of raw data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    The adjudication method used for establishing the ASIS (ground truth) was a form of 2+1 rule:

    • The anatomic site was considered infected if both primary reference NAAT test results were positive.
    • The anatomic site was considered not infected if both primary reference NAAT test results were negative.
    • If there was discordance between the two primary reference tests, an additional (tiebreaker) NAAT was performed. In this case, agreement of 2 out of 3 (two primary reference tests + tiebreaker) determined the ASIS result.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study is an analytical and clinical performance evaluation of an in vitro diagnostic (IVD) nucleic acid amplification test (NAAT), not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance study was done. The study directly compares the results of the Xpert CT/NG test (algorithm only) to the established ASIS ground truth for each specimen. There is no human-in-the-loop component mentioned for the interpretation of the Xpert CT/NG results themselves.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth used was Anatomic Site Infected Status (ASIS), which was determined by a composite reference method involving multiple nucleic acid amplification tests (NAATs). This is a type of laboratory-based diagnostic ground truth, not pathology, expert consensus on images, or long-term outcomes data.

    8. The sample size for the training set:

    The document does not provide information on the sample size for a training set. This is typical for traditional in vitro diagnostic (IVD) submissions like this one, as the assay development and validation generally don't utilize "training sets" in the same way machine learning algorithms do. The studies described are for validation of the final device.

    9. How the ground truth for the training set was established:

    Not applicable, as a distinct training set and its ground truth establishment are not discussed in this document, which focuses on the validation of the Xpert CT/NG assay.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1