510(k) Data Aggregation

    K Number
    K221436

    Validate with FDA (Live)

    Date Cleared
    2023-06-29

    (408 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K212031, K103767, K173770

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended to provide heated and humidified CPAP to spontaneously breathing premature and full term neonates and infant patients (up to a weight of 10 kg) who require breathing support, due to conditions associated with prematurity (such as Respiratory Distress Syndrome) or other conditions where CPAP is required or desired. It is intended for use within a hospital clinical environment such as a NICU (Neonatal Intensive Care Unit) and PICU (Pediatric Intensive Care Unit), within the limits of its stated technical specifications.

    Device Description

    The subject devices of this 510(k) are the F&P 950 Bubble CPAP Breathing Circuit Kits (950N60J and 950N62J). The F&P 950 Bubble CPAP Breathing Circuit Kits are single patient use and designed for use with the F&P 950 Respiratory Humidifier. The F&P 950 Bubble CPAP (continuous positive airway pressure) Breathing Circuit Kits provide breathing support to spontaneously breathing neonates and infants. The F&P 950 Bubble CPAP breathing circuit kits deliver heated and humidified respiratory gas through an inspiratory limb to the infant via a nasal interface. An expiratory limb connects to a Bubble CPAP generator to generate CPAP. The device is intended to be operated at input gas flows of 4 - 15 L/min with available CPAP levels of 3 - 10 cmH2O.

    AI/ML Overview

    This document outlines the acceptance criteria and supporting studies for the F&P 950 Bubble CPAP Breathing Circuit Kits.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the F&P 950 Bubble CPAP Breathing Circuit Kits are primarily established through compliance with recognized medical device standards and comparison to a predicate device. The performance data section summarizes various tests conducted to demonstrate this compliance.

    Since the document does not present specific quantitative acceptance criteria with corresponding performance metrics in a direct table format, I will reconstruct a table based on the information provided, inferring acceptance as "compliance" or "demonstrated safety/effectiveness" where specific numerical targets are not explicitly stated for the "acceptance criteria" column. The "reported device performance" will refer to the conclusion of the tests.

    Acceptance Criteria CategoryRegulatory/Standard-Based Acceptance CriteriaReported Device Performance
    Intended Use EquivalenceTo deliver heated and humidified respiratory gases to provide CPAP to spontaneously breathing premature and full-term neonates and infants up to 10kg, in a hospital clinical environment (NICU/PICU).Equivalent to Predicate: Subject and predicate devices are both intended to provide CPAP to patients.
    Patient Population EquivalenceNeonatal and infant patients weighing <10 kg.Identical to Predicate: Neonatal and infant patients weighing <10 kg.
    Operating Principle EquivalenceBreathing circuit kit contains a humidification chamber, inspiratory limb, expiratory limb, pressure relief manifold, and Bubble CPAP generator. (For 950N62J, includes Hudson extensions).Identical to Predicate: The subject device's operating principle is described as identical to the predicate (with the addition of Hudson extensions for 950N62J).
    CPAP Range3 – 10 cmH2OIdentical to Predicate: 3 – 10 cmH2O.
    Specified Flow Range4 – 15 L/minIdentical to Predicate: 4 – 15 L/min.
    Over-pressure protection for patient safetyYes – 17 cmH2OIdentical to Predicate: Yes – 17 cmH2O.
    Ambient operating conditions20°C – 26°CIdentical to Predicate: 20°C – 26°C.
    ReusabilitySingle UseIdentical to Predicate: Single Use.
    Duration of Use14 Days (for subject device)Different to Predicate (but accepted): The subject device has been tested to a 14-day duration, which is longer than the predicate's 7 days, indicating demonstrated capability for extended use.
    Shelf Life3 yearsIdentical to Predicate: 3 years.
    Storage conditions-10°C – 50°CIdentical to Predicate: -10°C – 50°C.
    BiocompatibilityCompliance with ISO 10993-1:2018, ISO 18562 (Parts 1, 2, 3, 4), and related standards for external communicating gas pathway (dry and humidified), permanent contact (> 30 days).Compliant: Testing demonstrated an appropriate biocompatibility profile for the device, confirming compliance with specified ISO standards.
    Electrical Safety & Electromagnetic Compatibility (EMC)Compliance with ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-2:2014 + AMD1:2020 Ed 4.1, and AIM Standard 7351731 Rev. 2.00 2017-02-23.Compliant: Testing demonstrated appropriate electrical safety and electromagnetic compatibility profile, confirming compliance with specified standards.
    Human Factors & UsabilityConducted in accordance with FDA's "Applying Human Factors and Usability Engineering to Medical Devices" guidance.Safe and Effective: Validation study demonstrated the device to be safe and effective for intended users, uses, and environments according to applicable guidance.
    Humidification Output, Thermal Overshoot, Surface Temp.Compliance with ISO 80601-2-74:2017.Compliant: Performance testing for these parameters was conducted in line with ISO 80601-2-74:2017 to demonstrate substantial equivalence. (Implies compliance, though specific figures are not provided in this summary).
    Resistance to Flow, Compliance, Gas LeakCompliance with ISO 5367:2014.Compliant: Performance testing for these parameters was conducted in line with ISO 5367:2014 to demonstrate substantial equivalence. (Implies compliance).
    Comparative System Level Pneumatic PerformanceDemonstrated equivalence.Demonstrated: Testing conducted to demonstrate substantial equivalence. (Implied successful demonstration).

    2. Sample size used for the test set and the data provenance
    The document does not explicitly state the sample sizes (number of devices or tests performed per device type) for the non-clinical tests (biocompatibility, electrical safety, human factors, bench/performance testing). It generally refers to "testing" or "studies" in the plural.

    The data provenance is from non-clinical tests performed by the manufacturer, Fisher & Paykel Healthcare Limited, to evaluate the device against specified international standards and to demonstrate equivalence to a predicate device. This is prospective testing of the manufactured device components and systems. No country of origin for the data is explicitly mentioned beyond the manufacturer's location in New Zealand, but the tests are conducted against universally recognized international standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not provided in the document. The studies listed are primarily non-clinical engineering and performance tests, not clinical studies requiring expert ground truth for diagnostic or clinical outcomes. For the Human Factors testing, "intended users" would have been involved, but their number and specific qualifications are not detailed.

    4. Adjudication method for the test set
    Not applicable. The tests described are objective, standard-based compliance evaluations or measurements, not clinical studies requiring expert adjudication of subjective outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is a mechanical breathing circuit kit, not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This device is a mechanical breathing circuit kit; there is no algorithm involved.

    7. The type of ground truth used
    For the non-clinical tests, the "ground truth" is defined by the requirements and methodologies outlined in the referenced international standards (e.g., ISO, ASTM, ANSI, IEC). For example, a successful biocompatibility test means the materials met the specified criteria of the ISO 10993 series. A successful electrical safety test means the device operated within the limits defined by IEC 60601-1-2.

    8. The sample size for the training set
    Not applicable. This device is a mechanical medical device, not an AI/machine learning product that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable. As above, this device does not utilize AI and therefore has no training set or associated ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220703

    Validate with FDA (Live)

    Date Cleared
    2023-06-16

    (463 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The F&P 950 Respiratory Humidifier is intended to provide heat and humidity to respiratory gases delivered to patients. It is for use in a hospital or long-term care facility by a health professional.

    F&P 950 Breathing Circuit Kits: 950A80J, 950A81J, 950A82J, 950A61J, 950A61J, 950A40J, AA451SU
    The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended for delivery of heated humidified respiratory gases to adult, adolescent and child patients, within the limits of its stated technical specifications.

    F&P 950 Breathing Circuit Kits: 950P81J, 950P40J
    The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended for delivery of heated humidified respiratory gases to neonatal, infant, child and adolescent patients of its stated technical specifications.

    F&P 950 Breathing Circuit Kits: 950N80J, 950N81J, 950N61J, 950N40J
    The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended for delivery of heated humidified respiratory gases to neonatal, infant and child patients, within the limits of its stated technical specifications.

    Device Description

    The F&P 950 Respiratory Humidifier is designed to warm and humidify air after it leaves a commercially available flow source and to maintain the air in a warmed and humidified condition to the point of delivery into the patient via heated tubing.

    The F&P 950 Respiratory Humidifier is made up of a heaterbase, sensor cartridge and expiratory heater wire adapter with breathing circuit kits to support the use with adult, pediatric and neonatal patients. Breathing circuit kits enable the use of the F&P 950 Respiratory Humidifier with clinical procedures such as invasive ventilation and nasal high flow.

    The device is a programmable electrical medical device that uses sensors throughout the system to humidify the gas. Gas travels through the chamber where it is humidity is requlated through monitoring the gas temperature exiting the chamber to achieve the target humidity level. The gas is then transported to the patient through a heated delivery tube. The gas temperature exiting the heated delivery tube is monitored to minimize the formation of condensate.

    AI/ML Overview

    The document is a 510(k) Summary for the F&P 950 Respiratory Humidifier and its accessory breathing circuit kits. It outlines the device's indications for use and compares its technological characteristics to a primary predicate device (F&P MR850 Respiratory Humidifier) and several secondary predicate devices for the various breathing circuit kits. The summary also details the non-clinical tests performed to demonstrate substantial equivalence.

    Here's the requested information based on the provided text, focusing on acceptance criteria and supporting studies:

    1. Table of acceptance criteria and the reported device performance:

      Acceptance CriteriaReported Device Performance
      Delivered humidity level (Bypassed airway)>33mg/L for flow < 60 L/min
      Delivered humidity level (Non-bypassed airway)>12mg/L for flow < 120 L/min
      Time to reach set temperature<30min
      Alarms (auditory)Exceed 45 dbA @1m
      Auditory alarm pause120 seconds
      Ambient operating conditions (Adult)18-26°C
      Ambient operating conditions (Neonatal and Pediatric)20-26°C
      Storage conditions (Hardware)-20°C - 60°C
      Service life (Hardware)7 years
      shelf life for circuit kits3 years (Note: some predicate devices had 5 years; for the subject device, this is stated as "Equivalent - The subject device falls within the secondary predicate device shelf life duration," implying 3 years meets the criteria for substantial equivalence, though less than some predicates).
      Storage conditions for circuit kits-10°C – 50°C (Note: some predicate devices had -30°C – 50°C; for the subject device, this is stated as "Equivalent - The subject device falls within the secondary predicate device storage condition range.")
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      The document does not explicitly state the sample sizes for specific performance tests (e.g., number of devices tested, number of runs for humidification accuracy). It refers to testing being "in line with" or conforming to various ISO and ASTM standards, which would define appropriate sample sizes for those tests. The data provenance (country of origin, retrospective/prospective) is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      This information is not applicable to the type of device and studies described. The studies are non-clinical (bench/performance testing, biocompatibility, electrical safety, software verification, human factors), not clinical studies involving expert interpretation of data like imaging.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      This information is not applicable as the studies described are non-clinical performance and safety tests, not clinical studies involving expert adjudication of findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for interpretative medical devices, often those incorporating AI, and is not applicable to a respiratory humidifier.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

      Not applicable. The F&P 950 Respiratory Humidifier is a hardware device with embedded software, not an AI algorithm intended for standalone performance evaluation in the context of interpretation tasks. Performance testing was done on the device as a whole system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      For the non-clinical performance tests, "ground truth" is established by adherence to recognized standards:

      • Humidification output, thermal overshoot, surface temperature: ISO 80601-2-74:2017
      • Resistance to flow, compliance, gas leak: ISO 5367:2014
      • Electrical safety: ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012
      • EMC: IEC 60601-1-2:2014 + AMD1:2020 Ed 4.1 and AIM Standard 7351731 Rev. 2.00 2017-02-23
      • Alarms: IEC 60601-1-8:2006 and A1:2012
      • Biocompatibility: ISO 10993-1:2018 and ISO 18562-1:2017
      • Software verification and validation: FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
      • Human Factors Testing: FDA's Guidance for Industry and FDA Staff, "Applying Human Factors and Usability Engineering to Medical Devices."

      The "ground truth" is essentially the specified requirements and acceptance criteria outlined in these internationally recognized standards and guidance documents.

    8. The sample size for the training set:

      Not applicable. This device does not appear to involve machine learning or AI algorithms that require a "training set" in the conventional sense of supervised learning. The software verification and validation is for embedded software controlling the device's functions.

    9. How the ground truth for the training set was established:

      Not applicable, as there is no "training set" described for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222197

    Validate with FDA (Live)

    Date Cleared
    2023-04-06

    (258 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173770, K131895

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    F&P Optiflow Junior 2: The Fisher & Paykel Healthcare Optiflow Junior 2 nasal cannula is a single use nasal cannula intended for use with a nasal high flow therapy (NHF) system to deliver heated and humidified nasal high flow therapy to spontaneously breathing patients. This product is designed for use in hospital environments and must be prescribed by a physician. The intended pediatric subpopulations targeted for use of the F&P Optiflow Junior 2 Nasal Cannula range includes: Neonates, birth up to 1 month of age, Infants, 1 month up to 2 years of age, Children, 2 years up to 12 years of age.

    F&P Optiflow Junior 2+: The Fisher & Paykel Healthcare Optiflow Junior 2+ nasal cannula is a single use nasal cannula intended for use with a nasal high flow therapy (NHF) system to deliver heated and high flow therapy to spontanously breathing patients. This product is designed for use in hospital environments and must be prescribed by a physician. The intended pediatric subpopulation targeted for use on the F&P Optiflow Junior 2+ Nasal Cannula includes: Infants. 1 month up to 2 years of age, Children, 2 years up to 12 years of age.

    F&P Optiflow Junior 2 HM Cannula: The Fisher & Paykel Healthcare Optiflow Junior 2 nasal cannula is a single use nasal cannula intended for use with a nasal high flow therapy (NHF) system to deliver heated and high flow therapy to spontaneously breathing patients. This product is designed for use in long term care environments and must be prescribed by a physician. The intended pediatric subpopulations targeted for use of the F&P Optiflow Junior 2 Nasal Cannula range includes: Infants, 1 month up to 2 years of age, Children, 2 years up to 12 years of age.

    F&P Optiflow Junior 2+ HM Cannula: The Fisher & Paykel Healthcare Optiflow™ Junior 2+ nasal cannula is a single use nasal cannula intended for use with a nasal high flow (NHF) therapy system to deliver heated and high flow therapy to spontanously breathing patients. This product is designed for use in long term care environments and must be prescribed by a physician. The intended pediatric subpopulation targeted for use on the F&P Optiflow Junior 2+ Nasal Cannula includes: Infants, 1 month up to 2 years of age, Children, 2 years up to 12 years of age.

    Device Description

    The F&P Optiflow Junior 2 / 2+ Nasal Cannula Interface Range are single-use patient interfaces that are intended to deliver heated and humidified Nasal High Flow therapy to spontaneously breathing neonates, infants and children. It is intended to be prescription-only and provided in a non-sterile state.

    The F&P Optiflow Junior 2 Product Codes are OJR410 (XS), OJR414 (M), OJR416 (L) OJR416HM (L), OJR418 (XL), and OJR418HM (XL).

    The F&P Optiflow Junior 2+ Product Codes are OJR520 (XXL) and OJR520HM (XXL).

    Optional Kits:
    Ventilator Transition Kits: The F&P Optiflow Junior 2 and 2+ Ventilator Transition Kits allow the use of the F&P Optiflow Junior 2 and F&P Optiflow Junior 2+ cannula with approved Fisher & Paykel Healthcare breathing circuits when connected to a ventilator to deliver heated and humidified nasal high flow (NHF) therapy. This kit contains the F&P Optiflow Junior 2 and 2+ cannula (subject device) and a 12F/15M Adaptor.

    Blender Transition Kits: The F&P Optiflow Junior 2 and 2+ Blender Transition Kits allow the use of F&P Optifilow Junior 2 and 2+ cannula with approved Fisher & Paykel Healthcare breathing circuits when connected to a blender heated and humidified nasal high flow (NHF) therapy to spontaneously breathing patients who require breathing support. This kit contains F&P Optiflow Junior 2 and 2+ cannula (subject device), a 12F/15M Adaptor and a Pressure Relief Manifold (cleared in K173770).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided document.

    It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed clinical trial results for novel devices. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study, standalone algorithm performance, or specific effect sizes for human improvement with AI assistance are not applicable here as this is a medical device (nasal cannula) and not an AI/ML powered device.


    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly define "acceptance criteria" against numerical targets in the way an AI/ML device would. Instead, substantial equivalence is demonstrated through a comparison of the subject device's (F&P Optiflow Junior 2 / 2+ Nasal Cannula Interface Range) characteristics and performance against a predicate device (F&P Optiflow Junior Nasal Cannula, K162553). The "acceptance" is that the device has similar technological characteristics, intended use, and performs safely and effectively as demonstrated through non-clinical testing.

    Here's a table summarizing the comparison, with "Reported Device Performance" reflecting how the subject device aligns with or deviates from the predicate, implying acceptability.

    Acceptance Criteria (based on predicate)Reported Device Performance (F&P Optiflow Junior 2 / 2+ Nasal Cannula)Comment / Justification
    ClassificationIdentical: Matches predicate
    ManufacturerFisher & Paykel Healthcare Ltd.Identical
    Device RegulationClass II, 21 CFR 868.5450Identical
    Product CodeBTTIdentical
    Classification PanelAnesthesiologyIdentical
    Intended UseIdentical: Use with a nasal high flow therapy system to deliver heated and humidified nasal high flow therapy to spontaneously breathing patients.
    Indications for UseEquivalent: Indications for Use statements for each subject device model are equivalent when compared to the predicate device.
    Patient PopulationExpanded: Intended for pediatric populations (infants, children). Subject device has additional sizes to accommodate neonates, which expands the age range compared to the predicate (which was 1 month to 12 years).
    Patient AcuitySpontaneously breathing patientsIdentical
    Patient MonitoringAppropriate patient monitoringIdentical
    Operating EnvironmentDifferent: Subject device is for hospital and long-term care environments. Predicate was for hospital and home environments. (This difference is noted but not presented as an impediment to substantial equivalence).
    ReusabilitySingle useIdentical
    DurationSeven daysIdentical
    Range of Cannula SizesExpanded: Available in six sizes (XS, S, M, L, XL, XXL) compared to predicate's two sizes (L, XL). Additional sizes accommodate broader patient range including neonates.
    Specifications (Flow Rates)Similar/Expanded Flow Rates: L and XL sizes have equivalent flow rates for AIRVO 2 system. XXL size has increased maximum flow rate for MR850 specific to the larger pediatric patients. Overall, a broader range of cannulas and associated flow rates.
    Ambient Operating Temperature18 – 26 °CIdentical
    Shelf-LifeThree yearsIdentical
    SterilityDevice not provided sterileIdentical
    Storage Temperature-10°C to +50°CIdentical
    Performance TestingMeets Applicable Standards: Tested to ISO 5356-1, ISO 10993-1, ISO 18562-1, ISTA 2A, ASTM F1980-16, IEC 62366-1. Additional performance tests (Leak Testing, Condensate Lavage, Retention System, Nasal Prong Stability, Tubing, Connector, Human Factors, Shelf Life, Transport, Accuracy of Delivered Flow).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes non-clinical performance testing rather than a clinical study with a "test set" of patients. Therefore, specific sample sizes for patient data or data provenance (country of origin, retrospective/prospective) are not provided as it's not relevant to this type of submission. The "test set" would refer to the physical devices and materials tested in the various engineering and biocompatibility evaluations. These tests are conducted by the manufacturer, Fisher & Paykel Healthcare Ltd. (New Zealand).


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to a 510(k) submission for a non-AI medical device like a nasal cannula. "Ground truth" in this context would come from recognized standards, engineering specifications, and established biocompatibility principles, not from expert review of patient data.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods are typically used in clinical studies for interpretation of complex diagnostic results to establish a consensus ground truth, which is not relevant for the engineering and biocompatibility tests conducted for this device.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a nasal cannula, not an AI-powered diagnostic or assistive tool. No human "readers" or AI assistance are involved in its primary function, and therefore, an MRMC study or effect size related to AI improvement is irrelevant.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is not an algorithm or AI system.


    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the non-clinical tests conducted, the "ground truth" is established by:

    • International and National Standards: e.g., ISO and ASTM standards for dimensions, materials, biocompatibility, packaging, and shelf life.
    • Engineering Specifications: Internal design specifications and performance requirements for the cannula's physical properties, leak rates, flow rates, and connection integrity.
    • Risk Management Processes: Identified hazards and associated mitigations based on expected device performance.

    8. The sample size for the training set

    This is not applicable. There is no "training set" in the context of a non-AI medical device submission.


    9. How the ground truth for the training set was established

    This is not applicable as there is no "training set."