510(k) Data Aggregation

    K Number
    K203546

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-02-12

    (70 days)

    Product Code
    Regulation Number
    882.5892
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173442

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    gammaCore Sapphire (non-invasive vagus nerve stimulator) is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. gammaCore is indicated for:

    • · The preventive treatment of migraine headache in adolescent (age 12 and older) and adult patients.
    • · The acute treatment of pain associated with migraine headache in adolescent (age 12 and older) and adult patients.
    • · Adjunctive use for the preventive treatment of cluster headache in adult patients.
    • · The acute treatment of pain associated with episodic cluster headache in adult patients.
    Device Description

    The gammaCore Sapphire (gammaCore) is a multiuse, handheld, rechargeable, portable device consisting of a rechargeable battery and signal-generating and amplifying electronics, with a slide control switch for user/operator control of the signal amplitude (relative range, 0-40 continuous).
    The gammaCore Sapphire:

    • . Includes a charging station incorporated into the "clam shell" storage case connected to a power adapter for charging of the device as necessary by the end user.
    • Provides visible (light and display) and audible (beep) feedback regarding device and stimulation . status.
    • . Allows for multiple stimulations or doses; each stimulation or dose lasts 120 seconds, after which the device automatically turns off unless turned off earlier by the user/operator. Note: One dose is defined as one stimulation cycle lasting 120 seconds (2 minutes).
    • . Delivers up to a fixed number of doses within a 24-hour period; once the maximum daily number of doses has been reached, the device will not deliver any more doses until the following 24-hour period.
    • Indicates on the display the number of remaining doses available in a 24-hour period. .
    AI/ML Overview

    The provided text describes a 510(k) summary for the gammaCore Sapphire device, which is an external vagal nerve stimulator for headache. This submission seeks to expand the indications for use to include the adolescent population for the preventive and acute treatment of migraine headache.

    Based on the information provided, here's a breakdown regarding acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance

      The document does not explicitly state acceptance criteria in terms of numerical performance metrics (e.g., sensitivity, specificity, accuracy) because this is a 510(k) submission for an expanded indication of an already cleared physical medical device, not a new AI/ML-driven diagnostic or prognostic device. The "acceptance criteria" in this context are related to demonstrating substantial equivalence to a predicate device and safety and effectiveness in the new population. The substantial equivalence comparison table (Table 1) acts as the primary "performance" comparison.

      Acceptance Criteria (Implied)Reported Device Performance (Subject Device vs. Predicate)
      Intended UseSame
      Indication for UseExpanded to adolescents (age 12+) for preventive and acute migraine treatment. Supported by extrapolation from adult data and a small adolescent study.
      Rx vs OTCNo change (Prescription use)
      Treatment Protocol for Migraine (Acute & Preventive)No change compared to adults (same 120-second stimulation cycle, number of stimulations, and daily schedule)
      Patient-Contacting MaterialsNo change (SS, ABS-PC, SignaGel electrode gel)
      Electrical ClassificationNo change (UL 60601-1 Class III Type BF Applied Part)
      Waveform/FrequencyNo change (Sinusoidal wave, symmetrical biphasic 5000-Hz pulses at 25 Hz)
      Maximum OutputNo change (30V, 60 mA peak)
      Load ImpedanceNo change (450-550 ohms)
      Power SupplyNo change (3V LiFePo4 battery)
      Service LifeNo change (3 Years from date of manufacture)
      ControlsNo change (Control slide)
      Output RegulationNo change (Device software and control slide)
      Device Status DisplayDifferences in user interface that do not impact safety or effectiveness.
      Battery ChargerNo change (Qi-compatible wireless charger in clam shell storage case)
      RFID Refill/Reload CapabilityNo change (Allows refilling/reloading)
      Device Diagnostics (Bluetooth)No change (Provides diagnostics for manufacturer)
      Alarm SignalsNo change (Start-up, Session complete, Errors/depleted battery, No doses left, Expired/no days left, Low battery, Reloading error, Card error)
      Display/MessageNo change to core output display, minor user interface differences considered not to impact safety/effectiveness.
    2. Sample sizes used for the test set and the data provenance

      For the expansion of indications to adolescents, the primary "test set" for safety and effectiveness relied heavily on:

      • Extrapolation from adult clinical data: This is clearly stated, meaning the adult data from prior studies (PRESTO, PREMIUM, EVENT) served as a large part of the basis.
      • Small adolescent study: "a small study was performed evaluating the use of gamma Core Sapphire in the acute treatment of migraine in adolescent population. As described in Grazzi et al¹, in nine (9) adolescents, 46.8% of treated migraine attacks were considered successfully treated and did not require any rescue medication."
      • Sample size for adolescent study: 9 adolescents.
      • Data provenance: The Grazzi et al. study is a published peer-reviewed paper (Neurol Sci. 2017 May). The document does not specify the country of origin for the Grazzi study, but typically these are international clinical trials. It is a prospective study as it evaluates the use in a group of adolescents.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

      This information is not provided in the 510(k) summary. For a device like nVNS, the efficacy is typically self-reported by patients (e.g., pain reduction, headache freedom) and assessed by clinicians based on established diagnostic criteria for migraine. The Grazzi et al. paper would contain specifics on patient selection and outcome assessment, but that detail is not within this 510(k) summary. Given it's a treatment device, "ground truth" would be clinical outcomes (e.g., headache resolution) rather than an expert-adjudicated label on an image.

    4. Adjudication method for the test set

      Not applicable in the context of this device. Adjudication methods like 2+1 or 3+1 are common in image-based AI studies where expert consensus is needed to establish a definitive diagnosis for a test image. Here, clinical outcomes are measured after device use.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      No MRMC comparative effectiveness study was done, as this is not an AI-assisted diagnostic device. It is a physical neuromodulation device. Therefore, no information on human reader improvement with AI assistance is applicable or provided.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

      Not applicable. The gammaCore Sapphire is a physical device that delivers therapy and is used by a human patient. It does not operate as a standalone algorithm or AI.

    7. The type of ground truth used

      The ground truth for the effectiveness of the device in the expanded indication (adolescents) is based on:

      • Clinical Outcomes Data: Self-reported pain relief and resolution of migraine symptoms, as well as the absence of requiring rescue medication. The Grazzi et al. study reported "46.8% of treated migraine attacks were considered successfully treated and did not require any rescue medication."
      • Adult Clinical Trial Results: The prior 510(k)s (K173442, K191830) relied on outcomes from multi-center, randomized, double-blind, sham-controlled clinical studies (PRESTO, PREMIUM, EVENT) for adults.
    8. The sample size for the training set

      Not explicitly applicable in the context of this device because it's not an AI/ML algorithm that is "trained" on data in the traditional sense. The device's technological characteristics are fixed. The clinical studies (PRESTO, PREMIUM, EVENT, and the Grazzi et al. study) provide the evidence base that informed the safety and effectiveness, rather than a "training set" for an algorithm. The total sample sizes for these studies are not fully detailed in this 510(k) summary but would be found in the referenced prior 510(k) documents and publications. The adult studies would have had considerably larger sample sizes than the 9 adolescents mentioned.

    9. How the ground truth for the training set was established

      Again, "training set" and "ground truth" establishment in the AI/ML sense are not directly applicable. The "ground truth" for the effectiveness shown in the supporting clinical studies (for both adults and adolescents) was established through clinical trial methodology:

      • Randomized, double-blind, sham-controlled study designs comparing the device to a sham control.
      • Measurement of defined clinical endpoints, such as pain reduction, headache freedom, and reduction in headache frequency.
      • Reporting of adverse events to assess safety.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182369

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-27

    (88 days)

    Product Code
    Regulation Number
    882.5892
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173442

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    gammaCore Sapphire (non-invasive vagus nerve stimulator) is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. gammaCore is indicated for:

    • · Adjunctive use for the preventive treatment of cluster headache in adult patients.
    • · The acute treatment of pain associated with episodic cluster headache in adult patients.
    • · The acute treatment of pain associated with migraine headache in adult patients.
    Device Description

    The gammaCore Sapphire (gammaCore) is a multiuse, handheld, rechargeable, portable device consisting of a rechargeable battery and signal-generating and amplifying electronics, with a slide control switch for user/operator control of the signal amplitude (relative range, 0-40 continuous).

    The gammaCore Sapphire:

    • . Includes a charging station incorporated into the "clam shell" storage case connected to a power adapter for charging of the device as necessary by the end user.
    • . Provides visible (light and display) and audible (beep) feedback regarding device and stimulation status.
    • . Allows for multiple stimulations or doses; each stimulation or dose lasts 120 seconds, after which the device automatically turns off unless turned off earlier by the user/operator. Note: One dose is defined as one stimulation cycle lasting 120 seconds (2 minutes).
    • . Delivers up to a fixed number of doses within a 24-hour period; once the maximum daily number of doses has been reached, the device will not deliver any more doses until the following 24-hour period.
    • Indicates on the display the number of remaining doses available in a 24-hour period. ●
    AI/ML Overview

    The provided documentation describes the gammaCore Sapphire (non-invasive vagus nerve stimulator), which is indicated for:

    • Adjunctive use for the preventive treatment of cluster headache in adult patients.
    • The acute treatment of pain associated with episodic cluster headache in adult patients.
    • The acute treatment of pain associated with migraine headache in adult patients.

    The information primarily focuses on demonstrating substantial equivalence to predicate devices rather than defining specific acceptance criteria for performance benchmarks against a gold standard for a diagnostic AI device. However, based on the clinical study presented, we can infer a primary efficacy outcome as an "acceptance criteria" and the reported device performance from that study.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Primary Endpoint)Reported Device Performance (SoC + nVNS group vs. Control)
    Significantly greater reduction in the mean number of Cluster Headache (CH) attacks per week.Study Outcome: Mean reduction of 5.9 CH attacks per week for SoC + nVNS vs. 2.1 for Control.
    Therapeutic Gain: 3.9 fewer attacks per week (95% CI: 0.5, 7.2; P=0.02).
    Higher 50% responder rates (proportion of subjects with ≥50% reduction in CH attacks).Study Outcome: 40% responder rate for SoC + nVNS vs. 8.3% for Control (P<0.001).
    Reduction in the use of abortive medications (Sumatriptan and Oxygen).Study Outcome: Reduction observed in SoC + nVNS group from baseline (7.2 to 5.9 for Sumatriptan, 17.3 to 13.9 for Oxygen). Control group showed smaller or no reduction.
    Clinically meaningful improvements in Quality of Life (QoL) outcomes (EQ-5D-3L Index, VAS, HIT-6 scores).Study Outcome: Statistically significant improvements in several QoL measures for SoC + nVNS group compared to control (e.g., EQ-5D-3L Index score change of 0.113 vs -0.047; P=0.011).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Study):
      • Sample Size:
        • Safety Population: 48 subjects in the nVNS + SoC group and 49 subjects in the Control group (total 97 subjects).
        • mITT (modified intent-to-treat) Population: Numbers vary slightly depending on the specific QoL measure evaluated (e.g., 35-37 for SoC + nVNS, 40-46 for Control).
      • Data Provenance:
        • Country of Origin: Europe (multicenter, across 10 sites).
        • Retrospective or Prospective: Prospective, randomized, controlled study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The device is a therapeutic device (vagus nerve stimulator) for headache treatment, not a diagnostic AI device requiring expert-established ground truth for image or data interpretation. The "ground truth" in this context refers to the clinical outcomes and patient-reported data (e.g., number of headache attacks, abortive medication use, quality of life scores) observed and recorded during a clinical trial. These outcomes were assessed by clinical study personnel (e.g., investigators, nurses) and self-reported by patients, rather than adjudicated by independent experts in the same way a diagnostic image would be.

    4. Adjudication Method for the Test Set

    As this is a therapeutic device trial, standard clinical trial methodology was used. There isn't a stated adjudication method in the sense of multiple experts reviewing and reaching a consensus on individual patient outcomes or data points being "adjudicated." The outcomes (headache frequency, medication use, QoL scores) were derived from patient diaries and questionnaires, and then statistically analyzed. The study was described as "randomized, controlled," implying proper blinding where feasible to minimize bias in outcome reporting, though explicit details on blinding of outcome assessors weren't provided.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed to evaluate the diagnostic accuracy of imaging or AI algorithms when interpreted by multiple human readers, often with and without AI assistance. This device is a therapeutic device, not an interpretative diagnostic tool.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The gammaCore Sapphire is a physical medical device (vagus nerve stimulator) that directly delivers therapy, not a standalone algorithm. Therefore, "standalone" algorithm performance without human interaction is not applicable. The device's performance is intrinsically tied to its use by a human patient.

    7. The Type of Ground Truth Used

    The "ground truth" for evaluating the efficacy of the gammaCore Sapphire (for the preventive treatment of cluster headache) was based on patient-reported clinical outcomes and objective measures of disease activity, including:

    • Change in mean number of cluster headache attacks per week.
    • 50% responder rates (proportion of patients experiencing a ≥50% reduction in attacks).
    • Reduction in abortive medication use (sumatriptan and oxygen).
    • Quality of life outcomes (EQ-5D-3L Index score, EQ-5D-3L VAS score, HIT-6 score).

    8. The Sample Size for the Training Set

    The provided documentation does not detail a "training set" in the context of an AI algorithm. The device's efficacy was established through a clinical trial in which the treatment group received the active device. There is no mention of a pre-market "training set" for an AI component.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of an AI algorithm "training set," this question is not applicable based on the provided text. The device's efficacy was established through the clinical study described, where the "ground truth" was the observed clinical improvement in the patient population.