510(k) Data Aggregation

    K Number
    K221846

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-11

    (17 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K171866

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Artix Balloon Guiding Sheath is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Artix Balloon Guiding Sheath is also indicated for use as a conduit for retrieval devices.
    The Artix Balloon Guiding Sheath is intended for use in the peripheral vasculature.

    Device Description

    The Artix Balloon Guiding Sheath is a single-use, over-the-wire system designed to facilitate the insertion and guidance of an intravascular catheter into a selected peripheral blood vessel and act as a conduit for retrieval devices. A compliant balloon mounted at the sheath's distal tip provides temporary vascular occlusion during angiographic procedures. The Artix Balloon Guiding Sheath is packaged with the following components: Artix Balloon Guiding Sheath (8 Fr, 65 cm or 105 cm) Two 8 Fr Introducer Dilators (0.014" and 0.035" guidewire compatibility) Balloon Inflation Syringe, 1 mL Large Bore Syringe, 30 mL 3-way Stopcock

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on what is present and noting what is not explicitly detailed.

    Acceptance Criteria and Reported Device Performance

    The document states: "Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications." However, it does not explicitly list the quantitative acceptance criteria for each test. It only lists the tests performed and claims that the results met the criteria.

    Acceptance Criteria (Quantitative)Reported Device Performance (Quantitative)
    Not explicitly detailed in the provided text for individual tests. The general statement is that "all acceptance criteria were met."Not explicitly detailed in the provided text for individual tests. The general statement is that "all acceptance criteria were met."

    Study Information:

    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated for each non-clinical test. The document lists the types of tests performed (e.g., Pouch Seal Peel Test, Simulated Use Cycling, Balloon Fatigue Test, Sheath Kink to Failure, etc.) but does not provide the number of units tested for each.
      • Data Provenance: The tests are non-clinical, likely performed in an Inari Medical lab or by a contract testing organization. The document does not specify a country of origin for the data or whether it was retrospective or prospective, as these terms are more applicable to clinical studies. These are laboratory/benchtop tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not applicable to the non-clinical (benchtop) tests described. Ground truth for these types of tests would be established by predefined engineering specifications, industry standards (e.g., ISO, AAMI), and the performance characteristics of the predicate device.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically used in clinical studies or image interpretation studies where human judgment is involved in establishing ground truth. For non-clinical tests, the criteria are objective and measurable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The document explicitly states: "Neither animal testing nor clinical testing were required for the determination of substantial equivalence." Therefore, no MRMC study, human reader improvement analysis, or AI assistance was involved since this submission is for a medical device (Balloon Guiding Sheath) and not an AI/Software as a Medical Device (SaMD).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable, as this is a physical medical device, not an algorithm or SaMD.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical tests, the "ground truth" or reference for comparison would be:
        • Predetermined engineering specifications: Based on the device's design requirements.
        • Relevant industry standards: Such as ISO 10993-1 for biocompatibility and ISO 11135:2014/Amd 1:2018 and AAMI TIR 28:2016 for sterilization.
        • Performance of the predicate device: To demonstrate substantial equivalence in key performance characteristics relevant to safety and effectiveness.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/machine learning device; therefore, there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203670

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-02-16

    (62 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160254, K171866

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Extra Large Check-Flo introducers and guiding sheaths are intended to introduce therapeutic devices into the vasculature.

    Device Description

    Cook Incorporated's Extra Large Check-Flo® Introducer is a single-use, sterile, disposable product that is used to introduce therapeutic or diagnostic devices into the vasculature. The Extra Large Check-Flo® Introducer consists of an introducer sheath, Check-Flo® hemostatic valve assembly, and dilator. The sheath is constructed with a hemostatic valve that accepts a large range of French sizes while preventing blood reflux and air aspiration. The subject device, the Extra Large Check-Flo® Introducer, is manufactured in diameters of 20.0, 22.0 and 24.0 French with sheath lengths of 25, 40, or 65 centimeters. The dilator is manufactured in lengths 16 centimeters longer than the sheath (41, 56, and 81 centimeters, respectively). The diameter of the distal endhole of the dilator measures 0.035 inch. The introducer sheath and dilator are matched as a unit to provide a smooth transition.

    AI/ML Overview

    Based on the provided text, the document is a 510(k) Premarket Notification for a medical device called the "Extra Large Check-Flo® Introducer." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as would be required for a PMA (Premarket Approval) application.

    Therefore, the information typically requested in your prompt regarding acceptance criteria for AI/ML device performance (e.g., sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance, ground truth establishment) is not applicable to this document. This document describes the acceptance criteria and study that proves a physical medical device (a catheter introducer) meets performance specifications, not an AI/ML-based diagnostic or therapeutic device.

    However, I can extract the acceptance criteria and the performance results for the physical device as described in the provided text.

    Here's the relevant information from the document:

    1. A table of acceptance criteria and the reported device performance:

    The document lists various tests performed and states that the "Test results met predetermined criteria" or "The predetermined acceptance criterion was met." Specific numerical acceptance criteria values are generally not provided in this summary, but the success of meeting them is stated.

    Test PerformedAcceptance Criteria (as described)Reported Device Performance (as described)
    BiocompatibilityPer ANSI AAMI ISO 10993-1All materials are biocompatible.
    Tensile TestingPer BS EN ISO 11070Test results met predetermined criteria.
    Compatibility TestingPer BS EN ISO 11070Test results met predetermined criteria.
    Integrity TestingSheath tip no visible damage (cracks, splits, rollback); Sheath patent and unwrinkled upon dilator withdrawalTest results met predetermined criteria.
    Rollback TestingPer BS EN ISO 11070, Annex AThe predetermined acceptance criterion was met.
    Radiopacity EvaluationPer ASTM F640-12Test results met predetermined criteria.
    Dilator Removal Force TestingMet specification and comparable to reference device (K171866)The force to remove the dilator from the sheath met the specification and was comparable to the reference device.
    Dimensional TestingEffective dimensions derived from manufacturing specificationsEffective product dimensions derived from the manufacturing specifications were confirmed.
    Simulated Use TestingReliability in a clinical simulated use scenarioSubject device tested through an anatomical model to test the subject device's reliability in a clinical simulated use scenario. (Implied successful performance)
    Valve LeakagePer BS EN ISO 11070Tested in accordance with the acceptance criterion established in BS EN ISO 11070. (Implied successful performance)

    Regarding the other points of your prompt (2-9), these are not applicable to the information provided in this 510(k) summary for a physical medical device:

    • 2. Sample sized used for the test set and the data provenance: Not applicable. This refers to a physical device, not an AI/ML model with test data. The number of physical units tested per criterion is not specified.
    • 3. Number of experts used to establish the ground truth... and the qualifications of those experts: Not applicable. Ground truth as typically understood in AI/ML (e.g., expert labels on images) is not relevant here. Device performance is measured by physical tests according to standards.
    • 4. Adjudication method: Not applicable. Device performance is determined by adherence to engineering specifications and international standards, not expert adjudication of diagnostic outcomes.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done...: Not applicable. This is for AI/ML diagnostic or assistive devices, not for a catheter introducer.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. No algorithm involved.
    • 7. The type of ground truth used: Not applicable. Device performance is measured against engineering specifications and test standards (e.g., tensile strength, dimensions, biocompatibility), not clinical "ground truth" derived from patient outcomes or expert consensus in the diagnostic sense.
    • 8. The sample size for the training set: Not applicable. This is a physical device, not an AI/ML model that undergoes training.
    • 9. How the ground truth for the training set was established: Not applicable. See point 8.

    In summary: The provided document is a 510(k) summary for a physical medical device. It demonstrates performance by stating that the device passed a series of engineering and biocompatibility tests in accordance with established industry standards and internal specifications, thus supporting its substantial equivalence to a predicate device. It does not contain information related to AI/ML model performance or associated clinical study designs.