Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K181111

    Validate with FDA (Live)

    Date Cleared
    2018-05-17

    (20 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K163375, K111520, K123270, K050438, K170011, K124004

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC™ POWEREASETM System.

    Device Description

    The subject Navigated INFINITY™ drill bit is a sterile, single-use instrument that can be operated manually or under power. This instrument is intended to be used when implanting components of the INFINITY™ OCT System. The Navigated INFINITY™ Drill bit is also compatible with the StealthStation® and IPC® POWEREASE® Systems. Note: The subject devices are not intended to support occipital screw placement.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving a device meets them. The document is an FDA 510(k) premarket notification for "Navigated INFINITY™ Instruments," detailing its indications for use, comparison to a predicate device, and sterilization information.

    Specifically, the document states:

    • No acceptance criteria or study details: The "Performance Data" section (VII) only mentions "Sterilization" and that "The addition of the sterile drill bit does not create a new 'worst case' and it was adopted in to the existing product family as detailed in the sterilization rationale provided in this submission since the design, material, manufacturing process, and manufacturing facilities of the sterile subject devices are equivalent to that of the referenced product family. The sterilization justification supports an SAL of 10-6. Based on the information provided in this submission, Medtronic believes that adequate evidence has been provided to show that we have fully validated the sterilization parameters." This indicates that the main "performance data" presented is a sterilization justification, not a study with acceptance criteria for device performance.
    • Focus on substantial equivalence: The document's goal is to demonstrate substantial equivalence to previously cleared devices, primarily based on identical indications, intended use, technology, materials, dimensions, and design features, with the only technological difference being that the subject device is sterile and single-use, while the predicate was non-sterile and single-use. This type of submission often relies on a comparison to a predicate device rather than extensive new performance studies if the changes are minor and do not introduce new questions of safety or effectiveness.

    Therefore, I cannot populate the requested table and information as the necessary details are not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173338

    Validate with FDA (Live)

    Date Cleared
    2018-01-18

    (87 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K163375, K111520, K123270, K050438, K170011, K124004

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC® POWEREASE® System.

    Device Description

    The Navigated INFINITY™ Instruments are non-sterile, reusable and single use instruments that can be operated manually or under power. These instruments are intended to be used when implanting components of the INFINITY™ OCT System. The Navigated INFINITY™ Instruments are also compatible with the StealthStation® and IPC® POWEREASE® Systems.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the "Navigated INFINITY™ Instruments":

    It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study report for new device approval. As such, the information provided for some of your requested points is limited or not applicable in the context of a 510(k).


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Maintain Navigation AccuracyNavigation Accuracy Analysis Confirmed navigated instrument accuracy
    Instrument Functionality under Expected Use ConditionsAnatomical Simulated Use Confirmed instrument functionality under expected use conditions
    Navigation System Functionality under Expected Use ConditionsNavigation Simulated Use Confirmed navigation system functionality under expected use conditions
    Accurate Reflection of CAD Models in Application SoftwareCAD Model Evaluation Verified that the CAD models are accurately reflected in the application software
    Proper Mating of Instruments with Appropriate DevicesImplant/Instrument Mating Conditions Verified that the instruments can be assembled with the appropriate devices according to their intended use
    Spine Tools Package Meets Interface Needs of Spine Application SoftwareSpine Tools Package Functional Testing Verified that the Spine Tools package has met the required interface needs of the spine application software

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not specify the sample size (e.g., number of instruments tested, number of simulated cases) for any of the performance tests.
      • Data Provenance: Not specified. Given the nature of a 510(k) and the types of tests listed (simulated use, accuracy analysis, CAD evaluation, functional testing), it is highly probable that the data provenance is retrospective (using internal testing data) and likely in-house laboratory or engineering testing, rather than clinical data from a specific country of origin.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. For the types of engineering and functional tests described (e.g., navigation accuracy, instrument functionality, CAD evaluation), "ground truth" would likely be established through engineering specifications, validated measurement systems, and perhaps expert observation from R&D engineers, rather than clinical experts like radiologists.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not provided and is unlikely to be relevant for the non-clinical, performance-based tests described in the 510(k) summary. Adjudication methods like 2+1 are typically used for clinical endpoints, especially in image interpretation.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. The device (Navigated INFINITY™ Instruments) is a surgical navigation instrument, not an AI-powered diagnostic or assistive tool for "human readers." Therefore, this type of study and its associated effect size are not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The term "standalone" and "algorithm only" are typically used for AI/software devices. This device is a physical surgical instrument that works with existing navigation systems (StealthStation™ System, IPC® POWEREASE® System). The performance tests listed (e.g., Navigation Accuracy Analysis, Anatomical Simulated Use) assess the instrument's accuracy and functionality within the intended surgical navigation system, which inherently involves human interaction (the surgeon). Therefore, a "standalone algorithm only" performance is not applicable in the context of this device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the tests described:
        • Navigation Accuracy Analysis: Ground truth would likely be established through pre-defined engineering specifications and validated measurement systems (e.g., comparing navigated position to physically measured, known positions).
        • Functional/Simulated Use Tests: Ground truth would be defined by the device's design specifications and intended functionality (i.e., whether it performs as designed under various conditions).
        • CAD Model Evaluation: Ground truth would be the original CAD models themselves and the expectation of their accurate representation in the software.
        • No expert consensus, pathology, or outcomes data is explicitly mentioned or seems relevant for these specific engineering tests.
    7. The sample size for the training set:

      • This question is not applicable. The "Navigated INFINITY™ Instruments" are physical surgical instruments, not a machine learning model that requires a training set. The underlying navigation systems (StealthStation™) might have calibration data or algorithms, but the instruments themselves do not have a "training set" in the sense of AI.
    8. How the ground truth for the training set was established:

      • This question is not applicable as there is no "training set" for these physical instruments.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1