510(k) Data Aggregation

    K Number
    K252093

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-16

    (105 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K243040, K210545, K163329, K223544

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Light Therapy Mask is an over-the-counter device intended for the treatment of full-face wrinkles and mild to moderate inflammatory acne using LED light therapy.

    • Red light: Treatment of full-face wrinkles.
    • Blue light: Treatment of mild to moderate inflammatory acne.
    • Red+Infrared Light: Treatment of full-face wrinkles.
    • Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne.
    Device Description

    The subject device Light Therapy Mask is a home use wearable LED phototherapy device whose purpose is to produce an even and narrow-band of light for the treatment of aesthetic indications including facial wrinkles and acnes.

    The subject device consists of a mask body unit that contains light emitting diodes (LEDs), a controller, straps and USB-C charging cable. And the device is powered by built-in rechargeable lithium battery on the controller.

    The LEDs generate 3 kinds of light which include Red light (wavelength 635nm±5nm), Blue light (wavelength 465nm±5nm), Infrared light (wavelength 850nm±5nm). A controller is connect to the mask body unit to control the device, such as turn on/off the device. And the straps used for securing the mask unit to the body part.

    The device have vibration function, and user can adjust the vibration intensity. Vibration is for general relaxation purposes.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243978

    Validate with FDA (Live)

    Device Name
    LED Facial Mask
    Date Cleared
    2025-06-20

    (179 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K162652, K163329, K162489

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LED Facial Mask is an over-the-counter (OTC) device intended for the following uses:

    For Red Mode(s): The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.

    For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.

    For Blue Mode: The device emits light in the blue region of the spectrum and is specifically indicated for the treatment of mild to moderate acne on the face.

    For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

    Device Description

    The LED Facial Mask is a device used for treatment of full-face wrinkles, mild to moderate acne on the face, and providing topical heating. LEDs housed inside the device emit light onto the face. These LEDs generate blue, red, and infrared wavelengths.

    This product is composed of the main device, controller, fastening straps, storage bag, blindfold, and charging cable.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to an LED Facial Mask, which is categorized as a Class II medical device. The information provided outlines the manufacturing, intended use, and comparison to predicate devices, but it does not contain the specific performance data one would expect for a study proving the device meets acceptance criteria.

    The document primarily focuses on demonstrating substantial equivalence to already cleared predicate devices based on technological characteristics and intended use. The "Performance Data" section lists various recognized consensus standards to which the device was tested (e.g., electrical safety, electromagnetic compatibility, usability, biological evaluation), rather than clinical study results establishing efficacy.

    Therefore, I cannot create the detailed table of acceptance criteria and reported device performance, nor can I answer questions about sample size, ground truth establishment, expert adjudication, or MRMC studies, as this information is not present in the provided FDA clearance letter.

    It explicitly states: "Clinical/animal testing was not performed for the proposed device as part of the submission." This confirms the absence of efficacy studies that would typically define acceptance criteria for performance outcomes.

    The acceptance criteria for this device, as implied by the FDA 510(k) process for a Class II device, are largely related to safety, performance against engineering standards, and substantial equivalence to legally marketed predicate devices. The clearance is based on the premise that if the new device is sufficiently similar to existing cleared devices in terms of its technology and intended use, and meets relevant safety standards, it does not require new clinical efficacy data.

    Here's a breakdown of what can be inferred from the document regarding acceptance criteria and the "study" (non-clinical testing) that proves the device meets them:


    Inferred Acceptance Criteria and Device "Performance" (Non-Clinical)

    Since no clinical efficacy studies were performed, the "acceptance criteria" are not related to a specific clinical outcome (e.g., X% reduction in wrinkles, Y% clearance of acne). Instead, they relate to safety, electrical performance, and biocompatibility, as demonstrated through testing against consensus standards.

    Acceptance Criterion (Inferred from Standards)Reported Device Performance (Implied by Clearance)
    Electrical Safety (IEC 60601-1, -2-57, -2-83)Device meets general requirements for basic safety and essential performance, particular requirements for non-laser light source equipment, and home light therapy equipment. Implies safe power delivery, insulation, and fault protection.
    Electromagnetic Compatibility (IEC 60601-1-2)Device meets requirements for electromagnetic disturbances. Implies suitable electromagnetic emissions and immunity to interference.
    Usability (IEC 60601-1-6)Device meets usability requirements. Implies the device is designed for safe and effective user interaction.
    Home Healthcare Environment Suitability (IEC 60601-1-11)Device meets requirements for use in the home healthcare environment. Implies robustness and safety for lay users.
    Photobiological Safety (IEC 62471)Device meets photobiological safety standards. Implies emitted light is safe for the eyes and skin under specified use conditions.
    Software Life Cycle Processes (IEC 62304)Device software development followed medical device software life cycle processes. Implies software is engineered to be safe and reliable.
    Biocompatibility (ISO 10993-1, -5, -10, -23)Materials in contact with the user are biologically evaluated against standards for cytotoxicity, skin sensitization, and irritation. Implies materials are safe for human contact.
    Wavelength AccuracyBlue: 410 nm, 460 nm; Red: 610 nm, 630 nm, 660 nm; NIR: 830 nm, 850 nm, 880 nm. (Measured against specified ranges)
    Intensity/Irradiance (mW/cm²)Red1 (Y) Mode: 36.0 mW/cm²; Red2 (R) Mode: 19.0 mW/cm²; Blue (B) Mode: 6.1 mW/cm²; NIR Mode: 45.0 mW/cm²; Red1 (Y)+NIR Mode: 81.0 mW/cm²; Red2 (R)+NIR Mode: 64.0 mW/cm². (Measured against specified output)
    Substantial Equivalence to Predicate DevicesDevice has similar intended use, technological characteristics (light source, anatomical location, wavelength, intensity/irradiance), and safety profile to previously cleared devices.

    Study Details (Based on the Provided Document)

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable in the context of clinical data for efficacy. The "test set" here refers to the device itself being tested against engineering and safety standards. There is no patient sample size.
      • Data Provenance: The tests conducted are in vitro (bench testing) and in silico (software analysis) or conducted on the physical device itself, not on human subjects. Therefore, there's no country of origin for patient data, nor is it retrospective or prospective clinical data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The "ground truth" for the engineering and safety tests is defined by the consensus standards themselves (e.g., a certain voltage tolerance, a specific emission limit). The "experts" would be the engineers and technicians performing the tests and verifying compliance with the standards, not clinical experts establishing ground truth for medical conditions.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This concept pertains to clinical studies where human interpretation of medical data needs to be harmonized. For engineering tests, results are typically objective measurements against a standard.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study is for AI/CADe devices and involves human readers reviewing cases. This device is an LED facial mask, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This applies to AI/CADe devices.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this clearance is adherence to recognized consensus standards for safety, electrical performance, photobiological safety, and biocompatibility, as well as demonstrated technical similarity and intended use alignment with predicate devices. There is no clinical ground truth (e.g., biopsy results, clinician's diagnosis) involved in this substantial equivalence determination.
    7. The sample size for the training set:

      • Not applicable. There is no AI model or "training set" for this device.
    8. How the ground truth for the training set was established:

      • Not applicable.

    In summary, the provided FDA 510(k) clearance document for the LED Facial Mask demonstrates that the device was cleared based on its substantial equivalence to existing devices and its adherence to a series of non-clinical safety and performance standards. No clinical efficacy studies were conducted or required for this particular clearance, as explicitly stated in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213285

    Validate with FDA (Live)

    Device Name
    easy Claire
    Date Cleared
    2022-03-03

    (153 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K163329

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The easy Claire is an over-the-counter medical device intended for the treatment of full face wrinkles.

    Device Description

    The easy Claire is a device that allows the emission of 630nm LED light and 850nm IRED light on the face, which induces photobiological effects for the treatment of facial wrinkles. The easy Claire is an over-the-counter device and consists of a collection of 99 LEDs (630nm) and 99 IREDs (850nm) for the treatment of facial wrinkles. Users place the lightweight mask over the face and use the touch switch button to operate the easy Claire. The device will automatically turn off after treatment. To prevent irradiation of LED lights to the eye during the treatment, easy Claire has a protective eye-shield that blocks light from LEDs. The device is powered by the internal rechargeable lithium-ion battery which is recharged by the specified external adapter with Input AC 100-240V, 50/60Hz, and Output DC 5V, 2A.

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria or a study that proves the device meets the acceptance criteria in the manner requested (i.e., in terms of AI/algorithm performance against a ground truth and human reader performance).

    The "510(k) Summary" (pages 3-5) describes the "easy Claire" device, which is a light-based mask for the treatment of full-face wrinkles. The summary primarily focuses on establishing substantial equivalence to a predicate device (Rejuvalite MD) and a reference device (Pulsaderm Wrinkle Mask).

    The "Performance Data" section (page 5) lists a series of non-clinical tests conducted to demonstrate the safety and effectiveness of the device. These tests align with various IEC and ISO standards related to medical electrical equipment, electromagnetic disturbances, usability, home healthcare environments, photobiological safety, and biocompatibility (in vitro cytotoxicity, irritation, and skin sensitization). Risk management was also conducted according to ISO 14971.

    There is no mention of:

    • Acceptance criteria related to algorithm performance (e.g., sensitivity, specificity, AUC).
    • A test set with a specific sample size for evaluating an algorithm or human readers.
    • Data provenance (country of origin, retrospective/prospective).
    • Experts establishing ground truth or their qualifications.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Training set sample size or how its ground truth was established.

    This document is a 510(k) summary, which typically demonstrates substantial equivalence of a new device to an existing legally marketed predicate device, rather than providing detailed clinical study results of a novel AI-powered diagnostic or treatment device. The easy Claire product appears to be a physical light therapy device, not an AI/software-as-a-medical-device (SaMD) that would require the types of performance studies outlined in the prompt's request.