Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K162878

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Embryo Transfer Catheter is intended for introduction of in vitro fertilized embryo(s) into the uterine cavity.

    The stylet is intended to be used with Vitrolife's Embryo Transfer Catheters to assist in uterine access of the guide during an embryo transfer procedure.

    Device Description

    Vitrolife´s Embryo Transfer Catheter (ETC) Assortment consists of three embryo transfer catheters (ETCs), one pre-curved variant (Model 17500, 230 mm) and two straight variants (REFs 17501, 17502) in different lengths (180 mm and 230 mm); and two stylets (Stylets) (REFs 17510, 17511) in different lengths (180 mm and 230 mm). ETC and Stylet are sterile, single-use devices used to deliver in vitro fertilized embryos to the uterine cavity.

    Embryo Transfer Catheter consists of:

    • A guide, also referred as outer sheath. The guide is used to navigate through the cervix canal. The guide has a distance marking, a stopper and a rounded tip to facilitate proper positioning. The guide is available in two different variants:
      • Pre-curved - stiff
      • Straight - soft and malleable
    • A catheter, also referred as inner catheter. The catheter is loaded with the embryo(s) in a small volume of transfer medium and inserted through the guide to gently deposit the embryo(s) into the uterine cavity. The transfer catheter is soft, flexible and approximately 50 mm longer than the guide. The tip of the catheter has an echogenic marking to enable ultrasound guidance.

    Embryo Transfer Catheter Stylet consists of:

    • A plastic-coated metal wire, which is malleable. The stylet is an accessory that can be used together with Vitrolife's guides to make them stiffer.

    The 17500 variant includes a stylet. The 17501 and 17502 variants do not include stylets, but the stylet can be purchased separately.

    AI/ML Overview

    The FDA Clearance Letter for Vitrolife Sweden AB's Embryo Transfer Catheter (ETC) Assortment primarily describes the non-clinical performance testing conducted to demonstrate substantial equivalence to a predicate device. This letter does not describe an AI/ML-based device, nor does it detail studies involving human experts, ground truth adjudication, or MRMC studies typically associated with AI/ML medical devices.

    Therefore, many of the requested details regarding acceptance criteria and study information (particularly points 2-9 related to AI/ML device testing) cannot be extracted from the provided text.

    However, I can provide the acceptance criteria and reported performance for the non-clinical tests that were performed, as outlined in the document.

    Summary of Device Acceptance Criteria and Performance (Based on Non-Clinical Testing):

    The acceptance criteria for this device are based on its physical properties, sterility, biocompatibility, and functionality, demonstrating its substantial equivalence to an existing (predicate) device. The studies described are primarily bench performance tests, material tests, and sterilization/packaging validations.

    Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
    SterilizationSterilization Validation (E-Beam)Compliance with ISO 11137-1:2006/(R)2015 and ISO 11137-2:2013Performed according to standards. (Implied compliance for clearance.)
    Packaging IntegrityVisual InspectionCompliance with ASTM F1886/F1886M-16Met predetermined acceptance criteria.
    Dye PenetrationCompliance with ASTM F1929-15Met predetermined acceptance criteria.
    Seal StrengthCompliance with ASTM F88/F88M-23Met predetermined acceptance criteria.
    TransportationTransportation Simulation TestingCompliance with ASTM 4169-22 and subsequent package integrity testingMet predetermined acceptance criteria.
    BiocompatibilityCytotoxicityNon-cytotoxic (per ISO 10993-5)Non-cytotoxic.
    SensitizationNon-sensitizing (per ISO 10993-10)Non-sensitizing.
    IrritationNon-irritating (per ISO 10993-23)Non-irritating.
    EndotoxinBacterial Endotoxin Assay (BEA)< 20 EU/device (per USP <85> and USP <161>)< 20 EU/device.
    EmbryotoxicityMouse Embryo Assay (MEA)≥80% embryos developed to expanded blastocyst at 96h (One-Cell MEA) (per 2021 FDA guidance)≥80% embryos developed to expanded blastocysts within 96 hours.
    Bench Performance (General)Visual Inspection (device, packaging, labels)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    Dimensional VerificationMet predetermined acceptance criteriaMet predetermined acceptance criteria.
    Compatibility Testing (catheter, guide, stylet)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    Usable Length Verification (assembled guide and catheter)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    Echogenic Marking Verification by UltrasoundMet predetermined acceptance criteriaMet predetermined acceptance criteria.
    Luer Lock Connection VerificationCompliance with ISO 80369-7:2021Met predetermined acceptance criteria.
    Aspiration, Flush, Leak, Flowrate Tests (catheter)Compliance with EN 1618Met predetermined acceptance criteria.
    Distance Marking Location and DurabilityMet predetermined acceptance criteriaMet predetermined acceptance criteria.
    Curvature Verification (Pre-curved guide)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    Dislodgement Force (positioning stopper)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    Tensile Strength Testing (catheter)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    Echogenic Marking Visibility Test (catheter)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
    **Bench Performance (Specific) **Tip Drop Angle of Simulated Loaded Catheter when held horizontallyMet predetermined acceptance criteria (with mitigation)"All test articles met the predetermined acceptance criteria in the tests, except for 'tip drop when held horizontally.'" The device labeling includes a precaution.

    Regarding the specific questions asked for an AI/ML device:

    1. A table of acceptance criteria and the reported device performance: Provided above for the non-clinical tests described.
    2. Sample sized used for the test set and the data provenance: Not applicable. These were non-clinical, benchtop, and laboratory tests. No "test set" in the context of patient data or clinical images is mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these physical/chemical tests is established by laboratory standards and measurements, not expert human interpretation of data.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the non-clinical tests, ground truth was derived from standardized testing methods, material specifications, and validated laboratory procedures (e.g., spectrophotometry for cytotoxicity, bacterial growth for endotoxin, visual inspection, dimensional measurements, mechanical testing).
    8. The sample size for the training set: Not applicable. This is not an AI/ML device, so there is no "training set."
    9. How the ground truth for the training set was established: Not applicable.

    In conclusion, the provided FDA clearance letter pertains to a conventional medical device (Embryo Transfer Catheter) and not an AI/ML artificial intelligence medical device. Therefore, the detailed requirements for AI/ML device testing (e.g., data sets, expert review, MRMC studies) are not found in this document. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the rigorous non-clinical, laboratory, and benchtop performance testing that ensured the device's safety and effectiveness compared to a legally marketed predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1