Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182337

    Validate with FDA (Live)

    Device Name
    HYBRID Guidewire
    Manufacturer
    Date Cleared
    2018-10-04

    (37 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161803, K022357

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HYBRID Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in the coronary arteries.

    Device Description

    HYBRID Guidewire is a coiled wire that is designed to fit inside a percutaneous microcatheter for the purpose of directing the catheter through a blood vessel. It is used in association with catheters to enable insertion and easier navigation into the intracranial vascular branches for diagnostic or therapeutic use. Guidewires constitute an important element that conditions the success of the operation. They can be pushed and torqued to move forward through patient's tortuous anatomy.

    HYBRID Guidewire consists of a proximal coated stainless steel core wire and a distal coated nitinol core wire. The distal core wire is tapered at the distal tip and is contained within a 92/8 platinum/tungsten and stainless steel coil for maximum radiopacity. The distal tip comes in various tip designs; it can be straight or curved distal tip (forming a double angle).

    The coil section of the HYBRID Guidewire and the distal stainless steel section is coated with hydrophilic coating, while the proximal stainless steel section is coated with Polytetrafluoroethylene (PTFE). The purpose of the surface coatings is to provide lubricity when HYBRID Guidewire is passed through compatible microcatheters. A torquer and an introducer are also provided with the device.

    AI/ML Overview

    The HYBRID Guidewire is a Class II device intended for general intravascular use, including neuro and peripheral vasculature, to facilitate the selective placement of diagnostic or therapeutic catheters. It is not intended for use in coronary arteries.

    Here's a breakdown of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual and Dimensional InspectionExternal surface free from process and surface defects; dimensions (overall length, radiopaque length, outer diameter, distal shape) per product specification.All test samples passed testing.
    Catheter Compatibility & Simulated UseIn vitro performance under simulated use conditions using a worst-case tortuous ICA vessel model.All test samples passed testing.
    Coating Integrity and AdherenceCoating quality (after dyeing and simulated use) inspected under a microscope.All test samples passed testing.
    Tensile StrengthMinimum force to break the guidewire meets ISO 11070. Tensile strength comparable to the predicate device.All test samples passed testing. Tensile strength was comparable to predicate devices.
    Corrosion ResistanceNo visual evidence of corrosion after immersion in sodium chloride solution and boiling distilled/deionized water.All test samples passed testing.
    Torque StrengthNumber of turns-to-failures comparable to the predicate device.All test samples passed testing. The torque strength of the subject device is equivalent to the predicate device.
    TorqueabilityRotational input to output ratio comparable to the predicate device.All test samples passed testing. The torqueability value of the subject device is equivalent to the predicate device.
    Fracture Resistance and Flexing TestResistance to damage by flexing and resistance to fracture per ISO 11070.All test samples passed testing.
    Tip FlexibilityForce required to deflect the distal tip (45° and 90°) comparable to the predicate device.All test samples passed testing. The tip flexibility of the subject device is equivalent to the predicate device.
    Particulate Matter CharacterizationParticulate matter in injections quantified after simulated use comparable to the predicate device.All test samples passed testing. Particulate matter characterization was comparable to predicate device.
    Design Verification and Packaging ValidationDevice design and packaging design meet product specification requirements at t=1 year time-point after exposure to 1x EtO sterilization.All test samples passed testing.
    Packaging Integrity (Visual Inspection)Packaging free of damage; all seals free of incomplete seals, voids, channels, wrinkles, foreign material, and/or over-heated sections. Labels legible, intact, no damage or delamination.All test samples met the acceptance criteria.
    Packaging Integrity (Bubble Leak Test)No leaks per ASTM F2096-11.All test samples met the acceptance criteria.
    Packaging Integrity (Seal Strength)Meets seal strength per ASTM F88/F88M-15.All test samples met the acceptance criteria.
    Animal Testing (GLP)In vivo performance in an acute porcine model: Trackability, handling, radiopacity, and microcatheter compatibility comparable to predicate devices.All test samples passed testing. Trackability and handling of the guidewire, radiopacity, and compatibility with microcatheter were comparable to predicate device.
    Biocompatibility (Cytotoxicity)Non-cytotoxic (percent viability of test article not less than 70% compared to cells exposed to control).The percent viability of the test article was 110.6%. Non-cytotoxic.
    Biocompatibility (Sensitization)Did not induce delayed sensitization.Did not elicit sensitization response.
    Biocompatibility (Irritation)Non-irritant.Non-irritant.
    Biocompatibility (Acute Systemic Toxicity)No evidence of significant systemic toxicity or mortality.Non-toxic.
    Biocompatibility (Pyrogenicity)Non-pyrogenic (rabbit temperature rise not exceeding 0.5°C).Non-pyrogenic.
    Biocompatibility (Hemolysis)Non-hemolytic (mean hemolytic index for direct contact and extract not exceeding 2%).Non-Hemolytic. (Mean hemolytic index for direct contact was 0.71%; for extract was 0.61%).
    Biocompatibility (Thrombogenicity)Minimal thrombus formation; similar thromboresistance characteristics as control devices.Passed. Minimal thrombus formation associated with test devices; neither animal's clotting abilities were compromised. Similar thromboresistance characteristics to control devices.
    Biocompatibility (Complement Activation)Statistically similar to negative control data.Passed. Statistically similar to the reference negative control data (p > 0.050).
    Biocompatibility (Partial Thromboplastin Time)Within the same thrombogenicity category (minimal activator); responses considered similar to predicate.Passed. Test article and predicate percent plasma control values fell within the same thrombogenicity category (minimal activator). Responses considered similar.
    Biocompatibility (Platelet and Leukocyte Counts)Test article comparable to reference control and comparison article (not statistically significant, p > 0.05).Passed. Test article results for leukocyte and platelet counts showed 100.6-105.6% of negative control. Not statistically significant (p > 0.05) when compared to reference material or comparison article. Comparable to both.
    Shelf Life (Accelerated Aging)All acceptance criteria met at T=1 year accelerated aging, no new questions of safety or effectiveness raised. Performs as intended to Design Specification.All acceptance criteria were met. Concluded that the HYBRID Guidewire will perform as intended to the Design Specification. Labeled for 1-year shelf life.
    Sterilization ValidationAchieves a minimum Sterility Assurance Level (SAL) of 10-6 in accordance with ISO 11135:2007.The validation study demonstrated that the sterilization process and equipment are capable of reliably and consistently sterilizing the devices to a minimum SAL of 10-6. Fractional cycles, half cycles, and full cycles met the acceptance criteria.
    EO and ECH ResidualsResidual traces of Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) below limits specified in ISO 10993-7:2008.The residual traces of EO and ECH for the worst-case device were well below the limits specified in ISO 10993-7.
    Bacterial Endotoxin LevelsBelow 2.15 EU/device, in accordance with FDA guidance, USP <85>, and European Pharmacopeia BET 2.6.14.< 2.15 EU/device.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the exact sample sizes for each bench test. However, it indicates "All test samples" in the results for most bench tests, implying an adequate number of samples were tested to demonstrate compliance.

    For the Animal Testing (GLP), it refers to an "acute porcine model" but does not specify the number of animals used. The provenance of the data is from these controlled laboratory (preclinical) studies, as implied by "Performance Bench Testing and Animal Testing." These are prospective studies conducted specifically for regulatory submission.

    For Biocompatibility testing, various in vitro and in vivo tests were performed:

    • Cytotoxicity: L-929 mouse fibroblast cells.
    • Sensitization: Guinea pig (not specified number but implies multiple for statistical significance).
    • Irritation: Rabbits (not specified number but implies multiple for statistical significance).
    • Acute Systemic Toxicity: Mice (not specified number but implies multiple for statistical significance).
    • Pyrogenicity: Rabbits (not specified number but implies multiple for statistical significance).
    • Hemocompatibility (Hemolysis, Thrombogenicity, Complement Activation, PTT, Platelet and Leukocyte Counts): Human blood or canine models. For thrombogenicity, it mentions "neither animal's clotting abilities," suggesting at least two animals were used, likely canine.

    The provenance for biocompatibility testing is also from controlled laboratory (preclinical) studies, conducted prospectively for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information (number and qualifications of experts) is typically relevant for studies involving subjective human interpretation, such as clinical trial imaging reads or pathology reviews. For a medical device like a guidewire, the "ground truth" is established through objective, quantifiable engineering and biological tests against established standards (e.g., ISO, ASTM) and comparison to predicate devices, rather than expert consensus on individual "cases." Therefore, this section is not directly applicable to the provided data. The "experts" are implicitly the engineers and scientists conducting and analyzing these performance and biocompatibility tests according to their respective standard operating procedures and regulatory guidelines.

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies when multiple human readers interpret data, and there's a need to resolve disagreements to establish a definitive "ground truth." Since the provided data details bench and animal testing, which rely on objective measurements and established protocols, an adjudication method in this sense is not applicable. The "adjudication" is inherent in the test methodology (e.g., meeting a specified numerical threshold, visual inspection against a clear defect standard).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This section is not applicable as the HYBRID Guidewire is a physical medical device, not an AI/Software as a Medical Device (SaMD) that involves human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This section is not applicable as the HYBRID Guidewire is a physical medical device, not an AI/SaMD. The performance described is inherent to the device itself.

    7. The Type of Ground Truth Used

    The ground truth for the HYBRID Guidewire's performance is established by:

    • Established Performance Requirements: Meeting engineering specifications that are commonly accepted for guidewires (e.g., tensile strength, torque strength, flexibility).
    • International Standards: Compliance with relevant ISO (e.g., ISO 11070, ISO 10993) and ASTM standards.
    • Comparison to Predicate Devices: Demonstrating that the subject device's performance is "comparable" or "equivalent" to legally marketed predicate devices, which have a history of safe and effective use.
    • Biological Endpoints: For biocompatibility, the ground truth is the absence of adverse biological reactions (e.g., cytotoxicity, sensitization, irritation, pyrogenicity, significant thrombus formation, etc.) as defined by the applicable ISO 10993 standards.

    8. The Sample Size for the Training Set

    This section is not applicable as the HYBRID Guidewire is a physical device, not a machine learning model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable as the HYBRID Guidewire is a physical device, not a machine learning model requiring a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1