Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K221093

    Validate with FDA (Live)

    Date Cleared
    2022-10-14

    (184 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K181277, K181617, K161439

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-CUBE 70 and X-CUBE 90 diagnostic ultrasound systems are intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic: Adult Cephalic: Trans-raginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Trans-esoph(Cardiac), Peripheral Vessel(PV); and Urology(including prostate).

    And, the imaging modes are 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode: High PRF Doppler mode: Tissue Doppler Imaging(TDI) Mode: 3D/4D mode.

    The X-CUBE 70 and X-CUBE 90 are intended to be used in a hospital or medical clinic.

    Device Description

    The X-CUBE 70 and X-CUBE 90 products are general purpose ultrasound imaging system for medical diagnosis assistance.

    These products are used as an aid tool to diagnosis, such as a commonly used ultrasound diagnostic device.

    Also X-CUBE 70 and X-CUBE 90 have the same operating principles, intended use, risk grade and design/manufacturing characteristics as reference/predicate devices.

    This systems platform provide patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, image quality.

    AI/ML Overview

    The provided text describes the regulatory clearance of the ALPINION Medical Systems X-CUBE 70 and X-CUBE 90 diagnostic ultrasound systems. It does not contain information about acceptance criteria and a study proving a device meets these criteria in the context of an AI/ML-driven medical device, as the device itself is a diagnostic ultrasound system and not an AI/ML product.

    However, the document lists several standards and regulations that the device conforms to to establish its safety and effectiveness relative to predicate devices. These can be interpreted as the high-level criteria for substantial equivalence.

    Here's a breakdown of the information that is available, reframed to address your request where possible, and noting where information is explicitly not present for an AI/ML context:

    1. A table of acceptance criteria and the reported device performance:

    The document does not specify performance acceptance criteria in terms of metrics like sensitivity, specificity, or accuracy that would be typical for an AI/ML device study. Instead, it states the device's conformance to various safety and performance standards. The "New Technologies" are features of the ultrasound system itself, not standalone AI algorithms with performance metrics.

    Acceptance Criteria (Standards Conformed To)Reported Device Performance (Conformance)
    IEC60601-1:2005 (General Requirements for Safety)Conforms
    IEC60601-1-2:2014 (Electromagnetic Compatibility)Conforms
    IEC60601-2-37:2007 (Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment)Conforms
    AAMI/ANSI/ISO10993-1:2009 (Biological Evaluation of Medical Devices - Biocompatibility)Conforms
    AAMI/ANSI/ISO14971:2007 (Application of Risk Management to Medical Devices)Conforms
    AIUM MUS, Third edition (Medical Ultrasound Safety)Conforms
    NEMA UD 2-2004 (Acoustic Output Measurement Standard)Conforms
    NEMA UD 3-2004 (Real Time Display of Thermal and Mechanical Acoustic Output Indices)Conforms
    21 CFR 820 (Quality System Regulation)Conforms
    ISO 13485 (Medical devices — Quality management systems)Conforms

    2. Sample size used for the test set and the data provenance:

    • Not applicable / Not provided. The document states: "The subject of this premarket submission, X-CUBE 70 and X-CUBE 90, did not require clinical studies to support substantial equivalence." This means there was no specific clinical test set used to evaluate performance in the way an AI/ML device would be tested. The evaluation was based on conformance to engineering and safety standards and comparison to existing predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not provided. As no clinical study or test set for an AI/ML algorithm was conducted, there was no need for experts to establish ground truth in this context.

    4. Adjudication method for the test set:

    • Not applicable / Not provided. No clinical test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable / Not provided. This is not an AI-assisted device, so no such study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable / Not provided. This is an ultrasound imaging system, not a standalone AI algorithm.

    7. The type of ground truth used:

    • Not applicable / Not provided. As no clinical evaluation study requiring ground truth was performed, this information is not available. The "ground truth" for this submission was based on the technical specifications and validated performance against established standards for ultrasound devices, as well as comparison to predicate devices.

    8. The sample size for the training set:

    • Not applicable / Not provided. This is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable / Not provided. This is not an AI/ML device.

    In summary, the provided document is a 510(k) clearance letter for a diagnostic ultrasound system. It outlines the regulatory process for demonstrating "substantial equivalence" to existing devices, primarily through adherence to safety and performance standards and a comparison of technological characteristics. It is not a submission for an AI/ML product and therefore does not contain the specific types of acceptance criteria and study details you would expect for such a device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213523

    Validate with FDA (Live)

    Date Cleared
    2022-01-28

    (86 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211300 X-CUBE 90, K181277 E-CUBE 12, K150773 E-CUBE 15, K181617 E-CUBE 8, K161439 E-CUBE 11

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-CUBE i8, X-CUBE i9 diagnostic ultrasound systems are intended for use by, or by the order the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-raginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Trans-esoph. (Cardiac); Peripheral Vessel(PV); and Urology(including prostate).

    And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode.

    The X-CUBE i8. X-CUBE i9 are intended to be used in a hospital or medical clinic.

    Device Description

    X-CUBE i8 and X-CUBE i9 products are an ultrasound imaging system for medical diagnosis. This system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

    Signal Mode:
    2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler(CF) Mode, Power Doppler(PD) Mode, Directional PD mode, Pulsed Wave Doppler(PWD) Mode, Continuous Wave Doppler(CWD) Mode, High PRF Doppler mode, Tissue Doppler Imaging(TDI) Mode

    Combination Mode:
    B/Color Doppler, B/PWD, B/Color Doppler/PWD

    Acoustic output track: Track 3

    AI/ML Overview

    This document is a 510(k) premarket notification for an ultrasound imaging system. It demonstrates substantial equivalence to a predicate device rather than providing a detailed study that proves the device meets specific performance acceptance criteria for an AI/ML-driven medical device. Therefore, it does not contain the information requested in your prompt regarding acceptance criteria tables, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details for an AI/ML system.

    The "Summary of Non-Clinical Tests" and "Summary of Clinical Tests" sections explicitly state:

    • Non-Clinical: "X-CUBE i8 and X-CUBE i9 have been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards." This refers to general device safety and performance based on standards like IEC 60601-1, NEMA UD 2, etc., not AI/ML performance.
    • Clinical: "The subject of this premarket submission, X-CUBE i8 and X-CUBE i9, did not require clinical studies to support substantial equivalence." This means no human-in-the-loop or standalone clinical AI/ML performance studies were conducted or presented for this submission.

    Therefore, I cannot fulfill your request to extract the specific information about acceptance criteria and a study proving AI/ML device performance from the provided document, as the document does not contain this type of information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211299

    Validate with FDA (Live)

    Device Name
    X-CUBE 70
    Date Cleared
    2021-08-09

    (103 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181277, K150773, K181617, K161439, K173713, K201632

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-CUBE 70 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).

    And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 70 is intended to be used in a hospital or medical clinic.

    Device Description

    X-CUBE 70 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

    AI/ML Overview

    The information provided about the X-CUBE 70 ultrasound system primarily pertains to its technical specifications and regulatory compliance for medical device safety standards, rather than performance against specific clinical acceptance criteria or a study with clinical endpoints.

    The document states: "The subject of this premarket submission, X-CUBE 70, did not require clinical studies to support substantial equivalence." This means that the device was deemed substantially equivalent to a predicate device (X-CUBE 70, K200450) based on non-clinical tests rather than clinical performance trials demonstrating specific acceptance criteria.

    Therefore, many of the requested details about acceptance criteria derived from a clinical study, sample size, expert ground truth, and comparative effectiveness studies are not available in the provided text.

    Based on the provided information, I can extract the following:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes the X-CUBE 70 as being in conformance with various safety and performance standards for ultrasound equipment. The "acceptance criteria" here relate to meeting these standards for safety and technical specifications, as opposed to clinical performance benchmarks.

    Acceptance Criteria (General Categories based on Non-Clinical Tests)Reported Device Performance
    BiocompatibilityConfirmed
    Acoustic OutputConforms to NEMA UD2, UD3 and AIUM Medical Ultrasound Safety
    Thermal SafetyConforms to IEC 60601-1 and IEC 60601-2-37
    Electrical SafetyConforms to IEC 60601-1 and IEC 60601-2-37
    Electromagnetic Compatibility (EMC)Conforms to IEC 60601-1-2
    Mechanical SafetyConfirmed
    Quality Management System (QMS) adherenceConforms to 21 CFR 820 and ISO 13485
    Imaging Modes and Clinical ApplicationsSupports listed modes (2D(B), Harmonic, M, Color M, Anatomical M, Color Flow Doppler, Power Doppler, Directional PD, Pulsed Wave Doppler, Continuous Wave Doppler, High PRF Doppler, Tissue Doppler Imaging, 3D/4D) and clinical applications (Fetal, Abdominal, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal, Cardiac, Peripheral Vessel, Urology) across various transducers.
    Dimensions and Weight(See detailed table, e.g., Weight: 90kg, Height: 1325/1560 mm, Width: 554 mm, Depth: 815 mm)
    Electrical Power(See detailed table, e.g., Voltage: 100-120V~, 200-240V~; Frequency: 50-60 Hz; Power: Max. 700VA)

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable as no clinical study with a "test set" of patient data was conducted or referenced for performance evaluation in this submission. The evaluation was based on non-clinical testing against industry standards.
    • Data Provenance: Not applicable for a clinical test set. The data provenance for the non-clinical tests would be from laboratory testing according to the listed standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable, as no clinical ground truth assessment by experts was explicitly mentioned in the context of performance testing for this submission. The compliance was assessed against safety and performance standards by the manufacturer and reviewed by the FDA.

    4. Adjudication method for the test set:

    • Not applicable due to the absence of a clinical test set requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. The X-CUBE 70 is a diagnostic ultrasound system and the submission does not describe any AI component or MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable, as the device is an ultrasound system, not a standalone algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the device's acceptable performance was established through adherence to recognized medical device safety and performance standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, AAMI/ANSI/ISO10993-1, AIUM MUS, NEMA UD 2, NEMA UD 3) and comparison to predicates for substantial equivalence. There is no mention of clinical ground truth (e.g., pathology, outcomes data).

    8. The sample size for the training set:

    • Not applicable. The document does not describe any machine learning or AI component requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable, as no training set is mentioned in this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211300

    Validate with FDA (Live)

    Device Name
    X-CUBE 90
    Date Cleared
    2021-08-09

    (103 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181277, K150773, K181617, K161439, K173713, K201632

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-CUBE 90 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiatric); Peripheral Vessel(PV); and Urology(including prostate).

    And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 90 is intended to be used in a hospital or medical clinic.

    Device Description

    X-CUBE 90 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

    AI/ML Overview

    The provided text describes the Alpinion Medical Systems Co., Ltd. X-CUBE 90 ultrasound system and its equivalence to predicate devices, but it does not include information about acceptance criteria or a study proving device performance in the context of AI/ML.

    The document is a 510(k) summary for a traditional medical device (ultrasound system), not an AI/ML-powered device. Therefore, it focuses on demonstrating substantial equivalence to previously cleared ultrasound systems based on technical characteristics, indications for use, and compliance with general medical device safety standards.

    Here's what can be extracted from the document, with the understanding that the request for AI/ML-specific details cannot be fully met from the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative clinical performance acceptance criteria or report device performance in terms of metrics like sensitivity, specificity, or AUC, which are common for AI/ML devices. Instead, it demonstrates performance by comparing its features and characteristics to legally marketed predicate devices, asserting that it is as safe and effective.

    The acceptance criteria are implicitly based on compliance with recognized medical device standards and having equivalent specifications and functionalities to the predicate devices.

    Acceptance Criterion (Implicit)Reported Device Performance (Implicit)
    Compliance with IEC 60601-1 (General Safety)X-CUBE 90 conforms to IEC60601-1:2005(Third Edition)+CORR.1:2006+CORR.2:2007+A1:2012
    Compliance with IEC 60601-1-2 (EMC)X-CUBE 90 conforms to IEC60601-1-2:2014
    Compliance with IEC 60601-2-37 (Ultrasonic Specific Safety)X-CUBE 90 conforms to IEC60601-2-37:2007/AMD1:2015
    Biocompatibility of patient contact materialsTransducer materials and other patient contact materials are biocompatible (conforming to AAMI/ANSI/ISO10993-1:2009(R)2013)
    Compliance with risk management standardsConforms to AAMI/ANSI/ISO14971:2007/(R)2010
    Compliance with Acoustic Output Measurement StandardConforms to NEMA UD 2-2004(R2009)
    Compliance with Real Time Display of Thermal and Mechanical AIConforms to NEMA UD 3-2004(R2009)
    Indications for Use (equivalent to predicate)The X-CUBE 90 has the same clinical applications and imaging modes as its primary predicate (K200449 X-CUBE 90).
    Technical Specifications (dimensions, electrical power, etc.)Specifications are comparable or slightly improved to predicate devices, not affecting safety or effectiveness.

    Regarding the sections specific to AI/ML devices (points 2-9 in your request), the provided text does not contain this information as it describes a conventional ultrasound system, not an AI/ML powered one.

    Therefore, the following cannot be answered from the provided document:

    • 2. Sample size used for the test set and the data provenance: Not applicable, as no AI/ML test set is mentioned.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    • 8. The sample size for the training set: Not applicable, as no AI/ML training set is mentioned.
    • 9. How the ground truth for the training set was established: Not applicable.

    Summary of Non-Clinical and Clinical Tests (from the document):

    • Summary of Non-Clinical Tests: The X-CUBE 90 was evaluated for biocompatibility, acoustic output, and thermal, electrical, electromagnetic, and mechanical safety. It was found to conform to various medical device safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, AAMI/ANSI/ISO10993-1, AAMI/ANSI/ISO14971, AIUM MUS, NEMA UD 2, NEMA UD 3). Quality management system measures were also applied, including risk management, requirements reviews, design reviews, component verification, integration review, performance testing, safety testing, and design validation.
    • Summary of Clinical Tests: The submission explicitly states: "The subject of this premarket submission, X-CUBE 90, did not require clinical studies to support substantial equivalence." This is typical for 510(k) submissions of conventional devices demonstrating equivalence to existing ones.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200450

    Validate with FDA (Live)

    Device Name
    X-Cube 70
    Date Cleared
    2020-07-10

    (137 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K150773, K181617, K161439, K173713

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-CUBE 70 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal: Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiatric); Peripheral Vessel(PV); and Urology(including prostate).

    And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode: 3D/4D mode. The X-CUBE 70 is intended to be used in a hospital or medical clinic.

    Device Description

    X-CUBE 70 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification stating that the Alpinion Medical Systems X-CUBE 70 ultrasound system is substantially equivalent to legally marketed predicate devices. It specifically outlines the indications for use, comparison to predicate devices, and a summary of non-clinical tests.

    This document focuses on the substantial equivalence of the new device (X-CUBE 70) to existing predicate devices, primarily based on technical characteristics and compliance with safety standards, rather than proving performance against specific acceptance criteria through a clinical or algorithmic study.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample size, ground truth, and expert reconciliation cannot be extracted from this particular document, as it describes a regulatory submission for a medical imaging device, not a performance study of an AI/algorithm-driven device.

    However, I can provide information based on what is available in the document:


    Acceptance Criteria and Device Performance (Not directly addressing AI/algorithmic performance)

    This section is interpreted in the context of the device's functional and safety characteristics, as the document details its equivalence to predicate ultrasound systems, not an AI component with specific performance metrics.

    Acceptance Criteria CategoryReported Device Performance (Summary from doc)
    Indications for Use (Clinical Applications)The X-CUBE 70 is intended for various clinical applications including Fetal, Abdominal (renal & GYN/pelvic), Pediatric, Small Organ (breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional and Superficial), Cardiac (adult & pediatric), Peripheral Vessel, and Urology (including prostate). The tables show equivalency (mark 'N' for new or 'P' for previously cleared) to predicate devices for these applications across different transducers.
    Imaging ModesSupports 2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler (CF) Mode, Power Doppler (PD) Mode, Directional PD mode, Pulsed Wave Doppler (PWD) Mode, Continuous Wave Doppler (CWD) Mode, High PRF Doppler mode, Tissue Doppler Imaging (TDI) Mode, and 3D/4D mode. Also supports combination modes like B/Color Doppler, B/PWD, B/Color Doppler/PWD. These modes are shown as equivalent to predicate devices.
    Physical Dimensions & WeightWeight: 90kg. Height: 1325/1560 mm. Width: 554 mm. Depth: 815 mm. (These are compared against predicate devices, generally falling within similar ranges).
    Electrical PowerVoltage: 100-120V~, 200-240V~. Frequency: 50-60 Hz. Power: Max. 700VA. (These are also compared against predicate devices and found to be standard).
    Imaging Functions (Advanced Features)Includes features like Xpeed™, Full SRI™, Spatial Compounding Image (SCI), Frequency Compounding Image (FCI), Panoramic, Stress Echo, Cube Strain™, Live HQ ™, Needle Vision™ / Needle Vision™ Plus, Elastography, Cube view™, Contrast Enhanced Ultrasound (CEUS), Cube Note, B-STIC, Point Shear Wave Elastography (PSWE), Microvascular Imaging (MVI), Free Angle MSV, AnySlice™, Volume Analysis. Most are common with predicate devices.
    AccessoriesCompatible with various accessories like color/B/W printers, DVD-RW, foot switch, wireless LAN, various biopsy guide kits, ECG modules, ultrasound gel, and several common disinfectants.
    Safety and ComplianceEvaluated for biocompatibility, acoustic output, thermal, electrical, electromagnetic, and mechanical safety. Conforms to voluntary standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, AAMI/ANSI/ISO 10993-1, AAMI/ANSI/ISO 14971, AIUM MUS, NEMA UD 2-2004(R2009), NEMA UD 3-2004(R2009). Also complies with 21 CFR 820 and ISO 13485 (Quality Management System measures).

    Since this document is a 510(k) submission for an ultrasound system, not an AI or algorithm-driven device that requires clinical performance studies to demonstrate accuracy in a diagnostic task, the following points are generally not applicable or stated as not required by the document:

    1. Sample size used for the test set and the data provenance: Not mentioned as no specific performance test set (in the context of AI/algorithm accuracy) was required. The substantial equivalence relies on technical specifications and safety profile similarity to existing devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic accuracy (e.g., presence/absence of disease based on expert consensus) is not part of this type of submission for an ultrasound machine.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an ultrasound imaging system, not an AI-assisted diagnostic tool being evaluated for reader improvement.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device (ultrasound system), not a standalone algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" here is the adherence to safety standards and performance characteristics comparable to predicate devices.
    7. The sample size for the training set: Not applicable, as this is not an AI/ML device that requires a training set for a diagnostic algorithm.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of Non-Clinical Tests (from document):
    The device underwent non-clinical tests to assess biocompatibility, acoustic output, thermal, electrical, electromagnetic, and mechanical safety. It was found to conform to applicable medical device safety standards and voluntary standards (listed in the table above).

    Summary of Clinical Tests (from document):
    The document explicitly states: "The subject of this premarket submission, X-CUBE 70, did not require clinical studies to support substantial equivalence." This reinforces that the focus of this 510(k) submission is on technical and safety equivalence to predicate devices, not on a new clinical claim or an AI/algorithmic performance claim.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200449

    Validate with FDA (Live)

    Device Name
    X-Cube 90
    Date Cleared
    2020-07-10

    (137 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K150773, K181617, K161439, K173713

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-CUBE 90 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).

    And, in the imaging modes of 2D(B) mode; Harnonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 90 is intended to be used in a hospital or medical clinic.

    Device Description

    X-CUBE 90 product is an ultrasound imaging system for medical diagnosis. This innovative system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

    AI/ML Overview

    The provided document is an FDA 510(k) premarket notification for the Alpinion Medical Systems Co., Ltd. X-CUBE 90 ultrasound system. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy through clinical studies for novel devices.

    Therefore, the acceptance criteria and study detailed here are focused on demonstrating that the new device (X-CUBE 90) performs equivalently to previously cleared predicate devices and complies with relevant safety and performance standards. It does not involve a traditional clinical study with patient data in the sense of an Artificial Intelligence (AI) device seeking de novo authorization or PMA.

    Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a "table of acceptance criteria" related to specific performance metrics for the ultrasound imaging capabilities beyond what is implied by the comparison to predicate devices and adherence to standards. Instead, substantial equivalence is demonstrated by a comparison of features, technical specifications, and compliance with recognized safety standards. The "performance" is implicitly deemed equivalent if these factors are comparable to or better than the predicate devices.

    Here's a breakdown of the comparison presented in the "Determination of Substantial Equivalence: Comparison table with Predicate devices" on page 25:

    FeatureProposed X-CUBE 90 PerformancePredicate E-CUBE 12 (K181277) Performance (and other Reference Devices)
    Indications for UseFetal, Abdominal (renal & GYN/pelvic), Pediatric, Small Organ (breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Cardiac (adult & pediatric), Peripheral Vessel (PV), Urology (including prostate).All listed indications are present ("√") for the predicate E-CUBE 12, suggesting that the X-CUBE 90 offers the same range of clinical applications.
    Dimensions & WeightWeight: 90kg, Height: 1325/1560 mm, Width: 554 mm, Depth: 815 mmWeight: 94 kg, Height: 1,420/1,520 mm, Width: 590 mm, Depth: 895 mm (Comparable physical characteristics to predicate devices).
    Electrical PowerVoltage: 100-120V~, 200-240V~, Frequency: 50-60 Hz, Power: Max. 700VAVoltage: 100-120V~, 200-240V~, Frequency: 50/60 Hz, Power: Max. 600VA (The X-CUBE 90 has slightly higher max power, but this would be evaluated against safety standards).
    Imaging Modes2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler (CF) Mode, Power Doppler (PD) Mode, Directional PD mode, Pulsed Wave Doppler (PWD) Mode, Continuous Wave Doppler (CWD) Mode, High PRF Doppler mode, Tissue Doppler Imaging (TDI) Mode, 3D/4D mode.All listed imaging modes are present ("√") for the predicate E-CUBE 12, indicating functionally equivalent imaging capabilities.
    Imaging FunctionsXpeed™, Full SRI™, Spatial Compounding Image (SCI), Frequency Compounding image (FCI), Panoramic, Stress Echo, Cube Strain™, Live HQ ™, Needle Vision™ / Needle Vision™ Plus, Elastography, Cube view™, Contrast Enhanced Ultrasound (CEUS), Cube Note, B-STIC, Point Shear Wave Elastography (PSWE), Microvascular Imaging (MVI), Volume Advance™ (Free Angle MSV, AnySlice™, Volume Analysis).Many of these advanced imaging functions (Xpeed, Full SRI, SCI, FCI, Panic, Stress Echo, Cube Strain, Live HQ, Needle Vision, Elastography, Cube view, CEUS, Cube Note, B-STIC, Volume Advance) are also present in the predicate devices, though some (like PSWE, MVI) are new to this device and would be evaluated for safety and effectiveness on their own, or by comparison to similar functions.
    Safety StandardsIEC60601-1, IEC60601-1-2, IEC60601-2-37, AAMI/ANSI/ISO10993-1, AAMI/ANSI/ISO14971, AIUM MUS, NEMA UD 2-2004(R2009), NEMA UD 3-2004(R2009).The document states the X-CUBE 90 "has been found to conform to applicable medical device safety standards" and "comply with voluntary standards as detailed in this premarket submission." The predicate device also conforms to these same standards, indicating equivalent safety profiles.

    Acceptance Criteria (Implicit from Substantial Equivalence):

    The acceptance criteria for this 510(k) submission are implicitly defined by demonstrating that the X-CUBE 90 is "substantially equivalent" to legally marketed predicate devices (primarily K181277 E-CUBE 12). This means:

    • Same intended use: The X-CUBE 90 has the same indications for use as the predicate device.
    • Same technological characteristics: The device uses the same fundamental technology (ultrasound, various imaging modes and functions).
    • No new questions of safety or effectiveness: Any differences in technological characteristics do not raise new questions of safety or effectiveness. This is primarily addressed through compliance with established safety and performance standards.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "The subject of this premarket submission, X-CUBE 90, did not require clinical studies to support substantial equivalence."

    This means there was no dedicated clinical test set of patient data in the typical sense for demonstrating diagnostic accuracy or efficacy. The "study" for this submission primarily consists of non-clinical performance testing, compliance with international safety and performance standards, and comparisons of technical specifications and features to predicate devices.

    Therefore, there is no sample size for an an imaging test set, and no data provenance relevant to patient-derived images.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As there was no clinical study involving a test set of patient data to establish ground truth diagnoses, this information is not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there was no clinical test set for which ground truth adjudication was required.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was conducted. This device is an ultrasound system, not an AI-assisted diagnostic tool that would typically be evaluated for reader improvement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical ground truth was established for a test set. The "ground truth" for this type of submission is adherence to recognized performance standards and demonstrated technological equivalence to already cleared devices.

    8. The sample size for the training set

    Not applicable. This device is an ultrasound imaging system, not an AI/machine learning model whose performance is enhanced through training on a dataset.

    9. How the ground truth for the training set was established

    Not applicable, as there was no training set for an AI/machine learning model.

    In summary, this 510(k) submission for the X-CUBE 90 ultrasound system primarily relies on non-clinical testing and comparison to predicate devices, rather than clinical studies with patient data and ground truth establishment, to demonstrate substantial equivalence. The "acceptance criteria" are met by demonstrating that the device's technical specifications, imaging modes, and overall performance fall within the established safety and effectiveness profile of already cleared ultrasound systems, and by proving compliance with relevant international and national medical device standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1