Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203169

    Validate with FDA (Live)

    Date Cleared
    2020-11-20

    (28 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K153429, K171013, K133705

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

    This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.

    The system is indicated for head, whole body, cardiac X-ray Computed Tomography applications in patients of all ages.

    The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

    Device Description

    The Revolution Ascend CT system is head and whole body CT system incorporating the same basic fundamental operating principles as the predicate device. It is composed of a gantry, patient table, operator console, host computer, and power distribution unit (PDU), and interconnecting cables. The system also includes image acquisition and reconstruction hardware/software, general system software, accompanying documents, and associated accessories, interconnections. Its materials and construction are identical to our existing marketed products.

    Identical to the predicate, Revolution Ascend generates cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions modes. Revolution Ascend's Intended Use and Indications for Use remain identical to those of the predicate device.

    Revolution Ascend includes virtually all the available features of the predicate device Revolution Maxima. Compared to the predicate, the changes incorporated into Revolution Ascend are primarily to introduce a widended bore gantry for easy handling of large patient, trauma examinations, interventional procedures and radiotherapy planning, and addition of other existing features already available from GE's other CT systems. These ported features include Auto Pilot workflow enabled by Deep learning based patient Auto Positioning, Intelligent Protocoling enabled by Machine Learning, Smart Plan and Auto Prescription all integrated into the modern software platform and GUI adopted from Revolution CT, and cardiac feature Auto Gating and as well as Interventional feature 3D Guidance.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Computed Tomography (CT) system, Revolution Ascend, seeking substantial equivalence to a predicate device, Revolution Maxima. This document primarily focuses on demonstrating the new device's equivalence to an already cleared device rather than proving its performance against a new set of clinical acceptance criteria through a standalone study with human readers or specific AI performance metrics.

    Therefore, the information required for a comprehensive answer regarding acceptance criteria and a study proving a device meets these criteria (especially for a medical AI/CADe device) is largely not present in this document. The submission is for a new iteration of a CT scanner, not a novel AI-powered diagnostic tool requiring specific clinical performance validation for its AI components against a defined ground truth.

    However, I can extract the information that is implicitly or explicitly stated, and highlight where the requested information is absent or not applicable to this type of submission.


    Acceptance Criteria and Device Performance (Implicit):

    Since this is a 510(k) for substantial equivalence to a predicate CT system, the "acceptance criteria" are primarily that the new device, Revolution Ascend, performs as safely and effectively as the predicate device, Revolution Maxima, and other previously cleared GE CT systems for specific features. The performance is assessed through non-clinical bench testing, image quality (IQ) and dose evaluation using phantoms, and verification/validation testing.

    Acceptance Criteria Category (Implicit from 510(k) context)Reported Device Performance (as stated in document)
    Overall Safety & Effectiveness"GE Healthcare believes that the Revolution Ascend is as safe and effective, and performs in a substantially equivalent manner to the predicate device Revolution Maxima (K192686)."
    Compliance with Standards"The Revolution Ascend has completed testing and in compliance with AAMI/ANSI ES 60601-1 and IEC60601-1 Ed. 3.1 and its associated collateral and particular standards, 21 CFR Subchapter J, and NEMA standards XR 25, XR 26, and XR 28." "Revolution Ascend remains compliant with IEC 60601-1 Ed. 3.1 and associated collateral and particular standards, IEC 61223-3-5, NEMA XR25, XR26, and 21 CFR Subchapter J performance standards."
    Functional Equivalence"ldentical to the predicate, Revolution Ascend generates cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions modes. Revolution Ascend's Intended Use and Indications for Use remain identical to those of the predicate device." "The changes described above do not change the fundamental control mechanism, operating principle, energy type, and do not change the intended use from the predicate device Revolution Ascend."
    Image Quality & Dose Performance"The performance and image quality specifications are substantially equivalent to the predicate." "IQ and dose evalauition include: Test using standard IQ, QA and ACR phantoms for standard conditions as well as challenging conditions such as with phantoms simulating large patients. Performance testing in accordance with IEC 61223-3-5 ed 2. 3D guidance test with phantoms simulating interventional conditions." "Non-clinical bench test results demonstrated the subject device performs equivalently to the predicate device."
    Software Level of Concern"The substantial equivalence was also based on software documentation for a 'Moderate' level of concern device."

    Regarding the Study Proving the Device Meets Acceptance Criteria:

    The document describes non-clinical testing for substantial equivalence, not a clinical study designed to prove new performance claims or the efficacy of novel AI features in a clinical setting with human readers.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: No specific number of "cases" or "patients" for a clinical test set is mentioned. The testing involves "standard IQ, QA and ACR phantoms for standard conditions as well as challenging conditions such as with phantoms simulating large patients" and "3D guidance test with phantoms simulating interventional conditions." This indicates laboratory/bench testing using physical phantoms, not a dataset of patient images.
      • Data Provenance: Not applicable as clinical data are not the primary basis for performance evaluation in this submission. The tests are "non-clinical bench test results."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The "ground truth" for non-clinical phantom testing involves established physical properties, measurements, and engineering specifications, not expert clinical interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as no human interpretation or adjudication of a test set is described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done, nor is it described. This submission is for a CT system, not an AI/CADe device requiring direct clinical performance evaluation in synergy with human readers. While the device includes "Intelligent Protocoling enabled by Machine Learning" and "Auto Positioning by Deep Learning," these appear to be workflow/control features, not diagnostic AI features needing MRMC studies for reader performance improvement for a 510(k) submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No standalone performance study of a diagnostic algorithm is detailed. The performance assessment is focused on the CT system's image quality and dose output, verified through phantom studies and engineering testing, ensuring it's equivalent to the predicate.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the non-clinical testing, the "ground truth" is based on the known physical properties of the phantoms, established metrics for image quality and dose (e.g., in accordance with IEC 61223-3-5), and design specifications. There's no clinical ground truth (e.g., pathology, expert consensus) involved.
    7. The sample size for the training set:

      • The document mentions "Intelligent Protocoling enabled by Machine Learning" and "Auto Positioning by Deep Learning." However, it does not provide any details about the training data size, composition, or provenance for these AI features. As these are described as "workflow features" and integral to the CT system's operation (rather than standalone diagnostic AI tools with independent performance claims), such detail is typically not required for a 510(k) of a CT scanner. They are presented as existing, ported features or minor enhancements that don't alter the fundamental operating principles or intended use.
    8. How the ground truth for the training set was established:

      • Not described/provided in the document. (See point 7).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1