510(k) Data Aggregation

    K Number
    K183697

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-15

    (135 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K153225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station). In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multiparameter devices for the display on those devices.

    The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.

    The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospital-type facilities. The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.

    The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal pation conditions in hospitals and hospitals and hospital-type facilities.

    The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions in hospital-type facilities.

    The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities.

    The optional Nomoline Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatric and infant patients.

    The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in hospitals, hospital-type facilities, mobile, and home environments. The NIBP module is designed to measure blood pressure for patient population described in the following table:

    Patient Population Approximate Age Range Newborn (neonate) Birth to 1 month of age Infant 1 month to 2 years of age Child 2 to 12 years of age Adolescent 12-21 years of age Adult 21 years of age and older

    The optional Masimo Centroid O2 is intended for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

    The optional Masimo Centroid O2 is indicated for the continuous monitoring of functional arterial oxygen saturation of hemoglobin (SpO2) and pulse rate (PR) for use with adult, pediatric and neonatal patients during both no motion and motion conditions and for patients who are well or poorly perfused in hospital-type facilities and home environments.

    Devices with Masimo technology are only to be used with Masimo sensors and cables.

    Device Description

    The subject device, Masimo Rad-97 System and Accessories (Rad-97 product family), features a touchscreen display that continuously displays numeric values for the measured monitoring parameters. The Rad-97 product family can be operated on AC power or internal rechargeable battery.

    The subject device (Rad-97 product family) is substantially the same as the predicate (Rad-97 product family) cleared under K180046, and has the same indications for use. The Rad-97 comprises of the same measurement technologies as cleared in the predicate, which includes the Masimo rainbow SET technology, capnography technology, and noninvasive blood pressure (NIBP) technology. These technologies enable the Rad-97 product family to provide noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Perfusion Index (Pi), Pleth Variability Index (PVi), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa), and/or optional capnography parameters or optional noninvasive blood pressure (NIBP) parameters.

    The subject of this submission, is the inclusion of Masimo Centroid 02 sensor to be used with the Rad-97 and similar devices that utilize Masimo SET technology.

    The Centroid O2 is a family of wearable, battery operated, sensors that support Masimo SET technology. The Centroid O2 uses wireless communication to a host device (e.g., Rad-97, Radical-7) in order to support the continuous monitoring of Masimo SET parameters (SpO2 and PR). Centroid O2, similar to the previously cleared RD SET sensors (K180046), are intended for the noninvasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

    The Centroid O2 consists of both disposable and reusable components. The disposable components include the sensor and wrist strap: the reusable component contains the hardware to transfer the measured data wirelessly. Centroid 02 uses a battery as a power source for the sensor and to enable measurement data to be transferred wirelessly to a host device (e.g. Rad-97, Radical-7).

    AI/ML Overview

    The provided text describes the Masimo Rad-97 and Accessories device and the study conducted to ensure its accuracy. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are presented as "Accuracy (ARMS)*" for SpO2, with the reported performance values reflecting the device's accuracy within those ranges.

    FeatureAcceptance CriteriaReported Device Performance (Centroid O2)
    SpO2, no motion70-100%, 2%, adults/pediatrics/infantsBias: -0.5, Adjusted Precision: 1.2, Adjusted RMS: 1.3 (corresponds to 2% ARms)
    70-100%, 3%, neonates(Not explicitly detailed for neonates in the clinical study results, but implied to meet this)
    SpO2, motion70-100%, 3% adults/pediatrics/infants/neonatesBias: -0.9, Adjusted Precision: 2.0, Adjusted RMS: 2.2 (corresponds to 3% ARms)
    SpO2, low perfusion70-100%, 2%, adults/pediatrics/infants(Clinical study results provided focus on no motion and motion conditions; low perfusion not explicitly detailed in the results table)
    70-100%, 3%, neonates(Clinical study results provided focus on no motion and motion conditions; low perfusion not explicitly detailed in the results table)
    Pulse rate, no motion25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates(Clinical study focused on SpO2 accuracy; PR accuracy not detailed in the provided clinical study results)
    Pulse rate, motion25-240 bpm, 5 bpm, adults/pediatrics/infants/neonates(Clinical study focused on SpO2 accuracy; PR accuracy not detailed in the provided clinical study results)
    Pulse rate, low perfusion25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates(Clinical study focused on SpO2 accuracy; PR accuracy not detailed in the provided clinical study results)

    *ARMs accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/- ARMs of the reference measurements in a controlled study.

    The clinical study explicitly reports on "SpO2 No Motion Accuracy" with an Adjusted RMS of 1.3, which corresponds to an ARMs value of 2% during no-motion conditions. For "SpO2 Motion Accuracy," the Adjusted RMS is 2.2, corresponding to an ARMs value of 3% during motion conditions. These reported values directly align with the acceptance criteria for SpO2 for adults/pediatrics/infants during no motion and for all patient populations during motion.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document states, "a clinical study was performed... on healthy adult volunteers." However, the exact sample size (number of healthy adult volunteers) used for the test set is not explicitly mentioned in the provided text.

    The data provenance is prospective, as it was a "clinical study" performed specifically "to establish the accuracy (ARMs) of the Centroid O2 sensors." The country of origin of the data is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document states that the comparison was made "in comparison to blood measurements from a laboratory CO-Oximeter." This implies that the ground truth for SpO2 measurements was established using a laboratory CO-Oximeter, which is a highly accurate reference method. It does not involve human experts in establishing the ground truth directly for SpO2 measurements in this context. The reliance is on the accuracy of the laboratory instrument.

    4. Adjudication Method for the Test Set

    Since the ground truth for SpO2 measurements was based on a laboratory CO-Oximeter, there was no human adjudication method (e.g., 2+1, 3+1) used for establishing the ground truth for the test set. The CO-Oximeter serves as the objective reference standard.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a Multi Reader Multi Case (MRMC) comparative effectiveness study in the provided text. The device is a patient monitor, and the study focuses on the accuracy of the device's measurements (SpO2, PR, etc.) compared to a reference standard, not on human interpretation or the impact of AI on human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone study was done. The clinical study described "evaluate[d] the sensor's performance for no motion and motion conditions, in the range of 70% to 100% SaO2, in comparison to blood measurements from a laboratory CO-Oximeter." This is a direct assessment of the device's (and its underlying algorithm's) performance without human intervention in the measurement process itself.

    7. The Type of Ground Truth Used

    The type of ground truth used for the clinical study was laboratory-measured SaO2 (functional oxygen saturation of arterial hemoglobin) from a CO-Oximeter. This is an objective physiological measurement from a reference device.

    8. The Sample Size for the Training Set

    The provided text does not mention the sample size for the training set. The clinical study described is for validation/testing, not for training. The document states that the Masimo Rad-97 and Accessories (Centroid O2) uses "Masimo SET technology," which is an established technology, implying that the underlying algorithms would have been developed and potentially trained on prior datasets, but details of such training sets are not included here.

    9. How the Ground Truth for the Training Set Was Established

    Since the document does not provide information on a training set, it does not describe how the ground truth for a training set was established. The clinical study focuses on evaluating the Centroid O2 sensor's performance against a reference (laboratory CO-Oximeter) during no-motion and motion conditions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162603

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-26

    (249 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140188, K151644, K153225

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥ 40 kg and on pediatrics > 5 kg and < 40 kg in healthcare environments.

    Device Description

    The Masimo Regional Oximetry System (O3 System) includes the O3 Sensors and the O3 Module. The O3 System measures hemoglobin under the sensor, allowing clinicians to continuously and accurately determine the absolute and trend measurements of regional blood oxygenation saturation in the tissue (rSO2). The O3 Sensors includes optical components that collect physiological signals. The O3 Module includes Masimo technology for processing those signals which resulted in regional oximetry (rSO2) measurements. In turn, these measurements are displayed on the Host/Backboard device.

    The O3 Sensor is a single-patient use adhesive sensor and is supplied non-sterile. The O3 Sensor, comprising of an emitter and two detectors, is applied to the patient's forehead at one end. The other end of the sensor is connected to a patient cable which in turn is connected to the 03 Module. Up to two 03 Sensors can be connected to each 03 Module and both sensors can be connected to a patient.

    The O3 Module includes Near InfraRed Spectroscopy (NIRS) technology. The O3 Sensor uses multiple wavelengths in the range of near infrared wavelengths to measure light absorption in the tissue. The O3 Sensors and O3 Module make up the O3 System for the monitoring of absolute regional hemoglobin oxygen saturation of blood (rSO2) under the sensors. The O3 System does not have its own power. The O3 Module is powered by connecting to a Host/Backboard Device such as the Root Monitoring System (Root in turn is powered by either AC power or internal rechargeable batteries.

    The O3 System provides the following key measurements:

    • Regional Oxygenation (rSO2): Regional tissue oxygenation level in the deep tissue local to the sensor site, including cerebral tissue
    • Delta Baseline (Abase): Relative difference in rSO2 with respect to baseline rSO2
    • Area Under the Limit (AUL index): Index that quantifies the duration (amount of time the patient stays below rSO2 low alarm limit) and depth (refers to the gap between the patient's rSO2 level and the rSO2 low alarm limit) of patient's stay below the user-defined rSO2 low alarm limit (LAL)
    • Delta SpO2 (ΔSpO2): The difference between SpO2 and rSO2. The source of SpO2 is from peripheral SpO2 measurement (using pulse oximeter).
    AI/ML Overview

    The provided document describes the Masimo O3 Regional Oximeter System, specifically focusing on the O3 Pediatric Sensor, and its equivalence to predicate devices (K160526 and K082327). The document outlines non-clinical and clinical testing performed to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for accuracy are presented as ARMS (Accuracy Root Mean Square) values.

    FeatureAcceptance Criteria (Specification)Reported Device Performance (O3 Pediatric Sensor)
    rSO2 Measurement Accuracy
    Adults ≥ 40kg (using O3 Large Sensor - Predicate)Absolute ARMS, 4% for SavO2 of 45%-85%N/A (this is predicate performance)
    Trending ARMS, 3% for SavO2 of 45%-85%N/A (this is predicate performance)
    Pediatrics ≥ 5 kg and < 40 kg (O3 Pediatric Sensor)Absolute ARMS, 5% for SavO2 of 45%-85%5% for SavO2 of 45%-85%
    Trending ARMS, 3% for SavO2 of 45%-85%3% for SavO2 of 45%-85%

    (Emphasis added to highlight the performance of the device under review, the O3 Pediatric Sensor)

    2. Sample size used for the test set and the data provenance

    • Sample Size for Absolute ARMS (Pediatric): 33 male and female pediatric patients.
    • Sample Size for Trending ARMS (Pediatric): Not explicitly stated as a distinct number of patients. The trending study used the O3 Large Sensor as a reference and collected rSO2 data from both the O3 Large Sensor and O3 Pediatric Sensor. It's implied this was performed on a subset or all of the 33 pediatric patients, but the exact number for the trending analysis isn't isolated.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates the study involved "pediatric patients," suggesting human clinical data. It is a prospective study, as blood draws were taken "before and after his/her medical procedure" for the absolute ARMS, and trending data was collected.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for the rSO2 measurements was established by "blood draws (arterial and jugular blood)" which were then analyzed to determine blood oxygen saturation. This is a direct physiological measurement, considered objective, and does not involve human expert interpretation in the same way imaging studies might. Therefore, the concept of "number of experts" with specific qualifications for ground truth establishment (like radiologists) is not applicable here. The accuracy is against a direct physiological laboratory standard, not an expert consensus.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Since the ground truth for both absolute and trending accuracy was based on direct blood measurements (physiological reference), there was no adjudication method involving human experts for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is an oximeter, which provides a direct measurement (rSO2) rather than an interpretation that would be assisted by AI. The document describes a comparison of the device (O3 Pediatric Sensor) against a physiological reference and against another sensor (O3 Large Sensor for trending), not human reader performance with or without AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance evaluation for accuracy (Absolute ARMS and Trending ARMS) is a standalone (algorithm only) performance evaluation. The device provides the rSO2 measurements directly, based on its NIRS technology and signal processing. The accuracy is reported for the device's measurements compared to a reference standard (blood draws or another sensor). While a human user reads the display, the accuracy itself is of the device's output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used was physiological measurements (blood draws) for absolute accuracy. Specifically, "Blood draws (arterial and jugular blood) were taken for each subject." For trending accuracy, the O3 Large Sensor's measurements (which were previously established via blood reference) served as the comparative ground truth.

    8. The sample size for the training set

    The document does not specify a separate training set sample size. This is typical for a device submission that primarily focuses on hardware and embedded algorithms, where the "training" (calibration and algorithm development) often occurs internally during the device development phase, and the submission focuses on the validation or test data. The "Clinical Testing" section describes the validation study for the new O3 Pediatric Sensor.

    9. How the ground truth for the training set was established
    As the document does not explicitly mention a separate "training set," it does not describe how ground truth for such a set was established. It describes the ground truth for the clinical validation (test) set, which was established through blood draws (arterial and jugular blood), considered the reference standard for blood oxygen saturation. It's reasonable to infer that the underlying scientific principles and initial calibrations would have relied on similar, well-established physiological measurement techniques.