    K Number
    K242752

    Validate with FDA (Live)

    Date Cleared
    2025-05-28

    (258 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143369

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Iradimed Corporation's MRidium 3870 MRI Infusion Pump System is indicated for general hospital or clinical use by medical professionals whenever it is required to infuse patients with intravenous fluids where a large magnetic field could be present such as Magnetic Resonance Imaging (MRI). The MRidium 3870 MRI Infusion pump can also be used during transport within the clinical environment as part of the periprocedural care before or after the MRI exam. The MRidium 3870 MRI Infusion Pump can be used with up to a 3.0 Tesla MRI system when placed as close as clinically possible to the MR system, without being placed within the magnet's bore, up to a 15,000 Gauss magnetic field line.

    The pump and infusion sets are designed to deliver saline and IV medication. The system is useful in the administration of fluids requiring precisely controlled infusion rates. The system can operate in primary, loading, bolus, or KVO delivery mode.

    The dedicated Iradimed 1056, 1058, and 1059 Infusion Sets for the MRidium 3870 MRI Infusion Pump are intended for single patient use up to 6 hours.

    The device is intended for Adult and Pediatric patients, including Neonates. Careful clinical considerations involving drug and dosing should always be applied especially with regard to neonatal subpopulations (i.e., low and extremely low birth weight).

    This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.

    Device Description

    The MRidium 3870 MRI Infusion Pump System is indicated for general hospital or clinical use by medical professionals whenever it is required to infuse patients with intravenous fluids where a large magnetic field could be present such as Magnetic Resonance Imaging (MRI). The MRidium 3870 MRI Infusion Pump can also be used during transport within the clinical environment as part of the periprocedural care before or after the MRI exam. The MRidium 3870 MRI Infusion Pump can be used with up to a 3.0 Tesla MRI system allowing placement as close as physically possible to the MRI scanner without being placed within the magnet's bore, up to a 15,000 Gauss magnetic field line. The magnetic content of the Pump is such that there is no hazard of magnetic attraction, up to 3T (30,000 gauss).

    The pump unit is designed with an integral single peristaltic pump channel utilizing an ultrasonic (non-magnetic) motor. The integral infusion set mounting channel is horizontally oriented in the front of the pump with position detection sensors and graphical user feedback to aid the user in the correct infusion set loading process. The main assembly of the pump unit contains the controls, display, power supply, battery, and processor/memory functions suitable to meet all system requirements.

    A remote display (3875) is also available as an option for independent viewing and control from the adjacent MRI Control Room area (Zone III).

    The Dose Rate Calculator feature allows the user to set up a patient's infusion rate based upon user selected parameters, including volume to be infused, dose, concentration, weight, and/or time. The Dose Rate Calculator feature also provides a Drug Library, allowing the user to program a patient's infusion protocol from selected parameters, including volume to be infused, dose, concentration, weight and/or time with soft and hard limits for each drug. These Drug Library inputs can only be created or modified by the Iradimed in-house clinical team based on input and validation from the healthcare facility. Once programmed, the library is digitally signed requiring proper verification by a software 'key' within the 3870 MR IV pump upon loading. Loading of the drug library requires access to the Password protected Service Menu.

    The Dose Error Reduction System (DERS) feature allows user-facilities to provide input to custom names, and doses, with hard and soft limits for use in the Drug Library. The Drug Library supports care area specific infusion protocols for primary, bolus, loading dose, and KVO delivery modes, retrievable by drug/protocol name. The library, as prescribed by the healthcare facility, is programmed with nominal starting values for: Dose, Concentration, and Time. Also, hard limits (maximum and minimum allowable) and soft limits (high and low limits that require a user confirmation to exceed) for Dose, Concentration, Time and Patient Weight can be programmed.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the MRidium 3870 MRI Infusion Pump System focuses on demonstrating substantial equivalence to a predicate device (MRidium 3860+ MRI Infusion Pump/Monitoring System) through non-clinical performance testing. It does not describe a study involving human readers or the establishment of ground truth by expert consensus, as might be typical for AI/ML-driven medical device submissions.

    Therefore, I will extract information related to the device's functional performance, which serves as its "acceptance criteria," and the "study" (non-clinical testing) that proves it meets these criteria. Many of the requested points (e.g., number of experts, adjudication methods, MRMC studies, sample sizes for training sets) are not applicable to this type of device and submission.

    Here's the breakdown based on the provided document:

    Acceptance Criteria and Study Proving Device Performance for MRidium 3870 MRI Infusion Pump System

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes performance specifications and indicates that the device meets them, primarily by demonstrating substantial equivalence to the predicate and through various verification tests. The "acceptance criteria" are implied by the specifications themselves, and the "reported device performance" is that it complies with these specifications and has been tested to perform as intended.

    Characteristic / Acceptance CriteriaReported Device Performance (as stated in comparison or test summary)
    Flow Rate Range0.4 to 1000 mL/hr in 0.1 mL/hr increments
    Flow Rate Accuracy+/- 5% 1.0 mL/hr to 1000 mL/hr; +/- 10% 0.4 ml/hr to <1.0 mL/hr (during 6-hour Infusion Set life)
    Primary Volume to Be Infused (VTBI) Range0.1 to 1000 mL
    Total Volume Infused (VI) Range0.1 to 99999 mL
    Keep Vein Open (KVO) Rate RangeProgrammed in Drug Library – OFF, Rate (0.4 mL/hr to 20 mL/hr), Continue Primary
    Patient Line (downstream) Back-Pressure Range+300 to -100 mmHg
    Downstream Occlusion Detection Range1 to 10 PSI (6.9 to 68.8 kPa), user-adjustable
    Occlusion Detection (no flow) TimeVariable, detailed in Operation Manual (REF 1234) table based on IV Set, Flow Rate, and Occlusion Pressure Alarm Limit Setting. (Improved labeling over simple predicate statement).
    Maximum Post-Occlusion Bolus Volume0.7 mL max
    Air-in-Line Detector Threshold(s)> 100 uL (+20%)
    Audible Alarm Range Pump (at 1 Meter)Minimum: 65 dBA; Maximum: > 85 dBA @ 1 meter
    Pump Battery Operating Time> 8 Hours at 125 mL/hr Rate
    Pump Battery Charge Time< 5 hours to 95% capacity
    Pump MRI Conditions: B0 Field / Gauss-Line Limit≤ 15,000 gauss
    Operating Temperature Range+15○C to +30○ C (+59○F to +86○ F)
    Operating Relative Humidity Range30% to 80% RH, non-condensing
    Ingress ProtectionIPX2
    Wireless EncryptionEncrypted for security (Improvement over predicate)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document indicates that non-clinical bench testing was performed. Specific numerical sample sizes (e.g., how many pumps were tested, how many infusion sets) are not explicitly stated in the provided text. The testing categories include:

      • Verification Testing of Product Requirements (e.g., flow rate accuracy, essential performance, reliability, air-in-line, occlusion, alarms, software V&V, cybersecurity, battery, EMC/EMI, environmental, shock/vibration, electrical safety, mechanical, biocompatibility, ingress protection, wireless co-existence, standards testing).
      • Human Factors Validation Testing.
      • MRI Environment Conditions of Use Testing (magnetic displacement, spurious emissions, essential performance during MRI scans including flow accuracy, alarms, unintended bolus, wireless communication).
    • Data Provenance: The data is from non-clinical bench testing performed by Iradimed Corporation (the manufacturer) as part of their 510(k) submission. No information about country of origin of "data" in the sense of patient data is relevant here, as it's not a clinical study on patients. It is implicitly prospective testing, designed to demonstrate compliance for the submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable. The device is an infusion pump, and its performance is evaluated against engineering and safety standards, not against human expert interpretations of medical images or patient outcomes. The "ground truth" for this device's performance verification comes from established engineering principles, metrology, and consensus standards (e.g., AAMI TIR101 for flow accuracy, IEC 60601 series for safety).

    4. Adjudication Method for the Test Set

    This section is not applicable. As the testing is non-clinical and pertains to physical and electrical performance, there is no need for expert adjudication in the traditional sense of clinical or image-based studies. Performance is measured against predefined specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This section is not applicable. MRMC studies are typically performed for AI/ML-driven diagnostic devices where human readers provide interpretations (e.g., radiologists reading medical images) and their performance with and without AI assistance is compared. This device is an infusion pump, not a diagnostic AI tool.

    6. Standalone Performance Study (i.e., algorithm only without human-in-the-loop performance)

    This is largely applicable to the extent that the device's functional performance (e.g., flow rate, alarm detection, battery life) is evaluated independently of human interaction for its core functions in a controlled bench setting. The "Summary of Performance Testing (Non-Clinical)" details various tests (e.g., flow rate accuracy, essential performance, reliability, air-in-line, occlusion performance, alarms, cybersecurity) that represent the device's standalone operation.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Engineering specifications and design requirements: The device is designed to meet specific measurable performance parameters (e.g., flow rate accuracy, pressure detection thresholds).
    • Consensus standards: Performance is verified against recognized industry standards (e.g., AAMI TIR101 for flow accuracy, IEC 60601 series for safety and alarms).
    • Bench testing measurements: Physical measurements using calibrated equipment are the "ground truth" to confirm the device's output against its specifications.

    8. The Sample Size for the Training Set

    This section is not applicable. This is a hardware medical device (an infusion pump) and not an AI/ML device that undergoes "training" on a "training set" of data. Its "training" is in its design, engineering, and manufacturing processes, culminating in verification and validation testing.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reason as point 8.