510(k) Data Aggregation

    K Number
    K251784

    Validate with FDA (Live)

    Device Name
    Hushd Pro Z-Link
    Date Cleared
    2025-10-24

    (135 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hushd Pro Z-Link device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.

    Device Description

    The Hushd Pro Z-Link device series are removable intraoral patient-specific device intended for the treatment of snoring and mild to moderate obstructive sleep apnea. The device consists of custom-made maxillary and mandibular splints that fit over the upper and lower teeth, and are connected via a pair of detachable rigid connectors. When linked, the splints function as a mandibular repositioner by holding the lower jaw in a forward position during sleep, thereby preventing the tongue and soft tissues from collapsing into the airway.

    Each device consists of one maxillary and one mandibular splint, designed and manufactured using 3D printing technology. The splints incorporate a pair of integrated studs that are located around the canine area for the maxillary splint and around the first molar area for the mandibular splint. These studs interface with slots on the detachable connectors to securely link the two splints together. This configuration permits the mandible to advance from the baseline position, offering improved patient comfort.

    Titration is achieved by replacing the connectors with another pair of a different size. The connectors are available in 1.0 mm incremental lengths, allowing progressive mandibular advancement according to patient tolerance and clinical need. A standard set includes six pairs of connectors, ranging from -1 mm to +5 mm protrusion, while an extended set ranging from +6 mm to +10 mm is available upon request, providing a maximum mandibular protrusion of 10 mm.

    The connectors are not elastic and do not contain adjustment mechanisms such as pistons, screws, straps, or repositioning elastics. The rigid connectors are manufactured with medical grade nylon PA12 using commercial injection molding process and are marked for easy identification.

    The splints of this device are manufactured using selective laser sintering (SLS) 3D printing with EOS PA2200 polyamide (nylon) PA12 material. This material offers excellent mechanical properties, long-term stability, and has a history of safe use in oral medical devices. The manufacturing process includes mechanical or chemical polishing, cleaning, and rigorous quality control prior to packaging.

    The Hushd Pro Z-Link is available for prescription only and designed based on optical or physical impressions taken by a licensed healthcare provider. The splints and connectors are designed exclusively by the device manufacturer and its authorized partners to ensure consistent quality and performance.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243752

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-17

    (286 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K222127, K143244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults

    Device Description

    The Double Tube Herbst appliance consists of upper and lower base appliances connected on each side by a telescopic apparatus that advances the mandible. Each telescopic apparatus is loosely but securely attached to wire loops formed at the ends of retentive anchors embedded in the acrylic on the outer (buccal) surfaces of the upper and lower base appliances. A telescopic apparatus consists of a double tube telescopic assembly that is attached to the back end of the upper wire loop; and a rod that is attached to the front end of the lower wire loop. The double tube telescopic assembly consists of a short internally threaded tube welded in parallel onto the back of a long straight tube in which the rod slides. The short internally threaded tube is adjustably attached to the back end of the upper wire loop by engaging the externally threaded lower end of a connector, the upper end of which ends in a helix that is twisted around the upper wire loop. The rod that slides in and out of the long tube is loosely but securely attached to the wire loop on the front end of the lower retentive anchor. The lengths of the telescopic assemblies can be adjusted in 1/2 mm increments along a range of 15 mm by disengaging the telescopic components, rotating the double tube assembly around the connecting arm, and sliding the rod back into the long tube to lock in the adjustment.
    The Double Tube Herbst appliance is worn while sleeping to keep the mandible in a forward position prescribed by the treating dentist to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The customized appliance is inserted and removed by the patients every night, and it is adjusted by the prescribing dentist.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252374

    Validate with FDA (Live)

    Device Name
    Nylon flexTAP(R)
    Date Cleared
    2025-09-11

    (43 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K972061, K181482, K143244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nylon flexTAP is intended to reduce or alleviate nighttime snoring and mild to moderate obstructive sleep apnea, OSA for adults, 18 years of age and older.

    Device Description

    The Nylon flexTAP is a custom-fit oral device intended to reduce or alleviate nighttime snoring and mild to moderate obstructive sleep apnea (OSA). The Nylon flexTAP device consists of 3 components:

    1. Upper Tray that fits on the upper teeth
    2. Lower Tray that fits on the lower teeth.
    3. An Adjustment Mechanism attached to the lower tray that fits into an Adjustment Post on the upper tray.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K230532

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-07-12

    (135 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended to reduce or alleviate snoring, mild to moderate Obstructive Sleep Apnea (OSA) while sleeping in adults.

    Device Description

    The RADx White Night and RADx White Night-p Intraoral Appliance for Snoring and Mild to Moderate Sleep Apnea are removable intraoral devices used for treating snoring and mild to moderate obstructive sleep apnea. It consists of two custom fabricated adjustable rods that fit separately over the upper and lower teeth. The device functions as a mandibular repositioner, maintaining the lower jaw in a forward position during sleep. This mechanical protrusion acts to increase the patient's pharyngeal space, improving their ability to exchange air during sleep. The device is a prescription device customized for each patient and has an adjustment mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The device can be adjusted only by the user and / or the dentist in increments of 0.5 mm.

    The custom oral appliance is checked for proper and comfortable fit as well as the appropriate initial progressive arm attachment. The dentist will provide the patient with subsequent progressive arms from 1.0 mm up to 5.0mm in 0.5mm increments.

    There are 9 different arms each identified with a number for each side of the jaw. The arms are also identified as either "L" for left or "R" for right.

    Variants / Configurations

    There are two variants of the device.

    The RADx White Night has removable progressive arms with the female housing active as the receptacle for the arms.

    The RADx White Night-p was designed with the upper housing and arm as one piece. In this version, the fixed upper tray-arm combination are the progressive parts of the entire apparatus.

    AI/ML Overview

    The provided document is a 510(k) summary for the RADx Intraoral Appliance for Snoring & Sleep Apnea. It describes the device and its compliance with regulatory standards, focusing on biocompatibility. However, it explicitly states that "Human clinical studies were not deemed necessary to evaluate the safety or effectiveness of the subject device." Instead, the substantial equivalence to predicate devices and material biocompatibility testing are the bases for its clearance.

    Therefore, many of the requested items related to clinical study performance and ground truth cannot be answered from this document.

    Here's what can be extracted based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    Since no clinical performance study was conducted for this device, a direct comparison of acceptance criteria to reported device performance metrics like sensitivity, specificity, accuracy, etc., is not available in the document. The performance testing focused on biocompatibility.

    Acceptance Criteria (from predicate/standards)Reported Device Performance (RADx Intraoral Appliance)
    Biocompatibility:
    Non-CytotoxicNon-Cytotoxic (for BioMed Clear and dental LT Clear V2 Resins)
    No cytotoxic potentialNo cytotoxic potential (for BioMed Clear and dental LT Clear V2 Resins)
    Non-sensitizerNon-sensitizer (for BioMed Clear and dental LT Clear V2 Resins)
    Non-IrritantNon-Irritant (for BioMed Clear and dental LT Clear V2 Resins)
    Non-Irritant (oral mucosal)Non-Irritant (for BioMed Clear and dental LT Clear V2 Resins)
    No evidence of systemic toxicityNo evidence of systemic toxicity (for BioMed Clear and dental LT Clear V2 Resins)
    Macroscopic reaction not significant compared to negative controlMacroscopic reaction was not significant as compared to the negative control article (for BioMed Clear and dental LT Clear V2 Resins)
    Non-ToxicNon-Toxic (for BioMed Clear and dental LT Clear V2 Resins)
    Absence of PyrogensAbsence of Pyrogens (for BioMed Clear and dental LT Clear V2 Resins)
    Non-MutagenicNon-Mutagenic (for BioMed Clear and dental LT Clear V2 Resins)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    Not applicable. No clinical test set or data provenance from human subjects is mentioned as human clinical studies were not deemed necessary. The biocompatibility tests were likely performed in laboratories using standard animal models or in-vitro methods.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. No clinical test set requiring expert ground truth establishment was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. No clinical test set requiring adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an intraoral appliance, not an AI-powered diagnostic tool, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical intraoral appliance, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    For biocompatibility testing, the "ground truth" is established by the accepted scientific standards and methodologies defined in the cited ISO and USP standards (e.g., cell viability in cytotoxicity, skin reaction in sensitization, etc.). These are objective laboratory measurements, not expert consensus or clinical outcomes data in the context of device performance for snoring/sleep apnea.

    8. The sample size for the training set:

    Not applicable. No algorithm or AI model was developed for this device, so there is no training set.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203596

    Validate with FDA (Live)

    Date Cleared
    2021-04-20

    (132 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102909, K143244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Panthera Occlusal Appliance is indicated for protections from the forces and damage of parafunctions like bruxism, prevention of noise associated with bruxism, and for alleviation of temporomandibular joint, muscle and tension headache pains associated with temporomandibular disorders.

    Device Description

    The Panthera Occlusal Appliance is a removable intraoral device used for protecting teeth and restorations against the forces of bruxism. It consists of patient-specific splint that fit over the upper or lower teeth. This device functions as a protective barrier for teeth and restorations by creating physical separation between upper and lower tooth surfaces preventing tooth damage caused by bruxism (like grinding and clenching) and alleviating temporomandibular joint, jaw, and muscle and tension headache pains. There are various occlusal designs for this application, all with varying degrees of function and suitability depending on the patient's needs. The Panthera Occlusal Appliance is customized for each patient based on the clinician's prescription, and is to be worn every day during hours of sleep.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Panthera Occlusal Appliance. This document focuses on demonstrating substantial equivalence to previously marketed devices rather than proving the performance of a novel AI/software-driven medical device against specific acceptance criteria using a study.

    Therefore, the information required to answer the prompt regarding AI/software performance, such as specific acceptance criteria for algorithm performance, sample sizes for test/training sets, expert adjudication methods, MRMC studies, standalone performance, and ground truth establishment, is not present in the provided document.

    The document states:

    • "Human clinical study was not deemed necessary to support substantial equivalence." (page 7)
    • "The Panthera Occlusal Appliance does not use design dissimilar from the primary predicate and reference devices, does not use new technologies different from the primary predicate and reference devices, and does not deviate from the indications for use of the primary predicate and reference devices." (page 7)

    This indicates that the submission relies on demonstrating that the new device is fundamentally similar in design, technology, and indications for use to existing, legally marketed devices (predicates and references). The "bench testing" mentioned (page 7) is for physical properties and intended use, not for AI/algorithm performance.

    In summary, the provided document does not contain the information requested in the prompt, as it is a 510(k) for a physical dental appliance based on substantial equivalence, not an AI/software medical device requiring performance studies against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202651

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-02-12

    (151 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NOA SLEEP APNEA AND SNORING DEVICE is a mandibular advance device (MAD) indicated for mild to moderate obstructive sleep apnea (OSA) and to alleviate or reduce snoring in adults.

    Device Description

    The NOA Sleep Apnea and Snoring Device is a customized prescription intraoral mandibular advancement device for snoring and obstructive sleep apnea intended for single-patient use by those who are 18 years or older. The device consists of CAD/CAM fabricated maxillary and mandibular splints connected by symmetrical, twin-mated posts that are customizable to allow lateral jaw movement as well as jaw opening. The series of one or more mandibular splints, and one or more maxillary splints to be exchanged with other splints in the series according to the prescriber, to titrate protrusive jaw movement according to patient comfort and needs. The described mandibular advancement is intended to reduce snoring and mild to moderate obstructive sleep apnea by holding the mandible and base of tongue forward during sleep, to provide increased pharyngeal space to improve the patient's ability to exchange air and decrease air turbulence. The NOA Sleep Apnea and Snoring Device does not have any additional adjustment mechanisms to modify or maintain the mandibular position such as pistons, straps, repositioning elastics or screws, eliminating the need for utilizing external controlling components such as keys, screwdrivers or ligature ties. The NOA device is designed with lateral fins called cam and follower. Contact between the cam and the follower prevents any mandibular retrusion even with the opening of the mouth. The device design is configured according to the patient's specific dental arch, to maximize tongue space and to allow open, close and laterality movements during wear. These movements can be personalized by the doctor. Furthermore, the doctor can change the height of the cam in the prescription to limit the maximum mouth opening during wear. Orthoapnea NOA device designs can be Standard or Customized. A Standard device is designed with standard height in the lateral cams among all lower splints. If the practitioner provides a measurement of the antero-posterior mobility (Maximum retrusion), a Customized device is designed decreasing the height of the cam according to the degree of advancement. increases, the ability to open the mouth without mandibular retrusion decreases, making the cam shorter. The following additional device design features can be customized when ordering the prescription device: Advance regulation/titration sequence, Lateral movement, Device opening limit, Frontal opening of the device.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "NOA Sleep Apnea and Snoring Device." It outlines the device's technical characteristics, intended use, and the non-clinical testing performed to establish substantial equivalence to predicate devices.

    Crucially, this document does not contain any information about an AI/ML-based device or a study involving human readers or AI assistance. The device described is a physical intraoral mandibular advancement device. Therefore, the requested information regarding acceptance criteria, performance metrics, sample sizes for AI/ML models, expert ground truth establishment, MRMC studies, and standalone algorithm performance cannot be extracted from this document.

    The document explicitly states:

    • "No clinical testing was performed." (Page 6, Section E)
    • "Non-clinical testing was used to support the determination of substantial equivalence." (Page 6, Section E)

    Therefore, I cannot fulfill the request as it pertains to an AI/ML device and its validation. The document describes a traditional medical device (mandibular appliance) and its clearance process based on non-clinical bench testing and comparison to existing predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190236

    Validate with FDA (Live)

    Device Name
    O2Vent Optima
    Date Cleared
    2019-08-29

    (204 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143244 The Panthera Anti-Snoring Device – Panthera Dental Inc, K971794 EMA Appliance – Myerson LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The O2Vent Optima is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA). The device is indicated for use during sleep to aid in the treatment of these conditions.

    Device Description

    The O₂Vent Optima is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA).
    The O₂Vent Optima consists of the following parts:
    Upper tray (upper teeth)
    Lower tray (lower teeth)
    Connector bands
    The upper tray is fitted over the upper teeth, with the breathing port at the front leading to the airways on each side to the rear of the appliance. The upper nylon tray has two side protrusions (lugs) that engage a connector that when attached, extends to the lower tray and attaches to a similar lug stabilizing and advancing the lower jaw. The lower tray, is customized to the bottom teeth, which when attached to the same connector via a side protrusion (lug) preventing the tongue and soft tissues of the throat from collapsing and obstructing the airway. In order to adjust the trays to the specific needs of the user, various length bands (13-21mm) are used to move the lower jaw (i.e. adjust the trays) forward or backward. The shorter the connector band, the further the lower jaw is moved forwards. The O2Vent Optima is delivered fully assembled with the 19mm connector band as the recommended starting titration length for all new users. The upper and lower trays of the O2Vent Optima are made with PA2200 (i.e. Nylon). The straps which connect the upper and lower trays together are made of 100% thermoplastic polyurethane/cured elastomer.

    AI/ML Overview

    Here's a summary of the O2Vent Optima's acceptance criteria and the study information, based on the provided FDA 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance:

    The document provided (an FDA 510(k) summary) acts as a substantial equivalence determination, not a full clinical study report with explicit acceptance criteria and corresponding reported performance metrics (like sensitivity, specificity, accuracy, or specific reductions in AHI).

    Instead, the submission argues for substantial equivalence by comparing the technological characteristics of the O2Vent Optima to predicate and reference devices. The "performance" is implicitly demonstrated through this comparison, indicating that the device is "as safe, as effective and performs comparatively" to the predicate and reference devices that are already legally marketed.

    Therefore, the table below reflects the characteristics that were compared to establish substantial equivalence rather than numerical acceptance criteria and performance data.

    Feature ComparedO2Vent Optima CharacteristicsPrimary Predicate (O2Vent W) CharacteristicsReference Device 1 (Panthera Anti-Snoring Device) CharacteristicsReference Device 2 (EMA Appliance) CharacteristicsConclusion (Implicit Acceptance)
    Indications for UseReduce/alleviate snoring, mild to moderate OSA. For use during sleep.Reduce/alleviate snoring, mild to moderate OSA. For use during sleep.Reduce/alleviate snoring, mild to moderate OSATreatment of nasal respiratory dysfunction of OSA and snoring, where repositioning of mandible and opening bite increases airspace.Similar to predicates/references
    Product CodesLRKLRKLRKLRKSame for all
    Regulation21CFR 872.557021CFR 872.557021CFR 872.557021CFR 872.5570Same for all
    ClassificationClass IIClass IIClass IIClass IISame for all
    Use of DeviceRemovable intraoral device, single patient multiple use, prescription use only.Removable intraoral device, single patient multiple use, prescription use only.Removable intraoral device, multiple use, prescription use only.Removable intraoral device, single patient multiple use, prescription use only.Similar or same for all
    Target PopulationAdults 18+ who snore and/or have sleep apnea.Adults 18+ who snore and/or have sleep apnea.AdultsNot indicated (for this specific comparison)Same or similar
    Device FunctionalityRepositions lower jaw forward, acts by increasing pharyngeal space, permits mouth breathing via built-in airway, retains teeth using rigid trays (separate upper/lower).Repositions lower jaw forward, acts by increasing pharyngeal space, permits mouth breathing via built-in airway, retains teeth using rigid trays (separate upper/lower).Mandibular repositioner, maintains lower jaw forward, acts to increase pharyngeal space, retains teeth using customized splints (separate upper/lower).Advance mandible to anterior/inferior position, pulls tongue forward, increases airspace, decreases obstruction, upper/lower trays for tooth retention (separate).Description similar for all devices; Airway feature same as primary predicate, different from some references.
    Mandibular AdvancementAdjusted by attaching connectors of varying lengths.Adjusted using titration of adjustment screws.Adjusted via interlocking rods.Adjusted by attaching connectors of varying lengths.Same as reference devices; similar to predicate.
    AdjustmentClinician and patient adjustable.Clinician and patient adjustable.Clinician/patient can select shortened connecting rod.Clinician and patient adjustable.Same for all
    Design (CAD/CAM)Customized per patient (CAD/CAM, additive manufacturing, laser sintering of PA2200).Customized per patient (CAD/CAM, plastic laminates/acrylics for main trays, titanium component 3D printed via EBM).Customized per patient (CAD/CAM, selective laser sintering).Custom oral appliance available through dental labs.Same as for reference device; similar to predicate.
    Adjustment AccessoryReplacement/re-supply connectors (13-21 mm).NAReplacement/re-supply connectors (19-36 mm).Replacement/re-supply connectors.Similar to reference devices.
    SterilityNon-SterileNon-SterileNon-SterileNon-SterileSame for all
    Materials of Construction (Trays)Polyamide type 12 (Nylon).Medical grade titanium (air channel), dental plastic laminates and acrylics.Polyamide type 12 (Nylon).Proprietary thermoplastic polymer.Same as reference device; different from predicate due to additive manufacturing.
    Materials of Construction (Connectors)100% thermoplastic Polyurethane/cured elastomer.NASame as above per labeling (assumed same as trays).100% thermoplastic Polyurethane/cured elastomer.Same as reference devices.
    BiocompatibilityYes, passed all tests (ISO 10993 Parts 5 & 10).Not performed (materials identical to K160234).Statement in K143244: biocompatible based on similarity to predicate.Information could not be verified.Same as predicate and reference devices per available information.

    2. Sample Size Used for the Test Set and Data Provenance:

    The provided document (510(k) summary) does not describe a clinical study of the O2Vent Optima that would involve a "test set" and a specific sample size for performance evaluation. Instead, it relies on demonstrating substantial equivalence to already cleared devices through a comparison of technological characteristics. Therefore, there is no information on sample size or data provenance from a specific test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    As no specific clinical performance study with a test set is described, there's no information regarding experts used to establish ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable, as no described test set or clinical performance study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC study is mentioned or summarized in the provided document. The submission focuses on device characteristics, not reader performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study:

    Not applicable. The O2Vent Optima is an intraoral medical device, not an AI algorithm.

    7. Type of Ground Truth Used:

    Not applicable. The submission focuses on comparing the physical and functional characteristics of the device and its materials to predicate devices rather than validating an algorithm against a ground truth. Biocompatibility testing was done to established standards (ISO 10993).

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI algorithm.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable.