    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremittes, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography.

    Device Description

    The MobileDiagnost wDR 2.2 is a motorized mobile radiographic system consisting of a mobile base unit, a user interface consisting of Eleva Workspot combined with flat solid state X-ray detectors used to operate, generate, process and handle digital X- ray images. The MobileDiagnost wDR 2.2 integrates a new wireless portable detector (SkyPlate E). The family of SkyPlate detectors (Large and Small) have already been integrated into the MobileDiagnost wDR 2.0 based on the K141736 pre-market submission.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the MobileDiagnost wDR 2.2, a mobile X-ray system. The focus of the submission is to demonstrate substantial equivalence to a predicate device, the MobileDiagnost wDR 2.0 (K141895), with several modifications. The document does not contain a clinical study to prove the device meets acceptance criteria in terms of diagnostic effectiveness or a "performance study" in the typical sense of evaluating a new diagnostic algorithm's accuracy. Instead, the submission relies on demonstrating substantial equivalence through non-clinical verification and validation tests against established standards.

    Therefore, the requested information needs to be framed within this context of demonstrating substantial equivalence, rather than a traditional diagnostic performance study.

    Here's the breakdown of the information based on the provided text:

    Acceptance Criteria and Device Performance (within the context of Substantial Equivalence)

    The acceptance criteria are implicitly defined by the compliance with recognized international and FDA consensus standards and the outcome of the comparison to the predicate device, showing that modifications do not raise new questions of safety or effectiveness. The device performance is assessed against these standards and through direct comparison of technical characteristics to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission relying on technical changes and compliance with standards, the "acceptance criteria" are compliance with these standards and the "reported device performance" is the verification that these standards are met, and that the technical characteristics of the modified device are acceptably equivalent to the predicate.

    Feature Area / Acceptance CriteriaReported Device Performance (MobileDiagnost wDR 2.2)Conclusion (vs. Predicate)
    I. Compliance with International and FDA-recognized Consensus Standards & FDA Guidance DocumentsNon-clinical verification and validation tests demonstrate compliance with: - IEC 60601-1 (Edition 3.1) - IEC 60601-1-2 (Edition 4.0) - IEC 60601-1-3 (Edition 2.1) - IEC 60601-1-6 (Edition 3.1) - IEC 60601-2-54 (Edition 1.1) - IEC 62304 (Edition 1.0:2006) - IEC 62366 (Edition 1.0 2015) - ISO 14971 (Edition 2.0, 2007) - ISO 10993-1 (Edition 4.0 2009) - IEC 60601-2-28 (Edition 2.0 2010-03) - IEC 62220-1 (Edition 1.0 2015-03) - NEMA PS 3.1 - 3.20 (DICOM) - "Guidance for the Submission of 510(k)s for Solid State X-Ray Imaging Devices" (Sept 1, 2016) - "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) - "Pediatric Information for X-ray Imaging Device Premarket Notifications" (Nov 28, 2017)Meets acceptance criteria and is adequate for its intended use, demonstrating substantial equivalence. Non-clinical information deemed sufficient to support substantial equivalence.
    II. Technical Characteristics Equivalence (MobileDiagnost wDR 2.2 vs. MobileDiagnost wDR 2.0)
    Base Unit Type, X-ray Tube rotational capabilities, Mains Supply, Mode of Exposure, Available Exposure Methods, X-ray Absorber, Installation type, Readout Mechanism, Maximum X-ray Dose for Linear Response, Maximum Usable Dose, Maximum Lifetime Dose, Image Processing (Eleva Workstation), ADC Digitisation, Signal to Electronic Noise Ratio (SENR), Data Interface to Workstation, X-ray Tube type, material, maximum voltage, nominal focal spot, anode type, nominal anode input power, Generator configurations, Collimator operation mode, beam shape, External Connectivity (DICOM), Software Platform (Eleva WorkSpot with SkyFlow).Same/Identical to predicate.Identical; thus, demonstrating Substantial Equivalence (SE).
    Dimensions (overall, transport, source-floor distance)Minor differences in mm measurements.Differences do not impact safety or effectiveness. Thus, demonstrating SE.
    Detector ModelsAddition of SkyPlate E Large (Trixell 3543DR). Previously cleared SkyPlate Large/Small retained.Addition of SkyPlate E detector does not impact safety or effectiveness. Thus, demonstrating SE. All technical detector characteristics influencing image quality assessed per FDA guidance.
    Detector WeightMax 3.1 kg (vs. Max 3 kg for predicate).Difference has no impact on clinical workflow, safety, or effectiveness. Thus, demonstrating SE.
    Image Size (X-ray field)345.0 mm x 426.0 mm (vs. 344.8 mm x 421.2 mm for predicate).Difference does not impact clinical Image Quality, safety, or effectiveness. Thus, demonstrating SE.
    Pixel Size160 µm (vs. 148 µm for predicate).Difference of 12 µm pixel size does not impact image resolution "to an extent that can impact the clinical image quality," safety or effectiveness. Thus, demonstrating SE.
    Image matrix size (Number of pixels)2156 x 2662 pixels (vs. 2330 x 2846 pixels for predicate)."Infinitesimal change" and reduction in number of pixels due to 160 µm pixel size does not impact clinical Image Quality, safety, or effectiveness. Thus, demonstrating SE.
    Nyquist Frequency3.125 lp/mm (vs. 3.38 lp/mm for predicate).Difference does not impact clinical Image Quality, safety, or effectiveness. Thus, demonstrating SE.
    Modulation Transfer Function (MTF) & Detective Quantum Efficiency (DQE)Slightly different typical values reported (e.g., MTF at 1 lp/mm: 62% vs. 61%; DQE at 3 lp/mm: 22% vs. 29%).Differences do not impact clinical Image Quality, safety, or effectiveness. Thus, demonstrating SE.
    GridsAddition of new large grids for SkyPlate E. Previously cleared grids retained.Addition of new grids has no impact on clinical workflow, safety, or effectiveness. Thus, demonstrating SE.
    Column ConfigurationSliding Column (vs. Standard/Short Column for predicate).Sliding Column provides better viewing; its introduction does not impact safety or effectiveness. Thus, demonstrating SE.
    System Power ON/OFFKeyless user access (vs. Physical Key for predicate).Keyless user access provides authenticated access; its introduction does not impact safety or effectiveness. Thus, demonstrating SE.
    Exposure @ zero degree column positionExposures allowed (vs. No Exposure allowed for predicate).Allows examinations in space-constrained situations; difference does not impact safety or effectiveness. Thus, demonstrating SE.
    Exposure during ChargingExposures allowed (vs. Not allowed for predicate).Allows charging during exam preparation; exposure energy still drawn from generator batteries, not mains. Does not impact safety or effectiveness. Thus, demonstrating SE.
    Handles on CollimatorHandles can be used to move both Tube head and Collimator (vs. only Tube head for predicate).Provides ease of use; difference does not impact safety or effectiveness. Thus, demonstrating SE.
    Image Processing AlgorithmUNIQUE 2 (vs. UNIQUE for predicate).UNIQUE 2 provides improved image processing (reduced noise, improved contrast) and was already cleared under K182973. Upgrading does not alter clinical workflow, safety, or effectiveness. Thus, demonstrating SE.

    Study Details (as applicable for a Substantial Equivalence submission based on non-clinical data)

    The provided text clearly states: "The MobileDiagnost wDR 2.2 does not require clinical study since substantial equivalence to the primary currently marketed and predicate device MobileDiagnost wDR 2.0 (K141895- September 18, 2014) was demonstrated with the following attributes: Indication for use; Technological characteristics; Non-clinical performance testing; and Safety and effectiveness."

    This means there was no "study" in the sense of a clinical trial or a diagnostic performance study to evaluate accuracy, sensitivity, specificity, etc., with human or AI readers. The "study" here refers to the comprehensive non-clinical verification and validation process against relevant standards.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not applicable. No clinical test set of patient data was used for a diagnostic performance evaluation. The "tests" were non-clinical verification and validation tests on the device itself (e.g., electrical safety, electromagnetic compatibility, radiation protection, software validation, usability, risk management).
    • Data Provenance: Not applicable. No patient data (retrospective or prospective, from any country) was used for evaluating the device's diagnostic performance for the purpose of this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not applicable. There was no diagnostic test set requiring ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. There was no diagnostic test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is not an AI-assisted diagnostic product but rather an X-ray imaging system with updated components and software functionalities.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is an X-ray system, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable. There was no diagnostic performance study requiring ground truth. Ground truth for the non-clinical tests would be the established scientific/engineering principles, specifications, and regulatory standards.

    8. The sample size for the training set

    • Not applicable. This device is an X-ray system, not a machine learning model that requires a "training set" in the context of AI. The software (UNIQUE 2) was previously cleared (K182973), indicating its own development and validation, but details of its training are not provided here.

    9. How the ground truth for the training set was established

    • Not applicable, as there was no training set for the device itself. For the incorporated UNIQUE 2 image processing algorithm, its ground truth establishment (if it involved machine learning) would have been part of its original 510(k) (K182973), but those details are not in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170113

    Validate with FDA (Live)

    Device Name
    pixium 4343RCE
    Date Cleared
    2017-02-09

    (28 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K141736, K153318, K063781

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As a part of a radiographic system, the Pixium 4343RCE is intended to acquire digital radiographic images. The Pixium 4343RCE is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

    Device Description

    The pixium 4343RCE is a Stationary x-ray system that converts x-ray patterns into electrical signals. The signals are converted into visible images for use in medical diagnosis. In the device, a cesium iodide scintillator absorbs the input x-ray photons. The scintillator in turn emits visible spectrum photons that illuminate an array of photodetectors that create an electrical charge representation of the x-ray input. A matrix scan of the array converts the integrated charges into a modulated electrical signal.

    The detector is permanently installed and intended to be integrated into an x-ray system, where it constitutes an x-ray receptor for direct x-ray imaging. It is electrically powered by and connected with the x-ray system. The device is connected to the Philips Eleva Workspot to create a complete x-ray imaging chain, and it is intended to be used in Philips x-ray systems such as the DigitalDiagnost.

    AI/ML Overview

    The provided text describes the Philips pixium 4343RCE, a stationary x-ray system. The document is a 510(k) summary for premarket notification to the FDA, asserting substantial equivalence to a predicate device (Philips pixium 4343RC, K131483).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present specific numerical acceptance criteria (e.g., minimum DQE, MTF values that must be met) for the pixium 4343RCE, nor does it provide a direct side-by-side comparison of claimed performance data against such defined criteria. Instead, the acceptance criteria are implicitly stated as demonstrating similar performance to the predicate device and compliance with various standards. The "Discussion" column in the comparison table serves as the "reported device performance" against the implicitly accepted criterion of not negatively impacting safety or effectiveness compared to the predicate.

    Here's a table summarizing relevant performance metrics and the discussion provided, which indicates how the device meets (or is considered to meet) implied acceptance criteria of non-inferiority or slight improvement:

    FeaturePredicate Device pixium 4343RC (K131483)Proposed Device pixium 4343RCE (K170113)Discussion / Performance vs. Implied Acceptance Criteria
    Design Features
    X-Ray AbsorberCsI ScintillatorCsI ScintillatorIdentical; no impact to safety and effectiveness.
    Installation typeStationary, permanently installedStationary, permanently installedIdentical; no impact to safety and effectiveness.
    Readout MechanismThin Film TransistorThin Film TransistorIdentical; no impact to safety and effectiveness.
    Detector Size500 x 490 x 45.5 mm³500 x 490 x 45 mm³Similar, slightly smaller; no impact to safety or effectiveness.
    Detector Weight<14 kg11.7kg ± 0.85kgSimilar, slightly lighter; no impact to safety or effectiveness.
    Image Size (Pixel)2840 x 28742840 x 2874Identical; no impact to safety and effectiveness.
    Image Size (X-ray field)420 x 425 mm²420 x 425 mm²Identical; no impact to safety and effectiveness.
    Distance Image to Rim34.9 mm40mmSimilar, slightly greater; no impact to safety or effectiveness.
    Pixel Size148µm148µmIdentical; no impact to safety and effectiveness.
    Nyquist Frequency3.37 lp/mm3.37 lp/mmIdentical; no impact to safety and effectiveness.
    ADC Digitization16 bit16 bitIdentical; no impact to safety and effectiveness.
    SENR (@ 1 µGy)42.3 dB42.3 dBIdentical; no impact to safety and effectiveness.
    Max X-ray Dose for Linear Response50 µGy50 µGyIdentical; no impact to safety and effectiveness.
    Image Readout Duration1.6 s1.36 sSlightly faster; no impact to safety and effectiveness.
    Number of Modes23Additional default mode included; no impact safety and effectiveness.
    Exposure Window Durations1-8192 ms0-8191 msMinimal decrease; insignificant, no impact to safety or effectiveness.
    Sequence Time Detector6 s6 sIdentical; no impact to safety and effectiveness.
    Image Data16.3 MBytes16.3 MBytesIdentical; no impact to safety and effectiveness.
    Use w and w/o Radiographic Grid?YesYesIdentical; no impact to safety and effectiveness.
    Maximum Lifetime Dose100 Gy100 GyIdentical; no impact to safety and effectiveness.
    Warm-up Duration before Calibration4 h2 hDecreased; no impact to safety or effectiveness.
    Data Interface to Workstation100 Mbit/s Ethernet1 Gbit/s EthernetIncreased; no impact to safety or effectiveness.
    Power Consumption20.4WAvg: 8W. Max: 15WDecreased; no impact to safety or effectiveness.
    Cover Factor (Optical Fill Factor)60.3%63%Essentially the same; no impact to safety or effectiveness.
    Image Quality
    Modulation Transfer Function (MTF)1 lp/mm 64%, 2 lp/mm 32%, 3 lp/mm 17%, Nyquist 13% (3.37 lp/mm)1 lp/mm 62%, 2 lp/mm 35%, 3 lp/mm 19%, Nyquist 15% (3.37 lp/mm)Similar, with some improvements and one minimal decrease; no impact to safety and effectiveness.
    Detective Quantum Efficiency (DQE) at 2 µGy0.05 lp/mm 65%, 1 lp/mm 52%, 2 lp/mm 42%, 3 lp/mm 25%, 3.37 lp/mm 18%0.05 lp/mm 67%, 1 lp/mm 51%, 2 lp/mm 42%, 3 lp/mm 27%, 3.37 lp/mm 18%Similar, with one improvement and one minimal decrease; no impact to safety and effectiveness.

    The study that proves the device meets the acceptance criteria is detailed as "Non-clinical verification and validation tests." These tests cover the intended use, technical claims, requirement specifications, and risk management results. The document explicitly states: "Non-clinical verification and validation test results demonstrate that the pixium 4343RCE... Meets the acceptance criteria and is adequate for its intended use." The basis for this is "compliance with the aforementioned international and FDA-recognized consensus standards and device specific guidance document" and showing "identical or equivalent technological characteristics," "equivalent diagnostic capability," and no "new potential hazards or safety risks" compared to the predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a separate "test set" in the context of data used for a clinical performance evaluation, as it explicitly states that no clinical study was required. The non-clinical verification and validation tests are performed on the device itself (hardware and software) and against established technical specifications and standards (e.g., ISO, IEC, NEMA, FDA guidance). Therefore, there is no mention of a "sample size" of patient data or "data provenance" (country of origin, retrospective/prospective) for a clinical test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since no clinical study was conducted, there was no "test set" of patient data requiring expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable as no clinical test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a digital x-ray detector, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the device itself (the x-ray detector) as a standalone component without human interpretation being part of the primary evaluation. The "Summary of Non-clinical Performance Data" and the "Comparison" tables effectively serve as the standalone performance assessment. The device's performance metrics (MTF, DQE, SENR, image readout speed, etc.) are evaluated directly for the detector. The conclusion is that "the device is as safe, as effective, and performs as well or better than the predicate device" in terms of its technological characteristics and ability to produce images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance evaluation, the "ground truth" is based on technical specifications, international and FDA-recognized consensus standards, and device-specific guidance documents. For instance, the MTF and DQE values are compared against those of the predicate device, with the implicit ground truth being that comparable or improved values demonstrate equivalent performance.

    8. The sample size for the training set

    Not applicable. The document describes an x-ray detector, not an AI algorithm that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable as there is no training set mentioned for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141895

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-09-18

    (66 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K133259, K141381, K141736

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography.

    Device Description

    The MobileDiagnost wDR 2.0 system is a motorized mobile radiographic system consisting of a mobile base unit, and a user interface (computer, keyboard, display, mouse), combined with a flat solid state X-ray detector. It is used by the operator to generate, process and handle digital X-ray images. The MobileDiagnost wDR 2.0 integrates a new generation of wireless portable x-ray detectors (SkyPlate) to replace the former detector WPD FD-W17 cleared under the predicate submission (K111725).

    AI/ML Overview

    Unfortunately, based solely on the provided text, I cannot provide a detailed table of acceptance criteria and reported device performance with the specific metrics you requested (e.g., sample sizes, number of experts, adjudication methods, MRMC study details, ground truth types for test and training sets).

    The document is a 510(k) summary for the MobileDiagnost wDR 2.0, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a comprehensive report of a standalone clinical study with detailed performance metrics against predefined acceptance criteria.

    However, I can extract and infer some information that partially addresses your request, particularly concerning the non-clinical and "clinical image concurrence" studies mentioned.

    Here's an attempt to answer your questions based on the available text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily states that the device's non-clinical performance values are "basically equal or better than the predicate device." Specific acceptance criteria or targets are not explicitly listed in a table format in the provided text. The performance is reported in comparison to the predicate device.

    Metric (Non-Clinical)Acceptance Criteria (Inferred from "equal or better than predicate")Reported Device Performance (MobileDiagnost wDR 2.0)
    Modulation Transfer Function (MTF)Equal to or better than predicate (60% to 15%)Better than predicate, ranging from 61% to 14% (implies some values are better)
    Detective Quantum Efficiency (DQE)Equal to or better than predicate (66% to 22%)Equal to or better than predicate, ranging from 66% to 24% (implies some are better)
    Other Non-Clinical TestsCompliance with standards and satisfactory resultsComplies with listed international and FDA-recognized consensus standards (IEC, AAMI, ISO)

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-Clinical Tests: Sample sizes are not specified for the non-clinical tests (DQE, MTF, Aliasing, etc.). The provenance is not explicitly stated but is implicitly from laboratory testing of the device itself.
    • Clinical Image Concurrence Study: Sample size for the test set is not specified. The data provenance is not explicitly stated, but it involved "clinical images were collected and analyzed." It is a retrospective study ("clinical images were collected"). No country of origin is specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Non-Clinical Tests: Not applicable, as these are objective physical measurements of device characteristics.
    • Clinical Image Concurrence Study: The text mentions a "single blinded, concurrence study" to ensure images meet "user needs" and provide "equivalent diagnostic capability" to the predicate. The number of experts is not specified, nor are their specific qualifications (e.g., "radiologist with 10 years of experience"). It implies expert readers assessed the images.

    4. Adjudication Method for the Test Set

    • Non-Clinical Tests: Not applicable.
    • Clinical Image Concurrence Study: The term "concurrence study" is used, implying agreement among readers or against a standard. However, the specific adjudication method (e.g., 2+1, 3+1, none) is not detailed in the provided text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • The document describes a "single blinded, concurrence study" for clinical image analysis. While it compares the new device's images to the predicate, it doesn't explicitly state it was an MRMC comparative effectiveness study in the typical sense of measuring human reader improvement with AI vs. without AI assistance. This device is an X-ray system, not an AI diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply directly to this device's evaluation as presented. No effect size is mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This question is not directly applicable as the MobileDiagnost wDR 2.0 is an X-ray imaging system, not an AI algorithm. The performance evaluation focuses on the image quality and physical characteristics of the imaging system itself, not an algorithm's diagnostic capabilities in isolation. The non-clinical performance data represents the standalone performance of the device's imaging capabilities.

    7. The Type of Ground Truth Used

    • Non-Clinical Tests: The "ground truth" for these tests are objective physical standards and measurements (e.g., how a detector should perform according to physics principles and consensus standards).
    • Clinical Image Concurrence Study: The "ground truth" was established by expert assessment for "equivalent diagnostic capability" and meeting "user needs." This is a form of expert consensus on image quality and diagnostic utility, comparing images from the new device to those from the predicate.

    8. The Sample Size for the Training Set

    • Not applicable / Not specified. The document describes performance testing of an imaging device, not a machine learning algorithm that requires a training set. The "clinical images" mentioned were for validation, not training.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no mention of a training set for an algorithm in this context.

    In summary: The provided 510(k) summary focuses on demonstrating that the MobileDiagnost wDR 2.0 is substantially equivalent to its predicate device through non-clinical (technical) performance criteria and a clinical image concurrence study. It highlights compliance with recognized standards and claims that the new device's technical performance is "equal or better" than the predicate and that its images provide "equivalent diagnostic capability." However, it lacks the fine-grained details about sample sizes, expert qualifications, and specific adjudication methods for the clinical image study that would be typical for evaluating a new diagnostic AI algorithm.