    Reference Devices :

    K141112, K132430, K053116, K093720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterling Monorail PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, renal, and carotid arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon expandable and selfexpanding stents in the peripheral vasculature.

    Device Description

    The Sterling Balloon Dilatation Catheter is a Monorail brand rapid exchange catheter with a semicompliant balloon fixed at the distal tip. The balloon catheter has a coaxial shaft design. The outer lumen is used for inflation of the balloon, and the wire lumen permits the use of guidewires 0.014 in / 0.018 in to facilitate advancement of the catheter to and through the stenosis to be dilated. The product's catheter includes a tapered tip to facilitate advancement of the catheter to and through the stenosis. Two radiopaque marker bands (one proximal and one distal), in conjunction with fluoroscopy, enable accurate positioning of the balloon. The effective length of the balloon catheter is 200 cm; catheter markers indicate the exit of the dilatation catheter tip out of the guiding catheter (one at 90 cm and two at 100 cm).

    AI/ML Overview

    This document is a 510(k) summary for a medical device called the Sterling™ MONORAIL™ PTA Balloon Dilatation Catheter. It seeks to demonstrate substantial equivalence to a predicate device.

    The provided text does not contain information related to an AI/ML device or its performance criteria. Therefore, I cannot furnish the requested details regarding acceptance criteria, study methodologies for AI performance, sample sizes for test/training sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance.

    The document focuses on the bench testing performed to demonstrate substantial equivalence for this physical medical device (a balloon catheter), not a software or AI/ML product. The performance data section explicitly lists:

    • Sterilization Testing
    • Bench Testing (Dimensional Verification, Simulated Use, Balloon Inflation & Deflation Time, Catheter Bond Strength, Flexibility & Kink test, Torque Strength, Shaft & Bond Burst, Catheter Extension & Deflation, Packaging Testing, Design Validation Testing)

    These are typical tests for a physical medical device to ensure its mechanical properties, function, and safety, not for evaluating an AI algorithm's diagnostic or predictive capabilities.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201794

    Validate with FDA (Live)

    Device Name
    Jade
    Date Cleared
    2020-07-28

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K192318, K141112, K132430

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jade PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, poplited, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilation of balloon expandable and self-expanding stents in the peripheral vasculature.

    Device Description

    The JADE PTA Balloon Dilatation Catheters is now also available as an over-the-wire balloon catheter for peripheral indications. The balloon diameters range from 1.5mm to 6.0mm and balloon lengths range from 20mm to 240mm, with 90cm, 150cm and 200cm catheter lengths. The balloon material is made of a minimally compliant material with a rated burst pressure of 18 atmospheres for 5.0-6.0mm and 20atm for 1.5-4.0mm. Hydrophilic lubricious coatings are applied to the distal section of the catheter. The shaft of the catheter is composed of a proximal shaft and a distal shaft. The distal shaft is composed of a distal outer tube and tri-extrusion inner tube with a balloon welded to both tubes at the distal tip to aid in tracking through vasculature. The proximal shaft is composed of a proximal outer tube and tri-extrusion inner tube that are bonded to a female luer connector, with the proximal outer tube allowing for proximal pushability with a smooth transition to the distal shaft. The inner lumen of the catheter accepts a maximum 0.014 inch (0.36mm) guidewire. The guidewire enters the catheter tip and advances coaxially out the hub guidewire lumen. Two radiopaque marker bands are positioned within the balloon shoulder, and for balloons of working length 180-240mm, two more marker bands (four marker bands in total) are positioned in the middle of the balloon. One marked section, for the 90cm catheter working length, or two marked sections, for the 150cm and 200cm catheter working lengths, are located on the proximal shaft to indicate catheter position relative to the tip of the guiding catheter or introducer sheath. The Y-type hub is bonded on the proximal end of the catheter at the entrance to the inflation lumen and the guidewire lumen. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Jade PTA Balloon Dilatation Catheter, which is a medical device. This type of submission is for proving "substantial equivalence" to a legally marketed predicate device, rather than proving that an AI/algorithm-based device meets specific performance criteria.

    Therefore, the document does not contain the information requested regarding acceptance criteria and the study that proves an AI/algorithm device meets those criteria. The provided text details performance testing for a physical medical device (a balloon catheter) to demonstrate its safety and effectiveness, and its equivalence to a predicate catheter.

    Specifically, the document does not address any of the following points relevant to an AI/algorithm-based device study:

    • 1. A table of acceptance criteria and the reported device performance (for an AI/algorithm)
    • 2. Sample sized used for the test set and the data provenance
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    • 8. The sample size for the training set
    • 9. How the ground truth for the training set was established

    The performance data mentioned in the document (Visual Inspection, Marker Band Radiopacity, Dimensional Verification, etc.) are all related to the physical characteristics and functionality of the balloon catheter, not an AI algorithm.