Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K180830

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-12-20

    (265 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K171764, K140485

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Reuter Tip Deflecting Wire Guides are intended for use in curving or deflecting catheter tips for pulmonary angiography, selective angiography, translumbar aortography, repositioning of central venous catheter tips and other vascular and non-vascular applications.

    Reuter Tip Deflecting Wire Guides are intended for use with the Reuter Tip Deflecting Handle assembly for curving or deflecting catheter tips for pulmonary angiography, translumbar aortography, bronchography, bronchography, repositioning of central venous catheter tips and other vascular and non-vascular applications.

    Device Description

    The Reuter Tip Deflecting Wire Guides, subject of this submission, are Class II devices according to 21 CFR §870.1330; product code DQX (Wire, Guide, Catheter). The subject devices are available with outer diameters of 0.025 and 0.035 inches, lengths of 80 - 145 centimeters, and a curved tip.

    The subject device includes two configurations indicated by prefixes TDW and DTDW. Both configurations are constructed with stainless steel mandrils, coils, and internal handles. The coils cover the entire mandril and are secured using distal and proximal anchor points. The internal handle is permanently affixed to the proximal end of the wire guide and is designed to mate with the tip deflecting handle. The internal handle of the wire guide is not patient contacting.

    The Reuter Tip Deflecting Wire Guide configuration with the TDW prefix is a disposable wire guide designed to be used with the Reusable Reuter Tip Deflecting Wire Guide handle which is not subject of this submission.

    The Disposable Reuter Tip Deflecting Wire Guide configuration with the DTDW prefix is designed with a deflecting handle permanently affixed to the proximal end of the wire guide. The deflecting handle is manufactured from polycarbonate and stainless steel materials.

    The purpose of the handle is to operate in a retracting motion thereby causing the tip of the connected wire guide to deflect to the specified radius.

    The side-arm flushing adapter (Class II 510(k) exempt under product code DTL) is supplied with the Disposable Reuter Tip Deflecting Wire Guides as an accessory. The flushing adapter may be used to flush a catheter being used with the subject device during diagnostic and interventional procedures.

    The subject device is a packaged, sterile device intended for single patient use.

    AI/ML Overview

    This document is a 510(k) summary for the Reuter Tip Deflecting Wire Guide, which is a medical device. The information provided primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report of a new AI/software-based device. Therefore, many of the requested categories for AI/software studies are not applicable or cannot be extracted from this document.

    Here's an analysis based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    Test PerformedAcceptance CriteriaReported Device Performance
    Biocompatibility Testing (ISO 10993-1:2009)Predetermined acceptance criteriaMet the predetermined acceptance criteria
    Corrosion Testing (ISO 11070:2014 Annex B)Predetermined acceptance criteriaMet the predetermined acceptance criteria
    Flexing Test (ISO 11070:2014 Annex G)Predetermined acceptance criteriaMet the predetermined acceptance criteria
    Surface Examination (ISO 11070:2014 Section 8.5 & Annex G, approved protocol)Predetermined acceptance criteriaMet the predetermined acceptance criteria
    Fracture Testing (ISO 11070:2014 Annex F)Predetermined acceptance criteriaMet the predetermined acceptance criteria
    Tensile Testing of the Union of the Core Wire and Coil (ISO 11070:2014 Annex H)Applicable values of ISO 11070:2014, Annex HMet the predetermined acceptance criteria
    Tensile Testing of the Union between Deflecting Handle and Wire Guide (ISO 11070:2017, Annex H)Applicable values of ISO 11070:2017, Annex HNo explicit statement, but generally implies meeting criteria
    Dimensional and Catheter Compatibility Testing (approved study protocol)Predetermined acceptance criteriaMet the predetermined acceptance criteria
    Radiopacity Testing (ASTM F640-12)Predetermined acceptance criteriaMet the predetermined acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the sample size for each test. It mentions "The following tests were performed...", implying a certain number of units were tested for each criterion, but the exact count is not given.
    • Data Provenance: Not applicable. These are physical performance tests of a device, not data analysis tests. The testing would have been performed by Cook Incorporated or a contracted lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The "ground truth" for these physical device tests is defined by the objective industry standards (ISO, ASTM) and the manufacturer's predetermined acceptance criteria for physical properties. No human experts are establishing "ground truth" in the context of diagnostic interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is not an adjudication process for clinical or diagnostic data. The tests are objective measurements against defined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This document describes a physical medical device, not an AI software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This document describes a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable in the AI/software context. For this medical device, the "ground truth" or reference for evaluating performance is the adherence to established international standards (ISO, ASTM) and internal engineering specifications for physical and material properties.

    8. The sample size for the training set

    • Not applicable. This is a physical device, not an AI model. There is no concept of a "training set."

    9. How the ground truth for the training set was established

    • Not applicable. This is a physical device, not an AI model.

    Summary of the Study per the Document:

    The document describes a series of performance tests conducted on the Reuter Tip Deflecting Wire Guide. These tests were designed to demonstrate that the modified device (with new lengths, diameter, and handle configuration) meets applicable design and performance requirements and supports a determination of substantial equivalence to its predicate device. The tests covered various aspects including:

    • Biocompatibility
    • Corrosion
    • Flexing
    • Surface Examination
    • Fracture
    • Tensile strength of the core wire and coil union
    • Tensile strength of the handle and wire guide union
    • Dimensional accuracy and catheter compatibility
    • Radiopacity

    For all listed tests, the document explicitly states that the "predetermined acceptance criteria were met," indicating successful completion of the studies and compliance with the specified standards. The study's conclusions support that the modifications do not raise new questions of safety or effectiveness and that the device is substantially equivalent to the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171912

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-19

    (266 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K140485, K142393, K081337

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Double Flexible Tipped Wire Guides are used to facilitate the placement of devices during diagnostic and interventional procedures.

    Device Description

    The Double Flexible Tipped Wire Guides are Class II devices according to 21 CFR §870.1330; product code DQX (Wire, Guide, Catheter). The Double Flexible Tipped Wire Guides are wire guides with a flexible distal and proximal end. Double Flexible Tipped Wire Guides are intended to facilitate placement of devices used during diagnostic and interventional procedures.

    The Double Flexible Tipped Wire Guides, subject of this submission, have been modified from the predicate device, Heavy Double Flexible Tipped Wire Guide (K150802), to include additional lengths ranging from 30 to 260 centimeters, additional outside diameters ranging from 0.018 to 0.038 inches, and additional curve radii ranging from 0 to 3 millimeters.

    The Double Flexible Tipped Wire Guides are manufactured using a stainless steel core wire, and stainless steel coils. Some configurations include an exterior coating on the coils.

    The modifications to the Double Flexible Tipped Wire Guides have been created to provide additional options for physicians in selecting the appropriate wire guide during diagnostic and interventional procedures.

    The Double Flexible Tipped Wire Guides are packaged, sterile devices intended for single use.

    AI/ML Overview

    The provided text describes the Double Flexible Tipped Wire Guides. Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Fracture TestingMet pre-determined acceptance criteria (in accordance with Annex F of BS EN ISO 11070:2014).
    Tensile TestingMet pre-determined acceptance criteria (in accordance with Annex H of BS EN ISO 11070:2014).
    Flexing TestMet pre-determined acceptance criteria (in accordance with Annex G of BS EN ISO 11070:2014).
    Coating IntegrityDevice evaluated for flaking of the coating using magnification to 40X. (No explicit "met" statement, but implied acceptable based on overall conclusion).
    Torque Strength TestingCharacterization testing performed (in accordance with FDA Coronary and Cerebrovascular Guidewire Guidance (1995)). (No explicit "met" statement, but implied acceptable based on overall conclusion).
    Tip Flexibility (Tip Deflection) TestingCharacterization testing performed (in accordance with FDA Coronary and Cerebrovascular Guidewire Guidance (1995)). (No explicit "met" statement, but implied acceptable based on overall conclusion).
    Corrosion TestingRevealed no signs of corrosion that would affect functional performance. Met pre-determined acceptance criteria (in accordance with Annex B of BS EN ISO 11070:2014).
    Catheter Compatibility TestingMet pre-determined acceptance criteria (in accordance with BS EN ISO 11070:2014).
    Biocompatibility TestingMaterials and methods used to manufacture the device are non-toxic and met pre-determined acceptance criteria for their intended use (in accordance with BS EN ISO 10993-1:2009).
    Performance Testing on Aged DevicesMet pre-determined acceptance criteria (in accordance with BS EN ISO 11070:2014).
    Acute Performance Evaluation (Animal Model)All test articles evaluated met the pre-determined acceptance criteria for all performance parameters.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test mentioned (e.g., number of wire guides tested for fracture, tensile, etc.). It only states that the various tests were "performed" and that "all test articles evaluated" met acceptance criteria for the animal model.

    Data provenance is not explicitly stated as retrospective or prospective, nor are countries of origin for the data. The tests are described as being performed "to demonstrate that the Double Flexible Tipped Wire Guides met applicable design and performance requirements."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This document does not describe a study involving human experts establishing ground truth for a test set. The evaluation is focused on the physical and biological performance of the medical device itself through standardized engineering and biocompatibility tests, and an animal model.

    4. Adjudication method for the test set

    Not applicable. This was not a study involving human interpretation or adjudication of results. The acceptance criteria were met based on direct measurements and observations during testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (wire guides) for diagnostic and interventional procedures, not an AI-assisted diagnostic tool. Therefore, no MRMC comparative effectiveness study involving human readers and AI assistance was conducted or would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance of the device was established by:

    • Standardized Test Methods: Adherence to established international standards (BS EN ISO 11070:2014, BS EN ISO 10993-1:2009) and FDA guidance documents (FDA Coronary and Cerebrovascular Guidewire Guidance (1995)).
    • Pre-determined Acceptance Criteria: Specific criteria were set for each test, and the device's performance was compared against these criteria.
    • Acute Performance Evaluation in an Animal Model: This constitutes a form of "ground truth" reflecting real-world physiological conditions for device functionality.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171764

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-09

    (268 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140485, K120119, K122468, K142393

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fixed Core Wire Guides are intended to facilitate placement of devices used during diagnostic and interventional procedures.

    Device Description

    The Fixed Core Wire Guides, subject of this submission, are Class II devices according to 21 CFR §870.1330; product code DQX (Wire, Guide, Catheter). The subject device consists of a core mandril, a spring coil anchored over the mandril, and an inner safety wire. The Fixed Core Wire Guides are available with outside diameters ranging from 0.015 inches to 0.052 inches and lengths ranging from 15 centimeters to 480 centimeters. The flexible tip portion of the wire guides is either straight or J-tipped. The Fixed Core Wire Guides will be uncoated or coated with Polytetrafluoro-ethylene (PTFE). The Fixed Core Wire Guides are packaged, sterile devices intended for single use.

    AI/ML Overview

    The document describes the premarket notification (510(k)) for the Fixed Core Wire Guides. However, it does not pertain to an AI/ML medical device. Therefore, much of the information requested in your prompt (e.g., patient data, training/test sets, expert consensus, MRMC studies, standalone performance, etc.) is not applicable to this submission, as it is for a physical medical device (wire guides).

    The "study that proves the device meets the acceptance criteria" refers to the various engineering and biological tests conducted on the physical device.

    Here's an attempt to answer the prompt based on the provided document, acknowledging the non-AI nature of the device:


    Device: Fixed Core Wire Guides (K171764)
    Device Type: Physical Medical Device (Catheter Guide Wire)
    Regulation: 21 CFR 870.1330 (Catheter Guide Wire)

    Given that this is a 510(k) submission for a physical medical device (wire guide) and not an AI/ML-driven device, many of the standard questions regarding AI/ML acceptance criteria and study design are not relevant or applicable. The "acceptance criteria" here refer to the passing thresholds for various physical and biological performance tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide specific quantitative acceptance criteria values or detailed performance results for each test. Instead, it states that "The pre-determined acceptance criteria were met" for all listed tests.

    Test CategoryAcceptance Criteria Description (as stated)Reported Device Performance (as stated)
    BiocompatibilityTested in accordance with ISO 10993-1:2009.The pre-determined acceptance criteria were met.
    Corrosion TestingTested in accordance with Annex B of ISO 11070:2014.The pre-determined acceptance criteria were met.
    Flexing TestTested in accordance with Annex G of ISO 11070:2014.The pre-determined acceptance criteria were met.
    Fracture TestingTested in accordance with Annex F of ISO 11070:2014.The pre-determined acceptance criteria were met.
    Tensile TestingTested in accordance with ISO 11070:2014, Annex H.The pre-determined acceptance criteria were met.
    Tip FlexibilityCharacterization testing performed in accordance with the FDA Coronary and Cerebrovascular Guidewire Guidance (1995).(No explicit statement of "met criteria," but characterization was successful)
    Torque Strength TestingCharacterization testing performed in accordance with the FDA Coronary and Cerebrovascular Guidewire Guidance (1995).(No explicit statement of "met criteria," but characterization was successful)
    Acute Performance EvaluationEvaluated in an animal model in accordance with an approved study protocol.The pre-determined acceptance criteria were met.

    2. Sample size used for the test set and the data provenance

    For a physical device, "sample size" typically refers to the number of device units tested. The document does not specify the exact sample sizes (number of units) used for each of the performance tests (e.g., biocompatibility samples, number of wires for tensile testing, etc.).

    • Data Provenance: The tests are laboratory and animal studies, not human clinical data.
      • Biocompatibility, Corrosion, Flexing, Fracture, Tensile, Tip Flexibility, Torque Strength: Performed in a laboratory setting.
      • Acute Performance Evaluation: Performed in an animal model.
    • Retrospective or Prospective: These would be considered prospective tests performed specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. For physical device performance testing, "ground truth" is established by the test procedures themselves (e.g., measuring force, observing breakage, biological reactions) and assessed by qualified laboratory personnel and engineers, not clinical experts for diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept is relevant for human interpretation of data, not for material or mechanical performance testing of a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is for AI/ML devices where human readers interact with AI. This device is a physical wire guide.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This pertains to AI/ML software performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this physical device, the "ground truth" for demonstrating performance is derived from:

    • Standardized industry and international test methods (e.g., ISO, FDA guidance documents).
    • Quantitative measurements (e.g., force, elongation, cycles to failure).
    • Qualitative observations (e.g., absence of corrosion, successful access in animal model).
    • Biocompatibility assessments based on established biological endpoints.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of AI/ML or statistical modeling from patient data. The "training" for such a device would be the design and manufacturing processes.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1