510(k) Data Aggregation

    K Number
    K172188

    Validate with FDA (Live)

    Date Cleared
    2017-10-06

    (78 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K170177, K142465, K133324

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.

    The Aquilion Prime SP has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

    Device Description

    The Aquilion Prime SP TSX-303B/1 is an 80-row CT System that is intended to acquire and display cross-sectional volumes of the whole body, including the head. This system is based upon the technology and materials of previously marketed Toshiba CT systems.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Toshiba Aquilion Prime SP, TSX-303B/1, v8.4. It outlines modifications to a previously cleared CT system. While the document mentions various performance evaluations and studies, it does not contain specific acceptance criteria tables nor detailed study designs that definitively "prove" the device meets acceptance criteria in the format of a typical peer-reviewed clinical study. Instead, it focuses on demonstrating substantial equivalence to a predicate device through engineering and performance testing.

    However, I can extract and infer information about the testing and performance as described in the document.

    Missing Information:

    • A clear table of acceptance criteria for specific performance metrics. The document describes improvements but doesn't explicitly state "acceptance criteria" values met.
    • Detailed sample sizes for all tests.
    • Specific data provenance for all tests (e.g., country of origin, retrospective/prospective).
    • Number and qualifications of experts for all ground truth establishment.
    • Adjudication methods.
    • MRMC comparative effectiveness study details (effect size of human reader improvement with AI).
    • Standalone algorithm performance (the device is a CT system, not an algorithm in the AI sense).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    Based on the provided text, here's what can be extracted and inferred:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria. Instead, it describes performance improvements and that the modified system "demonstrates equivalent or slightly improved image quality characteristics." The performance evaluations are primarily focused on physical parameters and dose reduction, not diagnostic accuracy in the way an AI algorithm might be assessed against clinical endpoints.

    Performance MetricReported Device Performance (Aquilion Prime SP, TSX-303B/1, v8.4)Implied Acceptance Criterion (relative to predicate)
    Spatial ResolutionEvaluated; demonstrated equivalent or slightly improved image quality.Equivalent or improved
    Axial Slice Thickness/Slice Sensitivity ProfileEvaluated; demonstrated equivalent or slightly improved image quality.Equivalent or improved
    CT Number Magnitude/UniformityEvaluated; demonstrated equivalent or slightly improved image quality.Equivalent or improved
    Noise PropertiesEvaluated; demonstrated equivalent or slightly improved image quality.Equivalent or improved
    Low Contrast Detectability (LCD)Evaluated; demonstrated equivalent or slightly improved image quality.Equivalent or improved
    Contrast-to-Noise Ratio (CNR)Evaluated; demonstrated equivalent or slightly improved image quality.Equivalent or improved
    Dose Reduction (with AIDR 3D Enhanced)51% to 75% dose reduction supported while preserving LCD and high contrast spatial resolution.Not explicitly stated, but demonstrated within range
    Dose Reduction (with PURE ViSION Optics)20%-31% quantitative dose reduction.Not explicitly stated, but demonstrated within range
    LCD Improvement (Head, PURE ViSION Optics)Range 13%-19% improvement.Not explicitly stated, but demonstrated improvement
    LCD Improvement (Body, PURE ViSION Optics)Range 15%-22% improvement.Not explicitly stated, but demonstrated improvement
    Noise Reduction (PURE ViSION Optics)13% noise reduction at the same dose.Not explicitly stated, but demonstrated improvement
    Diagnostic Quality of ImagesProduces images of diagnostic quality for head, chest, abdomen, and peripheral exams.Diagnostic quality maintained

    2. Sample size used for the test set and the data provenance

    • Sample Size for Physical Performance Tests: Not explicitly stated. The tests involved "model observer studies" using MITA-FDA LCD Head and MITA-FDA LCD Body phantoms, implying a phantom-based test set rather than patient data.
    • Sample Size for Image Review: "Representative diagnostic images" were obtained. The exact number is not specified.
    • Data Provenance: Not specified. Phantoms for performance tests. Clinical images for diagnostic quality assessment (implicitly from a clinical setting, but no country of origin or retrospective/prospective status is mentioned).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: One.
    • Qualifications of Expert: An "American Board Certified Radiologist." Further details on experience (e.g., years) are not provided.
    • Role: This radiologist "reviewed" the "representative diagnostic images" to confirm they were of "diagnostic quality."

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable or not specified in detail. The document states a single American Board Certified Radiologist reviewed images. There is no mention of consensus or multi-reader adjudication for this informal review of diagnostic quality. For the quantitative performance metrics (dose reduction, LCD, noise), these were based on phantom studies and model observer analysis, not human reader adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. The document does not describe a MRMC comparative effectiveness study. This submission is for a CT system itself, not an AI-assisted diagnostic tool in the typical sense of showing improved human reader performance. The "AI" mentioned (AIDR 3D Enhanced, SEMAR) refers to image processing algorithms within the CT system to improve image quality or reduce artifacts, not a separate AI application for diagnosis or interpretation assistance that would warrant an MRMC study comparing human readers with and without its aid.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, in a way. The "performance testing" of the modified system, including spatial resolution, CT number, noise properties, LCD, and CNR, as well as the quantitative dose reduction and LCD/noise improvement studies using phantoms and model observers, represent a standalone evaluation of the system's technical image quality parameters. These are inherent algorithmic and hardware performance metrics of the CT scanner, not dependent on human interpretation for their measurement.

    7. The type of ground truth used

    • For Quantitative Performance: Model observer studies using MITA-FDA LCD Head and MITA-FDA LCD Body phantoms. These phantoms represent a controlled, objective ground truth for physical image quality parameters.
    • For Diagnostic Quality: The subjective assessment of an "American Board Certified Radiologist" confirming images were of "diagnostic quality." This is expert opinion/consensus for a qualitative judgment rather than a definitive "ground truth" like pathology.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable / Not provided. This document describes a 510(k) submission for a CT scanner, not a machine learning algorithm that requires a "training set" in the conventional sense. While there might be internal development and validation data, it's not discussed as a distinct "training set" within this regulatory context.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment for Training Set: Not applicable / Not provided, as there is no described training set for an AI algorithm in the context of this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160587

    Validate with FDA (Live)

    Date Cleared
    2016-06-09

    (100 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.

    The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician.

    Device Description

    Aquilion ONE (TSX-305A/3) V7.3 is a whole body multi-slice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. This device captures cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. This system is based upon the technology and materials of previously marketed Toshiba CT systems.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Aquilion ONE (TSX-305A/3) V7.3:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document is a 510(k) summary for a premarket notification for a Computed Tomography X-ray System. It is not a clinical study report with specific acceptance criteria directly tied to a diagnostic performance metric (like sensitivity or specificity) of a disease-detecting AI algorithm. Instead, it demonstrates substantial equivalence to a predicate device, focusing on technical specifications and image quality for general diagnostic use.

    Therefore, the "acceptance criteria" here relate to demonstrating that the new device performs acceptably for its intended use and is equivalent to the predicate. The "performance" is primarily a comparison of technical specifications and image quality metrics against the predicate.

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
    Intended UseThe device is capable of acquiring and displaying cross-sectional volumes of the entire body, including the head, with the capability to image whole organs in a single rotation (e.g., brain, heart, pancreas). These volume sets should be usable for specialized studies by trained physicians. (Identical to predicate)Aquilion ONE (TSX-305A/3) V7.3 has identical Indications for Use as the predicate Aquilion ONE Vision, TSX-301C/1-8, V7.0. It is a whole-body multi-slice helical CT scanner for acquiring and displaying cross-sectional volumes and whole organs.
    Technical Specifications (Substantial Equivalence)Technical specifications should be comparable to the predicate device, or any differences should not raise new questions of safety and effectiveness. (e.g., gantry rotation speed, view rate, detector, pitch factor, FOV, wedge filter types, X-ray tube voltage/current, image reconstruction time, helical reconstruction method, metal artifact reduction, patient couch type, size, weight capacity, gantry opening, gantry tilt angle, minimum area for installation, area finder. Also, existing cleared software options being implemented should function as previously cleared.)Similarities:- View rate: Maximum 2910 views/s (same)- Detector: 896 channels x 320 rows (same)- Pitch factor: Range 0.555 to 1.575 / 0.555 to 1.5 (very similar)- FOV: 240/320/500mm / 180/240/320/400/500mm (subject has slightly reduced range, but still within typical diagnostic needs)- Metal artifact reduction: SEMAR (Volume, Helical, ECG gated) / SEMAR (Volume, Helical) (subject has added ECG gated capability)- Gantry opening size: 780 mm (same)- All previously cleared software options are listed as "no change" in functionality, with some having "workflow improvements" (e.g., Lung Volume Analysis, surESubtraction Lung, MyoPerfusion, Dual Energy System Package, 4D Airways Analysis) which are enhancements rather than regressions.Differences (addressed through testing or not raising new concerns):- Gantry Rotation Speed: 0.35s (Optional max 0.275s) for subject vs. 0.275s (Standard or optional) for predicate. This indicates a minor hardware difference, likely addressed by showing image quality is maintained.- Wedge filter types: Two types for subject vs. Three for predicate. This is a minor design change.- X-ray tube voltage/current: Max 72kW (Optional Max 90kW) for subject vs. Max 90kW (for one model) or Max 72kW (for others) for predicate. Comparable.- Image reconstruction time: Up to 80 images/s for subject vs. Up to 50 images/s for predicate. Improvement in subject device.- Helical reconstruction method: 20 rows or more: TCOT+ for subject vs. 16 rows or more: TCOT+ for predicate. Improvement in subject device (more rows).- Patient Couch Type and related dimensions/weights: Various configurations/differences between subject and predicate models, indicating design variations but within expected functional range.- Gantry tilt angle: ±30° for subject vs. ±22° for predicate. Improvement in subject device.- Minimum area for installation: Smaller for subject (27m² vs 37.2m²). Improvement in subject device.- Area finder: Optional for subject vs. NA for predicate. New feature on subject device.
    Image QualityImage quality metrics (spatial resolution, CT number magnitude/uniformity, noise properties, low contrast detectability/CNR performance) should meet established specifications and be comparable to the predicate device. Images obtained should be of diagnostic quality.CT image quality metrics performed using phantoms demonstrated that the subject device is substantially equivalent to the predicate device with regard to: spatial resolution, CT number magnitude/uniformity, noise properties, and low contrast detectability/CNR performance. Representative diagnostic images (head, chest, abdomen/pelvis, extremity, cardiac) were also reviewed and demonstrated diagnostic quality.
    Safety and Standards ComplianceThe device must be designed and manufactured under Quality System Regulations (21 CFR 820, ISO 13485) and conform to applicable performance standards for ionizing radiation-emitting products (21 CFR, Subchapter J, Part 1020). It must also comply with various IEC, NEMA, and other relevant standards.The device is designed and manufactured under QSR and ISO 13485. It conforms to applicable performance standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020] and numerous international standards including IEC60601-1 series, IEC60601-2 series, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25 and NEMA XR-26.
    Software ValidationSoftware documentation must comply with FDA guidance for a Moderate Level of Concern, and validation testing should be successfully completed.Software Documentation for a Moderate Level of Concern was included. Successful completion of software validation is cited in the conclusion.
    Risk ManagementRisk analysis should be conducted.Risk analysis was conducted.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Description: The "test set" for this submission primarily consists of:
      • Phantoms: Used for evaluating CT image quality metrics (spatial resolution, CT number, noise, low contrast detectability). The number and specific types of phantoms are not explicitly stated but are typically standard phantoms used in CT performance testing.
      • Representative Diagnostic Images: Clinical images covering various body regions (head, chest, abdomen/pelvis, extremity, cardiac). The number of cases/patients is not specified.
    • Data Provenance: The document does not explicitly state the country of origin for the diagnostic images. Given Toshiba Medical Systems Corporation is based in Japan and Toshiba America Medical Systems, Inc. is in the US, the data could originate from either region or a combination. The document also does not specify if the data was retrospective or prospective. However, for a 510(k) clearance based on substantial equivalence, particularly for a hardware/software update to a CT scanner, diagnostic images are often retrospectively collected or acquired as part of internal validation.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Number of Experts: One (1) expert is explicitly mentioned.
    • Qualifications of Experts: An "American Board Certified Radiologist." No specific years of experience are stated. This expert reviewed the representative diagnostic images to confirm diagnostic quality.

    4. Adjudication Method for the Test Set:

    • The document describes a single American Board Certified Radiologist reviewing images to confirm diagnostic quality. This indicates no formal adjudication method involving multiple readers (like 2+1 or 3+1) was used for this specific part of the evaluation. The assessment of image quality from phantoms would not typically involve expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document is for a general-purpose CT scanner system, not an AI-specific diagnostic tool that assists human readers. Therefore, there is no mention of an effect size for human reader improvement with or without AI assistance.

    6. If a Standalone (Algorithm-Only Without Human-in-the-Loop Performance) Study Was Done:

    • No, a standalone performance study in the context of an AI algorithm was not done. The Aquilion ONE (TSX-305A/3) V7.3 is a complete CT system where the "algorithm" refers to the image reconstruction and processing capabilities, which are inherent to the device's function. The study validates the overall system's ability to produce diagnostic images, not a separate AI algorithm's diagnostic accuracy. The performance is assessed on the system output.

    7. The Type of Ground Truth Used:

    • For the phantom studies, the "ground truth" is typically known physical properties of the phantoms (e.g., known dimensions, densities, contrast levels).
    • For the representative diagnostic images, the "ground truth" for confirming "diagnostic quality" is based on the expert opinion/consensus of an American Board Certified Radiologist. This is a form of expert consensus, albeit from a single expert in this stated context. There is no mention of pathology or outcomes data being used as ground truth for this submission.

    8. The Sample Size for the Training Set:

    • The document does not specify a separate "training set" sample size. This submission is for a medical imaging device (CT scanner) rather than a deep learning AI algorithm that undergoes distinct training. The underlying algorithms for image reconstruction and processing (e.g., TCOT+, SEMAR) are developed and refined through engineering and iterative testing, but not typically in the same "training set" paradigm as AI for diagnostic interpretation. The software validation is mentioned, which refers to standard software development lifecycle testing.

    9. How the Ground Truth for the Training Set Was Established:

    • As a "training set" in the context of AI development is not explicitly mentioned as relevant to this submission, the establishment of ground truth for a training set is not applicable/described. The "ground truth" during the development of a CT scanner's image reconstruction algorithms would typically involve engineering specifications, physical models, and potentially early clinical data used for empirical tuning and validation, but not a formally labeled training set in the AI sense.
    Ask a Question