510(k) Data Aggregation

    K Number
    K150407

    Validate with FDA (Live)

    Date Cleared
    2015-05-15

    (86 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K131942

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioFlo Midline Catheter is indicated for short term access (< 30 days) to the peripheral venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and the sampling of blood and blood products. Therapies not appropriate for midline catheters include continuous vesicant therapy, parenteral nutrition, infusates with pH less than 5 or greater than 9, and infusates with an osmolarity greater than 600 mOsm/L. Maximum Power Injection Flow Rates:

    • 3F Single Lumen, 20cm: 2 mL/sec .
    • . 4F Single Lumen, 20cm: 6 mL/sec
    • . 5F Single Lumen, 20cm: 6 mL/sec
    • 5F Dual Lumen, 20cm: 6 mL/sec .
    Device Description

    The BioFlo Midline Catheter is a short term (< 30 days) non-valved peripheral venous access devices between 3 to 10 inches in length (8 to 25 cm). Midlines are usually placed in an arm vein such as the basilic, brachial or cephalic and the tip ends below the level of the axillary line. Midline catheters are longer than peripheral IV catheters which are generally 1 to 3 inches long and shorter than peripherally inserted central catheters (PICC) which extend into the superior vena cava. This device provides an alternative to short peripheral IVs and PICCs for certain treatments.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the BioFlo Midline Catheter. It outlines the device's characteristics, intended use, and a comparison with predicate and reference devices, along with performance data from non-clinical bench testing. However, the document does not describe a clinical study in the format requested, nor does it provide the detailed acceptance criteria and study results for a device that relies on AI or machine learning.

    The acceptance criteria mentioned are general performance aspects for a medical device (intravascular catheter), and the "study" referred to is a series of non-clinical bench tests. The concept of "AI performance," "human reader improvement with AI," "standalone algorithm performance," "ground truth," and "training set" are not applicable to this particular device and its regulatory submission as described.

    Therefore, many of the requested fields cannot be filled based on the provided text.

    Here's an attempt to extract relevant information from the document, acknowledging the limitations:


    1. Table of acceptance criteria and the reported device performance:

    The document lists various non-clinical bench tests performed. These tests implicitly define "acceptance criteria" through their successful completion, demonstrating that the device meets relevant standards for its intended use. Exact quantitative acceptance criteria are not explicitly stated in numerical form within this summary, but the successful completion of the tests implies meeting the required thresholds.

    Acceptance Criteria (Implied by successful test)Reported Device Performance (Implied by successful test)
    Power Injection Flow Rate requirementsMet
    Static Burst Strength requirementsMet
    Multiple Power Injections requirementsMet
    Gravity Flow Rate requirementsMet
    Catheter Length specificationsMet
    Priming Volume specificationsMet
    Midline IdentificationMet
    Dimensional Verification (ID, OD, Length)Met
    Catheter Kink/Flex ResistanceMet
    Alcohol CompatibilityMet
    Catheter Marking & Identification/RadiopacityMet
    Tensile Testing (Catheter and Assembly)Met
    Compatibility Testing (with insertion accessories)Met

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified. The document only mentions "non-clinical bench testing." For bench tests, sample sizes can vary significantly depending on the specific test and standard requirements, but these are typically physical units, not patient data sets.
    • Data provenance: Not applicable in the context of clinical or image-based data. The tests are non-clinical bench tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is a physical medical device (catheter) undergoing bench testing, not a diagnostic or AI-driven device requiring expert-established ground truth from patient data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This concept is relevant for clinical studies or AI evaluations involving human readers and ground truth, not for the described bench tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-based device, and no MRMC study was performed or mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the context of patient data ground truth. The "ground truth" for bench testing is defined by the technical specifications and performance requirements outlined in the referenced standards (e.g., ISO 10555-1, ISO 594-2).

    8. The sample size for the training set:

    • Not applicable. This device does not involve an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This device does not involve an AI algorithm.

    Summary of the Study (Bench Testing) that Proves the Device Meets Acceptance Criteria:

    The study proving the BioFlo Midline Catheter meets acceptance criteria consists of a series of non-clinical bench tests. These tests were conducted in accordance with several FDA Guidance Documents and international standards, including:

    • FDA's "Guidance on Premarket Notification [510(k)] Submissions for Short-Term and Long-Term Intravascular Catheters"
    • EN ISO 10555-1:2013 - "Intravascular Catheters - Sterile and Single-Use Catheters - Part 1: General Requirements"
    • EN ISO 10555-3:2013 - "Intravascular Catheters Sterile and Single-Use Catheters Part 3: Central Venous Catheters"
    • ISO 594-2:1998 - "Conical Fittings with a 6% (Luer) Taper for Syringes, Needles, and Certain Other Medical Equipment - Part 2: Lock Fittings"
    • EN ISO 10993-1:2009 - "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process"

    The specific tests performed included: Power Injection Flow Rate, Static Burst Strength, Multiple Power Injections, Gravity Flow Rate, Catheter Length, Priming Volume, Midline Identification, Dimensional Verification, Catheter Kink/Flex Resistance, Alcohol Compatibility, Catheter Marking & Identification/Radiopacity Testing, Tensile Testing, and Compatibility Testing with insertion accessories.

    The submission concludes that "successful results of non-clinical bench testing" demonstrate that the proposed device is substantially equivalent to the predicate device. This implies that the device met all the required performance specifications and safety criteria as defined by the referenced standards and guidance documents.