510(k) Data Aggregation

    K Number
    K203396

    Validate with FDA (Live)

    Device Name
    DEKA SMARTPERIO
    Date Cleared
    2021-08-23

    (278 days)

    Product Code
    Regulation Number
    878.4810
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K124003, K200234, K952006

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DEKA SmartPerio system is intended to perform the following types of intraoral procedures: soft tissue dental, general, oral maxillofacial, and cosmetic surgery. The DEKA SmartPerio is intended for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact fiber-optic delivery system. The device is intended to be used in dentistry. The following are the oropharyngeal indications for use: Abscess Incision and Drainage Apthous Ulcers Treatment Biopsies Excision and Incision Crown lengthening Hemostatic assistance Fibroma Removal Frenectomy Frenotomy Gingival Incision and Excision Gingivectomy Gingivoplasty Operculectomy Oral Papillectomy Tissue retraction for Impression Vestibuloplasty. Selective ablation of enamel (first degree) caries Exposure of unerupted / partially erupted teeth Implant recovery Lesion (tumor) removal Leukoplakia Pulpotomy Pulpotomy as adjunct to root canal therapy Removal of filling material such as gutta percha or resin as adjunct treatment during root canal retreatment Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment level and tooth mobility tooth mobility. Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium.)

    Device Description

    The DEKA SmartPerio is a Nd:YAG laser device for soft-tissues intraoral treatments. The DEKA SmartPerio system delivers laser through an optical fiber that is guided to the target tissue with the aid of an handpiece and a tip. The DEKA Smartperio device consists of: An AC/DC power supply unit CPU controller LASER source Cooling system User interface with LCD touch screen Beam delivery system Laser activation is controlled by footswitch.

    AI/ML Overview

    The provided document is a 510(k) summary for the DEKA SmartPerio laser device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving the device meets specific acceptance criteria with quantifiable performance metrics.

    Therefore, much of the requested information cannot be extracted directly from this document. However, I can provide what is available, noting where information is absent.

    1. A table of acceptance criteria and the reported device performance

    This document does not specify quantitative acceptance criteria typically seen in performance studies for new medical devices, nor does it report specific device performance metrics in detail beyond a comparison of technical specifications with a predicate device.

    Instead, the document's primary "acceptance criteria" appear to be meeting electrical safety and electromagnetic compliance standards, as well as demonstrating substantial equivalence to a legally marketed predicate device.

    Acceptance Criteria (Inferred from regulatory context)Reported Device Performance (Summary from comparison)
    Substantial Equivalence:
    - Indications for useDEKA SmartPerio's indications are a subset of the predicate device's indications.
    - Technological Characteristics (Laser Wavelength, Energy per Pulse, Pulses per Second, Pulse duration, Average Power, Aiming Beam wavelength, Aiming beam power, Delivery system, Power Requirements, Weight, Dimensions)Most are identical or similar, with differences noted as not affecting safety and effectiveness.
    Safety Standards Compliance:
    - AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, cl:2009/(R)2012 and a2:2010/(R)2012 (Medical Electrical Equipment)Compliant
    - IEC 60601-1-2 Ed. 4 (Electromagnetic disturbance)Compliant
    - IEC60601-2-22 Ed 3.1 (Laser equipment safety)Compliant
    - IEC 60825-1 Ed. 3.0 (Safety of laser products)Compliant
    Biocompatibility:Subject device relies on biocompatibility of already FDA-cleared accessories.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document explicitly states "Clinical Performance Data: None" and "Non-Clinical Performance Data" only refers to compliance with electrical and laser safety standards. This implies there was no "test set" of patient data in the sense of a clinical trial or performance study. The data provenance and sample size for such a test set are therefore not applicable or not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable, as no clinical test set or ground truth adjudication using experts is mentioned.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set or ground truth adjudication is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The DEKA SmartPerio is a laser surgical instrument, not an AI-powered diagnostic or assistive device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The DEKA SmartPerio is a physical medical device (laser surgical instrument), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable, as no clinical performance study requiring a "ground truth" (e.g., for diagnostic accuracy) was performed or reported.

    8. The sample size for the training set

    Not applicable, as the DEKA SmartPerio is a physical laser device and not an AI or machine learning algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as the DEKA SmartPerio is a physical laser device and not an AI or machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160549

    Validate with FDA (Live)

    Date Cleared
    2017-04-07

    (403 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K124003

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "VELASI -15D" is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

    Device Description

    Medical diode laser systems are a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source. The medical diode laser systems for this unit is GaAlAs diode bar, and the wavelength is 1470nm of VELAS II-15D. It features impact structure, high efficiency and long lifetime. The beam divergence angle is too large lead to poor beam quality. In order to better meet the needs of Laser Diode practical applications, Laser Diode carried beam shaping is very necessary. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber. To start the laser unit, turn the main switch ON and turn the key switch clockwise to "I" position. The power indicator will turn green immediately with the system fans working. At the same time, the LCD screen lights up. After the parameters setting is finished, press the "Ready" button and the system will remind you to wear the protective eyewear (protective wavelength is from 800nm to 1700mm). At this time when you press down the footswitch, the laser will emit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, specifically a Medical Diode Laser System (VELAS II -15D). It focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than defining specific performance acceptance criteria for the novel device itself.

    Therefore, the document does not explicitly state specific acceptance criteria in terms of numerical performance thresholds (e.g., accuracy, sensitivity, specificity) for the device's clinical efficacy in treating reflux of saphenous veins. Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence to the predicate device, meaning the new device should perform as safely and effectively as the predicate.

    The "study that proves the device meets the acceptance criteria" refers to the non-clinical testing performed to demonstrate adherence to relevant safety and performance standards, and a comparison presented in the "Summary of Substantial Equivalence" table.

    Here's an analysis based on the information provided, though it will not contain specific performance metrics for clinical efficacy as none are present.


    1. Table of Acceptance Criteria and the Reported Device Performance

    As specific clinical performance acceptance criteria (e.g., success rate of vein closure, reduction in vein diameter) are not provided in this regulatory submission for the novel device, the "acceptance criteria" are inferred from the comparison to the predicate device, primarily focusing on technical specifications and safety standards.

    Acceptance Criterion (Inferred from Substantial Equivalence)Reported Device Performance (VELAS II -15D)
    Indications for Use"The 'VELAS II -15D' is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities."
    Laser TypeDiode laser
    Wavelength$1470nm \pm 10nm$
    Output power1-15W
    Operation modeCW, single pulse, repeat pulse
    Pulse width10ms-2.5s
    Pulse repetition rate0.2Hz-50Hz
    Application / Light delivery systemRecommended disposable sterile fiber (K124003, MED-Fibers, Inc.), bare fiber, 3m, 600μm fiber core diameter, NA ≥0.22, SMA905 connector, single use.
    Aiming BeamDiode laser of 635/532nm, power max. < 5mW, adjustable brightness.
    Laser Class4
    Operation interfaceColor LCD touch screen
    Power Supply100-240VAC, 50/60Hz, 350VA
    Safety classificationClass I Type B
    Waterproof levelIPX1 (device) / IPX8 (footswitch)
    Standards ComplianceIEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, FDA Guidance for Software.
    Non-sterileFiber is sterilized by EO.
    Microprocessor ControlYes
    Substantial Equivalence Conclusion"The proposed device of VELAS II -15D has the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device."

    Comparison to Predicate (K082225):

    • Indications for Use: Same.
    • Laser Type: Same.
    • Components: Similar (Proposed device does not equip with fiber and handpiece initially, but recommends a specific disposable fiber set; predicate includes these components directly). The submission argues this does not affect safety and effectiveness.
    • Wavelength: Similar ($1470nm \pm 10nm$ vs $1470nm \pm 30nm$). Proposed device has better accuracy.
    • Output power: Same (1-15W).
    • Operation mode: Same (CW, single pulse, repeat pulse).
    • Pulse width: Similar (10ms-2.5s vs 0.01s-60s). Proposed device's range is contained within the predicate, deemed sufficient for intended use.
    • Pulse repetition rate: Similar (0.2Hz-50Hz vs 0.01Hz-50Hz). Deemed not to affect safety and effectiveness.
    • Application / Light delivery system: Similar. Proposed device is more detailed, fiber core diameter range is contained within predicate's general description, deemed sufficient for intended use.
    • Aiming Beam: Similar. Proposed device's power range contains predicate's, deemed sufficient for intended use.
    • Laser Class: Same (4).
    • Operation interface: Same (Color LCD touch screen).
    • Power Supply: Similar (350VA vs 450VA). Rated power difference not affecting safety and effectiveness.
    • Safety classification: Same (Class I Type B).
    • Dimensions & Weight: Similar, proposed device is bigger and heavier. Not noted to affect safety or effectiveness.
    • Waterproof level: Same (IPX1 device, IPX8 footswitch).
    • Standard Compliance: Same listed standards.
    • Non-sterile: Same (Fiber sterilized by EO).
    • Microprocessor Control: Same (Yes).

    2. Sample size used for the test set and the data provenance

    The document specifies non-clinical tests (compliance with standards) and software verification. It does not describe a clinical study or a test set in the sense of patient data. Therefore, information regarding "sample size used for the test set" and "data provenance" (country of origin, retrospective/prospective) is not applicable or not provided in this regulatory submission, which focuses on substantial equivalence based on technical specifications and international standards compliance.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As no clinical test set/study is described, the concept of establishing ground truth by experts is not applicable or not provided in this document. The "ground truth" for the device's technical specifications and safety is established by adherence to international standards and internal quality assurance procedures, as overseen by the manufacturer's technical staff and verified through testing.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set or study requiring adjudication is described.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a surgical laser system, not an AI-assisted diagnostic tool or an imaging device involving "human readers." No MRMC study was mentioned or required for this type of submission.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a laser surgical instrument, meaning it inherently has a "human-in-the-loop" (the surgeon) performing the procedure. No standalone algorithm performance is relevant or described.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the technical performance aspects, the "ground truth" relies on:

    • International standards compliance: Demonstrated by testing against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
    • Software verification: According to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
    • Predicate device's established safety and effectiveness: The core of the 510(k) submission is to argue that the new device is "substantially equivalent" to an already approved predicate device, implying its "ground truth" for safety and effectiveness is comparable to that of the predicate, which has presumably demonstrated its clinical efficacy through other means (e.g., initial clinical trials for its approval, or long-standing safe use).

    No specific clinical "outcomes data" or "expert consensus" related to a new discrete clinical study for this specific device is provided in the document.


    8. The sample size for the training set

    Not applicable. This submission concerns a physical medical device (laser system), not an AI/machine learning model that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/machine learning model described.