    Reference Devices :

    K123907

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    General Total Knee Arthroplasty (TKR) Indications:

    • · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
    • · Post-traumatic loss of knee joint configuration and function.
    • · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    • · Revision of previous unsuccessful knee replacement or other procedure.
    • Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques.

    Additional Indications for Posterior Stabilized (PS) Components:

    • · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
    • · Absent or non-functioning posterior cruciate ligament.
    • · Severe anteroposterior instability of the knee joint.

    The Triathlon® Pro PS Femoral Components are intended for cemented use only.

    Device Description

    The subject Triathlon® Pro PS Femoral Component is a line extension to the existing Triathlon® Total Knee System and will be a modified version of the predicate Triathlon® PS Femoral Components. The subject component is sterile, single use, and available in sizes 1-8 and in right and left configurations. The femoral component is manufactured from Cobalt-Chromium alloy, which meets ASTM material standard F2886-17 as well as F75-18 standard requirements in its final finished form. It is intended for cemented use only.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the "Triathlon® Total Knee System - Triathlon® Pro Posterior Stabilized (PS) Femoral Component". It focuses on establishing substantial equivalence to previously cleared devices rather than defining new acceptance criteria or conducting a study proving the device meets new acceptance criteria.

    Therefore, the information you requested about acceptance criteria and a study proving the device meets them, particularly regarding AI assistance, multi-reader multi-case studies, and detailed ground truth methodologies, is not applicable to this submission. This is a medical device clearance based on substantial equivalence, not a performance study of a diagnostic AI algorithm.

    Here's a breakdown of why many of your questions cannot be answered from the provided text:

    • No specific "acceptance criteria" are provided in the context of device performance metrics, as this is a comparison to established predicates. The acceptance is based on demonstrating substantial equivalence in design, materials, and non-clinical performance characteristics.
    • No "study that proves the device meets the acceptance criteria" in terms of clinical outcomes or diagnostic performance is present. Instead, the sponsor performed non-clinical lab testing and engineering analysis to demonstrate that the new device performs similarly to predicate devices.
    • The device is a knee implant (femoral component), not an AI diagnostic tool. Therefore, questions related to AI effects, human reader improvement, and expert ground truth for diagnostic accuracy are not relevant.

    However, I can extract information related to the non-clinical testing performed to support substantial equivalence:

    1. A table of acceptance criteria and the reported device performance

    Since specific numerical acceptance criteria (e.g., "sensitivity > 90%") for AI performance are not defined or applicable here, I will instead list the types of non-clinical tests performed and the conclusion drawn.

    Acceptance Criteria Type (as demonstrated via non-clinical testing for substantial equivalence)Reported Device Performance / Conclusion
    Material Characterization (CoCr per ASTM F2886-17 and ASTM F75-18)Demonstrated compliance with specified material standards.
    Femoral Component Fatigue TestingPerformance data supports substantial equivalence (specific values not provided in this summary, but implied acceptable from conclusion).
    Wear Testing (per ISO 14243-3:2014)Performance data supports substantial equivalence (implied acceptable from conclusion).
    Characterization of Tibiofemoral Contact Area/Contact StressPerformance data supports substantial equivalence (implied acceptable from conclusion).
    Characterization of Patellofemoral Contact Area/Contact StressPerformance data supports substantial equivalence (implied acceptable from conclusion).
    Characterization of Articulating/Bearing Surface FinishPerformance data supports substantial equivalence (implied acceptable from conclusion).
    Characterization of Range of Motion and ConstraintPerformance data supports substantial equivalence (implied acceptable from conclusion).
    Biocompatibility (evaluated per ISO 10993-1:2018)Demonstrated compliance with biocompatibility standards.
    Shelf-life (validated per ISO 11607-1:2019, ISO 11607-2:2019, ASTM F1980-21)Demonstrated compliance with shelf-life validation standards.
    Bacterial endotoxin testing (BET) (ANSVAAMI ST72:2019 for pyrogenicity, endotoxin limit <20EU/Device)Demonstrated endotoxin limit of <20EU/Device.
    MR Compatibility (per FDA guidance May 2021, ASTM F2052-15, ASTM F2213-17, ASTM F2119-07(2013), ASTM F2182-19e2)Demonstrated MR safety and compatibility: Magnetically Induced Displacement Force, Magnetically Induced Torque, Image Artifact, and Heating by Radio Frequency Fields were evaluated according to standards.
    Overall ConclusionBased upon a comparison of the intended use, indications for use, design, material, sterilization method, biocompatibility, MR compatibility, technical and performance characteristics, and operational principles, the subject device is substantially equivalent to the cited predicate devices.

    Regarding the specific questions that are not applicable to this medical device submission:

    1. Sample size used for the test set and the data provenance: Not applicable. There is no "test set" of clinical cases for diagnostic performance. Non-clinical engineering tests were performed on device prototypes or specific material samples. The number of samples for each engineering test is not specified in the summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth was established by experts for a diagnostic performance test set.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No adjudication on clinical cases was performed.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of diagnostic accuracy. The "ground truth" for the non-clinical tests would be the established engineering and material standards (e.g., ASTM, ISO) that the device must meet to demonstrate specific physical properties or characteristics.
    7. The sample size for the training set: Not applicable. No AI model training set was used.
    8. How the ground truth for the training set was established: Not applicable. No AI model training set was used.