Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242190

    Validate with FDA (Live)

    Date Cleared
    2025-03-05

    (223 days)

    Product Code
    Regulation Number
    862.1205
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121214

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in serum, plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

    The DxC 500i Clinical Analyzer combines the DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The system is for in vitro diagnostic use only.

    The chemistry module of the DxC 500i Clinical Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (OC) material and other accessories. The immunoassay module of the DxC 500i Clinical Analyzer is an in-vitro diagnostic device used for the quantitative, semiquantitative, or qualitative determination of various analyte concentrations found in human body fluids.

    Device Description

    The Access Cortisol assay is a competitive binding immuno-enzymatic assay designed for use on Beckman Coulter's Access immunoassay analyzers in a clinical laboratory setting.

    The DxC 500i Clinical Analyzer is an integrated chemistry-immunoassay work cell that combines Beckman Coulter's DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The DxC 500i instrument has a single user interface and common point of entry for sample racks; the sample handling unit operates as a parallel processor and sample manager for both sides of the instrument. The DxC 500i operates in conjunction with the existing reagents, calibrators, controls, and system solutions for the AU and Access instrument families.

    AI/ML Overview

    The provided text describes the Beckman Coulter Access Cortisol assay on the DxC 500i Clinical Analyzer and its comparison to a predicate device. Here's a breakdown of the acceptance criteria and study information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Method ComparisonSlope criteria of 1.00 ± 0.12 (using Weighted Deming regression analysis when compared to predicate device)Serum: Slope = 0.974 (95% CI: 0.952 - 0.996) Urine: Slope = 1.002 (95% CI: 0.976 - 1.029)
    LinearityLinear throughout the analytical measuring range.Determined to be linear throughout the analytical measuring range (2.3 - 60.0 µg/dL).
    Imprecision (Repeatability & Total)Allowable imprecision of <12% at approximately 5 µg/dL and <10% for higher concentrations.Control L1 (approx 3 µg/dL): Instrument 1: Repeatability CV = 4.8%, Total CV = 6.4% Instrument 2: Repeatability CV = 5.6%, Total CV = 7.2% Control L2 (approx 17.7 µg/dL): Instrument 1: Repeatability CV = 3.2%, Total CV = 3.6% Instrument 2: Repeatability CV = 2.8%, Total CV = 4.1% Control L3 (approx 27 µg/dL): Instrument 1: Repeatability CV = 3.0%, Total CV = 3.7% Instrument 2: Repeatability CV = 2.9%, Total CV = 3.6% Serum Pool (approx 47-49 µg/dL): Instrument 1: Repeatability CV = 2.7%, Total CV = 3.8% Instrument 2: Repeatability CV = 3.0%, Total CV = 5.0%
    Detection CapabilityNot explicitly stated as acceptance criteria, but results are provided for LoB, LoD, and LoQ.Limit of Blank (LoB): ≤ 0.4 µg/dL Limit of Detection (LoD): ≤ 0.4 µg/dL Limit of Quantitation (LoQ) ≤ 20% within-lab CV: ≤ 0.8 µg/dL

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Method Comparison:
      • Serum: N = 104
      • Urine: N = 115
    • Imprecision: For each of the two instruments and four control levels/serum pools, N = 80 measurements were performed (this implies 20 days x 2 replicates x 2 runs, or similar, over the course of the study).
    • Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. It does refer to "patient samples" for the method comparison, suggesting real-world samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the device is an in-vitro diagnostic test for cortisol levels, not an imaging device requiring expert interpretation of results for ground truth. The "ground truth" for comparison and performance evaluation is generally established by the predicate device's measurements and reference methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for this type of in-vitro diagnostic device. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective human interpretation (e.g., radiology reads) where discrepancies between assessors need to be resolved. For quantitative laboratory tests, "ground truth" is typically established by comparing against established reference methods or predicate devices.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the device is an in-vitro diagnostic test for cortisol levels, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not directly applicable in the same way it would be for an AI algorithm. The Access Cortisol assay on the DxC 500i Clinical Analyzer is an automated in-vitro diagnostic system. The "standalone" performance is essentially what is being evaluated in the method comparison, linearity, imprecision, and detection capability studies, where the system itself generates the quantitative results without subjective human interpretation as part of the core measurement. Human operators are involved in running the instrument and interpreting the numerical results, but the measurement itself is automated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in these studies is established by:

    • Predicate Device Measurements: For the method comparison study, the Access Cortisol assay run on the predicate Access 2 Immunoassay System serves as the comparator or "reference" for evaluating the candidate device (Access Cortisol assay on the DxC 500i Clinical Analyzer).
    • Reference Materials/Established Protocols: For linearity, imprecision, and detection capability, the ground truth is often tied to known concentrations in control materials, spiked samples, and established analytical performance specifications derived from CLSI guidelines. The calibrators use "Human serum containing cortisol (purified chemical compound) at levels of 0 and approximately 2, 5, 10, 25, and 60 µg/dL," with traceability to "USP Reference Material."

    8. The sample size for the training set

    This information is not provided and is likely not explicitly documented in the submission for an IVD device of this nature. The instrument doesn't explicitly 'train' in the machine learning sense on a dataset. Its operational parameters are set during its development and validated through studies like those described.

    9. How the ground truth for the training set was established

    As there's no explicitly mentioned "training set" in the context of machine learning, this question is not directly applicable. The device's operational characteristics and performance specifications are established through rigorous analytical verification and validation using known standards, controls, and patient samples, rather than a "training set" in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221225

    Validate with FDA (Live)

    Date Cleared
    2022-11-10

    (196 days)

    Product Code
    Regulation Number
    862.1690
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    Access TSH (3rd IS) Assay, 510(k) Number K153651, Access 2 Immunoassay System, 510(k) Number K121214

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. This assay is capable of providing 3rd generation TSH results.

    The DxI 9000 Access Immunoassay Analyzer is an in vitro diagnostic device used for the quantitative, semiquantitative, or qualitative determination of various analyte concentrations found in human body fluids.

    Device Description

    Access TSH (3rd IS) assay is a two-site immunoenzymatic ("sandwich") assay. The Access TSH (3rd IS) reagent kit is in a liquid ready-to-use format designed for optimal performance on Beckman Coulter's immunoassay analyzers. Each reagent kit contains two reagent packs. Other items needed to run the assay include substrate, calibrators and wash buffer.

    The Dxl 9000 Access Immunoassay Analyzer is a fully automated, continuous, random-access sample processing and analysis instrument. The Dxl 9000 Access Immunoassay Analyzer uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

    AI/ML Overview

    The provided text describes the performance of the Access TSH (3rd IS) Assay on the DxI 9000 Access Immunoassay Analyzer. This is an in vitro diagnostic device, and the detailed information typically provided for AI/ML devices regarding ground truth establishment, expert qualifications, and MRMC studies is not directly applicable to this type of device.

    Here's a breakdown based on the information available:

    1. A table of acceptance criteria and the reported device performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Method ComparisonR² ≥ 0.90 and Slope 1.00 ± 0.10R²: 1.00 Slope: 1.06 (95% CI: 1.04, 1.07) Intercept: -0.019 (95% CI: -0.10, -0.0037)
    Imprecision (Within-laboratory/Total)Not explicitly stated as a single overall acceptance criterion, but implied to be within acceptable limits for a diagnostic assay.For TSH concentrations > 0.02 µlU/mL: - %CV ranged from 2.5% to 4.5%. For TSH concentrations ≤ 0.02 µlU/mL: - SD ranged from 0.0007 to 0.0014. (Detailed table with specific sample means and SD/CV values provided in the text)
    ReproducibilitySD ≤ 0.0038 for values ≤ 0.02 ulU/mL CV < 13.0% for values > 0.02 ulU/mLFor values ≤ 0.02 ulU/mL (Sample 1, mean 0.024): Overall Reproducibility SD = 0.0010 (meets criteria) For values > 0.02 ulU/mL: - Sample 2 (mean 0.37): Reproducibility CV = 4.3% (meets criteria) - Sample 3 (mean 4.8): Reproducibility CV = 3.7% (meets criteria) - Sample 5 (mean 12): Reproducibility CV = 3.8% (meets criteria) - Sample 4 (mean 46): Reproducibility CV = 3.4% (meets criteria)
    LinearityNot explicitly stated as a numerical criterion, but implies demonstration of accuracy across the measuring range.Linear throughout the analytical measuring interval of approximately 0.01 - 50.0 µIU/mL (mIU/L).
    Limit of Blank (LoB)Not explicitly stated as a numerical criterion, but implies a low detection threshold.0.002 µIU/mL
    Limit of Detection (LoD)Not explicitly stated as a numerical criterion, but implies a low detection threshold.0.003 µIU/mL
    Limit of Quantitation (LoQ)LoQ must be greater than or equal to LoD.Maximum LoQ determined was 0.001 µlU/mL, but reported as 0.003 µIU/mL to align with LoD, following CLSI EP17-A2 recommendation.

    2. Sample size used for the test set and the data provenance

    • Method Comparison Test Set: 111 serum samples.
    • Imprecision Test Set: 80 replicates per sample level across multiple runs/days.
    • Reproducibility Test Set: 75 replicates per sample level.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). Given it's an in vitro diagnostic device, these samples are typically laboratory-generated or clinical samples collected for analytical validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to the analytical performance studies conducted for this in vitro diagnostic device. The "ground truth" for these studies is established by the reference method (the predicate device for method comparison) or by known concentrations/spikes for other performance characteristics like linearity, LoB, LoD, LoQ, and imprecision.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used in clinical studies or for subjective interpretations of results, not for the analytical performance of an immunoassay system where quantitative measurements are directly compared.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated immunoassay analyzer, not an AI-assisted diagnostic tool that requires human interpretation or aids human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented as "Summary of studies" directly represent the standalone performance of the Access TSH (3rd IS) Assay on the DxI 9000 Access Immunoassay Analyzer. The device is an automated system, and its performance is evaluated directly without human interpretation in the analytical process.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Method Comparison: The predicate device, Access TSH (3rd IS) Assay on UniCel DxI 800 Immunoassay System (K153651), served as the reference or "ground truth" for comparative performance.
    • Imprecision, Reproducibility, Linearity, LoB, LoD, LoQ: "Ground truth" is established by the study design, using known concentrations, spiked samples, or statistical methods (e.g., CLSI guidelines) to define true values and assess the device's ability to measure them accurately and precisely.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set in the typical sense. Its performance is based on the inherent analytical characteristics of the reagents and instrumentation.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172787

    Validate with FDA (Live)

    Device Name
    Access hsTnl
    Date Cleared
    2018-06-14

    (272 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K121214

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Access hsTnl is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnl) levels in human serum and plasma using the Access 2 Immuoasay Systems to aid in the diagnosis of myocardial infarction (MI).

    Device Description

    The Access hsTnl and the Access 2 Immunoassay System compose the Access Immunoassay System for the quantitative determination of cardiac troponin I (cTnl) in human serum and plasma.

    The Access hsTnI reagent packs contain specific reagents for the in vitro diagnostic measurement of cTnl including:

    • R1a: Dynabeads* paramagnetic particles coated with mouse . monoclonal anti-human cTnl antibody suspended in TRIS buffered saline, with surfactant, bovine serum albumin (BSA). < 0.1% sodium azide, and 0.1% ProClin** 300.
    • R1b: 0.1 N NaOH .
    • R1c: TRIS buffered saline, surfactant, protein (mouse). < 0.1% sodium . azide, and 0.1% ProClin 300.
    • R1d: Sheep monoclonal anti-human cTnl alkaline phosphatase conjugate diluted in ACES buffered saline, with surfactant, BSA matrix, protein (bovine, sheep, mouse), < 0.1% sodium azide, and 0.25% ProClin 300.
    AI/ML Overview

    The provided text describes the Beckman Coulter Access hsTnI diagnostic device, an immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma to aid in the diagnosis of myocardial infarction (MI).

    Here's an analysis of the acceptance criteria and study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides a comparison of the new device (Access hsTnI) with its predicate device (Access AccuTnI+3 Reagent) and details multiple performance characteristics. The specific acceptance criteria themselves are not explicitly stated as pass/fail thresholds in a dedicated table, but rather implied through the performance data presented and comparison to the predicate. The "Summary of Studies" section details the studies conducted to establish the device's performance, which in effect serves as demonstration of meeting acceptance criteria for analytical and clinical performance.

    Implicit Acceptance Criteria and Reported Performance from the Text:

    Feature/CriterionPredicate Device (Access AccuTnl+3) PerformanceNew Device (Access hsTnl) Performance
    Intended UseAid in diagnosis of MIAid in diagnosis of MI
    Assay PrincipleChemiluminescent sandwich immunoassaySame
    Sample TypeSerum and heparinized plasmaSame
    Analytical Measuring Range0.02 ng/mL to 100 ng/mL (20 pg/mL to 100,000 pg/mL)2.0 pg/mL to 27,027 pg/mL
    Expected Results (99th Percentile URL)0.02 ng/mL with 95% CI of 0.01-0.05 ng/mLLithium Heparin Plasma: - Overall: 17.5 pg/mL (95% CI: 12.6–20.7 pg/mL) - Females: 11.6 pg/mL (95% CI: 8.4–18.3 pg/mL) - Males: 19.8 pg/mL (95% CI: 14.0–42.9 pg/mL) Serum: - Overall: 18.2 pg/mL (95% CI: 13.1–23.1 pg/mL) - Females: 11.8 pg/mL (95% CI: 8.7–18.7 pg/mL) - Males: 19.7 pg/mL (95% CI: 14.3–44.3 pg/mL)
    PrecisionTotal CV of ≤8% at >0.075 ng/mL; SD ≤0.006 at ≤0.075 ng/mL≤ 10% within-laboratory CV for concentrations ≥ 11.5 pg/mL; ≤ 1.15 pg/mL within-laboratory SD for concentrations < 11.5 pg/mL (Reported ranged from 2.6% to 7.4% for concentrations ≥ 11.5 pg/mL and 0.34 to 0.93 for concentrations < 11.5 pg/mL)
    Open Reagent Pack StabilityStable at 2 to 10°C for 56 daysStable at 2 to 10°C for 64 days
    Clinical Performance (Sensitivity, Specificity, PPV, NPV)Not detailed for predicateLithium Heparin Plasma (Overall, at Baseline): Sensitivity 86% (86/100), Specificity 90% (510/564), PPV 61% (86/140), NPV 97% (510/524) Serum (Overall, at Baseline): Sensitivity 84% (91/108), Specificity 91% (541/595), PPV 63% (91/145), NPV 97% (541/558) (Detailed results provided for different time intervals and by sex)
    High-dose Hook EffectNot detailed for predicateNo high-dose hook effect up to at least 2,000,000 pg/mL
    LinearityNot detailed for predicateLinear across the range of the assay (2.0 to approximately 27,027 pg/mL) in both serum and lithium heparin plasma.
    Dilution RecoveryNot detailed for predicateDilution recovery demonstrated across the range of the assay (2.0 to approximately 27,027 pg/mL) with 10-fold dilution yielding 100 ± 10% recovery for samples from 27,027 to 270,270 pg/mL.
    Limit of Blank (LoB)Not detailed for predicate0.8 pg/mL
    Limit of Detection (LoD)Not detailed for predicate2.0 pg/mL in both serum and lithium heparin plasma.
    Limit of Quantitation (LoQ)Not detailed for predicate4.1 pg/mL (CV ≤ 10%), 2.0 pg/mL (CV ≤ 20%).
    Analytical SpecificityNot detailed for predicateNo significant cross-reactivity found for tested compounds (defined as concentration shift > ± 2 SD; 10% for > 11.5 pg/mL, 2.30 pg/mL for ≤ 11.5 pg/mL).
    Interfering SubstancesNot detailed for predicateNo significant interference found for tested compounds at specified concentrations (defined as concentration shift > ± 2 SD; 10% for > 11.5 pg/mL, 2.30 pg/mL for ≤ 11.5 pg/mL).

    2. Sample size used for the test set and the data provenance

    • 99th percentile URL study:

      • Sample Size: 1089 subjects (595 females, 494 males for lithium heparin plasma; 595 females, 493 males for serum).
      • Provenance: Multicenter prospective study conducted at five geographically diverse locations throughout the United States.
    • Clinical performance study:

      • Sample Size: 1,851 evaluable subjects.
      • Provenance: Multicenter prospective study conducted for ED patients presenting with chest pain or equivalent ischemic symptoms suggestive of Acute Coronary Syndromes (ACS). Patients were enrolled from 14 geographically diverse, primary care hospital-associated emergency departments in the United States, reflecting regional, urban, suburban, and rural patient populations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Ground Truth Experts: An independent panel of expert physicians.
    • Qualifications of Experts: Not explicitly stated beyond "expert physicians." However, their role in adjudicating MI statuses using criteria consistent with the Universal Definition of Myocardial Infarction suggests expertise in cardiology and MI diagnosis. They were blinded to assay results and attending physicians' diagnoses.

    4. Adjudication method for the test set

    • Adjudication Method: "True MI statuses of all subjects were adjudicated by an independent panel of expert physicians using criteria consistent with the Universal Definition of Myocardial Infarction." The specific number of experts involved in the panel (e.g., 2+1, 3+1) is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This device is a laboratory immunoassay for measuring a biomarker (cardiac troponin I), not an imaging-based AI diagnostic tool requiring "human readers" or "AI assistance." Therefore, an MRMC comparative effectiveness study, in the context described, was not performed and is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is a standalone diagnostic assay (algorithm equivalent in this context), as it directly measures cTnI levels. Its performance (sensitivity, specificity, PPV, NPV) was evaluated without direct human intervention in the result generation or interpretation of the assay itself during the study. The study focuses on the diagnostic accuracy of the assay using the established 99th percentile URLs against an expert-adjudicated ground truth. However, the note explicitly states: "The Access hsTnl assay is not intended to be used in isolation; results should be interpreted in conjunction with other diagnostic tests and clinical information," indicating that clinical decision-making ultimately involves a human interpreting the result in context.

    7. The type of ground truth used

    • Ground Truth Type: Expert consensus. Specifically, "True MI statuses of all subjects were adjudicated by an independent panel of expert physicians using criteria consistent with the Universal Definition of Myocardial Infarction."

    8. The sample size for the training set

    • The document describes performance studies and reference range establishment. It does not mention a separate "training set" in the context of machine learning. The studies described are for validation and characterization of the assay and its clinical performance. The 99th percentile URL was established using 1089 healthy subjects, and the clinical performance was evaluated on 1,851 evaluable subjects. These represent the datasets used to demonstrate key performance metrics.

    9. How the ground truth for the training set was established

    • As noted above, the concept of a "training set" in the machine learning sense is not directly applicable here. For the clinical performance evaluation (which is the closest analog to a test set in ML terms for diagnostic accuracy), the ground truth for MI status was established by "an independent panel of expert physicians using criteria consistent with the Universal Definition of Myocardial Infarction."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1