Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171170

    Validate with FDA (Live)

    Date Cleared
    2017-05-25

    (34 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K121118, K132878, K141988, K153664

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M.U.S.T. Pedicle Screw System is intended for posterior non-cervical pedicle fixation (T1-S2/lium) and nonpedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis and failed previous fusion in skeletally mature patients.

    Device Description

    The M.U.S.T. Extension - Straight Connectors and Additional Screws are intended to be used as part of the M.U.S.T. Pedicle Screws System (K121118, K132878, K141988, K153664) for the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. Pedicle Screws System includes: cannulated or non-cannulated poly-axial pedicle screws (K121115and K132878), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K121115), straight and pre-bent rods (K121115), cross connectors (K132878), enhanced screws and rods designed for percutaneous surgery (K141988), and cannulated or non-cannulated reduction screws (K153664). The M.U.S.T. Additional Pedicle Screws (standard and enhanced pedicle) are made of Ti6Al4V ELI (ISO 5832-3, ASTM F136-13) and CoCrMo (ISO 5832-12, ASTM F1537-11), while the M.U.S.T. Additional Reduction Screws are made of Ti6A14V ELI (ISO 5832-3, ASTM F136-13). The screw shaft is color anodized to simplify the identification of the screw diameter. The pedicle screw has a dual lead thread to simplify the screw insertion and reduce the number of turns. The threads are designed with a cylindrical diameter. The M.U.S.T. Straight Connectors and Additional Screws introduces the following additional components: Pedicle screw, Enhanced Solid Pedicle screw, Enhanced Cannulated Pedicle screw, Reduction Pedicle screw, Reduction Cannulated Pedicle screw, Straight Cross Connector.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the M.U.S.T. Pedicle Screw System Extension - Straight Connectors and Additional Screws. It focuses on demonstrating substantial equivalence to previously cleared predicate devices through comparisons of technological characteristics and performance testing.

    Here's an analysis of the acceptance criteria and study information provided, structured as requested:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly list specific numerical acceptance criteria. Instead, it states that the acceptance criteria were "pre-defined" and "based on the standards, FDA guidance, and comparison to the predicate device system." The reported device performance is a general statement that the device "met all acceptance criteria" and "is substantially equivalent to the predicate devices."

    Acceptance Criteria CategorySpecific Acceptance Criteria (as implied)Reported Device Performance
    Mechanical Performance (Dynamic Compression Bending)Performance equivalent to predicate devices as determined by ASTM F1717-12, considering worst-case device.Met all acceptance criteria. Performance substantially equivalent to predicate devices.
    Mechanical Performance (Static Compression Bending)Performance equivalent to predicate devices as determined by ASTM F1717-12, considering worst-case device.Met all acceptance criteria. Performance substantially equivalent to predicate devices.
    Mechanical Performance (Static Torsion)Performance equivalent to predicate devices as determined by ASTM F1717-12, considering worst-case device.Met all acceptance criteria. Performance substantially equivalent to predicate devices.
    BiocompatibilityMaterial and manufacturing process identical to predicate devices, supporting biological safety.Deemed unnecessary to conduct additional testing as materials and manufacturing are identical to predicate devices.
    Technological CharacteristicsNew diameters and straight connector do not raise new questions of safety or effectiveness compared to predicates.A comparison evaluation shows no new risks associated with the subject device design.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document repeatedly mentions "the worst-case device" for each of the mechanical tests (Dynamic Compression Bending, Static Compression Bending, Static Torsion). This implies a very small sample size, potentially a single representative device or a small set of devices configured to represent the most challenging conditions. The exact number is not explicitly stated.
    • Data Provenance: The study is a laboratory-based mechanical testing study, not a clinical study involving human or animal data. Therefore, concepts like country of origin, retrospective, or prospective do not apply in the typical sense. It is an engineering verification study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable to this type of device and study. The "ground truth" for mechanical testing is established by engineering standards (ASTM F1717-12) and the inherent physical properties and design specifications of the device, rather than expert judgment or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable. Adjudication methods are typically used in clinical studies to resolve discrepancies in expert interpretations of data (e.g., imaging, clinical assessments). For mechanical testing, the "adjudication" is inherent in the adherence to established test standards and the objective measurement of physical properties.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not applicable to this device. This type of study evaluates the performance of AI-assisted diagnostic or classification systems in a clinical context, often involving human readers interpreting medical images. The M.U.S.T. Pedicle Screw System Extension is a physical orthopedic implant, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation (in the context of AI algorithms) is not applicable to this device. This device is a physical implant and does not involve any AI algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance data in this document is based on engineering standards and specifications. Specifically, the mechanical tests were performed according to ASTM F1717-12 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model. The acceptance criteria were based on these standards, FDA guidance, and comparison to the predicate device's established performance.

    8. The sample size for the training set

    This question is not applicable as there is no "training set" in the context of a physical device mechanical performance study. Training sets are relevant for machine learning algorithms.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1