510(k) Data Aggregation

    K Number
    K190779

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-06-25

    (90 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K120961

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
    2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
    3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50 cc for age < 80 years, 0-20 cc for age ≥ 80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
    Device Description

    The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever has a hydrophilic coating to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A torque device is provided with the Retriever to facilitate manipulation and retrieval. An insertion tool is provided to introduce the Retriever into a Microcatheter. The Insertion Tool is a sheath in which the Retriever comes preloaded. Once half the retriever's length is inserted into the microcatheter, the insertion tool is removed. Retrievers have a modified proximal end that permits attachment of the Abbott Vascular DOC Guide Wire Extension (REF 22260). Joining the Guide Wire Extension to the Retriever facilitates removal or exchange of a catheter while maintaining the Retriever position in anatomy. After the exchange has been completed, the extension can be detached.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Stryker Neurovascular Trevo XP ProVue Retriever. This document primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed clinical study with acceptance criteria for an AI device. The information requested in your prompt (e.g., acceptance criteria for device performance, sample sizes for test/training sets, expert qualifications, ground truth establishment methods, MRMC studies) is typically found in regulatory submissions for AI/ML-based medical devices, which assess algorithm performance.

    The provided document does not contain the specific type of information needed to fully answer your request regarding acceptance criteria and performance data for an AI-powered device.

    However, I can extract the information that is present concerning the Trevo Retriever, which is a mechanical thrombectomy device, and interpret it in the context of device performance, even if it's not an AI device.

    Let's break down what is available and what is not:

    What is Available (relevant to a physical medical device, not AI/ML):

    • Device Type: Neurovascular Mechanical Thrombectomy Device (a physical medical device, not an AI/ML algorithm).
    • Purpose: To restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke.
    • Performance Data - Non-Clinical:
      • Test: Simulated Use
      • Test Method Summary: "Simulated Use testing utilized a neurovascular model with a re-circulating water bath at 37℃ to simulate the human arterial circulation and to assess the device's ability to retrieve the clot and achieve recanalization."
      • Conclusion/Performance: "Simulated Use met acceptance criteria."
    • Performance Data - Clinical:
      • Data Source: Analysis of Real World Data (RWD) from the Trevo Retriever Registry.
      • Purpose: To assess safety or effectiveness on the combination use of Trevo Retriever and intermediate catheter aspiration approach.
      • Additional Support: Comprehensive review of relevant clinical literature.

    What is NOT available (and would be expected for an AI/ML device as per your prompt):

    • A table of acceptance criteria and reported device performance for an AI/ML algorithm. The document only states that "Simulated Use met acceptance criteria" without specifying the precise quantitative criteria or results.
    • Sample sizes used for the test set and data provenance (e.g., country of origin, retrospective/prospective). The "Simulated Use" refers to a bench test, not a patient test set, and the "Trevo Retriever Registry" is mentioned for clinical data, but no specifics about its use as a "test set" for performance evaluation against acceptance criteria are given.
    • Number of experts used to establish ground truth for the test set and their qualifications. This is relevant for AI/ML evaluation.
    • Adjudication method (e.g., 2+1, 3+1, none) for the test set.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size. This is specific to AI clinical utility.
    • If a standalone (algorithm only without human-in-the-loop performance) was done.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    Summary in relation to your prompt, given the provided document:

    Since the provided text describes a physical medical device (a thrombectomy device) and not an AI/ML algorithm, most of the requested points related to AI/ML software validation (like training set, ground truth experts, MRMC studies) are not applicable and thus not present in the document.

    The "acceptance criteria" mentioned in the non-clinical performance section only states that the "Simulated Use met acceptance criteria," but it does not specify what those criteria were (e.g., success rate, time to recanalization, etc.) nor does it provide the exact performance metrics achieved.

    Therefore, I cannot construct the table or provide the detailed answers to most of your questions based on the provided text, as it does not pertain to an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173352

    Validate with FDA (Live)

    Date Cleared
    2018-02-15

    (113 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K120961

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.

    2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (0-50 cc for age < 80 years, 0-20 cc for age ≥ 80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.

    Device Description

    The Trevo Retriever family, including the Trevo ProVue and Trevo XP ProVue Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever has a hydrophilic coating to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A torque device is provided with the Retriever to facilitate manipulation. The torque device is used to lock the core wire to the microcatheter during the procedure. Locking of the torque device to the wire allows the microcatheter and Retriever to be retracted as a system during clot retrieval. An insertion tool is provided to introduce the Retriever into a Microcatheter. The Insertion Tool is a sheath in which the Retriever comes preloaded. Once half the retriever's length is inserted into the microcatheter, the insertion tool is removed. Retrievers have a modified proximal end that permits attachment of the Abbott Vascular DOC Guide Wire Extension (REF 22260). Joining Guide Wire Extension to Retriever facilitates removal or exchange of a catheter while maintaining Retriever position in anatomy. After exchange has been completed, the extension can be detached.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Trevo Retriever, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a bulleted or numbered list with specific thresholds. Instead, the study aims to demonstrate superior clinical outcomes for the treatment arm compared to the control arm across several endpoints. The success thresholds are defined within the study design.

    Performance Metric (Acceptance Criteria)Reported Device Performance (Treatment Arm vs. Control Arm)
    Primary Effectiveness:
    90-day Utility Weighted Modified Rankin Scale (UW-mRS): Posterior probability the Trevo thrombectomy increases expected scores relative to medical management alone, with a threshold for success of at least 0.986.Mean UW-mRS: 5.5 (Treatment) vs. 3.4 (Control) Posterior mean treatment benefit, core-adjusted: 2.0 (95% credible interval 1.1 to 3.0) Probability of superiority: > 0.9999
    90-day Dichotomized mRS (0-2 vs. 3-6) (Functional Independence): Success based on the same model and hierarchical fashion as UW-mRS.Functional Independence (mRS 0-2): 48.6% (Treatment) vs. 13.1% (Control) Posterior treatment benefit, core-adjusted: 32.8% (95% credible interval 21.1% to 44.1%) Probability of superiority: > 0.9999
    Primary Safety:
    Incidence of stroke-related mortality at 90 days: No significant difference between study arms.Treatment Arm: 15.9% (17/107) vs. Control Arm: 18.2% (18/99) Difference: -2.3% [-12.6%, 8.0%] p-value = 0.7126 (No significant difference)
    Secondary Effectiveness:
    Early Neurological Response (Day 5-7/Discharge): Defined as NIHSS drop of ≥ 10 points or NIHSS score of 0 or 1.Treatment Arm: 47.7% vs. Control Arm: 19.2% Mean Absolute Difference: 28.5% (16.2%, 40.7%) Risk Ratio: 2.5 (1.6, 3.9) P-Value: <0.001 (Superiority for Treatment Arm)
    Revascularization Rates at 24 Hours: Presence of partial or complete recanalization.Treatment Arm: 76.6% (82/107) vs. Control Arm: 38.4% (38/99) Difference: 40.2% [27.1%, 51.5%] p-value: < 0.0001 (Superiority for Treatment Arm)
    Secondary Safety:
    Incidence of sICH (by ECASS III definition) within 24 hours post-randomization: No statistical difference between arms.Treatment Arm: 5.6% (6/107) vs. Control Arm: 3.0% (3/99) Difference: 2.6% [-2.9%, 8.1%] p-value = 0.5011 (No statistical difference)
    Incidence of Neurological Deterioration (≥ 4-point increase in NIHSS from baseline through Day 5-7/Discharge):Treatment Arm: 14.0% (15/107) vs. Control Arm: 26.3% (26/99) Difference: -12.2% [-23.1%, -1.4%] p-value = 0.0358 (Less frequently in Treatment Arm)
    All-Cause Mortality at 90 days: No significant difference between study arms.Treatment Arm: 18.7% (20/107) vs. Control Arm: 18.2% (18/99) Difference: 0.5% [-10.1%, 11.1%] p-value = 1.0000 (No statistical difference)
    Serious Adverse Events (Procedure-related and Device-related): No new safety risks identified, comparable across treatment groups.Vascular perforation: 0.9% (Treatment) Intramural arterial dissection: 1.9% (Treatment) Embolization to a new territory: 3.7% (Treatment) Full list in Table 11 & 12 No new safety risks identified, comparable risk profile to predicate.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: 206 subjects (107 in the Treatment Arm, 99 in the Control Arm).
    • Data Provenance: The DAWN™ study (IDE G130223) was a "multi-center, ... conducted in multiple countries." The document doesn't specify the exact countries, but it indicates an international scope. The study was prospective, randomized, open, masked endpoint trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document mentions "masked endpoint trial" and "CEC adjudication" (Clinical Events Committee). While it confirms the involvement of a committee for adjudication, it does not specify the exact number of experts or their specific qualifications (e.g., number of years of experience, subspecialty) for establishing the ground truth.

    4. Adjudication Method for the Test Set:

    The exact adjudication method (e.g., 2+1, 3+1) is not explicitly stated. However, it refers to "CEC adjudication," indicating that a Clinical Events Committee independently reviewed and determined certain outcomes (e.g., sICH, neurological deterioration, all-cause mortality, serious adverse events). This typically implies a consensus-based approach among committee members.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed as described in the document. The study was a randomized controlled trial comparing the Trevo thrombectomy device plus medical management to medical management alone. This is a direct patient outcome study, not a study of human readers' performance with or without AI assistance. The Trevo Retriever is a medical device for mechanical thrombectomy, not an AI diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This question is not applicable as the Trevo Retriever is a physical medical device (neurovascular mechanical thrombectomy device), not a standalone algorithm or AI. Its performance is directly measured in terms of patient outcomes and procedural metrics.

    7. Type of Ground Truth Used:

    The ground truth used in the DAWN study was based on patient outcomes data, specifically:

    • Clinical Outcomes: Modified Rankin Scale (mRS) at 90 days (utility-weighted and dichotomized).
    • Neurological Assessment: NIHSS scores (for early neurological response and deterioration).
    • Imaging Data: Recanalization rates (CTA or MRA at 24 hours), core infarct volume (for inclusion criteria and statistical adjustment).
    • Safety Endpoints: Stroke-related mortality, sICH (ECASS III definition), all-cause mortality, and adverse events (adjudicated by CEC).

    8. Sample Size for the Training Set:

    The document describes a clinical trial (DAWN study) to evaluate the device in general, not specifically an AI model with a separate training and test set. Therefore, there is no "training set" for an AI algorithm mentioned in the context of this device's premarket notification. The entire DAWN study cohort (206 subjects) served as the dataset for evaluating the device's effectiveness and safety.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no "training set" for an AI algorithm, this question is not applicable. The clinical trial data was gathered through the rigorous protocol of the DAWN study, including masked endpoint assessment and CEC adjudication for relevant clinical and safety outcomes, which served as the "ground truth" for evaluating the device's performance in comparison to standard medical management.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K150616, K143077, K133464, K132641, K122478, K120961

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trevo Retrievers are indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.

    Device Description

    The Trevo ProVue and XP ProVue Retrievers consist of a flexible, tapered core wire with a shaped self-expanding section at the distal end for clot capture and removal. Radiopaque platinum wires in the shaped section and radiopaque markers on the distal end allow fluoroscopic visualization. The Trevo Retrievers have a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retrievers. The Trevo Retrievers are delivered to the site of occlusion in the neurovasculature through a microcatheter. The torque device may be used to lock the core wire of the Trevo Retriever to the microcatheter during the procedure, allowing the Trevo Retriever and microcatheter to be retracted as a system through the guide catheter and removed from the body with captured clot.

    AI/ML Overview

    The acceptance criteria for the Trevo ProVue and XP ProVue Retrievers, a neurovascular mechanical thrombectomy device, are derived from the clinical study, bench testing, animal studies, and various regulatory requirements. The study supporting these criteria is primarily the MR CLEAN (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands) trial, with additional non-clinical and animal studies.

    Here's a breakdown of the acceptance criteria and reported device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state "acceptance criteria" for each performance metric with specific thresholds (e.g., "must achieve X%"). Instead, the assessment is based on demonstrating safety and effectiveness as compared to medical management alone, or meeting benchmarks set by non-clinical testing. Where specific targets are not given, the trial's statistically significant improvement or demonstration of comparable safety/effectiveness serves as the "meeting" of the criteria.

    CategoryAcceptance Criteria (Implied / Explicit)Reported Device Performance
    EffectivenessPrimary Effectiveness Endpoint: Clinically meaningful functional independence (mRS ≤ 2) at 90 days, demonstrating superiority over Medical Management (MM) alone.29.2% (28/96) of patients in the Trevo Retriever group achieved mRS ≤ 2 at 90 days, compared to 19.3% (48/249) in the MM group. Odds Ratio [95% CI]: 1.88 [1.07, 3.29], p-value = 0.014. (Demonstrated statistically significant superiority).
    Secondary Effectiveness Endpoint 1: Percentage of patients with no intra-cranial occlusion after 24 hours.77.5% (62/80) in the Trevo group achieved no intra-cranial occlusion, compared to 33.51% (65/194) in the MM group. (Demonstrated significantly higher rates of recanalization).
    Secondary Effectiveness Endpoint 2: Percentage of Trevo patients with recanalization (TICI ≥ 2a).81.3% (78/96) of Trevo patients achieved TICI ≥ 2a.
    Secondary Effectiveness Endpoint 3: Neurological outcomes assessed by NIHSS at 24 hours and 5-7 days post-randomization, demonstrating improvement compared to MM alone.NIHSS (24 Hours): Trevo Mean ± SD: 14.23 ± 9.23 (N=96) vs. MM Mean ± SD: 16.19 ± 7.77 (N=240). NIHSS (5-7 Days): Trevo Mean ± SD: 13.02 ± 12.16 (N=94) vs. MM Mean ± SD: 15.76 ± 11.49 (N=237). (Directionally favorable for Trevo, though statistical significance for these endpoints isn't explicitly stated).
    SafetyPrimary Safety Endpoint: All-cause mortality rate at 90 days, demonstrating non-inferiority or reduced mortality compared to MM alone.Trevo FDA Cohort: 13.3% (16/120) mortality. MM: 22.89% (57/249) mortality. (Lower mortality rate in Trevo group).
    Secondary Safety Endpoint 1: Proportion of patients with symptomatic intracranial hemorrhage (sICH) within 24 hours.Trevo FDA Cohort: 6.67% (8/120) sICH. MM: 4.42% (11/249) sICH. (Slightly higher incidence in Trevo group, but likely considered acceptable within the benefit-risk profile).
    Secondary Safety Endpoint 2: Proportion of patients with neurological deterioration within 5-7 days or discharge.Trevo FDA Cohort: 9.17% (10/109) . MM: 9.57% (22/230). (Comparable rates).
    Other Adverse Events: Monitor rates of various SAEs and adverse events to ensure an acceptable safety profile.Comprehensive reporting of SAEs (Table 10) and MedDRA coded AEs (Table 11) indicates generally comparable or acceptable rates across various categories, with some expected differences due to the invasive nature of the procedure (e.g., higher haemorrhage intracranial in Trevo, but also higher "other infection" and "other complication"). The overall benefit-risk determination concluded that benefits outweigh risks.
    BiocompatibilityDevice materials demonstrated to be biocompatible (ISO 10993-1:2009/AC:2010 compliance).Leveraged from prior submissions (K120961). Tests included hemocompatibility/coagulation, hemolysis, cytotoxicity, maximization test for delayed hypersensitivity, intracutaneous reactivity, acute systemic toxicity, and material-mediated rabbit pyrogenicity. (Met standard).
    SterilitySterility assurance level (SAL) of 10^-6 (ISO 11135-1:2007). Sterilant residuals per ISO 10993-7:2008. Endotoxin testing <2.15 EU/device (USP <161>).Achieved SAL of 10^-6. Sterilant residuals tested. Endotoxin testing met criterion. (Met standards).
    Shelf-Life24-month shelf-life supported by accelerated and real-time shelf-life testing, including packaging integrity and component functional performance.Supported by accelerated testing (Trevo XP ProVue) and real-time testing (Trevo ProVue), including: dimensional verification, tensile strength, radial force, tip flexibility, torque/tensile durability, retriever platinum wire and joint durability, radiopacity, deliverability, simulated use, coating integrity, and particulate testing. Packaging integrity tested to ASTM D4169, F1980, F1929, F2906, F88/M. (Met standard).
    Performance (Bench)Device performs as intended under anticipated conditions of use (Table 2 tests: Dimensional Verification, Tensile Strength, Radial Force, Tip Flexibility, Torque/Tensile Durability, Platinum Wire Joint Strength/Durability, Radiopacity, Deliverability, Particulate Evaluation, Coating Integrity Evaluation, Simulated Use).All listed bench tests were conducted and passed: Verified dimensions; recorded peak tensile force; constrained and released shaped section to record radial force; flexed distal tip to record compression/flex force; applied rotations and tensile cycles for durability; recorded peak tensile force for platinum wires; performed wrap/unwrapped and deploy/reload cycles; assessed radiopacity visually; measured force to push device through tortuous model; measured particulates; performed visual assessment of coating integrity before/after simulated use; simulated use in silicone neurovascular model with thrombus. (All tests passed, indicating performance as intended).
    Performance (Animal)Safety (vessel response) demonstrated in animal studies, including no arterial dissection, perforation, or thrombosis.Acute and chronic swine studies (up to 6 passes). Acute study: no evidence of vessel dissection, perforation, or thrombosis at Day 0. Chronic study: no evidence of dissection, perforation, or thrombosis at Day 0, and no stenosis, vessel irregularity, intimal flap or pseudoaneurysm at Day 30. Histopathology showed arterial healing consistent with routine catheterization. (Demonstrated safety in animal models).

    2. Sample Sizes Used for the Test Set and Data Provenance

    The primary clinical study supporting the device's expanded indication for the Trevo Retrievers was the MR CLEAN trial.

    • Test Set Sample Size:

      • TREVO FDA Cohort: 120 patients (from the IAT arm, met specific criteria) (N=120 for primary safety; N=96 for primary effectiveness after exclusions).
      • Medical Management (MM) Control Group: 249 patients (after exclusions for IV t-PA timing).
      • Total Patients in MR CLEAN (originally randomized): 500 patients (233 IAT, 267 control).
    • Data Provenance:

      • Country of Origin: The Netherlands (MR CLEAN was conducted in every endovascular hospital center in the Netherlands).
      • Retrospective or Prospective: Prospective, randomized, open-label, controlled, multicenter trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document mentions expert review for several endpoints:

    • Recanalization and Occlusion Assessment (Table 4 & 5):
      • Number of Experts: "Consensus review by up to three readers" for occlusion assessment (Table 4). "Consensus review by at least three readers" for TICI score recanalization (Table 5).
      • Qualifications: The document does not explicitly state the specific qualifications (e.g., years of experience, subspecialty) of these readers/experts. However, given the nature of the study (neurovascular stroke treatment), it is highly probable they were experienced neurologists, neuroradiologists, or stroke specialists involved in the study's clinical centers.

    4. Adjudication Method for the Test Set

    • Adjudication Method: "Consensus review by up to three readers" or "at least three readers" was used for evaluating arterial occlusion and recanalization (TICI scores). This implies a method where multiple experts independently (or collectively) assess the imaging and arrive at a joint decision. The specific process (e.g., if there were disagreements, how they were resolved beyond just "consensus") is not detailed beyond "consensus review."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement

    • No, a typical MRMC comparative effectiveness study, as often seen for diagnostic AI, was not performed in this context.
      • The MR CLEAN trial directly compared patient outcomes with and without the device (with medical management), rather than evaluating how human readers' diagnostic accuracy or treatment decisions improved with AI assistance.
      • The Trevo Retrievers are a mechanical device for thrombectomy, not a diagnostic AI algorithm intended to assist human readers in interpreting images. Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" does not directly apply to this device's clinical evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    • Yes, in essence, the "device" (Trevo Retrievers) was evaluated in a standalone manner as a treatment in the context of the clinical trial.
      • The MR CLEAN trial evaluated the performance of the neurovascular mechanical thrombectomy device (Trevo Retrievers) in conjunction with best medical management against best medical management alone. This is effectively a "standalone" evaluation of the device's therapeutic effect, rather than an "algorithm only" evaluation, as it's a physical medical device.
      • The device's performance is measured by its direct impact on patient outcomes when used by clinicians, not as an interpretive AI tool.

    7. The Type of Ground Truth Used

    • Clinical Outcomes/Expert Assessment (Blinded):
      • Primary Effectiveness (mRS): The 90-day mRS assessment was performed by an assessor who was blinded to the subject's treatment allocation. This is a high-quality method for establishing ground truth for functional independence.
      • Imaging-based Endpoints (Occlusion/Recanalization): Assessed by a consensus review by up to three/at least three readers (presumably domain experts) after 24 hours using CTA or MRA and the AOL/TICI scales.
      • Safety Endpoints: Based on collected adverse events, neurological assessments (NIHSS), and mortality data, which are objective clinical measures or physician-reported events.
      • Animal Studies: Safety ground truth established by intra-procedural angiography and histopathology assessments performed by experts.

    8. The Sample Size for the Training Set

    • Not Applicable in the traditional sense for an AI/algorithm: The information provided describes the regulatory approval of a physical medical device (mechanical thrombectomy retriever), not an AI algorithm. Therefore, there is no "training set" for an AI model.
    • The clinical data from the MR CLEAN trial serves as the pivotal clinical evidence supporting the device's effectiveness and safety, analogous to a validation set for an AI, but it's not a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As explained above, this is a physical medical device, not an AI algorithm, so there is no "training set" or ground truth establishment process for an AI model. The "ground truth" for the device's performance is established through rigorous clinical trials and non-clinical testing as described in the previous sections.