Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140178

    Validate with FDA (Live)

    Date Cleared
    2014-02-27

    (34 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K101997, K113695, K113219, K131677, K133289

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ellipse MAGEC Spinal Bracing and Distraction System is intended for skeletally immature patients less than 10 years of age with severe progressive spinal deformities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) associated with or at risk of Thoracic Insufficiency Syndrome (TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.

    Device Description

    The Ellipse Technologies, Inc. MAGEC Spinal Bracing and Distraction System is comprised of a sterile single use spinal rod that can be surgically implanted using appropriate Stryker® Xia fixation components (i.e. Pedicle screws, hooks and/or connectors). The system includes a non-sterile hand held External Remote Controller (ERC) that is used at various times after implant to non-invasively lengthen the implanted spinal rod. The implanted spinal rod is used to brace the spine during growth to minimize the progression of scoliosis. The titanium rod (Ti-6Al-4V ASTM F136) includes an actuator portion that holds a small internal magnet. The magnet in the actuator can be turned non-invasively by use of the ERC. Rotation of the magnet causes the MAGEC rod to be lengthened or shorten.

    The hand held non-invasive ERC is electrically powered. The ERC is placed over the patient's spine and then manually activated, which causes the implanted magnet to rotate and either lengthen or shorten the rod. Periodic lengthening of the rod is performed to distract the spine and to provide adequate bracing during growth to minimize the progression of scoliosis. Once the physician determines that the implant has achieved its intended use and is no longer required, the implant is explanted. Additional accessories for the MAGEC System include the MAGEC Manual Distractor and the MAGEC Wand Magnet Locator. The MAGEC Manual Distractor is a sterilizable, single use device, which is used in the operating room to test the device prior to implantation. The MAGEC Wand Magnet Locator is a non-sterile device which is used during the distraction procedure to locate the magnet within the MAGEC rod. The ERC is placed over this location on the child's back.

    AI/ML Overview

    The provided text describes the MAGEC® Spinal Bracing and Distraction System and its premarket notification (K140178) to the FDA. The submission focuses on establishing substantial equivalence to a predicate device, the Harrington Rod System, rather than providing detailed acceptance criteria and a study proving the device meets those criteria in the context of a diagnostic AI product.

    Therefore, much of the requested information (acceptance criteria, specific performance metrics, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance with specific metrics like sensitivity/specificity, and detailed ground truth establishment for training) cannot be extracted from the provided document as it pertains to a mechanical medical device rather than a diagnostic AI.

    However, I can extract information related to the device's performance evaluation.

    Here's the information that can be extracted and a clear statement about what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of numerically defined acceptance criteria for a diagnostic AI product. Instead, it describes performance in terms of mechanical and clinical outcomes for a spinal bracing system.

    CategoryAcceptance Criteria (Not explicitly defined as numerical targets for a diagnostic device)Reported Device Performance
    Non-Clinical TestingEquivalent mechanical performance to predicate devices (Harrington Rod System). Compliance with relevant mechanical, sterilization, biocompatibility, electrical safety, EMC/EMI, software, and magnetic field safety standards.- Static and Dynamic Mechanical Testing (ASTM F1717, ASTM F2627): "Results of these tests demonstrate the MAGEC Spinal Bracing and Distraction System can be expected to perform in a manner substantially equivalent to the predicate devices."- Shelf Life Packaging Validation (ISO 11607-1): Performed.- Sterilization (ANSI/AAMI/ISO 11137-2): "Verification that the gamma radiation sterilization process provides a sterility assurance level of 10-6."- Biocompatibility (ISO 10993-1): Performed.- Device functionality and verification: Performed.- Electrical Safety (IEC 60601-1): Performed.- Electrical Interference and Compatibility (EMC/EMI) (EN 60601-1-2): Performed.- Magnetic Field Safety (ICNIRP 2009): Performed.- Software (FDA Guidance (May 11, 2005)): Performed.
    In Vivo (Animal) StudySafe and efficient non-invasive distraction of the spine in an animal model.- Porcine Model (9 male Yucatan pigs): "Results of the in vivo porcine study demonstrates that the MAGEC System is safe and provides an efficient means of non-invasive distraction of the spine. No complications from distraction occurred." Attempted and actual distraction were recorded.
    Clinical PerformanceSafety and probable benefit, including deformity correction, continued growth, and reduced need for subsequent surgical procedures, similar to traditional growing rods.- Retrospective clinical study (outside US): "The results of the clinical study showed the MAGEC System provides the benefits of spinal deformity correction and continued growth, similar to that for traditional growing rods, without the need for regular surgical lengthening procedures in these children."- Specific endpoints assessed: Cobb angle correction, thoracic spine height increase, improvement in space available for lung (SAL), coronal and sagittal balance, reduction in number of subsequent surgical procedures, and weight gain.- Demonstrated ability for non-invasive adjustment to lengthen the implanted rod, allowing continued spinal growth and increased Thoracic Spine Height.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size:
      • In Vivo (Animal) Study: 9 male Yucatan pigs (randomly assigned to two groups: 1 (MAGEC rod) and 2 (sham)).
      • Clinical Study: Not explicitly stated, but described as a "retrospective clinical study for children who had either a primary or revision spinal bracing procedure using the MAGEC System."
    • Data Provenance:
      • In Vivo Study: Porcine model (animal study).
      • Clinical Study: "evaluated outside the United States" (retrospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The studies performed were primarily in vivo animal studies and a retrospective clinical study focusing on device performance and clinical outcomes, not on establishing ground truth for a diagnostic AI.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided as it is not relevant to the type of device and studies described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided as it is not relevant to the type of device (spinal bracing system) and evaluation presented. It's a study design typically used for diagnostic imaging AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided as it is not relevant to the type of device described, which is a physical, implantable medical device, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Non-Clinical and In Vivo Studies: Ground truth was based on physical measurements, functional specifications, and observation of physiological responses (e.g., distraction achieved, lack of complications) in controlled experimental settings.
    • Clinical Performance Data: Ground truth or "probable benefit" was assessed based on clinical outcomes data such as Cobb angle correction, thoracic spine height increase, improvement in space available for lung (SAL), coronal and sagittal balance, reduction in subsequent surgical procedures, and weight gain.

    8. The sample size for the training set

    This information is not provided as there is no mention of a "training set" in the context of an AI/algorithm-based device. The device is mechanical with non-invasive magnetic actuation.

    9. How the ground truth for the training set was established

    This information is not provided as there's no training set for an AI in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1