Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113501

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-12-05

    (10 days)

    Product Code
    Regulation Number
    862.3250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K113046

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CHECKCUP™ Multi-Drug Test is a one-step immunoassay for the qualitative detection of multiple drugs and drug metabolites in human urine at the following cutoff concentrations:

    TestCalibratorCut-off (ng/ml)
    AMPd-Amphetamine1000
    BARSecobarbital300
    BUPBuprenorphine10
    BZOOxazepam300
    COC 150Benzoylecgonine150
    COCBenzoylecgonine300
    MDMA3,4-methylenedioxymethamphetamine500
    MET 500d-Methamphetamine500
    METd-Methamphetamine1000
    MTDdl-Methadone300
    OPI 300Morphine300
    OPIMorphine2000
    OXYOxycodone100
    PCPPhencyclidine25
    PPXPropoxyphene300
    TCANortriptyline1000
    THC11-nor-A9-THC-9-COOH50

    The CHECKCUP™ Multi-Drug Test consists of two test formats: flat cup and round cup. The configurations of the CHECKCUP™ Multi-Drug Test consist of any combination of the drugs listed above. The CHECKCUP™ is used to obtain a visual, qualitative result and is intended for professional use only.

    This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmation methods.

    Device Description

    The Ameditech CHECKCUP™ Multi-Drug Test is a lateral flow immunoassay intended for professional (prescription) use only. The device is used for the qualitative determination of parent compound and/or major metabolites of drugs in human urine specimens, and results are read visually. The device will be made available in two formats that use identical test strips: Round Cup and Flat Cup. The modified device contains up to seventeen (17) assays that include fourteen (14) different drugs at various cutoff concentrations. The assays detect different illicit drugs and medications that are commonly abused. Once the urine sample is administered, negative results can be read as soon as 1 minute and positive results should be read between 5 to 10 minutes. The presence of the line on the test region indicates a negative result for the drug and the absence of a line on the test region indicates a positive result for the drug. A visible line is also present at the control region of the test strip. This line should always appear, regardless of the presence of drugs or metabolites in the urine sample. This means that a negative urine sample will produce both a test line and control line, and a positive urine sample will generate only a control line. The presence of control line serves as a built-in control, which demonstrates that the test is performed properly.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Ameditech CHECKCUP™ Multi-Drug Test, focusing on acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in terms of sensitivity, specificity, or accuracy percentages for the CHECKCUP™ Multi-Drug Test. Instead, it relies on demonstrating equivalence to predicate devices. The implicit acceptance criterion is that the modified device performs "as well or better than the predicate device" across various analytical performance characteristics.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    PrecisionEquivalent to predicate devices (K053175, K113046)"Confirms that the Ameditech CHECKCUP™ has equivalent performance compared with the predicate devices in terms of precision (reproducibility)."
    Analytical SensitivityEquivalent to predicate devices (K053175, K113046)"Confirms that the Ameditech CHECKCUP™ has equivalent performance compared with the predicate devices in terms of... analytical sensitivity."
    Analytical SpecificityEquivalent to predicate devices (K053175, K113046)"Confirms that the Ameditech CHECKCUP™ has equivalent performance compared with the predicate devices in terms of... analytical specificity."
    Interfering SubstancesEquivalent to predicate devices (K053175, K113046)"Confirms that the Ameditech CHECKCUP™ has equivalent performance compared with the predicate devices in terms of... interfering substances."

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states that no clinical studies were conducted for this special 510(k) submission (Section E). The performance characteristics were validated in the predicate submissions (K053175 and K113046) and are considered applicable to the modified device.

    The study described is a "simplified performance studies" and "analytical performance verification and validation testing." Details on sample sizes for these analytical studies are not provided in the given text.

    The data provenance is implied to be from Ameditech, Inc. (the submitter) and likely from internal laboratory testing, given the analytical nature of the studies discussed. There is no mention of country of origin of the data or whether it was retrospective or prospective, though analytical testing is typically prospective in nature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as no clinical studies were performed. The "ground truth" for the analytical studies would have been established by laboratory methods, not by expert consensus on patient cases.

    4. Adjudication Method for the Test Set

    This information is not applicable as no clinical studies were performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. The device is a qualitative lateral flow immunoassay intended for visual reading by a professional, not an AI system. No MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a lateral flow immunoassay that requires visual interpretation by a human professional. Therefore, the concept of "standalone (algorithm only)" performance is not applicable in the context of AI. The performance validated here is inherent to the chemical reactions and visual indicators on the test strip, which are read by a human.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    For the analytical performance studies (sensitivity, specificity, precision, interfering substances), the ground truth would have been established by reference laboratory methods (e.g., Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS) are mentioned as preferred confirmation methods for positive results, implying these or similar techniques would be used for analytical validation). The document explicitly states, "In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmation methods."

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. The device is a chemical immunoassay kit. Therefore, this concept is not applicable in the traditional sense. The development of the test strips and their performance characteristics would have been optimized through iterative laboratory testing and formulation, but not a "training set" as understood in AI/ML.

    9. How the Ground Truth for the Training Set Was Established

    As no "training set" in the AI/ML sense is mentioned or implied, this question is not applicable. The underlying principles of an immunoassay rely on antibody-antigen reactions, and the "ground truth" for developing such a device would be based on known concentrations of substances and established analytical reference methods.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1