510(k) Data Aggregation

    K Number
    K172754

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-02-09

    (150 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102330, K112760, K112813, K120632, K131120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO® maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.

    Each Cycle can sterilize non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle and Fast Non Lumen Cycle.

    The Non Lumen Cycle can sterilize: †

    Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes.

    t: The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg).

    The Fast Non Lumen Cycle can sterilize: *

    Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes.

    • The validation studies were conducted using a validation load consisting of one pouched instrument tray for a total weight of 11 lbs (5 kg).

    The Flexible Cycle can sterilize:

    Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations:

    1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. * The flexible endoscopes may contain either:
    • · A single lumen that is ≥ 1 mm internal diameter (ID) and ≤ 1050 mm in length
    • · Or two lumens with:
      • One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
      • And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
    • The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
    1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. ++ The flexible endoscope may contain either:

    • A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length

    • Or two lumens with:

    • One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
    • And the other lumen is ≥ 1 mm ID and ≤ 850 mm in length.

    ** The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).

    The Lumen Cycle can sterilize: †

    Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:

    • Single channeled devices with a stainless lumen that is > 0.77 mm ID and ≤ 500 mm in length
    • Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
    • · Triple channeled devices with stainless lumens that are either:
      • ≥ 1.2 mm ID and ≤ 275 mm in length
      • 1.8 mm ID and < 310 mm in length

    or

    • ≥ 2.8 mm ID and ≤ 317 mm in length
    • Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg).
    Device Description

    The V-PRO maX 2 Low Temperature Sterilization System is a new vaporized hydrogen peroxide sterilizer model to be added to the STERIS V-PRO family of sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1, Amsco V-PRO 1 Plus, V-PRO maX and V-PRO 60 Sterilizers.

    As with the predicate device (K162413), the V-PRO maX 2 Sterilizer has three preprogrammed cycles (the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle) with a fourth added (the Fast Non Lumen Cycle). The V-PRO maX 2 Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.

    The V-PRO maX 2 Sterilizer uses VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its four preprogrammed cycles all utilize a conditioning phase, a sterilize phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    AI/ML Overview

    The provided text describes the V-PRO® maX 2 Low Temperature Sterilization System, a medical device for sterilizing instruments. It is a 510(k) premarket notification, indicating it is intended to be substantially equivalent to a predicate device, not a new type of device requiring extensive clinical trials for efficacy. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to performance tests demonstrating substantial equivalence to the predicate device and the general efficacy of the sterilization process.

    Here's an breakdown based on your request, interpreting "acceptance criteria" as the required successful outcomes of the performance tests for substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test NameAcceptance Criteria (Implied)Reported Device Performance
    AOAC Sporicidal TestAll carriers must be sterile, demonstrating killing of spores.All 720 carriers processed using 3 lots of EOSL sterilant were sterile.
    Determination of D-value and Total Kill EndpointAchieve greater than a 12 log reduction of the most resistant organism within all cycles.Greater than a 12 log reduction of the most resistant organism is achieved within all cycles (Lumen Cycle, Non Lumen Cycle, Flexible Cycle and Fast Non Lumen) of the V-PRO maX 2 Sterilizer.
    1/2 Cycle Modified Total Kill Endpoint VerificationAll sterile results within the validation load for standard and at least one lower injection weight. Partial positives or all survive results seen at even lower injection weights.Modified total kill end point analysis was demonstrated for all four V-PRO maX 2 Sterilizer cycles. The standard injection weight of 2.1 g and at least one lower injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle. Partial positives or all survive results were seen at lower injection weights.
    1/2 Cycle Sterilization Verification of Cycle ClaimsEach cycle must reproducibly sterilize specified load configurations under worst-case conditions in 1/2 cycle.- The Lumen Cycle reproducibly sterilizes single, dual and triple lumen devices under worst case conditions in 1/2 Cycle.- The Flexible Cycle reproducibly sterilizes 1 x 990 mm flexible endoscope lumens under worst case conditions in 1/2 Cycle.- The Non Lumen Cycle reproducibly sterilizes non-lumened devices under worst case conditions in 1/2 Cycle.- The Fast Non Lumen Cycle reproducibly sterilizes non-lumened devices under worst case conditions in 1/2 Cycle.
    ½ Cycle Verification of Mated SurfacesDemonstrate sterile efficacy for mated surfaces under worst-case conditions.Sterile efficacy was demonstrated for mated surfaces under worst case conditions.
    Simulated Use TestVerify the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.Simulated use testing verified the ability of the V-PRO maX 2 Sterilizer cycles to sterilize medical devices under worst case processing conditions.
    In Use TestDemonstrate the ability of the sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.The in use investigation demonstrated the ability of the V-PRO maX 2 Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.
    BiocompatibilityDemonstrate biocompatibility of materials after processing.Cytotoxicity and residue analysis of 23 materials have demonstrated biocompatibility after processing in the V-PRO 60 Sterilizer (implicitly considered relevant for V-PRO maX 2).
    Medical Device Material CompatibilityDemonstrate compatibility of medical devices with the sterilization process after multiple cycles.Evaluation of medical devices after multiple cycles in the V-PRO maX 2 Sterilizer has demonstrated compatibility with 23 materials of construction.
    Final Process QualificationSuccessful qualification for all sterilizer cycles.The V-PRO maX 2 Sterilizer final process qualification was successful for all four (4) sterilizer cycles.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily discusses device performance and sterilization efficacy, not diagnostic accuracy where "test sets" of patient data are typically used. Instead, the "test set" here refers to the biological indicators (BIs) and test loads used in the sterilization efficacy studies.

    • AOAC Sporicidal Test: "All 720 carriers processed using 3 lots of EOSL sterilant were sterile." This implies a total of 720 biological indicators (carriers).
    • Other Efficacy Tests (D-value, Total Kill, 1/2 Cycle Verification, Simulated Use, In Use): These tests involve specific "validation loads" and "worst-case conditions."
      • Non Lumen Cycle: Validation load of two instrument trays for a total weight of 50 lbs (22.7 kg).
      • Fast Non Lumen Cycle: Validation load of one pouched instrument tray for a total weight of 11 lbs (5 kg).
      • Flexible Cycle (Load Config 1): Two flexible endoscopes, each packaged into a tray with silicone mat and light cord.
      • Flexible Cycle (Load Config 2): One flexible endoscope in a tray with silicone mat and light cord, plus an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).
      • Lumen Cycle: A maximum of 20 lumens per load, using a validation load of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg).
      • "Patient-soiled, clinically-cleaned, medical instruments" were used for the in-use test.

    Data Provenance: The studies were conducted by the device manufacturer (STERIS Corporation) internally as part of their premarket notification for the Food and Drug Administration (FDA) in the USA. These are laboratory and simulated/actual use tests, not human data. The context implies these are prospective tests designed to prove the device's performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    This information is not applicable in the context of device sterilization validation. "Ground truth" for sterilization is established by scientific principles of microbial inactivation (e.g., whether biological indicators containing a known number of resistant spores are killed), rather than expert consensus on interpretive tasks. The methodology for assessing sterility (e.g., incubation of biological indicators) relies on established microbiological techniques.

    4. Adjudication Method for the Test Set:

    Not applicable. Sterility tests involve objective outcomes (growth or no growth of microorganisms). There is no "adjudication" in the sense of resolving disagreements between human readers, as for an AI diagnostic tool.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data, and AI might assist in that interpretation. This document is about a sterilization device, which does not involve human interpretation of output in that manner.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device itself is a "standalone" sterilizer. Its performance is evaluated purely on its ability to sterilize medical instruments, without human intervention in the sterilization process once initiated. The tests performed (e.g., AOAC Sporicidal Test, D-value determination) are indeed standalone assessments of the device's efficacy. There is no "human-in-the-loop" concept for the primary function of sterilization.

    7. The Type of Ground Truth Used:

    The ground truth used for these performance tests is microbiological eradication/inactivation.

    • For the AOAC Sporicidal Test, D-value, and Total Kill Endpoint, the ground truth is the absence of viable microorganisms (specifically, highly resistant bacterial spores like Geobacillus stearothermophilus). This is determined by standard microbiological culture techniques (e.g., incubating biological indicators and observing for growth).
    • For the "In Use Test," the ground truth for successful sterilization of patient-soiled instruments would similarly be the absence of microbial growth from processed items.
    • For biocompatibility and material compatibility, the ground truth is established through standardized cytotoxicity assays and visual / functional assessment of materials following repeated processing.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical sterilizer and not an AI/machine learning model that undergoes "training." Its operational parameters are designed and validated based on scientific principles of sterilization, not statistical inference from a training dataset derived from "training examples."

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no "training set" in the context of this traditional device validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160433

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-06

    (141 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Lumen Cycle can sterilize:(a)

    • Lumened and non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices, including single, dual and triple channeled rigid endoscopes, with the following configurations: (a)
      • o single channeled devices with a stainless lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length
      • o dual channeled devices with stainless steel lumens that are > 0.77 mm ID and < 527 mm in length
      • o triple channeled devices with stainless steel lumens that are
        • ≥ 1.2 mm ID and ≤ 275 mm in length
        • ≥ 1.8 mm ID and ≤ 310 mm in length
        • or
        • 2.8 mm ID and < 317 mm in length

    • (a) The validation studies for all channel/lumen configurations were conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

    The V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Non Lumen Cycle can sterilize: (0) Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes and non-luments with stainless steel or titanium diffusion-restricted areas such as the hinged portion of forceps or scissors.

    • (b) The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 50 lbs.

    The V-PRO maX Low Temperature Sterilization System's Flexible Cycle can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of two load configurations:

      1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. (c) The flexible endoscopes may contain either:
      • · a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
      • · or two lumens with:
        • · one lumen that is > 1 mm ID and < 990 mm in length
        • · and the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
      • (c) The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
      1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffiusion-restricted areas such as the hinged portion of forceps or scissors.(d) The flexible endoscope can contain either:
      • · a single lumen that is > 1 mm ID and < 1050 mm in length
      • · or two lumens with:
        • one lumen that is ≥ 1 mm ID and ≤ 990 mm in length
        • · and the other lumen that is ≥ 1 mm and ≤ 850 mm in length
      • (d) The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (ff not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 lbs.
    Device Description

    The V-PRO 1 Plus and V-PRO maX Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Devices that meet the criteria described within the indications for use and composed of compatible materials are placed within the sterilization chamber and the appropriate cycle selected through the graphic user interface. The preprogrammed cycles all utilize a conditioning phase which prepares the chamber, a sterilize phase in which loaded devices are exposed to the sterilizing agent, VHP, at low pressure and an aeration phase to remove residual hydrogen peroxide. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    The V-PRO 1 Plus Low Temperature Sterilization System provides two cycle options, Lumen and Non Lumen. The sterilizer contains a 136 L welded aluminum chamber with two shelves for device loading, an onboard printer to provide cycle printouts, a cup interface where the VAPROX HC cups are loaded, a display for the graphic user interface and the components necessary to achieve the desired cvcle parameters (injection cylinder, vaporizer, vacuum pump and catalytic converter). The side panels of the sterilizer are composed of stainless steel.

    The V-PRO maX Low Temperature Sterilization System provides three cycle options, Lumen, Non Lumen and Flexible. The sterilizer contains a 136 L welded aluminum chamber with two shelves for device loading, an onboard printer to provide cycle printouts, a cup interface where the VAPROX HC cups are loaded, a display for the graphic user interface and the components necessary to achieve the desired cycle parameters (injection cylinder, vacuum pump and catalytic converter). The side panels of the sterilizer are composed of plastic.

    The purpose of this pre-market notification is to modify the maximum load weight for the Non Lumen Cycle from 19.65 to 50 lbs, to add titanium diffusion-restricted areas to the indications for use of the Non Lumen Cycle, and to remove the term "reusable" from indications for use statement.

    AI/ML Overview

    The FDA 510(k) summary for the V-PRO® 1 Plus and V-PRO® maX Low Temperature Sterilization Systems describes performance testing to support the substantial equivalence claim. Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the "PASS" result in the context of sterilization, meaning the device successfully sterilized the test articles under specified conditions. The document acts as a summary, so detailed quantitative acceptance criteria (e.g., specific log reduction of microorganisms) are not explicitly stated in this high-level summary but are inherent to sterilization validation.

    TestAcceptance Criteria (Implied)Reported Device Performance
    ½ Cycle Modified Total Kill Endpoint VerificationAll sterile results within validation load at standard and at least one lower sterilant injection weight. Partial positives/all survive at lower injection weights.PASS
    ½ Cycle Sterilization Verification of Cycle ClaimsReproducibly sterilizes non-lumened devices, mated surfaces, and device tray accessories contact sites under worst-case conditions in ½ Cycle.PASS
    Simulated Use TestVerified ability of Non Lumen Cycle to sterilize medical devices under worst-case processing conditions.PASS
    Final Process QualificationAll cycle specifications met through manual inspection of process parameter data.PASS
    Sterilizer Accessories PerformanceReproducibly sterilizes SCBIs under worst-case ½ cycle conditions; PIs exhibit passing color change; SCBIs negative for growth under full cycle conditions; Chemical indicators show complete color change; Test articles effectively sterilized when packaged as specified.PASS

    2. Sample Size for the Test Set and Data Provenance

    The document does not explicitly state a numerical sample size for the test set in terms of the number of items sterilized in total. However, it provides details on the load configurations used in the validation studies:

    • Lumen Cycle: Maximum of twenty (20) lumens per load, using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
    • Non Lumen Cycle: Validation load consisting of two instrument trays for a total weight of 50 lbs. (This is a modification from the previous 19.65 lbs of two instrument trays and two pouches).
    • Flexible Cycle (Config 1): Two flexible endoscopes (with light cord/mat) with no additional load.
    • Flexible Cycle (Config 2): One flexible endoscope (with light cord/mat) and additional non-lumened instruments (instrument tray and one pouch) for a total load weight of 24.0 lbs.

    The data provenance is not explicitly stated as retrospective or prospective, nor is a country of origin identified for the data. Given it's a pre-market notification for a medical device by the STERIS Corporation (Mentor, OH, USA), it is highly likely the studies were prospective, conducted by the manufacturer, and likely performed in the USA or under relevant international standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not mention the use of human experts to establish ground truth for the test set. Sterilization validation typically relies on objective measures, primarily biological indicators (BIs) containing highly resistant bacterial spores (e.g., Geobacillus stearothermophilus) and chemical indicators (CIs). The "ground truth" for sterilization is the complete inactivation of these spores, indicated by no growth.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the ground truth for sterilization is objective (growth/no growth of biological indicators), not based on human interpretation or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a sterilization system, not an AI-assisted diagnostic or interpretive device that would involve human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is a standalone system. The "algorithm" here is the pre-programmed sterilization cycle, and its performance is assessed directly against the ability to sterilize. There is no human intervention in the sterilization process itself; the human selects the cycle and loads the sterilizer. The performance data presented (e.g., ½ Cycle Modified Total Kill Endpoint Verification, Simulated Use Test) are indeed standalone performance of the device without human interpretation of the sterilization outcome itself.

    7. Type of Ground Truth Used

    The primary ground truth for sterilization efficacy is the inactivation of biological indicators (BIs). The document explicitly mentions:

    • "reproducibly sterilized under worst case ½ cycle conditions" for STERIS VERIFY® V24 Self-Contained Biological Indicator (SCBI).
    • "all SCBIs were negative for growth under worst case full cycle conditions."

    Additionally, chemical indicators (CIs) are used to confirm exposure to the sterilant, showing "complete color change."

    8. Sample Size for the Training Set

    Not applicable. This device is a physical sterilizer and its associated pre-programmed cycles, not a machine learning model that requires a "training set" in the conventional AI sense. The "training" of the cycles would refer to the developmental and engineering process of determining optimal parameters, which involves a different type of experimental design and validation than data-driven machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reasons as point 8. The "ground truth" for developing the sterilizer's parameters (e.g., sterilant injection, exposure time, pressure, temperature) would have been established through extensive empirical experimentation and adherence to sterilization standards (e.g., AAMI, ISO) to achieve a validated Sterility Assurance Level (SAL) of 10^-6 or better. This involves progressively testing parameters with biological indicators to determine the minimum conditions required for sterility.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140498

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-24

    (147 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K102330, K083097, K102394, K111810, K112760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-PRO 60 Low Temperature Sterilization System, with VAPROX® HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

    The V-PRO 60 Sterilizer's Lumen Cycle can sterilize:

    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Non-lumened devices including non-lumened rigid and semi-rigid endoscopes
    • Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:
      • single or dual lumen devices with stainless steel lumens that are ≥ 0.77 mm (~1/32") internal diameter (ID) and ≤ 410 mm (16-9/64") in length
      • triple lumen devices with stainless steel lumens that are
        • ≥ 1.2 mm (~3/64") ID and ≤ 275 mm (~10-55/64") in length
        • ≥ 1.8 mm (~5/64") ID and ≤ 310 mm (~12-13/64") in length
        • or ≥ 2.8 mm (~7/64") ID and ≤ 317 mm (12-31/64") in length

    The V-PRO 60 Sterilizer's Non Lumen Cycle can sterilize

    • Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    The V-PRO 60 Sterilizer's Flexible Cycle can sterilize:

    • One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:
      • single or dual lumen device with lumens that are ≥ 1 mm (~3/64") ID and ≤ 990 mm (38-63/64") in length
    Device Description

    The V-PRO 60 Low Temperature Sterilization System is a new vaporized hydrogen peroxide sterilizer model to be added to the STERIS V-PRO family of sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1, Amsco V-PRO 1 Plus and Amsco V-PRO maX Sterilizers. As with the predicate device (K102330), the V-PRO 60 Sterilizer has three preprogrammed cycles: the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle. The V-PRO 60 Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities. The V-PRO 60 Sterilizer uses VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its three preprogrammed cycles all utilize a conditioning phase, a sterilize phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.

    AI/ML Overview

    This document describes the V-PRO® 60 Low Temperature Sterilization System, a vaporized hydrogen peroxide sterilizer. The acceptance criteria are based on achieving sterility and maintaining material compatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied)Reported Device Performance
    Sterilization Efficacy
    AOAC Sporicidal TestAll carriers processed should be sterile.All 720 carriers processed using 3 lots of EOSL sterilant were sterile.
    Determination of D-value and Total Kill EndpointGreater than a 12 log reduction of the most resistant organism.Greater than a 12 log reduction of the most resistant organism is achieved within all cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) of the V-PRO 60 Sterilizer.
    ½ Cycle Modified Total Kill Endpoint VerificationAll sterile results within the validation load.Modified total kill end point analysis was demonstrated for all three V-PRO 60 Sterilizer cycles. The standard injection weight of 1.1 g and at least one lower injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle. Partial positives or all survive results were seen at lower injection weights.
    ½ Cycle Sterilization Verification of Cycle ClaimsReproducible sterilization of specified devices/lumens.• The Lumen Cycle reproducibly sterilizes single, dual and triple lumen devices under worst case conditions in ½ Cycle. • The Flexible Cycle reproducibly sterilizes 1 x 990 mm flexible endoscope lumens under worst case conditions in ½ Cycle. • The Non Lumen Cycle reproducibly sterilizes non-lumened devices under worst case conditions in ½ Cycle.
    ½ Cycle Verification of Mated SurfacesSterile efficacy for mated surfaces under worst case conditions.Sterile efficacy was demonstrated for mated surfaces under worst case conditions.
    Simulated Use TestSterilization of medical devices under worst case conditions.Simulated use testing verified the ability of the V-PRO 60 Sterilizer cycles to sterilize medical devices under worst case processing conditions.
    In Use TestSterilization of patient-soiled, clinically-cleaned instruments.The in use investigation demonstrated the ability of the V-PRO 60 Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.
    Device Compatibility & Integrity
    BiocompatibilityBiocompatibility after processing.Cytotoxicity and residue analysis of 23 materials have demonstrated biocompatibility after processing in the V-PRO 60 Sterilizer.
    Medical Device Material CompatibilityCompatibility with materials of construction after multiple cycles.Evaluation of medical devices after multiple cycles in the V-PRO 60 Sterilizer has demonstrated compatibility with 23 materials of construction.
    Process Control & Verification
    Final Process QualificationSuccessful qualification for all cycles.The V-PRO 60 Sterilizer final process qualification was successful for all three (3) sterilizer cycles. All three lots of CI exhibited complete color change. All three SCBI PIs exhibited a passing color change and all SCBIs were negative for growth. Manual inspection of the process parameter data confirmed that all cycle specifications were met.
    Conformity to Standards (Safety and EMC)Compliance with EN 61010-1:2001 and EN61326-1:2006.The V-PRO 60 Low Temperature Sterilization System has been tested for conformity and is certified to EN 61010-1:2001 and EN61326-1:2006.

    Study Proving Device Meets Acceptance Criteria:

    The document summarizes the "Summary of Nonclinical Tests" performed on the V-PRO 60 Low Temperature Sterilization System to demonstrate its substantial equivalence to the predicate device, the V-PRO maX Low Temperature Sterilization System. The studies aim to prove the device's ability to sterilize various medical devices effectively and safely.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document provides specific sample sizes for particular tests:

    • AOAC Sporicidal Test: "All 720 carriers processed using 3 lots of EOSL sterilant."
    • Validation Loads:
      • Lumen Cycle: "a maximum of twelve (12) lumens per load... consisting of one instrument tray with instrument organizers and mat and two pouches for a total weight of 11 lbs (5.0 kg)."
      • Non Lumen Cycle: "a validation load consisting of one instrument tray with instrument organizers and mat and one pouch for a total weight of 12 lbs (5.4 kg)."
      • Flexible Cycle: "one flexible endoscope, packaged into a tray with silicone mat, instrument organizers and light cord (if not integral to scope) and no additional load."
    • Biocompatibility and Material Compatibility: "23 materials" were analyzed.

    Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective. However, based on the nature of medical device clearance submissions to the FDA (United States Food and Drug Administration), these are typically prospective laboratory and simulated-use studies conducted by the manufacturer or authorized testing facilities, aiming to meet regulatory requirements prior to market entry.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not specify the "number of experts" or their "qualifications" used to establish ground truth. The studies involve standardized microbiological and engineering tests rather than expert interpretation of results in a clinical setting. For sterilization studies, the ground truth is typically established by laboratory methods such as growth/no growth of biological indicators (e.g., Geobacillus stearothermophilus) following exposure to the sterilization process.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests described are objective performance evaluations (e.g., sterility, log reduction, material compatibility) rather than interpretation-based assessments that would require an adjudication method like 2+1 or 3+1.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The device is a sterilizer, and its performance is evaluated through objective physical and biological tests, not through human reader interpretation of cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies described are standalone performance evaluations of the sterilization device. There is no "human-in-the-loop" interaction in the output or interpretation of the sterilization cycle results beyond operating the device and reading standard indicators. The device itself is the "algorithm only" in the context of its function.

    7. The Type of Ground Truth Used:

    The primary type of ground truth used would be:

    • Microbiological Negativity/Positivity: For sterilization efficacy tests (AOAC Sporicidal, D-value/Total Kill, ½ Cycle Modified Total Kill Endpoint, ½ Cycle Sterilization Verification, Simulated Use, In Use), the ground truth is established by culturing biological indicators or test organisms and observing the presence or absence of growth. "Sterile" or "no growth" is the ground truth for effective sterilization.
    • Physical/Chemical Analysis: For biocompatibility and material compatibility, the ground truth is established through laboratory analyses (e.g., cytotoxicity assays, residue analysis, visual inspection for degradation) and quantitative measurements.
    • Process Parameter Verification: For Final Process Qualification, the ground truth is established by objective measurements of process parameters (time, temperature, pressure, sterilant injection weight) and the observed color change of chemical indicators.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical device (sterilizer) and not an AI/machine learning model that would require a 'training set' of data in the conventional sense. The development of the sterilizer's cycles and parameters is based on engineering principles and iterative testing, not on data training in the way an algorithm learns from a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no "training set" in the context of an AI/machine learning algorithm for this device. The physical and biological principles governing sterilization, along with extensive engineering development and testing, define the performance of the sterilizer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131120

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-16

    (85 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K083097, K102394, K111810, K102330, K112760

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amsco V-PRO I, V-PRO I Plus and V-PRO maX Low Temperature Sterilization Systems, with VAPROX® HC Sterilant, are vaporized hydrogen peroxide sterilizers intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.
    The Amsco V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Lumen Cycle, the subject of this submission, can sterilize:

    • Lumened and non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:
      • single channeled devices with a stainless lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length
      • dual channeled devices with stainless steel lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
      • triple channeled devices with stainless steel lumens that are either
        • ≥ 1.2 mm ID and ≤ 275 mm in length
        • ≥ 1.8 mm ID and ≤ 310 mm in length
        • or
        • ≥ 2.8 mm ID and ≤ 317 mm in length
    • The validation studies for all channel/lumen configurations were conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
      The Amsco V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Non Lumen Cycle, cleared under K083097, K102394 and K111810, can sterilize:
    • Non-lumened instruments including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
      • The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 Ibs.
        The Amsco V-PRO maX Low Temperature Sterilization System's Flexible Cycle, cleared under K102330 and K112760, can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of two load configurations:
      1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.
        The flexible endoscopes may contain either:
    • a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
    • or two lumens with:
      • one lumen that is ≥ 1 mm ID and ≤ 998 mm in length
      • and the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
    • The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
      1. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors.
        The flexible endoscope can contain either:
    • a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
    • or two lumens with:
      • one lumen that is ≥ 1 mm ID and ≤ 998 mm in length
      • and the other lumen that is > 1 mm and ≤ 850 mm in length
    • The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 Ibs.
    Device Description

    The Amsco V-PRO 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems are vaporized hydrogen peroxide sterilizers. The V-PRO Sterilizers use VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). The pre-programmed V-PRO cycles all utilize a conditioning phase, a sterilization phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Amsco V-PRO 1, V-PRO 1 Plus, and V-PRO maX Low Temperature Sterilization Systems. It focuses on updating the indications for use statement for the Lumen Cycle.

    Here's the breakdown of the acceptance criteria and the study information based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria are implicitly defined by the successful demonstration of sterilization capability under various test conditions for the device's Lumen Cycle, especially for the modified stainless steel lumen claims. The "Conclusion" column in the table below explicitly states "PASS" for all tests, indicating that these criteria were met.

    TestAcceptance Criteria (Implied)Reported Device Performance
    ½ Cycle Modified Total Kill Endpoint VerificationDemonstrate a modified total kill endpoint analysis, showing all sterile results for specified Lumen Cycle claims at standard and lower injection weights, and survival at the lowest weight.Modified total kill endpoint analysis was demonstrated. The standard injection weight of 2.1 g and a lower injection weight of 1.457 g resulted in all sterile results for the Lumen Cycle modified stainless steel lumen claims. All survival results were shown at the lowest weight evaluated (0.121 g). Partial positives results were seen at the intermediate injection weights.
    ½ Cycle Sterilization Verification of Lumen ClaimsReproducibly sterilize the modified stainless steel lumen claims under worst-case conditions in a V-PRO Lumen ½ Cycle.The V-PRO Lumen Cycle reproducibly sterilizes the modified stainless steel lumen claims under worst case conditions in a V-PRO Lumen ½ Cycle.
    ½ Cycle Sterilization Verification of Double Pouched LumensReproducibly sterilize the modified stainless steel lumen claims under worst-case conditions in a V-PRO Lumen ½ Cycle in a double pouched configuration.The V-PRO Lumen Cycle reproducibly sterilizes the modified stainless steel lumen claims under worst case conditions in a V-PRO Lumen ½ Cycle in a double pouched configuration.
    Simulated Use TestVerify the ability of the V-PRO Lumen Cycle to sterilize the modified stainless steel lumen claims under worst-case processing conditions.Simulated use testing verified the ability of the V-PRO Lumen Cycle to sterilize the modified stainless steel lumen claims under worst case processing conditions.
    In Use TestDemonstrate the ability of the V-PRO Lumen Cycle to sterilize clinically cleaned, patient-soiled medical instruments representative of the modified stainless steel lumen claims.The in use investigation demonstrated the ability of the V-PRO Lumen Cycle to sterilize clinically cleaned, patient-soiled medical instruments representative of the modified stainless steel lumen claims.
    BiocompatibilityDemonstrate biocompatibility after processing in the V-PRO Lumen Cycle through residue analysis.Residue analysis evaluation has demonstrated biocompatibility after processing in the V-PRO Lumen Cycle.
    Validation of Accessories for use with the V-PRO Lumen Cycle Modified Lumen Claims (V-PRO Sterilization Trays)Successfully sterilize the modified stainless lumens claims in V-PRO Sterilization Trays under Lumen ½ cycle worst-case conditions.The modified stainless lumens claims were successfully sterilized in the V-PRO Sterilization Trays under Lumen ½ cycle worst case conditions.
    Validation of Accessories for use with the V-PRO Lumen Cycle Modified Lumen Claims (Vis-U-All Tyvek Pouches)Successfully sterilize the modified stainless lumens claims in Vis-U-All Tyvek Pouches under ½ cycle worst-case conditions.The modified stainless lumens claims were successfully sterilized in the Vis-U-All Tyvek Pouches under ½ cycle worst case conditions.

    Study Details

    Based on the provided text, the document describes nonclinical tests primarily focused on verifying the sterilization efficacy of the device.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: The exact number of individual devices or lumens tested for each specific test is not explicitly stated in numerical form. However, for the "Lumen Cycle," the validation studies for all channel/lumen configurations were conducted using a maximum of twenty (20) lumens per load. The validation load consisted of two instrument trays and two pouches for a total weight of 19.65 lbs.
      • Data Provenance: The document does not specify the country of origin of the data. It is a 510(k) submission to the US FDA, implying the studies were conducted to meet US regulatory requirements. The studies appear to be prospective validation studies conducted by the manufacturer, STERIS Corporation, for the purpose of demonstrating the device's efficacy.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not mention the use of experts or their qualifications for establishing the ground truth. Sterilization efficacy studies like these typically rely on microbiological testing (e.g., spore-kill assays) to establish an objective "ground truth" of sterility, rather than expert human interpretation.
    3. Adjudication method for the test set:

      • Adjudication methods (like 2+1, 3+1) are typically used in studies involving human readers or subjective assessments. This document describes objective performance testing (sterilization efficacy), so such adjudication methods are not applicable and therefore not mentioned.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a study on a sterilization device, not an AI-assisted diagnostic device that would involve human readers. Therefore, an MRMC study and related effect sizes are not relevant or described.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in a sense. The tests described are "standalone" performance evaluations of the sterilization system's ability to achieve sterility without direct human intervention during the sterilization cycle itself. The device is an automated sterilizer, so its performance is inherently "algorithm-only" in its operation. However, no "algorithm" in the typical AI sense is discussed.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth in these sterilization studies is microbiological sterility. The "½ Cycle Modified Total Kill Endpoint Verification" and "½ Cycle Sterilization Verification" tests indicate the use of biological indicators (spores) to definitively determine if sterilization was achieved (total kill/survival results). The "In Use Test" also mentions sterilization of "patient-soiled medical instruments," implying a microbiological assessment of sterility. Biocompatibility is assessed by "residue analysis."
    7. The sample size for the training set:

      • This document describes validation and verification studies for a physical sterilization device, not a machine learning model. Therefore, the concept of a "training set" for an algorithm is not applicable and not mentioned.
    8. How the ground truth for the training set was established:

      • As there is no training set for an AI algorithm, this information is not applicable.