Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K123185

    Validate with FDA (Live)

    Date Cleared
    2012-11-02

    (23 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams.

    Device Description

    DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-Mode, M-Mode, PW-Mode, CW mode, Color-Mode, Color M-Mode, Power/Dirpower Mode, TDI mode, 3D/4D mode, Elastography or the combined mode (i.e. B/M-Mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array with a frequency range of approximately 3 MHz to 10.0 MHz.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System. This type of submission is a statement to the FDA that the device is substantially equivalent to another legally marketed device, not a claim of superior performance or a detailed study of acceptance criteria against novel performance metrics.

    Therefore, the provided text does not contain the acceptance criteria or a study proving the device meets those criteria in the context of specific performance metrics for diagnosis or improved human reader performance with AI assistance. Instead, it focuses on demonstrating substantial equivalence to predicate devices and adherence to general safety and performance standards.

    Here's an analysis of the provided information relative to your requested points:


    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices and compliance with recognized safety and performance standards, rather than defining specific outcome-based acceptance criteria and reporting performance against them.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided as there is no clinical performance study detailed in this 510(k) summary. The document states "DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety." These are non-clinical hardware tests, not clinical performance studies involving a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable/not provided. As no clinical performance study is detailed, there's no mention of a test set, ground truth establishment, or experts for that purpose.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. No clinical performance study or test set requiring adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done, and there is no mention of AI assistance in this document. The device is a general-purpose diagnostic ultrasound system, not explicitly an AI-powered diagnostic tool. The purpose of this 510(k) is to establish substantial equivalence, not to demonstrate improved human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable/not provided. The device is a diagnostic ultrasound system, which inherently involves human operation and interpretation. There is no mention of a standalone algorithm or AI performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not applicable/not provided. Since no clinical performance study with a test set is detailed, there's no mention of the type of ground truth used.

    8. The sample size for the training set

    This information is not applicable/not provided. The device is an ultrasound system. There is no mention of a "training set" in the context of an algorithm or AI.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. As there's no training set mentioned in the context of an algorithm or AI, there's no corresponding information on how ground truth would be established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121807

    Validate with FDA (Live)

    Device Name
    V5MS TRANSDUCER
    Date Cleared
    2012-07-10

    (20 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    V5Ms trans-esophageal echocardiograph (TEE) ultrasound transducer is intended primarily for cardiology applications.

    ACUSON Antares Ultrasound System: The Acuson Antares ultrasound imaging system is intended for the following applications: Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Intracardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal. Great Vessel, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    ACUSON CV70 Cardiovascular System: The CV70 ultrasound imaging system is intended for the following applications: Abdominal. Intraoperative, Small Parts, Transcranial, OB/GYN, Cardiac, Intracardiac, Transesophageal, Pelvic, Neonatal/Adult Gephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal, Great Vessel, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    ACUSON Cypress Ultrasound System: The Cypress platform is intended for use in the following applications: General Imaging and Cardiology for Fetal. Abdominal. Intraoperative (cardiac), Pediatrics, Neonatal Cephalic, Cardiac (adult, pediatric), Trans-esophageal, Peripheral Vessel, Intraluminal and Intra-cardiac applications, and intended uses as defined in the FDA quidance document. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes

    ACUSON S1000/S2000 Ultrasound System: The S2000™ ultrasound imaging systems are intended for the following applications: Fetal. Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal(conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging."

    ACUSON SC2000 Ultrasound System: The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications.

    ACUSON Sequoia Ultrasound System: The Siemens ACUSON Sequoia ultrasound imaging system is intended for the following applications: General Imaging and Cardiology for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breast, testes, thyroid and penis), Neonatal/Adult Cephalic, Cardiac (adult, pediatric, and neonatal), Trans-esophageal, Transrectal, Transvaginal, Peripheral Vessel, and Musculo-sceletal (superficial and conventional) applications, and intended uses as defined in the FDA quidance document. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

    ACUSON X300 Ultrasound System: The Siemens ACUSON X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides measurement of anatomical structures and analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients reqarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement: Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

    Device Description

    The V5Ms Transesophageal Transducer consists of a gastroscope control housing where nosepiece articulation and transducer rotation are controlled. A flexible transesophageal guide tube with a nosepiece containing the acoustic array extends from one end of the control housing and the system cable/connector extends from the other end. The acoustic array has 64 elements and rotates 180 degrees to provide imaging planes from transverse view to inverse transverse view. Rotation is powered by a motor in the control housing and is controlled by the operator using a switch button on the control housing for clockwise and counterclockwise rotation. Nosepiece articulation is achieved by manipulating a vertebrae section adjacent to the nosepiece through a series of control wires attached to knobs on the control housing. The nosepiece can be articulated in both the anterior/posterior and left/right directions.

    AI/ML Overview

    Here's a breakdown of the requested information, based on the provided document:

    Acceptance Criteria and Study Details for V5Ms Transesophageal Transducer

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) submission document does not explicitly state quantitative acceptance criteria or specific device performance metrics in the way one might expect for a new or significantly modified device. Instead, the submission relies heavily on substantial equivalence to a previously cleared predicate device (Siemens V5Ms Transesophageal Transducer, K052021). The key "performance data" presented is that "Performance data is not required for this modification as there is no change to software or hardware." This implies that the device is expected to perform identically to its predicate because the changes are limited to patient contact materials.

    Therefore, the acceptance criteria are implicitly met by demonstrating that the modifications do not negatively impact the established performance of the predicate device.

    Implicit Acceptance Criteria and Reported Performance (Based on Substantial Equivalence):

    Acceptance Criterion (Implicit)Reported Device Performance
    Safety and EffectivenessThe modified device is safe and effective for its intended uses, as demonstrated by its substantial equivalence to the predicate device (Siemens V5Ms Transesophageal Transducer, K052021). No changes to software or hardware, thus performance is maintained.
    BiocompatibilityMeets ISO10993-1 standards for cytotoxicity, sensitization, and irritation.
    Intended Use (across various ultrasound systems)The device is intended for the same cardiology applications and clinical applications (e.g., Transesophageal, Cardiac, various Doppler modes, 3D imaging, Harmonic Imaging) as the predicate when used with compatible ACUSON systems. The intended uses are identical to the predicate device.
    Technological CharacteristicsThe device possesses the same fundamental technological characteristics (e.g., phased array transducer, 64 elements, 180-degree rotation, nosepiece articulation) as the predicate, with modifications only to patient contact materials.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable (N/A). The document explicitly states, "Performance data is not required for this modification as there is no change to software or hardware." Therefore, there was no clinical performance test set used. The evaluation centered on non-clinical tests for biocompatibility and asserting substantial equivalence.
    • Data Provenance: N/A for clinical performance data. For biocompatibility testing, the data would typically originate from a laboratory (likely in the USA or a facility that adheres to international standards like ISO). The origin is not specified beyond "Nonclinical tests contained in this submission include biocompatibility testing per ISO10993-1 (cytotoxicity, sensitization, irritation)."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • N/A. As no clinical performance test set was conducted, there was no need for experts to establish ground truth. The primary "experts" involved in the submission would be Siemens' own regulatory and engineering teams, and subsequently, FDA reviewers.

    4. Adjudication Method for the Test Set

    • N/A. Since no clinical performance test set was conducted, no adjudication method was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. No MRMC study was conducted. The submission is a 510(k) for substantial equivalence based on minor modifications (patient contact materials) to an existing device, not a new clinical effectiveness study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • N/A. The device is a diagnostic ultrasound transducer, not an AI algorithm. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply.

    7. The Type of Ground Truth Used

    • N/A for clinical effectiveness. For biocompatibility, the "ground truth" is defined by the ISO10993-1 standard for cytotoxicity, sensitization, and irritation. This involves standardized biological tests.

    8. The Sample Size for the Training Set

    • N/A. This device is a hardware component (transducer) and does not involve machine learning or AI, and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • N/A. As there is no training set for this device, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113647

    Validate with FDA (Live)

    Date Cleared
    2011-12-16

    (4 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams.

    Device Description

    DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasound diagnostic system. Its function is to acquire and display ultrasound images in B-Mode, M-Mode, PW-Mode, CW mode, Color-Mode, Color M-Mode, Power/Dirpower Mode, TDI mode, 4D mode or the combined mode (i.e. B/M-Mode).This system is a Track 3 device that employs an array of probes that include linear array, convex array and phased array with a frequency range of approximately 3 MHz to 10.0 MHz.

    AI/ML Overview

    The provided text is a 510(k) summary for the Shenzhen Mindray Bio-medical Electronics Co., LTD DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System. This document focuses on demonstrating substantial equivalence to predicate devices and adherence to medical safety standards rather than establishing new performance criteria through clinical studies.

    Therefore, the document does NOT contain information regarding:

    • Specific acceptance criteria framed as quantitative metrics (e.g., sensitivity, specificity, accuracy).
    • Details of a study proving the device meets acceptance criteria in terms of clinical performance.
    • Sample sizes used for test or training sets for an AI/algorithm.
    • Data provenance, number or qualifications of experts, or adjudication methods for establishing ground truth related to clinical performance.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies or effect sizes for human readers with/without AI assistance.
    • Standalone algorithm performance studies.
    • The type of ground truth used as it relates to clinical performance studies.
    • The sample size for the training set or how ground truth was established for a training set.

    The document primarily outlines the device's technical specifications, intended uses, and lists applicable safety and electrical standards it conforms to. It states that the device has been evaluated for "acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety" and found to conform with applicable medical safety standards.

    However, based on the provided text, I can infer the "acceptance criteria" and "study" in a regulatory context, which is demonstrating substantial equivalence to predicate devices and compliance with recognized standards.

    Here's a breakdown of what can be extracted based on that understanding:

    1. Table of "Acceptance Criteria" and "Reported Device Performance":

    In the context of a 510(k) submission for a diagnostic ultrasound system like this, the "acceptance criteria" are not reported as specific performance metrics (like sensitivity/specificity) but as conformance to established safety and performance standards, and substantial equivalence to legally marketed predicate devices. The "reported device performance" is the statement of conformity.

    Acceptance Criteria (Regulatory Context)Reported Device Performance (as stated in the 510(k) Summary)
    Substantial Equivalence to Predicate Devices (similar technological characteristics, comparable safety & effectiveness, same intended uses)"DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices: Mindray DC-7 (K103583), Mindray M7 (K103677), GE Voluson E8 (K101236), Mindray DC-T6 (K110199), Sonosite M-Turbo (K101757), Siemens ACUSON S2000 (K112596), Siemens ACUSON SEQUOIA 512 (K063085), GE LOGIQ e (K102256). They have the similar technological characteristics, are comparable in key safety and effectiveness features, and have the same intended uses and basic operating modes as the predicate devices."
    Conformance with Applicable Medical Safety Standards"DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical safety standards. This device has been designed to meet the following standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, IEC 62304, IEC 62366, UL 60601-1, ISO14971, UD 2, UD 3 and ISO 10993-1."
    Quality System Compliance"The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems."
    Intended Uses are Consistent with Traditional Clinical Practices and FDA Guidelines"Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination."

    2. Sample size used for the test set and the data provenance:

    The document does not detail specific clinical studies with test sets, sample sizes, or data provenance (country, retrospective/prospective) for proving diagnostic performance. The evaluations mentioned are primarily for safety, electrical, and mechanical conformity.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as no described clinical performance study with a test set requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The device described is a general diagnostic ultrasound system, not an AI-powered diagnostic algorithm aiding human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is a diagnostic ultrasound system, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable in the context of clinical diagnostic performance. For technical evaluations, the "ground truth" would be the specifications and requirements defined by the applicable standards (e.g., standard measures for acoustic output, material properties for biocompatibility testing).

    8. The sample size for the training set:

    Not applicable, as no described AI/algorithm training process.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1